Published clinical literature (Parel et al., 2001); Published clinical literature (Fernández-Ruiz et al., 2021)
The sponsor utilized published clinical studies to support the substantial equivalence of additional zygomatic implant lengths by demonstrating clinical use in patient rehabilitation and patient-reported outcomes.
Clinical literature; Zygomatic implants; Patient rehabilitation; Quality of life
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Parel et al. (2001); Clinical case report/series
Patients with maxillary defects; Sample Size: 59 zygomatic implants
Not applicable for this study
Clinical rationale for use of remote bone sites and zygomatic implants
Fernández-Ruiz et al. (2021); Prospective randomized clinical study
Southern Implants ZAGA Zygomatic System implants are intended to be implanted in the upper jaw arch to provide support for fixed dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
ZAGA Zygomatic System consists of partially threaded root-form dental implants designed for placement into the zygoma to support fixed dental prostheses in edentulous or partially edentulous maxillae. Implants feature an external hexagonal abutment interface angled at 55°. Available in 'Round' and 'Flat' designs with varying lengths. Implants are placed by dental surgeons in a clinical setting. The device provides a stable anchor for prosthetic rehabilitation, potentially improving patient quality of life and satisfaction. The system is provided sterile for single-patient, single-use.
Clinical Evidence
Clinical evidence includes a case report and two published studies. Parel et al. (2001) reported on 59 zygomatic implants (25-60mm) for maxillary defect rehabilitation. Fernández-Ruiz et al. (2021) conducted a prospective randomized clinical study of 40 zygomatic cases (35-60mm) comparing patient quality of life and satisfaction against conventional All-on-Four implants.
Technological Characteristics
Unalloyed titanium (ASTM F67, Grade 4, UTS ≥ 900MPa). Partially threaded root-form design with grit-blasted endosseous surface. External hex interface with 55° head angulation. Sterilized via gamma irradiation. Complies with ISO 14801 for dynamic fatigue testing.
Indications for Use
Indicated for patients with partially or fully edentulous maxillae requiring support for fixed dental prostheses via implantation in the upper jaw arch. Appropriate for immediate loading given good primary stability and appropriate occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Southern Implants (Pty) Ltd Leith Cumming Acting Head of Regulatory Affairs and Quality 1 Albert Road Irene, Gauteng 0062 SOUTH AFRICA
Re: K222497
Trade/Device Name: Zaga Zygomatic System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: August 17, 2022 Received: August 18, 2022
# Dear Leith Cumming:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K222497
Device Name
ZAGA Zygomatic System
Indications for Use (Describe)
Southern Implants ZAGA Zygomatic System implants are intended to be implanted in the upper jaw arch to provide support for fixed dental prostheses in patially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
| Type of Use ( <i>Select one or both, as applicable</i> ) | <span> <span style="padding-right: 5px;"> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
### 510(k) Summary
## ZAGA Zygomatic System
#### Southern Implants (Pty) Ltd
#### K222497
#### November 9, 2022
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Southern Implants (Pty) Ltd<br>1 Albert Road<br>Irene, Gauteng, 0062 South Africa |
|-------------------|----------------------------------------------------------------------------------------------------------|
| | Telephone: +27 12 667 1046<br>Fax: +27 12 667 1029 |
| Official Contact | Leith C. Cumming<br>Acting Head of Regulatory Affairs and Quality<br>Email: leith.c@southernimplants.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | ZAGA Zygomatic System |
