Harvest Printable Resin is indicated for the fabrication of dental bases for full removable dentures, artificial teeth, dental bite splints and guards and try-in devices, i.e., denture base and teeth. The material is an alternative to traditional heat-curable and auto-polymerizing resins and is intended for professional dental work only. This material is intended to be used by dental lab technicians and approved by licensed practitioners before being provided to the patient.
Device Story
Harvest Printable Resin is a light-cured resin for additive manufacturing (3D printing) of dental appliances; specifically denture bases, artificial teeth, bite splints, and guards. The process involves using a CAD/CAM system and a digital light processing (DLP) printer to generate the device from a solid model, followed by post-processing in a UV light curing unit for final polymerization. The device is intended for use in dental laboratories by technicians, with final approval by licensed practitioners. It is optimized for use with specific Asiga and Sprintray printers and curing units. The resin serves as an alternative to traditional heat-curable and auto-polymerizing resins. By enabling 3D-printed fabrication, it allows for customized dental restorations, potentially improving workflow efficiency and precision in dental appliance manufacturing.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity per ISO 10993-1 and ISO 7405:2018) and mechanical property testing (flexural strength, flexural modulus, water sorption, and water solubility) per ISO 20795-1 and ISO 20795-2.
Technological Characteristics
Light-cured resin composed of methacrylates, urethane methacrylates, photo initiators, and pigments. Form factor is liquid resin for 3D printing. Requires CAD/CAM system, DLP printer, and UV curing unit. Complies with ISO 20795-1 and ISO 20795-2 for mechanical properties. Biocompatibility per ISO 10993-1 and ISO 7405:2018. No MMA monomers used.
Indications for Use
Indicated for fabrication of dental bases for full removable dentures, artificial teeth, dental bite splints, guards, and try-in devices. Intended for professional dental use by lab technicians, subject to approval by licensed practitioners.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
May 12, 2023
Harvest Dental Products, LLC Colleen Boswell Regulatory Affairs Consultant 905 Columbia Street Brea, California 92821
Re: K222489
Trade/Device Name: Harvest Printable Resin Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI, ELM, MQC Dated: April 25, 2023 Received: April 25, 2023
Dear Colleen Boswell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -2
For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K222489
Device Name
Harvest Printable Resin
#### Indications for Use (Describe)
Harvest Printable Resin is indicated for the fabrication of dental bases for full removable dentures, artificial teeth, dental bite splints and guards and try-in devices, i.e., denture base and teeth. The material is an alternative to traditional heat-curable and auto-polymerizing resins and is intended for professional dental work only. This material is intended to be used by dental lab technicians and approved by licensed practitioners before being provided to the patient.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below "HarvestDental" is the text "LABORATORY PRODUCTS" in a smaller, gray font.
# 510(k) Summary
- 1. Submitter:
Harvest Dental Products, LLC 905 Columbia Street Brea, California 92821
| Contact Person: | Colleen Boswell |
|-------------------|-----------------|
| Telephone Number: | (714) 674-7400 |
| Fax Number: | (714) 674-7402 |
Date Prepared: April 23, 2023
- 2. Device:
| Name of Device: | Harvest Printable Resin |
|-----------------|---------------------------------------------|
| Common Name: | Denture relining, repairing, rebasing resin |
| | Denture, plastic teeth |
| | Prescription mouthguard |
| Classification Name/ | Denture Relining, Repairing, or Rebasing Resin, per 21 |
|------------------------|---------------------------------------------------------|
| Device Classification/ | CFR 872.3760, II, EBI (Primary classification) |
| Device Class/ | Denture, Plastic, Teeth, per 21 CFR § 872.3590, II, ELM |
| Product Code: | Unclassified (Mouthguard), MQC |
- 3. Predicate Device:
| Primary Predicate: | <i>NextDent Denture/E-Denture</i> , Vertex-Dental B.V., K162572, Product Code EBI |
|----------------------|-----------------------------------------------------------------------------------|
| Secondary Predicate: | <i>E-Guard</i> , EnvisionTEC GmbH, K201173, Product Code MQC, EB |
#### 4. Device Description
Harvest Printable Resin is a light-cured resin developed for additive manufacturing (3D printing) of individual full denture bases, artificial teeth, bite splints and guards and try-in devices, i.e., denture base and teeth. After being utilized in a 3D stereolithographic printer to generate a dental device based on a solid model, the device is placed in a UV light curing unit for final polymerization.
