← Product Code [FAJ](/productcode/FAJ) · K222602

# Pusen Single Use Flexible Video Cystoscope System (Single Use Flexible Video Cystoscope: PC200-AS, PC200-AR, PC200-S and PC200-R; HD Medical Video Endoscope Image Processor: PV300) (K222602)

_Zhuhai Pusen Medical Technology Co., Ltd. · FAJ · Sep 26, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222602

## Device Facts

- **Applicant:** Zhuhai Pusen Medical Technology Co., Ltd.
- **Product Code:** [FAJ](/productcode/FAJ.md)
- **Decision Date:** Sep 26, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Pusen Single Use Flexible Video Cystem is intended to be used for endoscopic access to and examination of the lower urinary tract. The Pusen Single Video Cystoscope is intended to provide visualization via video processor and can be used with endoscopic accessories. This system is intended for use in a hospital environment or medical office environment.

## Device Story

System comprises single-use flexible video cystoscope and reusable HD medical video endoscope image processor (PV300). Cystoscope features distal tip with two LEDs and CMOS imaging sensor; connects to processor for real-time video display on LCD screen. Used by physicians for lower urinary tract (urethra/bladder) visualization and access; supports use of endoscopic accessories. Processor handles image signal processing. System provides visual examination capabilities to assist clinical decision-making in urology. Benefits include single-use sterility, reducing cross-contamination risk.

## Clinical Evidence

Bench testing only. No clinical or animal studies performed. Testing included biocompatibility (ISO 10993-1), electrical safety/EMC, mechanical/optical performance (ISO 8600), photobiological safety (IEC 62471), and luer taper testing (ISO 80369-7). Comparative image quality testing performed against predicate.

## Technological Characteristics

Flexible, single-use cystoscope; CMOS imaging sensor; LED illumination. Dimensions: 18 Fr insertion width, 380 mm working length, 2.3 mm working channel. Sterilization: Ethylene Oxide (EO) to 10⁻⁶ SAL. Powered by PV300 processor (line power/battery). Complies with ISO 8600, IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Ambu aScope 4 Cysto ([K193095](/device/K193095.md))

## Reference Devices

- EVIS EXERA II 180 SYSTEM ([K133538](/device/K133538.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

September 26, 2022

Zhuhai Pusen Medical Technology Co., Ltd. Changshen Wang Regulatory Affairs Director 5/F, Building 1, No 33, Ke Ji San Road High-tech Zone, Tangjiawan Town Zhuhai, Guangdong 519085 China

Re: K22602

> Trade/Device Name: Pusen Single Use Flexible Video Cystoscope System (Single Use Flexible Video Cystoscope: PC200-AS, PC200-AR, PC200-S and PC200-R; HD Medical Video Endoscope Image Processor: PV300) Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ Dated: August 25, 2022 Received: August 29, 2022

Dear Changshen Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K222602

#### Device Name

Pusen Single Use Flexible Video Cystem (Single Use Flexible Video Cystoscope: PC200-AR, PC200-AR, PC200-S and PC200-R; HD Medical Video Endoscope Image Processor: PV300)

#### Indications for Use (Describe)

The Pusen Single Use Flexible Video Cystem is intended to be used for endoscopic access to and examination of the lower urinary tract. The Pusen Single Video Cystoscope is intended to provide visualization via video processor and can be used with endoscopic accessories. This system is intended for use in a hospital environment or medical office environment.

| Type of Use (Select one or both, as applicable)                                                                                                                                           |                                                                                                                                                                                                |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;"> <span style="font-size: 20px;">☑</span> </div> <span>Prescription Use (Part 201. Subpart D)</span> </div> | <div style="display:flex; align-items:center;"> <div style="margin-right:5px;"> <span style="font-size: 20px;">☐</span> </div> <span>Over-The-Counter Use (21 CFR 201 Subpart C)</span> </div> |

> Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo consists of the word "PUSEN" in a stylized, sans-serif font, with each letter rendered in a light blue color. Below the logo, the company name "Zhuhai Pusen Medical Technology Co., Ltd." is written in a smaller, gray font.

# 510(k) Summary

# I. SUBMITTER

Submitter's name: Zhuhai Pusen Medical Technology Co, Ltd. Submitter's address: 5/F, Building 1, No 33, Keji San Road, High-tech Zone, Tangjiawan Town, 519085 Zhuhai, Guangdong, People's Republic of China.