|---------------------------|---------------------------|
| Common Name | Dental implant |
| Classification Name | Endosseous dental implant |
| Classification Regulation | 21 CFR 872.3640, Class II |
| Product Code | DZE |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
#### PREDICATE DEVICE INFORMATION The primary predicate device is K192651.
#### INDICATIONS FOR USE STATEMENT
Southern Implants ZAGA Zygomatic System implants are intended to be implanted in the upper jaw arch to provide support for fixed dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
#### SUBJECT DEVICE DESCRIPTION
The ZAGA zygomatic implants are partially threaded root-form dental implants with an external hexagonal abutment interface angled 55° at the head of the implant, and are threaded internally for attachment of mating abutments. The ZAGA zygomatic implants are provided in two designs: the ZAGA Round zygomatic implant (previously named the ZAGA Regular), coronally and apically threaded, diameter 4.3 mm (coronal) tapering to 3.4 mm (apical), in overall lengths ranging of 30 mm and the ZAGA Flat zygomatic implant (previously named the ZAGA Advanced), apically threaded, diameter 4.3 mm (coronal) tapering to 3.4 mm (apical), in overall lengths ranging from 30 mm, with a flat cut into the buccal-facing side of the implant body. Only the length 30, 32.5, 57.5 and 60 mm ZAGA Round zygomatic implants and length 55, 57.5 and 60 mm ZAGA Flat zygomatic implants are a subject of this submission.
{4}------------------------------------------------
All ZAGA zygomatic implants are manufactured from unalloyed titanium conforming to ASTM F67. The apically threaded portions of the implants have the identical aluminum oxide grit-blasted surface as the implants cleared in K192651. The subject device implants are manufactured in the same facilities using the same manufacturing processes as used for the Southern Implants predicate device previously cleared in K192651.
#### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility (referenced from K192651); comparison and features; sterilization validation according to ISO 11137-1, ISO 11137-2, ISO 17665-1, ISO TS 17665-2; bacterial endotoxin according to USP 39-NF34; sterile barrier shelf life (referenced from K192651); MR safety evaluation and MR Conditional labeling as per the recommendations of the FDA Guidance Document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment''; and static and dynamic compression-bending according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants (referenced from K192651 as the surgical protocol of the implants did not warrant the additional subject device lengths as a new worst-case for fatigue testing).
Non-clinical worst-case MRI review was performed to evaluate the metallic Straumann® ZAGA Zygomatic Implant System devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795; and Oriso, Kenta, Takuya Kobayashi, Makoto Sasaki, Ikuko Uwano, Hidemichi Kihara, and Hisatomo Kondo. "Impact of the static and radiofrequency magnetic fields produced by a 7T MR imager on metallic dental materials." Magnetic Resonance in Medical Sciences 15, no. 1 (2016): 26-33), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
Clinical data submitted, referenced or relied upon to demonstrate substantial equivalence include: a clinical case report demonstrating the use of 60mm length zygomatic implants in a patient, and two published articles utilizing, among other implants, 60mm length zygomatic implants for patient rehabilitation. These articles are summarized below:
- Parel et al. (Remote implant anchorage for the rehabilitation of maxillary defects. The Journal of . prosthetic dentistry 86.4 (2001): 377-381) presented a rationale for the use of remote bone sites and zygomatic implants for rehabilitating patients with maxillary defects. Parel et al. provide clinical data on the use of 59 zygomatic implants in lengths ranging from 25 to 60mm.
- Fernández-Ruiz et al. (Evaluation of quality of life and satisfaction in patients with fixed prostheses on . zygomatic implants compared with the all-on-four concept: a prospective randomized clinical study. International Journal of Environmental Research and Public Health 18.7 (2021): 3426) present a prospective randomized clinical study comparing patient quality of life and satisfaction with their prosthesis on zygomatic implants versus the All-on-Four with conventional implants. This study reported 40 zygomatic cases in which zygomatic implants of lengths 35 - 60mm were used.
#### EQUIVALENCE TO MARKETED DEVICE
Southern Implants (Pty) Ltd submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical device is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
#### K192651, ZAGA Zygomatic System, Southern Implants (Pty) Ltd.
A comparison of the technological characteristics of the subject devices and the primary predicate devices is provided in the following table.