Fabrication using Harvest Printable Resin requires an appropriate computer-aided Design and Manufacturing (CAD/CAM) system, a digital light processing (DLP) printer, and postprocessing light curing equipment. It has been optimized for use with Asiga (Max UV, Pro 4K65 and Pro 4K80) and Sprintray (Pro95 S and Pro55 S) printers and, therefore, may only be used in conjunction with these printers and their associated software systems. It is also only to be used in conjunction with Asiqa (Max UV, Pro 4K65 and Pro 4K80) and Sprintray (Procure and Procure 2) curing units for post-processing.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features the words "Harvest Dental" in a stacked format, with "Harvest" in gray and "Dental" in green. Above the text is a graphic of two green leaves, one larger than the other. Below the text, in smaller gray font, are the words "LABORATORY PRODUCTS".
The 3D printer is not included with the device.
## 5. Statement of Indications for Use:
Harvest Printable Resin is indicated for the fabrication of dental bases for full removable dentures, artificial teeth, dental bite splints and guards and try-in devices, i.e., denture base and teeth. The material is an alternative to traditional heat-curable and auto-polymerizing resins and is intended for professional dental work only. This material is intended to be used by dental lab technicians and approved by licensed practitioners before being provided to the patient.
#### 6. Summary of Technological Characteristics with the Predicate Device
The technological characteristics of the subject Harvest Printable Resin is similar to the predicate devices, NextDent Denture/E-Denture (K162572) and E-Guard (K201173). There are no substantial technical or functional differences between the Harvest Printable Resin and the predicate devices in terms of chemical composition, function and intended use. All are light-cured resins used in a 3D printer which prints the shape determined by a 3D stereolithographic drawing. See Table 1 below for technological characteristics and comparisons of the denture relining, rebasing resin, plastic denture teeth resin and prescription mouthguard and splints resin.
| | | | | Table 1: Comparison of Subject and Predicate Devices |
|--|--|--|--|-------------------------------------------------------|
|--|--|--|--|-------------------------------------------------------|
| Element | Harvest Printable Resin<br>(Proposed Device) | NextDent Denture/<br>E-Denture<br>(Primary Predicate) | E-Guard<br>(Secondary Predicate) | Comparison |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Harvest Dental Products, LLC | Vertex-Dental B.V. | EnvisionTEC GmbH | N/A |
| 510(k) | K222489 | K162572 | K201173 | N/A |
| Target Users | Dental laboratories | Healthcare<br>facility/hospital, dental<br>(technical) laboratory | Dental laboratories<br>and dental practices | Same |
| Common Name | Printable Resin | Printable Resin | Printable Resin | Same |
| Device Description | Harvest Printable Resin is<br>light-cured resin developed<br>for additive manufacturing<br>(3D printing) of individual full<br>denture bases, artificial<br>teeth, bite splints and guards<br>and try-in devices, i.e.,<br>denture base and teeth.