K222551 · Entreprise Premont, Inc. · FXX · Nov 10, 2022 · General, Plastic Surgery
Device Facts
Record ID
K222551
Device Name
Humask Pro Vision, Humask Pro Vision 3000
Applicant
Entreprise Premont, Inc.
Product Code
FXX · General, Plastic Surgery
Decision Date
Nov 10, 2022
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Humask Pro Vision 3000 mask is single use disposable intended to be worn in the operating room as well as dental, isolation, and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.
Device Story
Humask Pro Vision 3000 is a single-use, disposable, flat-pleated surgical mask featuring a transparent rectangular window. Designed for use in operating rooms, dental, and isolation settings by healthcare personnel and patients. Device acts as a physical barrier to prevent transfer of microorganisms, body fluids, and particulate material. Features include ear loops, a nose band, and a HuCare membrane (PTFE film laminated onto polypropylene spun-bound sheet) for filtration. Transparent window allows for visibility. Device is non-sterile. Provides protection through mechanical filtration and fluid resistance; benefits include reduced risk of cross-contamination and improved communication via the clear window.
Clinical Evidence
Bench testing only. Performance data includes Bacterial Filtration Efficiency (ASTM F2101), Particle Filtration Efficiency (ASTM F2299), Differential Pressure (EN 14683), Flammability (16 CFR Part 1610), Resistance to synthetic blood (ASTM F1862), and Biocompatibility (ISO 10993-5 Cytotoxicity, ISO 10993-10 Allergenicity and sensitization). Results met all specified criteria.
Technological Characteristics
Materials: Polypropylene spun-bound outer layers, HuCare membrane (PTFE film laminated to polypropylene spun-bound), 3ml steel wire noseband, spandex ear loops, polyethylene plastic window. Design: Flat, pleated surgical mask with rectangular transparent window (4.9 in x 1.7 in). Non-sterile. Standards: ASTM F1862, ASTM F2101, ASTM F2299, EN 14683, 16 CFR Part 1610, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for use by healthcare personnel and patients in operating rooms, dental, isolation, and medical settings to provide barrier protection against microorganisms, body fluids, and particulates.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
November 10, 2022
Entreprise Prémont Inc. Charlie Marchand Director of Product Development and Compliance 731 Boulevard Saint Laurent Est Louiseville, Quebec J5V 1J1 Canada
Re: K222551
Trade/Device Name: Humask Pro Vision 3000 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 15, 2022 Received: August 23, 2022
Dear Charlie Marchand:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222551
Device Name Humask Pro Vision 3000
Indications for Use (Describe)
The Humask Pro Vision 3000 mask is single use disposable intended to be worn in the operating room as well as dental, isolation, and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
I.
- SUBMITTER Humask 731 Boulevard Saint Laurent Est Louiseville, Quebec, J5V 1J1 Canada Contact person: Charlie Marchand Phone: 819-519-3255 ext.215 Date prepared: 31st of October 2022
- II. DEVICE Name of the device: Humask Pro Vision 3000 Common of usual name: Surgical Mask Classification name: Surgical Mask Regulatory Class: 2 Product Code: FXX Regulation: 21 CFR 878.4040
- III. PREDICATE DEVICE Communicator (K152561)- primary predicate The predicate has not been subject to a design-related recall.
- IV. DEVICE DESCRIPTIONS The Humask Pro Vision 3000 is a clear window surgical mask. A Medical grade face mask with a transparent window.
#### V. INDICATIONS FOR USE
The Humask Pro Vision 3000 mask is single use disposable intended to be worn in the operating room as well as dental, isolation, and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.
- VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technological characteristics of the Humask Pro Vision 3000 is highly analogous to the predicate device, the Communicator (K152561).
The similarities and differences are illustrated in the tables below:
| Characteristics | Humask Pro Vision 3000 |
|-----------------|------------------------|
| | Communicator (K152561) |
Humask Pro Vision 3000
{4}------------------------------------------------
| Indications for Use | The Humask Pro Vision 3000 | The Communicator Surgical |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Humask Pro Vision 3000<br>mask is single use disposable<br>intended to be worn in the<br>operating room as well as dental,<br>isolation, and other medical<br>procedures to protect both the<br>patient and healthcare personnel<br>from transfer of microorganisms,<br>body fluids, and particulate<br>material. | Facemask with Clear Window,<br>FM86000, is single use disposable<br>intended to be worn in the<br>operating room as well as dental,<br>isolation, and other medical<br>procedures to protect both the<br>patient and healthcare personnel<br>from transfer of microorganisms,<br>body fluids, and particulate<br>material |
| Prescription Only or<br>OTC | OTC | OTC |
The indications for use statement for the subject device is functionally the same as the predicate device.
