K222603 · Wallaby Medical, Inc. · DQY · Mar 2, 2023 · Cardiovascular
Device Facts
Record ID
K222603
Device Name
6F Wallaby Long Sheath
Applicant
Wallaby Medical, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Mar 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The 6F Wallaby Long Sheath is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
Device Story
6F Wallaby Long Sheath is a single-use, vascular catheter; single lumen, variable stiffness, composite design. Inputs: manual insertion into peripheral/neuro vasculature. Operation: catheter advanced over guidewire to target site; distal tip visible under fluoroscopy; hydrophilic coating reduces friction; semi-rigid proximal shaft transitions to flexible distal shaft for tortuous anatomy. Output: provides access channel for interventional devices. Used in clinical settings by physicians. Healthcare provider uses fluoroscopic visualization to guide placement; output facilitates delivery of therapeutic devices. Benefits: enables minimally invasive access to peripheral and neuro vasculature.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including visual/dimensional inspection, simulated use in tortuous anatomical models, delivery/retrieval force testing, tip stiffness, torque strength, coating integrity/lubricity, burst/leak testing, kink resistance, particulate testing, and radiopacity. Biocompatibility testing performed per ISO 10993-1:2018.
{0}------------------------------------------------
March 2, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Wallaby Medical David Ruiz Staff Quality Engineer 22901 Mill Creek Drive Laguna Hills, California 92653
# Re: K222603
Trade/Device Name: 6F Wallaby Long Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, QJP Dated: February 1, 2023 Received: February 1, 2023
# Dear David Ruiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K22603
Device Name 6F Wallaby Long Sheath
Indications for Use (Describe)
The 6F Wallaby Long Sheath is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
Type of Use (Select one or both, as applicable)
V Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K222603 510(k) Summary
As required by 21 CFR 807.92:
| Applicant: | Wallaby Medical<br>22901 Mill Creek Drive<br>Laguna Hills, CA 92653 |
|------------------------|-----------------------------------------------------------------------------------------------------|
| Contact: | David Ruiz |
| Phone Number: | (1)-951-818-7984 |
| Date Prepared: | 03/02/23 |
| Device Trade Name: | 6F Wallaby Long Sheath |
| Device Common Name: | Sheath, Access |
| Classification Name: | DQY (Catheter, Percutaneous) and QJP (Catheter, Percutaneous,<br>Neurovasculature), 21 CFR 870.1250 |
| Predicate Device Name: | Neuron MAX System (K111380) |
#### Device Description a.
The 6F Wallaby Long Sheath is a single-use, vascular catheter consisting of a single lumen, variable stiffness, composite catheter. The device has an inner diameter (ID) of 0.088 inch and outer diameter (OD) of 0.105 inch designed with three different working lengths (80 cm, 90 cm, and 100 cm) and two different tip configurations (straight and multipurpose curve). The distal tip of the 6F Wallaby Long Sheath is visible under fluoroscopy and the distal shaft of the catheter is designed with an external hydrophilic coating to reduce friction during use. The proximal end of the catheter incorporates a strain relief and a standard Luer adapter to facilitate the attachment of accessories. The 6F Wallaby Long Sheath has a semi-rigid proximal shaft which transitions into a flexible distal shaft to facilitate the advancement of the catheter in tortuous anatomy.
The 6F Wallaby Long Sheath is a non-active, surgically invasive device intended for short term use within the vasculature.
