← Product Code [GEX](/productcode/GEX) · K222555

# reepot Nd; YAG laser system (K222555)

_Ilooda Co,., Ltd. · GEX · Sep 23, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222555

## Device Facts

- **Applicant:** Ilooda Co,., Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Sep 23, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The reepot Nd;Y AG laser system in indicated for : the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 532nm Wavelength : - Tattoo removal: light ink (red, tan, purple, orange, skyblue, green) - Removal of Epidermal Pigmented Lesions - Removal of Minor Vascular Lesions including but not limited to telangiectasias - Treatment of Lentigines - Treatment of Cafe-Au-Lait - Treatment of Seborrheic Keratoses - Treatment of Post Inflammatory Hyper-Pigmentation - Treatment of Becker's Nevi, Freckles and Nevi Spilus

## Device Story

The reepot Nd:YAG laser system is a surgical laser device used in hospitals, clinics, and medical spas by trained professionals. It consists of a console, LCD control panel, footswitch, and VSLS handpieces (which include an integrated camera and LCD display). The system delivers 532nm laser energy via an articulated arm to perform soft tissue incision, ablation, and coagulation. The operator uses the LCD interface to control parameters; the integrated handpiece camera/display assists in visualization during procedures. The device provides therapeutic benefits by enabling precise tissue removal and vascular/pigmented lesion treatment. Clinical decision-making is guided by the physician's assessment of the patient's dermatological condition.

## Clinical Evidence

No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1, alongside software verification and validation and biocompatibility assessment.

## Technological Characteristics

Nd:YAG laser medium; 532nm wavelength; 5-20ns pulse width; 1-10Hz repetition rate; 4mm/6mm spot size; 635nm diode aiming beam; articulated arm optical delivery; LCD touch screen interface; 220-230VAC power. Complies with IEC 60601-1, 60601-1-2, 60601-2-22, and 60825-1 standards.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- CuRAS Nd:YAG Laser ([K173038](/device/K173038.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

September 23, 2022

Ilooda Co., Ltd. % Do Hyun Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-ru Seoul, Seoul 06210 Korea, South

Re: K222555

Trade/Device Name: reepot Nd; YAG laser system Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 7, 2022 Received: September 9, 2022

Dear Do Hyun Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K222555

Device Name

reepot Nd:Y AG laser system

Indications for Use (Describe)

The Reepot laser system is indicated for the incision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.

532nm Wavelength :

- Tattoo removal: light ink (red, tan, purple, orange, skyblue, green)
- Removal of Epidermal Pigmented Lesions
- Removal of Minor Vascular Lesions including but not limited to telangiectasias
- Treatment of Lentigines
- Treatment of Cafe-Au-Lait
- Treatment of Seborrheic Keratoses
- Treatment of Post Inflammatory Hyper-Pigmentation
- Treatment of Becker's Nevi, Freckles and Nevi Spilus

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for ilooda. The logo consists of a stylized graphic above the word "ilooda". The graphic is made up of several vertical lines of varying heights, arranged in two rows. The word "ilooda" is written in lowercase letters.

#### I. General Information [21 CFR 807.92(a) (1)]

| Applicant/Submitter: | Ilooda Co., Ltd.<br>9F, 25, Deokcheon-ro 152beon-gil, Manan-gu,<br>Anyang-si, Gyeonggi-do, Republic of Korea<br>Yun-Jung HA (yjha@ilooda.com) / RA Manager |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:      | Do Hyun Kim, BT Solutions, Inc.<br>Unit 904, Eonju-ro 86gil 5,<br>Gangnam-gu, Seoul 06210, Korea.<br>Tel: +82-2-538-9140<br>Email: ceo@btsolutions.co.kr   |
| Preparation Date:    | August 17, 2022                                                                                                                                            |

#### II. Names [21 CFR 807.92 (a) (2)]

| Trade or Proprietary Name reepot Nd: YAG laser system    |                                                                        |                                                |                                                                                                       |
|----------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Common Device<br>Name(s) and Regulatory<br>Class         | Product<br>Code(s)                                                     | Classification<br>Panel                        | Regulation                                                                                            |
| Laser Powered Surgical<br>Instruments (&<br>Accessories) | GEX                                                                    | General & Plastic<br>Surgery Panel, 79<br>(SU) | § 878.4810, Laser surgical<br>instrument for use in general<br>and plastic surgery and<br>dermatology |
| Class II                                                 | Surgical Powered Lasers and Delivery Devices/Hand piece<br>Accessories |                                                |                                                                                                       |

#### III. Predicate Devices [21 CFR 807.92(a) (3)]

| K #     | Predicate Device   |
|---------|--------------------|
| K173038 | CuRAS Nd:YAG Laser |

#### IV. Product Description [21 CFR 807.92(a) (4)]

The reepot Nd;Y AG laser system is comprised of the following major components:

- 1) Laser system console
- 2) LCD control panel
- 3) VSLS handpieces (included camera and LCD display)
- 4) Footswitch.
- 5) Accessories

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for ilooda. The logo consists of a stylized graphic above the word "ilooda". The graphic is made up of vertical lines of varying lengths, arranged in a pattern that resembles a sound wave or equalizer display. The word "ilooda" is written in a lowercase sans-serif font.