| | Humask Pro Vision 3000 | PREDICATE Communicator (K152561) | Similarities/ differences |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|---------------------------|
| Principle of Operation | Protective mask | Protective mask | SAME |
| Single use | Yes | Yes | SAME |
| Mask Style | Surgical mask- flat, pleated | Surgical mask- flat, pleated | SAME |
| Design/ Const. | | | |
| Materials | Polypropylene spun bond PTFE laminated Polypropylene 3ml steel wire (Noseband) Spandex (ear loops) Polyethylene plastic window | Polypropylene bicomponent spun bond Melt blown Plastic window Aluminum (Noseband) | DIFFERENT |
| Design features | Ear loops | Ear loops | SAME |
Humask Pro Vision 3000
{5}------------------------------------------------
| | Humask Pro Vision<br>3000 | PREDICATE<br>Communicator<br>(K152561) | Similarities/<br>differences |
|-------------------------------------|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| | Nose band<br>Transparent window | Nose band<br>Transparent window | |
| Transparent<br>window shape | Rectangular | Diamond | DIFFERENT |
| Transparent<br>window<br>dimensions | 4.9 in x 1.7 in | 4.75 in x 1.35 in<br>(tapering to 0.5 width) | DIFFERENT |
| How provided | Non-sterile | Non-sterile | SAME |
| Biocompatibility | ISO 10993 | ISO 10993 (presumed) | SAME |
| Performance<br>Data | ASTM F1862<br>ASTM F2101<br>ASTM F2299<br>EN 14683<br>16 CFR PART 1610<br>ISO 10993-5<br>ISO 10993-10 | (Presumed - standard<br>requirements)<br>ASTM F1862<br>ASTM F2101<br>ASTM F2299<br>EN 14683<br>16 CFR PART 1610<br>ISO 10993-5<br>ISO 10993-10 | SAME |
The performance test data demonstrating compliance with those requirements outlined in FDA Guidance Surgical Masks Premarket Notification Submissions, May 2003.
The technological characteristics of the subject device are similar to the technological characteristics of the predicate. The fundamental difference is the nature of the filter media used within the mask and the size and shape of the window. The Predicate product's media filter is a melt blown material which is a polypropylene spun bound product electrostatically charged to trap particles coming through the filter. The HuCareTM membrane is comprised of (polytetrafluorethylene) PTFE film laminated onto a polypropylene spun bound sheet. The HuCare™ membrane is also smooth and non-pilling as well as hydrophobic which plays a role in the resistance to body fluids and evacuation of humidity that can accumulate within the mask throughout use. The size and shape of the window in this mask is also a differentiation point between the predicate mask and the Humask Pro Vision 3000 mask: The predicate product's window is in the shape of a diamond (4.75 in x 1.35 in (tapering to 0.5 width)) reaching its widest point in the middle of the mask. The Humask Pro Vision 3000's window is rectangular (4.9 in x 1.7 in) and longer than the predicate.
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Outer layers are polypropylene spun bound material
{6}------------------------------------------------
- Mask type is based on a flat pleated mask
- . Window material is clear see-through plastic
- Ear loops and nose band for secure and comfortable wear
- Mask contains inner filter layer ●
The following differences exist between the subject and predicate devices:
- Window shape (Humask Pro Vision 3000 is rectangular and predicate is . diamond shaped)
- . Inner filter material (Humask Pro Vision 3000 uses HuCare membrane (laminated PTFE) and predicate uses melt blown)
### VII. PERFORMANCE DATA
The following performance data support the substantial equivalence of the Humask Pro Vision 3000.
- Bacterial filtration efficiency (ASTM F2101) ●
- . Particle filtration efficiency (ASTM F2299)
- Differential Pressure (EN 14683) .
- Flammability (16 CFR PART 1610) ●
- Resistance to synthetic blood (ASTM F1862) ●
- ISO 10993-5 Cytotoxicity ●
- ISO10993-10 Allergenicity and sensitization ●
The subject devices met all specified criteria and did not raise new safety or performance questions. Based on the performance testing, the Humask Pro Vision 3000 was found to have a safety and effectiveness profile that is similar to the predicate device.
### VIII. CONCLUSIONS
The Humask Pro Vision 3000 is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.