#### b. Indications for Use
The 6F Wallaby Long Sheath is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
#### C. Predicate Comparison
The predicate device for the 6F Wallaby Long Sheath is the Neuron MAX System cleared under K111380. The table below describes the technological differences between the 6F Wallaby Long Sheath and the Neuron MAX System:
| Device Name | Predicate Device:<br>Neuron MAX System | Subject Device:<br>6F Wallaby Long Sheath | Rationale for Difference |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K111380 | K222603 | |
| Classification | Class II, DQY | Class II, DQY and QJP | QJP product code added |
| Device Name | Predicate Device:<br>Neuron MAX System | Subject Device:<br>6F Wallaby Long Sheath | Rationale for Difference |
| Indications for<br>Use | The Neuron MAX System is<br>indicated for the introduction of<br>interventional devices into the<br>peripheral, coronary, and neuro<br>vasculature. | The 6F Wallaby Long Sheath is<br>indicated for the introduction<br>of interventional devices into<br>the peripheral and neuro<br>vasculature. | The 6F Wallaby Long Sheath is<br>not indicated for use in the<br>coronary vasculature. |
| Materials<br>Shaft | | | |
| Extrusions | Outer layer:<br>Polyurethane, polyamide<br><br>Inner layer:<br>PTFE | Outer layer:<br>Neusoft (polyurethane),<br>PEBAX, polyamide, PET<br><br>Inner layer:<br>PTFE | Both device materials are<br>biocompatible, designed to be<br>used in the vasculature. |
| Wire<br>Reinforcement | Stainless Steel | Stainless Steel and Nitinol | |
| Components | | | |
| Hub | Polyamide | Polyamide | SAME |
| Coating | Hydrophilic Coating | Hydrophilic Coating | The subject device materials are<br>biocompatible and designed to<br>be used in vasculature. |
| Strain Relief | Stainless Steel | Pebax | biocompatible and designed to<br>be used in vasculature. |
| Colorant | Clear/Natural or Blue | Clear/Natural or Purple | The subject device colorants<br>are biocompatible. |
| Marker Band | C-cut Pt/Ir Band | C-cut Pt/Ir Band | SAME |
| Tip<br>Configuration | Straight and multipurpose | Straight and multipurpose | SAME |
| Accessories | | | |
| Dilator | HDPE | HDPE | SAME |
| Rotating<br>Hemostasis<br>Valve (RHV) | Not included | Polycarbonate, Silicone,<br>Silicone Lubricant | The RHV materials are<br>biocompatible. |
| Dimensions | | | |
| Shaft | | | |
| Proximal OD | 0.110 in | 0.105 in | Both devices are evaluated to<br>achieve proper placement<br>during advancement. |
| Distal OD | 0.105 in | 0.105 in | achieve proper placement<br>during advancement. |
| Proximal ID | 0.088 in | 0.088 in | SAME |
| Distal ID | 0.088 in | 0.088 in | SAME |
| Effective<br>Length | 80-100 cm | 80-100 cm | SAME |
| Coating Length | 12 cm | 14 cm | Both devices are evaluated to<br>achieve proper placement<br>during advancement. |
| Accessories | | | |
| Dilator ID | 0.039 in Min | 0.039 in Min | SAME |
| Dilator OD | 0.085 in Max | 0.087 in Max | Both devices are evaluated for<br>the insertion and removal of<br>the dilator from the sheath. |
| Device Name | Predicate Device:<br>Neuron MAX System | Subject Device:<br>6F Wallaby Long Sheath | Rationale for Difference |
| | | | |
| Packaging Material | | | |
| Pouch | PET/PE/Tyvek Pouch | PA/Tyvek pouch | Packaging materials are similar<br>and common for medical<br>devices. Both packaging<br>configurations maintain<br>sterility of the device<br>throughout the shelf life. |
| Tubing | HDPE | HDPE | SAME |
| Packaging Card | Polyethylene | Polyethylene | SAME |
| Display Carton | SBS Paperboard | SBS Paperboard | SAME |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | SAME |
| How Supplied | Sterile, Single Use | Sterile, Single Use | SAME |
| Shelf Life | 36 Months | 12 months | A 12-month shelf life was<br>validated for the subject<br>device. |
### Table 1: 6F Wallaby Long Sheath Technological Comparison to Neuron MAX System
{4}------------------------------------------------
{5}------------------------------------------------
#### d. Non-Clinical Performance Tests
To establish the substantial equivalence of the 6F Wallaby Long Sheath to the predicate Neuron MAX System and to meet the requirements of the risk analysis , non-clinical bench and biological compatibility testing was conducted and driven by the risk analysis. The testing and results are summarized below:
# Design Verification Testing - Bench
Performance testing was conducted. The results of the design verification and validation testing confirm that the 6F Wallaby Long Sheath conforms to the pre-defined acceptance criteria. Testing included:
| Test | Methods and Results |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Visual Inspection | The 6F Wallaby Long Sheath was evaluated to verify the visual inspection requirements were met. The device met all pre-defined acceptance criteria. |
