Life Spine ALIF Buttress Plate System

K222628 · Life Spine, Inc. · KWQ · Nov 16, 2022 · Orthopedic

Device Facts

Record IDK222628
Device NameLife Spine ALIF Buttress Plate System
ApplicantLife Spine, Inc.
Product CodeKWQ · Orthopedic
Decision DateNov 16, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Life Spine ALIF Buttress Plate System in conjunction with traditional rigid fixation is intended for use in spinal fusion procedures of the thoracolumbar to S1 spinal region as a means to maintain the relative position of weak bony tissue such as allografts or autografts. This device is not intended for load bearing applications.

Device Story

Life Spine ALIF Buttress Plate System consists of titanium alloy plates and screws; used in spinal fusion procedures (thoracolumbar to S1) to maintain position of bone grafts (allografts/autografts). Device functions as non-load-bearing fixation; used in conjunction with traditional rigid fixation. Operated by surgeons in clinical/OR settings. Benefits patient by stabilizing graft material during fusion process.

Clinical Evidence

No clinical data. Substantial equivalence based on design, material, and performance characteristics compared to predicate devices.

Technological Characteristics

Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. System includes plates and screws. Non-load-bearing spinal fixation orthosis. Single-use only.

Indications for Use

Indicated for patients undergoing spinal fusion procedures in the thoracolumbar to S1 spinal region requiring maintenance of relative position of weak bony tissue (allografts or autografts). Not for load-bearing applications.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 16, 2022 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Life Spine, Inc. Angela Batker RA/QA Manager 13951 S. Quality Drive Huntley, Illinois 60142 Re: K222628 Trade/Device Name: Life Spine ALIF Buttress Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: August 22, 2022 Received: August 31, 2022 Dear Angela Batker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Anne D. Talley -S for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K222628 Device Name Life Spine ALIF Buttress Plate System Indications for Use (Describe) The Life Spine ALIF Buttress Plate System in conjunction with traditional rigid fixation is intended for use in spinal fusion procedures of the thoracolumbar to S1 spinal region as a means to maintain the relative position of weak bony tissue such as allografts or autografts. This device is not intended for load bearing applications. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary ALIF Buttress Plate | Submitted By: | Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 | |-----------------------|----------------------------------------------------------------------------------------------------------------------------------| | 510(k) Contact: | Angela Batker<br>Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 | | Date Prepared: | August 24th, 2022 | | Trade Name: | Life Spine ALIF Buttress Plate System | | Common Name: | Appliance, Fixation, Spinal Intervertebral Body | | Classification: | KWQ, 21 CFR 888.3060 Spinal Intervertebral Body Fixation Orthosis,<br>Class II | | Primary Predicate: | Vail Buttress Plate System (K180755) | | Additional Predicate: | RCS Anterior Buttress Plate System (K092659)<br>Life Spine Sentry Lateral Plate System (K180166) | ### Device Description: The Life Spine ALIF Buttress Plate System consists of a variety of plates and screws to suit the individual pathology and anatomical conditions of the patient. All components are fabricated and manufactured from titanium alloy Ti-6A1-4V-ELI per ASTM F-136. All implants are intended for single use only and should not be reused under any circumstances. Do not use any of the ALIF Buttress Plate System components with components from any other system or manufacturer. The ALIF Buttress Plate System components should never be reused under any circumstances. # Indications for Use of the Device: The Life Spine ALIF Buttress Plate System in conjunction with traditional rigid fixation is intended for use in spinal fusion procedures of the thoracolumbar to S1 spinal region as a means to maintain the relative position of weak bony tissue such as allografts. This device is not intended for load bearing applications. {4}------------------------------------------------ K222628 Page 2 of 2 ### Technological Characteristics: The ALIF Buttress Plate System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing. ### Material: This submission seeks clearance of a device made from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. This this is the same material used in the predicate devices. ### Substantial Equivalence: The ALIF Buttress Plate System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and performance. ### Conclusion: The information presented demonstrates the substantial equivalency of the ALIF Buttress Plate System.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...