← Product Code [FXX](/productcode/FXX) · K222551

# Humask Pro Vision, Humask Pro Vision 3000 (K222551)

_Entreprise Premont, Inc. · FXX · Nov 10, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222551

## Device Facts

- **Applicant:** Entreprise Premont, Inc.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Nov 10, 2022
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Humask Pro Vision 3000 mask is single use disposable intended to be worn in the operating room as well as dental, isolation, and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.

## Device Story

Humask Pro Vision 3000 is a single-use, disposable, flat-pleated surgical mask featuring a transparent rectangular window. Designed for use in operating rooms, dental, and isolation settings by healthcare personnel and patients. Device acts as a physical barrier to prevent transfer of microorganisms, body fluids, and particulate material. Features include ear loops, a nose band, and a HuCare membrane (PTFE film laminated onto polypropylene spun-bound sheet) for filtration. Transparent window allows for visibility. Device is non-sterile. Provides protection through mechanical filtration and fluid resistance; benefits include reduced risk of cross-contamination and improved communication via the clear window.

## Clinical Evidence

Bench testing only. Performance data includes Bacterial Filtration Efficiency (ASTM F2101), Particle Filtration Efficiency (ASTM F2299), Differential Pressure (EN 14683), Flammability (16 CFR Part 1610), Resistance to synthetic blood (ASTM F1862), and Biocompatibility (ISO 10993-5 Cytotoxicity, ISO 10993-10 Allergenicity and sensitization). Results met all specified criteria.

## Technological Characteristics

Materials: Polypropylene spun-bound outer layers, HuCare membrane (PTFE film laminated to polypropylene spun-bound), 3ml steel wire noseband, spandex ear loops, polyethylene plastic window. Design: Flat, pleated surgical mask with rectangular transparent window (4.9 in x 1.7 in). Non-sterile. Standards: ASTM F1862, ASTM F2101, ASTM F2299, EN 14683, 16 CFR Part 1610, ISO 10993-5, ISO 10993-10.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Communicator ([K152561](/device/K152561.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.

November 10, 2022

Entreprise Prémont Inc. Charlie Marchand Director of Product Development and Compliance 731 Boulevard Saint Laurent Est Louiseville, Quebec J5V 1J1 Canada

Re: K222551

Trade/Device Name: Humask Pro Vision 3000 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 15, 2022 Received: August 23, 2022

Dear Charlie Marchand:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222551

Device Name Humask Pro Vision 3000

Indications for Use (Describe)

The Humask Pro Vision 3000 mask is single use disposable intended to be worn in the operating room as well as dental, isolation, and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

I.

- SUBMITTER Humask 731 Boulevard Saint Laurent Est Louiseville, Quebec, J5V 1J1 Canada Contact person: Charlie Marchand Phone: 819-519-3255 ext.215 Date prepared: 31st of October 2022
- II. DEVICE Name of the device: Humask Pro Vision 3000 Common of usual name: Surgical Mask Classification name: Surgical Mask Regulatory Class: 2 Product Code: FXX Regulation: 21 CFR 878.4040
- III. PREDICATE DEVICE Communicator (K152561)- primary predicate The predicate has not been subject to a design-related recall.
- IV. DEVICE DESCRIPTIONS The Humask Pro Vision 3000 is a clear window surgical mask. A Medical grade face mask with a transparent window.

#### V. INDICATIONS FOR USE

The Humask Pro Vision 3000 mask is single use disposable intended to be worn in the operating room as well as dental, isolation, and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.

- VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technological characteristics of the Humask Pro Vision 3000 is highly analogous to the predicate device, the Communicator (K152561).

The similarities and differences are illustrated in the tables below:

| Characteristics | Humask Pro Vision 3000 |
|-----------------|------------------------|
|                 | Communicator (K152561) |

Humask Pro Vision 3000

{4}------------------------------------------------

| Indications for Use         | The Humask Pro Vision 3000                                                                                                                                                                                                                                                                                            | The Communicator Surgical                                                                                                                                                                                                                                                                                                 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                             | The Humask Pro Vision 3000<br>mask is single use disposable<br>intended to be worn in the<br>operating room as well as dental,<br>isolation, and other medical<br>procedures to protect both the<br>patient and healthcare personnel<br>from transfer of microorganisms,<br>body fluids, and particulate<br>material. | Facemask with Clear Window,<br>FM86000, is single use disposable<br>intended to be worn in the<br>operating room as well as dental,<br>isolation, and other medical<br>procedures to protect both the<br>patient and healthcare personnel<br>from transfer of microorganisms,<br>body fluids, and particulate<br>material |
| Prescription Only or<br>OTC | OTC                                                                                                                                                                                                                                                                                                                   | OTC                                                                                                                                                                                                                                                                                                                       |

 The indications for use statement for the subject device is functionally the same as the predicate device.

