# PatCom Distal Chip Endoscope (K222587)

_H&A Mui Enterprises, Inc. · EOB · Jul 28, 2023 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222587

## Device Facts

- **Applicant:** H&A Mui Enterprises, Inc.
- **Product Code:** [EOB](/productcode/EOB.md)
- **Decision Date:** Jul 28, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.4760
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

The PatCom Distal Chip Endoscope is a video endoscope used for visualization to aid in diagnosis in the regions of mouth, nasal cavity, and upper airway. It is to be used only by healthcare professionals who have received adequate training in handling nasopharyngoscopes.

## Device Story

Portable video endoscope for visualization of mouth, nasal cavity, and upper airway; consists of grip section and insertion tube. Input: light and color information captured by distal image sensor. Processing: internal chip converts optical data to binary patterns; no external CCU required. Output: 720x1280p HD video projected onto computer monitor via USB connection. Used in healthcare facilities by trained professionals. Powered by 5V USB from laptop. Features manual angulation knob (up to 130 degrees). Plug-and-play functionality using standard UVC driver. Benefits: provides real-time visualization for diagnostic aid; improved maneuverability and patient comfort via smaller insertion tube diameter.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing only, including comparative imaging performance (FOV, DOF, resolution, noise, color, latency) and biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity).

## Technological Characteristics

Materials: Stainless steel (bending section), TPU (insertion tube cover), ABS (handle). Sensing: Distal CMOS chip. Energy: 5V DC via USB. Dimensions: 3.2mm insertion tube diameter, 340mm working length. Connectivity: USB 2.0/UVC. Sterilization: High-level disinfection (reusable). Software: Embedded firmware, UVC driver. Standards: ISO 8600-1/3/5, ISO 12233, ISO 15739, ISO 10993-1/5/10, IEC 60601-1, IEC 62304.

## Regulatory Identification

A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- Schoelly CMOS Video Nasopharyngoscope ([K132009](/device/K132009.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 28, 2023

H&A Mui Enterprises Inc. Tammy Mui Operations Manager 145 Traders Blvd. E., Unit #34 Mississauga. Ontario L4Z 3L3 Canada

## Re: K222587

Trade/Device Name: PatCom Distal Chip Endoscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOB Dated: June 27, 2023 Received: June 28, 2023

### Dear Tammy Mui:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Joyce C. Lin -S

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222587

Device Name PatCom Distal Chip Endoscope

Indications for Use (Describe)

The Patcom Distal Chip Endoscope is indicated when endoscopic visualization in the regions of mouth, nasal cavity, and upper airway is required.

**Type of Use (Select one or both, as applicable)**

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Mui Scientific. The logo consists of a circle with a dot in the center and three smaller circles around the perimeter. To the right of the logo is the text "Mui Scientific" in a bold, serif font.

Division of
H&A Mui
Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 www.muiscientific.com Email: mail@muiscientific.com Website:

July 28, 2023

# 510(k) Summary

Summary prepared by:

Contact Person: Tammy Mui Title: Operations Manager Manufacturer: H&A Mui Enterprises Inc., o/a Mui Scientific Address: 145 Traders Blvd. East, Unit #33-34, Mississauga, Ontario, Canada L4Z 3L3 Phone: (905) 890-5525 Fax: (905) 890-3523 Email: tammy.mui@muiscientific.com

Trade name: PatCom Distal Chip Endoscope Common name: Nasopharyngoscope Classification name: Nasopharyngoscope (flexible or rigid) (as per CFR 874.4760) Class: 2 Review Panel: Ear Nose & Throat Product Code: EOB

| Predicate Device                         | 510(k) Number | Manufacturer             |
|------------------------------------------|---------------|--------------------------|
| Schoelly CMOS Video<br>Nasopharyngoscope | K132009       | Schoelly Fiberoptic GmbH |

An agreement exists between the following companies: Note:

- Zhongshan Wesee Meditech Co., Ltd (the actual manufacturer of the endoscopes) 1.
- 2. H&A Mui Enterprises Inc, o/a Mui Scientific (the virtual manufacturer of the endoscopes for Canada, the United States, and the EU)
- PatCom Medical Inc (the exclusive distributor for Canada and the United States only) 3.

### Indications for Use:

The Patcom Distal Chip Endoscope is indicated when endoscopic visualization in the regions of mouth, nasal cavity, and upper airway is required.