| Dimensional Inspection | The 6F Wallaby Long Sheath was evaluated to verify the dimensional requirements for ID, OD, overall length, working length, and coating length were met. The subject device met all pre-defined acceptance criteria. |
| Simulated Use | The 6F Wallaby Long Sheath was evaluated under a simulated use in a representative tortuous anatomical model. Subject device compatibility with 6F and 8F catheters, the RHV, guidewire, and stent retriever were evaluated. The device performed as intended. |
| Physician Validation | The 6F Wallaby Long Sheath was evaluated under a simulated use in a representative tortuous anatomical model by physicians in comparison to the predicate device, including preparation and ease of assembly, 8F short sheath interaction, RHV Luer connection interaction, dilator interaction, compatibility with guidewire, guide and aspiration catheters, and kink resistance. The subject device performed as intended. |
| Delivery and Retrieval | The 6F Wallaby Long Sheath was evaluated under a simulated use in a representative tortuous anatomical model for delivery and retrieval forces in comparison to the predicate device. The forces required to deliver and retrieve the subject device with ancillary devices were compared to the predicate device, which demonstrated similar forces. |
{6}------------------------------------------------
| Test | Methods and Results |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tip Stiffness | The tip stiffness of the 6F Wallaby Long Sheath was evaluated and compared to the predicate device. The results demonstrated tip stiffness is similar to the predicate. |
| Tensile Strength and Elongation to Failure | The 6F Wallaby Long Sheath was evaluated to verify the tensile strength of the distal shaft and the proximal hub and met the minimum tensile requirement. The 6F Wallaby Long Sheath was evaluated to verify the shaft elongation to failure. The device met all predefined acceptance criteria. |
| Torque Strength | The 6F Wallaby Long Sheath was tested for torque to failure in a simulated use test model. Test results demonstrated the torque strength of the subject device is similar to the predicate device. |
| Coating Integrity | The 6F Wallaby Long Sheath coating integrity was inspected after the particulate testing and no visible defects or sign of irregularity were observed. |
| Coating Lubricity | The 6F Wallaby Long Sheath was evaluated for frictional forces in a simulated use test model and demonstrated similar results between the subject device and the predicate device. |
| Catheter Dynamic and Static<br>Burst (Pressure)<br>Leak (Liquid)<br>Leak (Air) | The 6F Wallaby Long Sheath was evaluated for leakage and burst and is compatible with accessories per ISO 10555-1. |
| Kink Resistance | The 6F Wallaby Long Sheath was evaluated for resistance to kinking around bends with clinically relevant radii at specific locations along the shaft and met the acceptance criteria. |
| Particulate | The 6F Wallaby Long Sheath was evaluated within a simulated use anatomical model to verify the amount of particulate generated. The number of particulates was comparable to the predicate. The device met the acceptance criteria. |
| Corrosion Resistance | The 6F Wallaby Long Sheath was evaluated for corrosion resistance per ISO 10555-1 and met the acceptance criteria. |
| Insertion and Retrieval Forces | The 6F Wallaby Long Sheath was evaluated for insertion and retrieval forces with a dilator. The results demonstrated similar forces for the subject device and the predicate device. |
| Radiopacity | The 6F Wallaby Long Sheath was evaluated for marker band visibility under fluoroscopy during the physician validation testing and demonstrated results similar to the predicate device |
# Design Verification Testing – Animal
No animal testing was deemed necessary to support the substantial equivalence of the 6F Wallaby Long Sheath.
### Sterilization and Shelf Life
The 6F Wallaby Long Sheath is sterilized using an Ethylene Oxide (EO) sterilization cycle. The sterilization cycle was verified to ensure a sterility assurance level (SAL) of 10 * in accordance with ISO 11135:2014, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.
Aging studies for the 6F Wallaby Long Sheath have established the device and its packaging remains functional for the 12-month shelf life. Aging studies for packaging integrity, seal strength, and device functionality were performed and met the acceptance criteria.
### Biocompatibility
Biocompatibility testing for the 6F Wallaby Long Sheath and accessories was performed in accordance with ISO 10993- 1:2018, Biological evaluation of medical devices – Evaluation and testing within a risk management process.