The system is intended to be used in user facilities such as hospitals, physicians' offices and medical spas.

#### V. Intended Use and Indications for Use [21 CFR 807.92(a) (5)]

The reepot Nd;Y AG laser system in indicated for : the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.

532nm Wavelength :

- Tattoo removal: light ink (red, tan, purple, orange, skyblue, green)
- Removal of Epidermal Pigmented Lesions
- Removal of Minor Vascular Lesions including but not limited to telangiectasias
- Treatment of Lentigines
- Treatment of Cafe-Au-Lait
- Treatment of Seborrheic Keratoses
- Treatment of Post Inflammatory Hyper-Pigmentation
- Treatment of Becker's Nevi, Freckles and Nevi Spilus

#### VI. Summary of Technical Characteristics [21 CFR 807.92(a)(6)]

| Table 1: Technical Comparison for the Q-switched Laser |  |  |
|--------------------------------------------------------|--|--|
|--------------------------------------------------------|--|--|

| Parameter                        | reepot Nd;YAG laser<br>system<br>(K222555)                                                                                                                                                                                                                                                                                                                                                                                                                                               | CuRAS Nd;YAG laser system<br>(K173038)                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code &<br>Regulation No. | GEX<br>21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | GEX<br>21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Laser Medium                     | Nd:YAG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Nd:YAG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Laser wavelength                 | 532nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 1064nm/532nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Output energy                    | Max 0.35J @532 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Max 1.6J @1064 nm<br>Max 0.4J @532 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Pulse width                      | 5-20ns                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 5-20ns                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Repetition Rate                  | 1-10Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 1-15Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Spot size                        | 4mm, 6mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 2mm-10mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Aiming beam                      | Diode 635nm 5mW                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Diode 635nm 5mW                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| User Interface                   | LCD touch screen                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | LCD touch screen                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Optical guide                    | Articulated arm                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Articulated arm                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Electrical<br>Requirements       | 220-230VAC,<br>50-60 Hz,                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 220-230VAC,<br>50-60 Hz,                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                  | The reepot Nd:YAG laser<br>system in<br>indicated for : the<br>incision,excision,<br>ablation, vaporization of soft<br>tissues for<br>general dermatology,<br>dermatologic and<br>general surgical procedures for<br>coagulation and hemostasis.                                                                                                                                                                                                                                         | The CuRAS Nd:YAG laser<br>system in<br>indicated for : the<br>incision,excision,<br>ablation, vaporization of soft<br>tissues for<br>general dermatology,<br>dermatologic and<br>general surgical procedures for<br>coagulation and hemostasis.                                                                                                                                                                                                                                       |
| Indications for<br>Use           | 532nm Wavelength :<br>- Tattoo removal: light ink<br>(red, tan,<br>purple, orange, skyblue,<br>green)<br>- Removal of Epidermal<br>Pigmented<br>Lesions<br>- Removal of Minor Vascular<br>Lesions<br>including but not limited to<br>telangiectasias<br>- Treatment of Lentigines<br>- Treatment of Cafe-Au-Lait<br>- Treatment of Seborrheic<br>Keratoses<br>- Treatment of Post<br>Inflammatory Hyper-<br>Pigmentation<br>- Treatment of Becker's Nevi,<br>Freckles<br>and Nevi Spilus | 532nm Wavelength :<br>- Tattoo removal: light ink (red,<br>tan,<br>purple, orange, skyblue, green)<br>- Removal of Epidermal<br>Pigmented<br>Lesions<br>- Removal of Minor Vascular<br>Lesions<br>including but not limited to<br>telangiectasias<br>- Treatment of Lentigines<br>- Treatment of Cafe-Au-Lait<br>- Treatment of Seborrheic<br>Keratoses<br>- Treatment of Post<br>Inflammatory Hyper-<br>Pigmentation<br>- Treatment of Becker's Nevi,<br>Freckles<br>and Nevi Spilus |

{5}------------------------------------------------

![ilooda](https://raw.githubusercontent.com/google/generative-ai-docs/main/site/en/examples/images/ilooda.png)
ilooda

#### VII. Performance Testing [21 CFR 807.92(b)(1)]

IEC 60601-1 Medical Electrical Equipment-Part 1: General Requirements for safety IEC 60601-1-2 Medical Electrical Equipment 1-2 General Requirements for basic safety and essential performance

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for ilooda. The logo consists of a stylized graphic above the word "ilooda". The graphic is made up of seven vertical bars of varying heights, arranged in a way that suggests sound waves or a digital equalizer. The word "ilooda" is written in a lowercase sans-serif font.

IEC 60601-2-22 Medical Electrical Equipment-Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment

IEC 60825-1 Safety of laser products-Part 1: Equipment Classification, requirements and user's guide

In addition software verification and validation testing was performed and biocompatibility was established.

#### Clinical Data [21 CFR 807.92(b) (2)] VIII.

Based on the similarities in the safety and effectiveness profiles of the subject, reference, and primary predicate devices, no animal or clinical studies were deemed needed to support this submission.

#### IX. Conclusion

The reepot Nd; Y AG laser system was found to be substantially equivalent to the predicate devices.

The reepot Nd; Y AG laser system shares the same or similar indications for use, similar design features, and functional features with, and thus is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K222555](https://fda-staging.innolitics.com/device/K222555)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).