<br>After being utilized in a 3D<br>stereolithographic printer to<br>generate a dental device<br>based on a solid model, the<br>device is placed in a UV light | NextDent Denture/ E-<br>Denture 3D-printing<br>material is a light-cured<br>resin indicated for the<br>manufacturing of denture<br>bases. The material is<br>used in a 3D printer,<br>which prints the shape<br>determined by a 3D<br>stereolithographic<br>drawing. After printing,<br>the printed product is<br>placed in a UV-light | The E-Guard system<br>combines a scanner<br>with design software,<br>the light-cured resin, a<br>3D printer and a<br>curing unit. These<br>components are used<br>together during the<br>manufacture of the<br>dental appliance<br>splint/bite guard. The<br>light-curing resin is<br>composed of acrylate/ | Same. The<br>resins are used<br>in the 3D<br>printing of<br>dental devices<br>including<br>denture bases,<br>bite splits and<br>guards.<br>Note: Artificial<br>teeth are 510(k)<br>exempt and the |
| Element | Harvest Printable Resin<br>(Proposed Device) | NextDent Denture/<br>E-Denture<br>(Primary Predicate) | E-Guard<br>(Secondary Predicate) | Comparison |
| | curing unit for final<br>polymerization.<br><br>Fabrication using <b>Harvest Printable Resin</b> requires an<br>appropriate computer-aided<br>Design and Manufacturing<br>(CAD/CAM) system, a digital<br>light processing (DLP)<br>printer, and post-processing<br>light curing equipment. It<br>has been optimized for use<br>with Asiga (Max UV, Pro<br>4K65 and Pro 4K80) and<br>Sprintray (Pro95 S and Pro55<br>S) printers and, therefore,<br>may only be used in<br>conjunction with these<br>printers and their associated<br>software systems. It is also<br>only to be used in<br>conjunction with Asiga (Max<br>UV, Pro 4K65 and Pro 4K80)<br>and Sprintray (Procure and<br>Procure 2) curing units for<br>post-processing.<br><br>The 3D printer is not<br>included with the device. | curing box for final<br>polymerization.<br><br>3D printer is not included<br>with the device. | methacrylate,<br>methacrylated<br>oligomers and<br>monomers, photo<br>initiators,<br>colorants/dyes and<br>absorbers. It is used<br>by dental laboratories<br>and dental practices<br>to make customized<br>bite splints, using the<br>3D-printer. The resin<br>is offered in lightproof<br>1kg PE bottles along<br>with a programmed<br>chip (referred to as<br>TAG), which is<br>required for use with<br>the 3D printer. The<br>TAG contains<br>information<br>identifying the resin<br>material, name and<br>amount. The E-Guard<br>resin is an alternative<br>material to heat-cured<br>and auto-polymerizing<br>resins. EnvisionTECs<br>Perfactory DDP<br>(Digital Dental Printer)<br>models designed for<br>use with the E-Guard<br>light cured resin are:<br>Perfactory DDP4 XL;<br>Perfactory Desktop<br>DDP plus; Perfactory<br>Desktop Pixera plus;<br>Perfactory DDP4 M;<br>Perfactory DDP Mini;<br>Perfactory DDP Mini<br>XL; Perfactory Vida 2;<br>Perfactory Vida 2 Hi-<br>RES; Perfactory Vida<br>HD cDLM; Perfactory<br>Vida cDLM; Perfactory<br>P4K 35, 62, 75, 90;<br>Perfactory | reason for no<br>predicate for this<br>indication. |
| Element | Harvest Printable Resin<br>(Proposed Device) | NextDent Denture/<br>E-Denture<br>(Primary Predicate) | E-Guard<br>(Secondary Predicate) | Comparison |