### Intended Use:

The PatCom Distal Chip Endoscope is a video endoscope used for visualization to aid in diagnosis in the regions of mouth, nasal cavity, and upper airway.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a circular diagram with a central dot and five smaller circles evenly spaced around the perimeter of a larger circle. The central dot is solid black, and the smaller circles are outlined in black. The larger circle is also outlined in black, creating a symmetrical and geometric design.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 www.muiscientific.com Email: mail@muiscientific.com Website:

It is to be used only by healthcare professionals who have received adequate training in handling nasopharyngoscopes.

As the indicated location of use is not considered a sterile environment: sterility of these medical devices is not required

This device is for transient use (under 24 hrs), and will come in direct contact with the patient's mucosal lining.

### Device Description:

The PatCom Distal Chip Endoscope is a portable video endoscope for visualization in the reqions of the mouth, nasal cavity, and upper airway. It is made up of the grip section and insertion tube. It requires connection via USB to a laptop computer, and gains power from the USB ports which supplies 5V of electricity with a maximum current of 0.5A. This is sufficient to power the endoscope. The endoscope does not need an external CCU for image processing, as the processing is done directly on the device on a small chip. It converts the light and color information received by the image sensor into binary numbers (patterns of zeros and ones). The USB Video Class (UVC) driver is a Microsoft-provided AVStream minidriver that provides driver support for USB Video Class devices. When a device uses UVC, it does not need to support their own driver, but instead works automatically with the system-supplied driver. It is compatible to all Windows and Mac computers, with a simple plug and play feature. The visualization from the distal chip is projected onto the computer monitor, at a resolution of 720px1280p high definition.

The insertion tube portion of the device is used for visualization within the natural orifices of the mouth and the nasal cavities and in the upper airway anatomy. The light emitted by the LED light source at the distal end of the tip is illuminated into the body cavity of the subject, and the image is displayed onto a screen.

The grip section of the device contains a control knob for moving the distal end of the insertion tube bend up or down, up to 130 degrees in either direction.

The proximal end of the grip section contains the USB connector, for plug and play with a computer, for power and to provide visualization of the camera chip found at the distal end of the insertion tube.

The PatCom Distal Chip Endoscope is substantially equivalent to the Schoelly CMOS Video Nasopharyngoscope, manufactured by Schoelly Fiberoptic GmbH. The Schoelly Nasopharyngoscope has already been approved for the US market since 2014. Similar to the PatCom Distal Chip Endoscope, their nasopharyngoscope also has a grip section, insertion tube section, and a manual knob for angulation of the tip, and are both used for visualization and diagnosis within the nose, mouth and throat.

The Schoelly Nasopharyngoscope has been selected as an equivalent device based on a detailed comparison of intended use, technological, imaging and biological characteristics. The intended use of PatCom Distal Chip Endoscope was considered the same as the predicate device, but small differences were identified for structure and technological characteristics. Rationales have been provided to ensure that these differences do not raise any additional safety or performance risks of the PatCom Distal Chip Endoscope compared with equivalent

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Mui Scientific. The logo consists of a circular graphic to the left of the text "Mui Scientific". The graphic is a circle with a dot in the center and five smaller circles around the perimeter.

Division of
H&A Mui
Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 Website: www.muiscientific.com Email: mail@muiscientific.com

device. Data generated from performance studies undertaken has also been used for comparison and to substantiate that it is safe to extrapolate data from equivalent devices. The main difference between the submitted device and the predicate device is the power supply. The distal end of the grip section contains the USB connector, for plug and play with a laptop computer, for power and to provide visualization of the distal chip found at the end of the insertion tube. I It requires connection via USB to a laptop computer, and gains power from the USB ports which supplies 5V of electricity with a maximum current of 0.5A. This is sufficient to power the endoscope. The device does not need an external CCU for image processing, as the processing is done directly on the device on a small chip. It converts the light and color information received by the image sensor into binary numbers (patterns of zeros and ones). The USB Video Class (UVC) driver is a Microsoft-provided AVStream minidriver that provides driver support for USB Video Class devices. When a device uses UVC, it does not need to support their own driver, but instead works automatically with the system-supplied driver. It is compatible to all Windows and Mac computers, with a simple plug and play feature. The visualization from the distal chip is projected onto the computer monitor, at a resolution of 720px1280p high definition. The predicate device, however, requires connection to a control unit for power, which in turn, is powered by connection to a 110V outlet.

