← Product Code [GCJ](/productcode/GCJ) · K222575

# HurriChem Device Kit (K222575)

_Thermasolutions, LLC · GCJ · Dec 6, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222575

## Device Facts

- **Applicant:** Thermasolutions, LLC
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Dec 6, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The ThermaSolutions HurriChem Device Kit is indicated for use in patients undergoing a laparoscopic procedure. It is designed to deliver sterile irrigation fluids to surgical sites during laparoscopic procedures and to lavage blood and tissue debris from the surgical site.

## Device Story

Device consists of stainless-steel irrigation wand and braided polyurethane tubing; connects to external standard injection pump. Wand features micro-opening nozzle; transforms incoming fluid into fine mist of microscopic droplets. Used in laparoscopic surgery; operated by trained physicians. Delivers irrigation to surgical site; enables lavage of blood/debris using minimal fluid volume. Does not provide suction; requires separate suction/evacuation devices if fluid removal is needed. Single-use, sterile, disposable device.

## Clinical Evidence

No clinical data. Safety and effectiveness supported by bench testing, including biocompatibility, sterilization validation, packaging validation, shelf life, transportation testing, and human factors studies.

## Technological Characteristics

Stainless-steel wand; braided polyurethane tubing. Mechanical fluid delivery via micro-opening nozzle to create mist. Powered by external injection pump. Single-use; ethylene oxide sterilized. No electrosurgery or suction capability.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- DANNIK Laparoscopic Suction Irrigation System ([K192643](/device/K192643.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 6, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

ThermaSolutions LLC Rhea Johny Regulatory Specialist 1889 Buerkle Road St. Paul, Minnesota 55110

Re: K222575

Trade/Device Name: HurriChem Device Kit Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: November 9, 2022 Received: November 9, 2022

Dear Rhea Johny:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Colin K. C

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2023

See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

Device Name HurriChem Device Kit

Indications for Use (Describe)

The ThermaSolutions HurriChem Device Kit is indicated for use in patients undergoing a laparoscopic procedure. It is designed to deliver sterile irrigation fluids to surgical sites during laparoscopic procedures and to lavage blood and tissue debris from the surgical site.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary for the HurriChem Device Kit

#### Contact Information

- ThermaSolutions LLC 1889 Buerkle Road St. Paul, MN 55110 USA Phone: +1.877.651.6100 Fax: Rhea Johny, Regulatory Specialist 4 March 2022
#### Information about the Device

Trade name: HurriChem Device Kit

Common name: laparoscopic irrigation

Classification name: Endoscope and accessories (21 CFR 876.1500, Product Code GCJ)

## Substantially Equivalent Device

Legally marketed device: DANNIK Laparoscopic Suction Irrigation System (K192643)

Common name: laparoscopic irrigation

Classification name: Endoscope and accessories (21 CFR 876.1500, Product Code GCJ)

{4}------------------------------------------------

### Indication for Use

The ThermaSolutions HurriChem Device Kit is indicated for use in patients undergoing a laparoscopic procedure. It is designed to deliver sterile irrigation fluids to surgical sites during laparoscopic procedures and to lavage blood and tissue debris from the surgical site.

## Description of the Device

ThermaSolutions' HurriChem Device Kit consists of the HurriChem Irrigation Wand and braided poly tubing. The HurriChem Device Kit is intended for use to deliver sterile irrigation fluids to surgical sites during laparoscopic procedures. It is designed to lavage blood and tissue debris from the surgical site. The kit is to be used in conjunction with standard injection pumps. The kit is to be used only by physicians trained in the use of this device. The device is packaged and sterilized for single use only, and should be discarded after use according to the policy of the facility.

The HurriChem Device Kit consists of a stainless-steel wand and a length of braided poly tubing. The one end of the tubing attaches to the stainless-steel wand, and the other end of the tubing is intended to attach to a standard injection pump (not part of the device). The sterile fluid is delivered from the injection pump through the poly tubing and into the stainless-steel wand. The wand has internal mechanics that deliver the fluid as microscopic droplets in a fine mist. The irrigation using fluid droplets allows for maximum lavage of the area with a minimal amount of fluid.

## Technological Characteristics of HurriChem Compared to Predicate Device

|                       | HurriChem            | DANNIK |
|-----------------------|----------------------|--------|
| Intended use:         | Laparoscopic surgery | Same   |
| Insertion via trocar: | Yes                  | Yes    |
| Sterilization:        | Ethylene Oxide       | Same   |
| Prescription only:    | Yes                  | Yes    |

{5}------------------------------------------------

| Disposable:                                 | Yes                     | Yes                     |
|---------------------------------------------|-------------------------|-------------------------|
| Single use:                                 | Yes                     | Yes                     |
| Biocompatibility                            | Yes                     | Yes                     |
| Electrosurgery capable:                     | No                      | No                      |
| Suction capable:                            | No                      | Yes                     |
| Fluid delivered by external<br>pump source: | Yes, not part of device | Yes, not part of device |
|                                             |                         |                         |
|                                             | <b>Materials</b>        |                         |
| wand/probe:                                 | Stainless steel         | Same                    |
| tubing:                                     | Braided polyurethane    | PVC                     |
| hand piece w/valve:                         | None                    | Polycarbonate           |
|                                             |                         |                         |

#### Summary of Technological Difference in Irrigation Droplet Size

Both devices receive irrigation fluid via delivery from an external mechanical source. The wand (HurriChem) or probe (DANNIK Suction Irrigation) directs the irrigation fluid as determined by the physician. The HurriChem delivers the irrigation as small droplets of fluid. The DANNIK Suction Irrigation delivers the irrigation as a larger stream of fluid.

The HurriChem device's mechanical conversion of the fluid into droplets permits a small volume of fluid to be used for irrigation. The stainless-steel wand in the HurriChem device kits contains a nozzle with a micro-opening because the small opening converts and delivers the irrigation fluid in the form of droplets. The Suction Irrigation has a larger opening that delivers the irrigation fluid as a stream of fluid.

{6}------------------------------------------------

### Summary of Technological Difference in Suction

The HurriChem Device Kit is indicated for use to deliver sterile irrigation fluids to surqical sites during laparoscopic procedure. The device is not designed nor intended to provide suction or removal of fluid. Evacuation of fluid, if desired by the clinician, may be performed with laparoscopic suction or evacuation devices, drains or catheters.

## Non-clinical Testing

A series of tests were performed to assess the safety and effectiveness of the HurriChem Device Kit to the predicate device. Biocompatibility, sterilization validation, packaging validation, shelf life and transportation testing were completed. Additionally, ThermaSolutions Engineering and Product Development teams evaluated fit, function, performance and equivalence of the device through a combination of human factors studies and bench testing. All the tests demonstrate that the device meets the requirements of its pre-defined acceptance criteria and intended use. The results of the non-clinical testing demonstrate that the HurriChem Device Kit is as safe and effective as the predicate device.

## Clinical Testing

No clinical trials were performed on the HurriChem Device Kit.

## Conclusion

The HurriChem Device Kit has the same intended use and application as the predicate device. The materials of the HurriChem Device Kit and the predicate device are equivalent. The difference in technical characteristics of the HurriChem's irrigation spray from DANNIK's irrigation stream does not raise different questions of safety and effectiveness. The non-clinical testing and performance evaluations show the ThermaSolutions HurriChem Device Kits is as safe and effective and meets the same expected performance as the DANNIK Suction Irrigation System. Therefore, the proposed HurriChem Device Kits is substantially equivalent to the predicate.

---

**Source:** [https://fda-staging.innolitics.com/device/K222575](https://fda-staging.innolitics.com/device/K222575)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