Phone: +86 756 688 0865

Contact Person: Ellen Wang Date Prepared: September 22, 2022

# II. DEVICE

Name of Device: Pusen Single Use Flexible Video Cystoscope System (Single Use Flexible Video Cystoscope: PC200-AS, PC200-AR, PC200-S and PC200-R; HD Medical Video Endoscope Image Processor: PV300) Common Name: Cystoscope and accessories 510(k) number: K222602 Classification Name: Endoscope and Accessories (21 CFR 876.1500) Regulatory Class: Class II Product Code: FAJ

# III. PREDICATE DEVICE

# Predicate device

Name of Device: Ambu aScope 4 Cysto 510(k) number: K193095 Classification regulation: Cystoscope and Accessories, Flexible/Rigid Product code: FAJ This predicate device has not been subject to a design-related recall.

## Reference device

Name of Device: EVIS EXERA II 180 SYSTEM 510(k) number: K133538 Classification regulation: Endoscope and Accessories Product code: NWB; FAJ This reference device has not been subject to a design-related recall.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo consists of the word "PUSEN" in a stylized, sans-serif font, with rounded corners and a light blue color. Below the logo is the company name, "Zhuhai Pusen Medical Technology Co., Ltd.," in a smaller, gray font.

## IV. DEVICE DESCRIPTION

The Pusen Single Use Flexible Video Cystoscope System includes a Single Use Flexible Video Cystoscope and an HD Medical Video Endoscope Image Processor.

The system is intended to be used by physicians for endoscopic access to and examination of the lower urinary tract in adults. This system is intended to provide visualization and can be used with endoscopic accessories. This system is intended for use in a hospital environment or medical office environment.

- The Pusen Single Use Flexible Video Cystoscope is provided sterile and has the following 4 models:

| Model    | Description of difference                                                 |
|----------|---------------------------------------------------------------------------|
| PC200-AS | 'A' means this model has function buttons; 'S' means standard deflection. |
| PC200-AR | 'A' means this model has function buttons; 'R' means reverse deflection.  |
| PC200-S  | 'S' means standard deflection.                                            |
| PC200-R  | 'R' means reverse deflection.                                             |

- The Pusen Single Use Flexible Video Cystoscope has two LEDs and CMOS imaging sensor at its distal tip. The Pusen Single Use Flexible Video Cystoscope needs to be connected with the HD Medical Video Endoscope Image Processor PV300 as a system. The Pusen Single Use Flexible Video Cystoscope is powered by the HD Medical Video Endoscope Image Processor PV300.
- The HD Medical Video Endoscope Image Processor PV300 is a reusable device, used to process the image signal from the Pusen Single Use Flexible Video Cystoscope and display the real-time video on its LCD screen, which enables visual examination of the lower urinary tract (the urethra and the bladder).

#### V. INDICATIONS FOR USE

The subject device: The Pusen Single Use Flexible Video Cystoscope System is intended to be used for endoscopic access to and examination of the lower urinary tract. The Pusen Single Use Flexible Video Cystoscope is intended to provide visualization via video processor and can be used with endoscopic accessories. This system is intended for use in a hospital environment or medical office environment.

The predicate device: Ambu® aScope™ 4 Cysto is a sterile, single-use, flexible cystoscope intended to be used for endoscopic access to and examination of the lower urinary tract. The Ambu® aScope™ 4 Cysto is intended to provide visualization via Ambu® displaying unit and can be used with endoscopic accessories. Ambu® aScope™ 4 Cysto is intended for use in a hospital environment or medical office environment. Ambu® aScope™ 4 Cysto is designed for use in adults.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo consists of the word "PUSEN" in a stylized, sans-serif font, with each letter rendered in a light blue color. Below the logo, the company name "Zhuhai Pusen Medical Technology Co., Ltd." is written in a smaller, gray font.

Traditional 510(k)

The indications for use are the same for the subject device and the predicate device. The subject and predicate device have the same intended use – to provide visualization in the lower urinary tract.