|                                                                                          | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                     |
|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Trade name                                                                               | Schoelly CMOS Video<br>Nasopharyngoscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | PatCom Distal Chip Endoscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                     |
| 510K holder                                                                              | Schoelly Fiberoptic GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Mui Scientific                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                     |
| 510K<br>number                                                                           | K132009                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | K222587                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                     |
| Manufacturer                                                                             | Schoelly Fiberoptic GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Mui Scientific                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                     |
| Indications<br>For Use                                                                   | The Schoelly CMOS Video<br>Nasopharyngoscope System may only<br>be used by persons with an appropriate<br>medical qualification and who are<br>acquainted with the rhino/laryngoscopic<br>technique. The endoscope is used for<br>endoscopic diagnosis within the nasal<br>lumens and airway anatomy, and is<br>intended to provide visualization via a<br>video monitor.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | The Patcom Distal Chip Endoscope is<br>indicated when endoscopic<br>visualization in the regions of mouth,<br>nasal cavity, and upper airway is<br>required.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                     |
| Environment<br>of Use                                                                    | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                     |
| Discussion                                                                               | In comparing the Indication for Use of the predicate and the proposed device, we<br>found they are general similar, but utilizing different terminology for the same<br>physical anatomy structures (lumens/cavity, airway anatomy/upper airway).<br>According to the description above, the clinical application of the Distal Chip<br>Endoscope and the equivalent device are essentially identical and does not pose<br>a significant issue of safety or efficacy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                     |
| Technical Specifications                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                     |
| Power Supply                                                                             | Endoscope is powered by a 5V d.c., obtained from the USB connection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                     |
| control unit. The control unit<br>is powered by standard<br>110V outlet via an IEC cable | with a laptop computer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                     |
| Field of View<br>angle                                                                   | 85°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 110°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                     |
| Direction of<br>view                                                                     | 0°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 0°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                     |
| Depth of Field                                                                           | 6 - 60mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 6-60mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                     |
| Insertion Tube<br>outer diameter<br>(mm)                                                 | 3.8mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 3.2mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                     |
| Instrument<br>Channel                                                                    | No channel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | No channel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                     |
| Configuration                                                                            | Light-Emitting (LED),<br>Endoscope UP/DOWN<br>angulation control knob,<br>Bending section                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Light-Emitting (LED), Endoscope UP/DOWN<br>angulation control knob, Bending section                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                     |
| Working length                                                                           | 300mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 340mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                     |
| Angulation<br>range                                                                      | UP130° Down130°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | UP130° Down130°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                     |
| Source of<br>examination<br>light                                                        | LED                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | LED                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                     |
| Software                                                                                 | Unknown level of concern<br>(possibly low); image<br>processing done on external<br>control unit; proprietary<br>driver required                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Low level of concern; image processing done<br>on internal chip in endoscope; standard USB<br>Video Class (UVC) driver                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                     |
| Discussion                                                                               | The biggest difference between the proposed and predicate device exists in<br>the aspect of power supply. While the predicate device requires power from a<br>control box that is directly connected to a 110V outlet, the submission device<br>requires only a small amount of power via USB from a laptop computer. The<br>electrical safety of the PatCom Distal Chip Endoscope is further supported<br>by the successful completion of the required electrical testing.   For software,<br>both contain image processing firmware and require a driver to connect<br>imaging to a monitor. The difference lies in where the image processing<br>chip is located, and what type of driver is required for the display monitor, of<br>which the PatCom Distal Chip Endoscope is simpler (with the chip directly<br>embedded within the endoscope) and easier to use (with the ability to plug<br>and play and use any standard UVC driver.   The other minor differences<br>noted reflect enhanced features, such as having a wider field of view angle<br>for better visualization, a smaller insertion tube outside diameter for safer<br>intubation and patient comfort, and longer working length to allow for<br>improved maneuverability.    Therefore, these minor differences are not<br>considered to raise different questions of safety and effectiveness compared<br>to the predicate device. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                     |
|                                                                                          | Performance<br>Characteristics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | The two devices have been submitted to an external accredited laboratory for<br>extensive comparative imaging testing, such as Field of View (FOV, Depth of<br>Field (DOF), Optimum Working Distance, Noise and Dynamic Range,<br>Geometric Distortion, Image Intensity Uniformity (IIU), Color Performance and<br>Latency Assessment for testing the Quality of the real-time video feed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                     |
|                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | The tests conducted were performed in compliance with the following<br>regulations:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                     |
|                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | ISO 8600-1 (2005-05-1): Optics and photonics-Medical endoscopes<br>and endotherapy devices-Part 1: General Requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                     |
|                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | ISO 8600-3 (2019-8): Determination of field of view and direction of<br>view of endoscopes with optics…

---

**Source:** [https://fda-staging.innolitics.com/device/K222587](https://fda-staging.innolitics.com/device/K222587)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).