{5}------------------------------------------------
| | Subject Device | Primary Predicate Device |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison | Additional lengths of ZAGA Zygomatic Implant System<br>Southern Implants (Pty) Ltd | K192651<br>ZAGA Zygomatic Implant System<br>Southern Implants (Pty) Ltd |
| Implant | | |
| Indications for Use Statement | Straumann® Zygomatic implants are intended to be implanted in the upper jaw arch to provide support for fixed dental prostheses in patients with partially or fully edentulous maxillae.<br>All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. | Straumann® Zygomatic implants are intended to be implanted in the upper jaw arch to provide support for fixed dental prostheses in patients with partially or fully edentulous maxillae.<br>All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. |
| Product Code | DZE | DZE |
| Intended Use | Functional and esthetic rehabilitation of the edentulous maxilla. | Functional and esthetic rehabilitation of the edentulous maxilla. |
| Reason for Predicate/Reference | Not applicable | Implant design |
| Implant Design | Partially threaded root-form implants for placement into the zygoma. | Partially threaded root-form implants for placement into the zygoma. |
| Implant Diameter | ZAGA Round: 4.3 mm (coronal) taper to 3.4 mm (apical)<br>ZAGA Flat: 4.3 mm (coronal with 3.35 mm flat) taper to 3.4 mm (apical) | ZAGA Round: 4.3 mm (coronal) taper to 3.4 mm (apical)<br>ZAGA Flat: 4.3 mm (coronal with 3.35 mm flat) taper to 3.4 mm (apical) |
| Implant Length | ZAGA Round: 30, 32.5, 57.5, 60 mm<br>ZAGA Flat: 55, 57.5, 60 mm | ZAGA Round: 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 55 mm<br>ZAGA Flat: 30, 32.5, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5 mm |
| Platform Angle, Relative to orthogonal to implant long axis | 55° angulation at head of implant | 55° angulation at head of implant |
| Implant Prosthetic Diameter | 4.07 mm | 4.07 mm |
| Implant Interface | External Hex | External Hex |
| Implant Material | Unalloyed titanium (ASTM F67) Grade 4, and UTS ≥ 900MPa (cold-worked) | Unalloyed titanium (ASTM F67) Grade 4, and UTS ≥ 900MPa (cold-worked) |
| Implant Endosseous Surface | Grit-blasted | Grit-blasted |
| How Provided | | |
| Sterility | Provided sterile | Provided sterile |
| Sterilization | Gamma irradiation | Gamma irradiation |
| Usage | Single-patient, single-use | Single-patient, single-use |
{6}------------------------------------------------
The Indications for Use Statement for the subject device is identical to the predicate device K192651.
The primary predicate device K192651 is for substantial equivalence of the subject device implant designs. The subject device implants have the same external hex connection, identical implant diameters, prosthetic platform angulation and diameter, material, endosseous surface, sterility status method, usage and compatible prosthetic components. The clinical evidence, in the form of a clinical case report and two publications, is for the substantial equivalence of the subject device implant lengths (see clinical performance data).
Substantial equivalence of the subject device implants in terms of biocompatibility is supported by the fact that materials are identical in formulation, processing, component interactions, and storage conditions to the predicate devices in K192651.
In support of substantial equivalence of the additional lengths of ZAGA zygomatic implants in terms of mechanical performance, dynamic compression-bending testing to ISO 14801 are referenced from K192651. Dynamic testing (referenced from K192651) was performed on worst-case subject device constructs. The additional lengths of the ZAGA zygomatic implants does not influence the results of this fatigue testing and so additional fatigue testing was not warranted for this submission. The results from the testing provided in K192651 are thus valid for the subject devices and demonstrates the fatigue performance of the subject device exceeds its indication.
#### CONCLUSION
The subject device and the predicate device have the same intended use and indications for use, have similar technological characteristics, and are made of the same materials. The subject device and the predicate device encompass a similar range of physical dimensions, including the same diameter of the implants, same prosthetic platform and same compatible abutment range. The differing total lengths of the subject device is supported by clinical data provided. The subject device and the predicate devices are packaged in similar materials and sterilized using the same methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.