| Indications For Use | Harvest Printable Resin is<br>indicated for the fabrication<br>of dental bases for full<br>removable dentures,<br>artificial teeth, dental bite<br>splints and guards and try-in<br>devices, i.e., denture base<br>and teeth. The material is an<br>alternative to traditional<br>heat-curable and auto-<br>polymerizing resins and is<br>intended for professional<br>dental work only. This<br>material is intended to be<br>used by dental lab<br>technicians and approved by<br>licensed practitioners before<br>being provided to the<br>patient. | NextDent Denture/ E-<br>Denture is a light-cured<br>resin indicated for the<br>fabrication of denture<br>bases fabricated in dental<br>laboratories, including full<br>and partial removable<br>dentures. The material is<br>an alternative to<br>traditional heat cured and<br>auto polymerization<br>resins. NextDent<br>Denture/ E-Denture is<br>intended exclusively for<br>professional dental work.<br>Fabrication of denture<br>bases with NextDent<br>Denture/ E-Denture<br>requires a computer-<br>aided and manufacturing<br>(CAD/CAM) system that<br>includes the following:<br>scanner, design software,<br>additive printer, and<br>post-cure unit. NextDent<br>Denture/ E-Denture is<br>compatible with the<br>following CAD/CAM<br>systems components -<br>3Shape D900 scanner,<br>3Shape Dental-System<br>2016-Premium design<br>software, EnvisionTEC<br>DDP 4 printer using<br>Perfactory software,<br>Rapidshape D30 printer<br>using NetFabb software,<br>Micraft 125Y printer using<br>MiiUtility MiiController<br>software, 3D systems<br>Figure 4 printer using 3D<br>Sprint software, Roland<br>DG DWP-80S Printer<br>using Ver1.1 software, | MicroPlusXL;<br>Perfactory<br>EnvisionOne cDLM.<br><br>E-Guard is a light-<br>cured resin. It is a<br>polymer used to<br>create removable<br>structures for<br>therapeutic<br>restorations, i.e., bite<br>guards/splints and<br>occlusal night<br>guards/splints using<br>the Additive<br>Manufacturing<br>process. The resin in<br>combination with a<br>scanner, printer, and<br>curing unit make up<br>the system. | The resins are<br>indicated for use<br>in printing<br>denture bases<br>and/or as<br>custom bite<br>splints and<br>guards.<br><br>Note: Artificial<br>teeth are 510(k)<br>exempt and the<br>reason for no<br>predicate for this<br>indication. |
| Element | Harvest Printable Resin<br>(Proposed Device) | NextDent Denture/<br>E-Denture<br>(Primary Predicate) | E-Guard<br>(Secondary Predicate) | Comparison |
| | | and NextDent LC-3DPrint<br>Box post-curing unit. | | |
| Curing Mechanism | Light-cure | Light-cure | Light-cure | Same |
| Technique | Additive Manufacturing<br>(3D printing) | Additive Manufacturing<br>(3D printing) | Additive<br>Manufacturing<br>(3D printing) | Same |
| Basic Chemical<br>Composition | Methacrylates, urethane<br>methacrylates, photo<br>initiators and pigments | Dimethacrylate-based<br>resins with photo-<br>initiator and pigments | Acrylates,<br>methacrylates,<br>methacrylated<br>oligomers and<br>monomers, photo<br>initiators,<br>colorants/dyes and<br>absorbers. | Similar - All<br>methacrylate-<br>based light-cure<br>resins with<br>colorants and<br>photo initiators. |
| Ultimate Flexural<br>Strength<br>(ISO 20795-1) | 81.6 MPa<br>≥ 65 MPa | 84 MPa<br>≥ 65 MPa | N/A | Same - Passed<br>ISO 20795-1<br>requirements |
| Ultimate Flexural<br>Strength<br>(ISO 20795-2) | 81.6 MPa<br>≥ 50 MPa | N/A | 79.4 - 85.1 MPa<br>≥ 50 MPa | Same - Passed<br>ISO 20795-2<br>requirements |
| Flexural Modulus<br>(ISO 20795-1) | 2,339 MPa<br>≥ 2,000 MPa | 2,383 MPa<br>≥ 2,000 MPa | N/A | Same - Passed<br>ISO 20795-1<br>requirements |
| Flexural Modulus<br>(ISO 20795-2) | 2,339 MPa<br>≥ 1,500 MPa | N/A | 2050-2130 MPa<br>≥ 1,500 MPa | Same - Passed<br>ISO 20795-2<br>requirements |
| Water Sorption<br>(ISO 20795-1) | 24.3 µg/mm³<br>≤ 32 µg/mm³ | 28 µg/mm³<br>≤ 32 µg/mm³ | N/A | Same - Passed<br>ISO 20795-1<br>requirements |
| Water Sorption<br>(ISO 20795-2) | 24.3 µg/mm³<br>≤ 32 µg/mm³…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.