{7}------------------------------------------------
#### Table 3: Biocompatibility Testing
| Test | Results | Conclusion |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Sheath (100 cm model) | | |
| MTT – L-929 Cytotoxicity Study | 1X MEM test extract showed no cytotoxic potential to<br>L-929 mouse fibroblast cells undiluted or at any<br>dilution; viability ≥ 70%; 80-91%. | Non-cytotoxic |
| ISO Intracutaneous Irritation | The difference between the average scores of the test<br>article extract and the vehicle control are: 0; 0. | Non-irritant |
| ISO Guinea Pig Maximization Sensitization | Test and control animals' responses are not greater<br>than "0". | Non-sensitizing |
| ISO Acute Systemic Toxicity | No abnormal clinical signs indicative of toxicity were<br>observed for 72 hours. All animals were alive at the<br>end of 72 hours and body weight changes were within<br>acceptable parameters. | Non-toxic |
| ISO Material Mediated Rabbit Pyrogen | No rabbit temperature rise ≥ 0.5°C. | Non-pyrogenic |
| Complement Activation - SC5b-9 Assays<br>with Sponsor-Supplied Comparison | Results within acceptable range as compared to the<br>comparator device. | The test article<br>complement activation<br>was similar to the<br>comparator device |
| ASTM Hemolysis - Direct Contact and<br>Extract Method | Blank corrected Hemolytic index: 0.4, 0.1. | Non-hemolytic |
| Platelet and Leukocyte counts | No ranges or levels outside an acceptable range and<br>comparable to Control Device. | Counts from subject<br>device are within<br>acceptable ranges and are<br>similar to the control |
| Partial Thromboplastin Time (PTT) | Test and predicate devices have similar performance. | The test article is not<br>considered an activator of<br>the intrinsic<br>coagulation pathway |
| Thromboresistance Evaluation | No adverse effects or clinical signs during test period<br>and no thrombus score > 3 for either test or control<br>device. | Thrombogenic risk<br>potential similar to<br>the control device |
| Dilator | | |
| MTT – L-929 Cytotoxicity Study | 1XMEM test extract showed no cytotoxic potential to<br>L-929 mouse fibroblast cells undiluted or at any<br>dilution; viability ≥70%; 79.8%. | Non-cytotoxic |
| ISO Intracutaneous Irritation | The difference between the average scores of the test<br>article extract and the vehicle control are: 0; 0. | Non-irritant |
| ISO Guinea Pig Maximization Sensitization | Test and control animals' responses are not greater<br>than "0". | Non-sensitizing |
| ISO Acute Systemic Toxicity | No abnormal clinical signs indicative of toxicity were<br>observed for 72 hours. All animals were alive at the<br>end of 72 hours and body weight changes were<br>within acceptable parameters. | Non-toxic |
| ISO Material Mediated Rabbit Pyrogen | No rabbit temperature rise ≥ 0.5°C. | Non-pyrogenic |
| Complement Activation - SC5b-9 Assays<br>with Sponsor-Supplied Comparison | Results within acceptable range as compared to the<br>negative reference material. | Non-activator of<br>complement system |
| ASTM Hemolysis - Direct Contact and<br>Extract Method | Blank corrected Hemolytic index: 0.3, 0.5. | Non-hemolytic |
| Thromboresistance Evaluation | No adverse effects or clinical signs during test period<br>and no thrombus score > 3 for either test or control device. | Thrombogenic risk potential similar to the control device |
| RHV | | |
| Cytotoxicity MEM Elution | Percent Cell Lysis: 0%<br>Cytotoxic Score: 0 | Non-cytotoxic |
| ISO Intracutaneous Irritation | The difference between the average scores of the test article extract and the vehicle control are: 0; 0. | Non-irritant |
| ISO Guinea Pig Maximization Sensitization | Test and control animals' response not greater than "0". | Non-sensitizing |
| ISO Acute Systemic Toxicity | No abnormal clinical signs indicative of toxicity were observed for 72 hours. All animals were alive at the end of 72 hours and body weight changes were within acceptable parameters. | Non-toxic |
| ISO Material Mediated Rabbit Pyrogen | No rabbit temperature rise ≥ 0.5°C. | Non-pyrogenic |
| ASTM Hemolysis - Direct Contact and<br>Extract Method | Blank corrected Hemolytic index: 0.0, 0.1. | Non-hemolytic |
{8}------------------------------------------------
# Clinical
No clinical testing was deemed necessary to support the substantial equivalence of the 6F Wallaby Long Sheath.
### Conclusion
The 6F Wallaby Long Sheath is substantially equivalent to the predicate Neuron MAX System based on the non-clinical testing results, as well as similar principles of operation, materials of construction, packaging, usability, and the indications for use. Any differences between the subject device and the predicate device do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.