|                        | Humask Pro Vision 3000                                                                                                         | PREDICATE Communicator (K152561)                                                  | Similarities/ differences |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|---------------------------|
| Principle of Operation | Protective mask                                                                                                                | Protective mask                                                                   | SAME                      |
| Single use             | Yes                                                                                                                            | Yes                                                                               | SAME                      |
| Mask Style             | Surgical mask- flat, pleated                                                                                                   | Surgical mask- flat, pleated                                                      | SAME                      |
| Design/ Const.         |                                                                                                                                |                                                                                   |                           |
| Materials              | Polypropylene spun bond PTFE laminated Polypropylene 3ml steel wire (Noseband) Spandex (ear loops) Polyethylene plastic window | Polypropylene bicomponent spun bond Melt blown Plastic window Aluminum (Noseband) | DIFFERENT                 |
| Design features        | Ear loops                                                                                                                      | Ear loops                                                                         | SAME                      |

Humask Pro Vision 3000

{5}------------------------------------------------

|                                     | Humask Pro Vision<br>3000                                                                             | PREDICATE<br>Communicator<br>(K152561)                                                                                                         | Similarities/<br>differences |
|-------------------------------------|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
|                                     | Nose band<br>Transparent window                                                                       | Nose band<br>Transparent window                                                                                                                |                              |
| Transparent<br>window shape         | Rectangular                                                                                           | Diamond                                                                                                                                        | DIFFERENT                    |
| Transparent<br>window<br>dimensions | 4.9 in x 1.7 in                                                                                       | 4.75 in x 1.35 in<br>(tapering to 0.5 width)                                                                                                   | DIFFERENT                    |
| How provided                        | Non-sterile                                                                                           | Non-sterile                                                                                                                                    | SAME                         |
| Biocompatibility                    | ISO 10993                                                                                             | ISO 10993 (presumed)                                                                                                                           | SAME                         |
| Performance<br>Data                 | ASTM F1862<br>ASTM F2101<br>ASTM F2299<br>EN 14683<br>16 CFR PART 1610<br>ISO 10993-5<br>ISO 10993-10 | (Presumed - standard<br>requirements)<br>ASTM F1862<br>ASTM F2101<br>ASTM F2299<br>EN 14683<br>16 CFR PART 1610<br>ISO 10993-5<br>ISO 10993-10 | SAME                         |

The performance test data demonstrating compliance with those requirements outlined in FDA Guidance Surgical Masks Premarket Notification Submissions, May 2003.

The technological characteristics of the subject device are similar to the technological characteristics of the predicate. The fundamental difference is the nature of the filter media used within the mask and the size and shape of the window. The Predicate product's media filter is a melt blown material which is a polypropylene spun bound product electrostatically charged to trap particles coming through the filter. The HuCareTM membrane is comprised of (polytetrafluorethylene) PTFE film laminated onto a polypropylene spun bound sheet. The HuCare™ membrane is also smooth and non-pilling as well as hydrophobic which plays a role in the resistance to body fluids and evacuation of humidity that can accumulate within the mask throughout use. The size and shape of the window in this mask is also a differentiation point between the predicate mask and the Humask Pro Vision 3000 mask: The predicate product's window is in the shape of a diamond (4.75 in x 1.35 in (tapering to 0.5 width)) reaching its widest point in the middle of the mask. The Humask Pro Vision 3000's window is rectangular (4.9 in x 1.7 in) and longer than the predicate.

At a high level, the subject and predicate devices are based on the following same technological elements:

- . Outer layers are polypropylene spun bound material

{6}------------------------------------------------

- Mask type is based on a flat pleated mask
- . Window material is clear see-through plastic
- Ear loops and nose band for secure and comfortable wear
- Mask contains inner filter layer ●

The following differences exist between the subject and predicate devices:

- Window shape (Humask Pro Vision 3000 is rectangular and predicate is . diamond shaped)
- . Inner filter material (Humask Pro Vision 3000 uses HuCare membrane (laminated PTFE) and predicate uses melt blown)

### VII. PERFORMANCE DATA

The following performance data support the substantial equivalence of the Humask Pro Vision 3000.

- Bacterial filtration efficiency (ASTM F2101) ●
- . Particle filtration efficiency (ASTM F2299)
- Differential Pressure (EN 14683) .
- Flammability (16 CFR PART 1610) ●
- Resistance to synthetic blood (ASTM F1862) ●
- ISO 10993-5 Cytotoxicity ●
- ISO10993-10 Allergenicity and sensitization ●

The subject devices met all specified criteria and did not raise new safety or performance questions. Based on the performance testing, the Humask Pro Vision 3000 was found to have a safety and effectiveness profile that is similar to the predicate device.

### VIII. CONCLUSIONS

The Humask Pro Vision 3000 is substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K222551](https://fda-staging.innolitics.com/device/K222551)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