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

| Item                                  | Subject device                                       | Predicate device                                |
|---------------------------------------|------------------------------------------------------|-------------------------------------------------|
| Trade name                            | Pusen Single Use Flexible Video<br>Cystoscope System | Ambu aScope 4 Cysto                             |
| 510(K) number                         | K222602                                              | K193095                                         |
| Scope type                            | Flexible                                             | Flexible                                        |
| Scope<br>reusability                  | Single-use                                           | Single-use                                      |
| Energy used                           | Powered by chargeable battery or<br>line power.      | Powered by chargeable<br>battery or line power. |
| Digital<br>video<br>technology        | CMOS                                                 | CMOS                                            |
| Illumination<br>source                | LED                                                  | LED                                             |
| Field of view                         | 90 °                                                 | 120 °                                           |
| Direction<br>of<br>view               | 30°                                                  | 0°                                              |
| Depth of field                        | 3~80 mm                                              | 5~50mm                                          |
| Maximum<br>insertion<br>portion width | 18 Fr                                                | 18 Fr                                           |
| Working length                        | 380 mm                                               | 390 mm                                          |
| Working<br>channel size               | 2.3 mm                                               | 2.2 mm                                          |
| Up/down<br>deflection                 | Up: 210°<br>Down: 210°                               | Up: 210°<br>Down: 120°                          |
| Suction                               | Provided                                             | Not Provided                                    |
| Sterility                             | Ethylene Oxide (EO)<br>SAI : 10⁻⁶                    | EO<br>SAI : 10⁻⁶                                |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo is in blue and features the word "PUSEN" in a stylized font. The text "Zhuhai Pusen Medical Technology Co., L" is located below the logo.

Traditional 510(k)

The subject and the predicate device have different dimensions (e.g., working length, channel size) and optical specifications (e.g., depth of field of view, direction of view). However, these differences do not raise different questions of safety and effectiveness.

# VII. PERFORMANCE DATA

The subject device has been verified for the safety and effectiveness based on the following performance data.

## Biocompatibility testing

The biocompatibility evaluation for the Pusen Single Video Cystoscope was conducted in accordance with the FDA final guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"" (issued on September 4, 2020) and international standard ISO 10993-1:2018 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. To support the contact classification as a surface device contacting mucosal membrane for a limited duration, the following endpoints were evaluated:

- Cytotoxicity
- Intracutaneous Irritation
- Sensitization

The subject device passed the relevant tests and complies with ISO 10993-1.

## Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the subject device system, and the system complies with the IEC 60601-1 and IEC 60601-2-18 standards for safety and the IEC 60601-1-2 standard for EMC.

## Software Verification and Validation Testing

Software verification and validation testing were conducted, and relevant software documentations were provided as recommended by FDA's Guidance for Industry and FDA staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

{7}------------------------------------------------

## Bench Performance Testing

#### Mechanical and Optical Performance

The subject device was designed to comply with applicable parts of ISO 8600. Optical measurements and physical performances tests were performed according to applicable part of ISO 8600 standard, include appearance, working length, working channel diameter, maximum insertion width, angle of deflection, field of view and direction of view.

Mechanical characteristics were tested and include flow rate of water, suction rate and tensile strength testing.

#### Image quality

Comparative testing related to image quality performances including color performance (color reproduction and color contrast enhancement), optical performance (resolution, depth of field, image intensity uniformity and distortion) tests were performed for the subject device and the predicate device to support substantial equivalence.

#### Photobiological safety

The LEDs in the subject device were tested according to the following FDA recognized standards:

· IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems.

#### Luer taper

Luer taper in the subject device was tested according to ISO 80369-7: 2021, Smallbore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications.

## Sterilization and shelf life

- A shelf life of 3 years for the Pusen Single Use Flexible Video Cystoscope was supported with accelerated aging of final finished devices, followed by simulated shipping distribution, package integrity testing, and product performance testing.
● The EO sterilization method has been validated to ISO 11135:2014. (only applicable to the Pusen Single Use Flexible Video Cystoscope)

● Reprocessing method was validated in accordance with the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling". (only applicable to the HD Medical Video Endoscope Image Processor)

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image displays the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo features the word "PUSEN" in a stylized, sans-serif font, with each letter rendered in a light blue color. Below the logo, the company's full name, "Zhuhai Pusen Medical Technology Co., Ltd," is written in a smaller, gray font.

Traditional 510(k)

#### Animal Study and Clinical study

No animal study or clinical study is included in this submission.

#### Summary of performance data

All tests were passed and all evaluation acceptance criteria were met.

## VIII. CONCLUSIONS

The performance data described above demonstrate that the Pusen Single Use Flexible Video Cystoscope System is as safe and effective as the predicate device and supports a substantial equivalence determination.

---

**Source:** [https://fda-staging.innolitics.com/device/K222602](https://fda-staging.innolitics.com/device/K222602)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).