OptiCross™ 18 is intended for intravascular ultrasound examination of peripheral vessels only. Intrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.
Device Story
OptiCross 18 is a sterile, short-rail intravascular ultrasound (IVUS) imaging catheter. It consists of an imaging core with a 30 MHz ultrasonic transducer and a catheter body with a distal imaging window. The device connects to a Motordrive Unit (MDU5 PLUS™) via an electro-mechanical interface. During peripheral vascular procedures, the catheter is flushed with heparinized saline to provide acoustic coupling. The transducer rotates to capture ultrasonic images of vessel interiors. A telescoping section allows 15 cm of linear movement for the imaging core, enabling longitudinal imaging. Radiopaque markers and depth markers assist in positioning. The device is used by physicians in clinical settings to visualize vessel morphology, aiding in the assessment of lesions and guiding interventional procedures. Output is displayed on the IVUS system, assisting clinical decision-making regarding treatment strategy.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including physical integrity, functionality, and performance evaluations (imaging robustness, catheter robustness, dimensional requirements, and interface compatibility).
Technological Characteristics
30 MHz ultrasonic transducer; short-rail catheter design; telescoping shaft for 15 cm linear movement; electro-mechanical connector for MDU5 PLUS™; radiopaque markers; heparinized saline acoustic coupling; sterile.
Indications for Use
Indicated for intravascular ultrasound examination of peripheral vessels in patients who are candidates for transluminal interventional procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue.
September 23, 2022
Boston Scientific Corporation Cassie Clark Regulatory Affairs Specialist II 1 Scimed Place Maple Grove, Minnesota 55311
Re: K222568
Trade/Device Name: OptiCross™ 18 Peripheral Imaging Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ, ITX Dated: August 23, 2022 Received: August 24, 2022
Dear Cassie Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222568
Device Name
OptiCross 18TM Peripheral Imaging Catheter
Indications for Use (Describe)
OptiCross™ 18 is intended for intravascular ultrasound examination of peripheral vessels only. Intrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Boston Scientific. The text "Boston" is on the top line, and the text "Scientific" is on the bottom line. The text is in a serif font and is blue.
# K222568 510(k) Summary
per 21 CFR §807.92
| Sponsor | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA | | | |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------------|--|
| Contact Name and<br>Information | Cassie Clark<br>One Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-1425<br>e-mail: Cassie.Clark@bsci.com | | | |
| Prepared by | Cassie Clark<br>August 23, 2022 | | | |
| Proprietary Name | OptiCross™ 18 Peripheral Imaging Catheter | | | |
| Common Name | Diagnostic Intravascular Catheter, Ultrasound Transducer | | | |
| Product Code | OBJ, ITX | | | |
| Classification | Catheter, Ultrasound, Intravascular (OBJ) has been classified as Class<br>II per 21 CFR 870.1200<br>Transducer Ultrasonic (ITX) has been classified as Class II per 21<br>CFR 892.1570 | | | |
| Predicate Device | OptiCross™ 18 Peripheral<br>Imaging Catheter | K160514 | June 22, 2016 | |
| Reference Devices | OptiCross™ 35 Peripheral<br>Imaging Catheter | K200733 | April 17, 2020 | |
| | OptiCross™ 6 Coronary<br>Imaging Catheter | K213593 | January 14, 2022 | |
{4}------------------------------------------------
# Device Description
The OptiCross 18 30 MHz Peripheral Imaging Catheter is a sterile, short rail imaging catheter.
It consists of two main assemblies:
- 1. Imaging Core
- 2. Catheter Body
The imaging core is composed of a hi-torque, flexible, rotating drive cable with a radial looking 30 MHz ultrasonic transducer at the distal tip. An electro-mechanical connector interface at the proximal end of the catheter makes the connection to the Motordrive Unit (MDU5 PLUS™) Instrument. The MDU5 PLUS-catheter interface consists of an integrated mechanical drive socket and electrical connection.
The catheter body is comprised of three sections:
- 1. Distal Imaging Window Lumen
- 2. Proximal Shaft Lumen
- 3. Telescoping Section
The distal imaging window lumen and proximal shaft lumen sections comprise the "working length" of the catheter, and the telescoping section remains outside of the guiding catheter.
The catheter body has a distal imaging window lumen with proximal exit at 1.6 cm from the distal end. A radiopaque (RO) marker is embedded in the catheter body at 0.5 cm from the distal tip. In addition, two insertion depth markers are located on the proximal shaft lumen at 90 cm and 100 cm from the distal tip to aid in estimating catheter position relative to the distal guide catheter tip. The proximal shaft lumen is attached to the telescoping section via a strain relief connection.
The telescoping shaft (section) allows the imaging core to be advanced and retracted for 15 cm of linear movement. The corresponding movement of the transducer occurs from the proximal end of the quidewire exit port to the proximal end of the distal imaging window lumen. The telescope section has proximal markers for lesion length assessment, consisting of a series of marks spaced 1 cm apart on the telescope body.
A flush port with a one-way check valve is used to flush the interior of the catheter body and maintain a flushed condition. The catheter must be flushed with heparinized saline prior to use, as this provides the acoustic coupling media required for ultrasonic imaging. The one-way check valve helps retain saline in the catheter during use.
### Indications for Use / Intended Use
OptiCross 18 is intended for intravascular ultrasound examination of peripheral vessels only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.
# Comparison of Technological Characteristics
The OptiCross™ 18 Peripheral Imaging Catheters incorporate substantially equivalent design, materials, fundamental technology, interface with the IVUS imaging system, sterilization process, and intended use as those featured in the predicate; OptiCross™ 18 Peripheral Imaging Catheter K160514.
Minor design modifications are being implemented to the telescope design and the imaging core design to enhance device robustness and ease of use of the telescope. These changes are intended to harmonize design features across BSC's imaging catheter portfolio with the same enhancements previously submitted and reviewed as part of K213593.
{5}------------------------------------------------
## Non-clinical Performance Data
Determination of substantial equivalence is based on an assessment of non-clinical performance bench testing data.
# Bench Testing
Bench testing was performed to evaluate physical integrity, functionality, and performance of the OptiCross™ 18 Peripheral Imaging Catheters. Performance criteria includes imaging robustness requirements, catheter robustness requirements, dimensional requirements, and interface with compatible devices.
# Clinical Testing
Performance testing from clinical studies is not required to demonstrate substantial equivalence of the OptiCross™ 18 Peripheral Imaging Catheters.
# Conclusion
Based on the indications for use, technological characteristics, and performance testing, the OptiCross™ 18 Peripheral Imaging Catheters have been shown to be appropriate for their intended use and are considered to be substantially equivalent to the predicate device; OptiCross™ 18 Peripheral Imaging Catheter K160514.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.