K222487 · Anika Therapeutics, Inc. · GDW · May 8, 2023 · General, Plastic Surgery
Device Facts
Record ID
K222487
Device Name
Anika Tissue Tack Fixation System
Applicant
Anika Therapeutics, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
May 8, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Anika Tissue Tack is indicated for the fixation of prosthetic or biologic material to soft tissues in various minimally invasive and open surgical procedures, such as rotator cuff repair.
Device Story
System consists of dart-shaped bioabsorbable tacks and a delivery instrument; used for fixation of prosthetic or biologic material to soft tissues during surgery (e.g., rotator cuff repair). Delivery instrument (stainless steel/high-temp plastic) inserts tacks into tissue. Single-use, sterile device. Surgeon operates instrument to secure materials; provides stable fixation during healing. Benefits include secure attachment of implants/biologics to soft tissue.
Clinical Evidence
No clinical testing was submitted or relied upon. Evidence consists of non-clinical bench testing, including mechanical retention strength (tensile/shear), degradation testing (mass loss, molar mass), insertion/removal evaluation, biocompatibility, and bacterial endotoxin testing.
Indicated for fixation of prosthetic or biologic material to soft tissues in minimally invasive and open surgical procedures, including rotator cuff repair.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 8, 2023
Anika Therapeutics, Inc. Shajunath Nirupama Sr. Regulatory Affairs Specialist 32 Wiggins Avenue Bedford, Massachusetts 01730
Re: K222487
Trade/Device Name: Anika Tissue Tack Fixation System Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: August 15, 2022 Received: August 17, 2022
Dear Shajunath Nirupama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows a digital signature. The name "Mark Trumbore" is written in a large font on the left side of the image. On the right side of the image, it says "Digitally signed by Mark Trumbore -S" and "Date: 2023.05.08 11:03:49 -04'00'".
Mark Trumbore Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222487
Device Name Anika Tissue Tack
Indications for Use (Describe)
The Anika Tissue Tack is indicated for the fixation of prosthetic or biologic material to soft tissues in various minimally invasive and open surgical procedures, such as rotator cuff repair.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the company "ANIKA". The logo consists of a blue circular graphic on the left, followed by the word "ANIKA" in blue capital letters. The circular graphic appears to be a stylized letter "A" or a similar abstract design. The overall design is clean and modern.
# 510(K) Summary
[As required by 21 CFR 807.92]
Date Prepared: April 7, 2022 510(k) Number: K222487
## Submitter Name
Anika Therapeutics, Inc. 32 Wiggins Avenue Bedford, MA 01730 Establishment No: 3007093114
## Contact Person
Shajunath Nirupama Sr. Regulatory Affairs Specialist Office: (781) 457-9230 Email: snirupama@anika.com
## General Information
## Table 1: General Information
| General Information of Subject Device | |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Trade Name | Anika Tissue Tack Fixation System |
| Common Name | Tissue Tack |
| 510(k) Submitter | Anika Therapeutics, Inc. |
| Class | II |
| Classification Name | Staple, Implantable |
| Regulation | 21 CFR 878.4750 |
| Product Code | GDW |
| Review Panel | General & Plastic Surgery |
| Primary Predicate & 510(K) | Rotation Medical Soft Tissue Staple-<br>K131637 |
| Reference Device & 510(K) | OptiFix™ AT Absorbable Fixation System<br>with Articulating Technology - K170278<br>TissueTak device - K203117 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "ANIKA" in blue, with a stylized blue logo to the left of the word. The logo appears to be a circular design with curved lines and a small dot inside. The font of the word "ANIKA" is a sans-serif typeface, and the letters are evenly spaced.
## Device Description
The Anika Tissue Tack Fixation System consists of dart shaped tacks and a delivery instrument. The tissue tacks are comprised of bioabsorbable 82/18 poly(lactic-co-glycolic acid) [PLGA] and dyed with D&C Violet #2. The tissue tacks are designed to provide stable fixation of a prosthetic material or biologics to soft tissues during the healing process and are used in conjunction with a delivery instrument from Anika. The delivery instruments are comprised of surgical grade stainless steels and high temperature plastics and are designed to deliver the tissue tacks. The tissue tacks and delivery instruments are provided sterile for single use only. The tacks are packaged in a caddy within a dual sterile seal configuration.
## Indication for Use
The Anika Tissue Tack Fixation System is indicated for the fixation of prosthetic or biologic material to soft tissues in various minimally invasive and open surgical procedures, such as rotator cuff repair.
## Substantial Equivalence Summary
Anika Therapeutics has demonstrated that for the purposes of the FDA's regulation of medical devices, the Anika Tissue Tack is substantially equivalent in indications for use, design principles and performance to the predicate device (Rotation Medical Soft Tissue Staple, K131637) and reference devices (OptiFix™ AT Absorbable Fixation System with Articulating Technology - K170278 and TissueTak device, K203117).
| Device Name | Anika Tissue Tack Fixation System<br>Subject Device | Rotation Medical Soft Tissue<br>Staple (RMST)<br>Predicate Device | Comparison |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| 510(k) No: | K222487 | K131637 | N/A |
| Product Code | GDW | GDW | Same |
| Indications for Use | The Anika Tissue Tack Fixation System is<br>Indicated for the fixation of prosthetic or<br>biologic material to soft tissues in various<br>minimally invasive and open surgical<br>procedures, such as rotator cuff repair. | The RMST Staple is intended for<br>fixation of prosthetic material to<br>soft tissues in various minimally<br>invasive and open surgical<br>procedures, such as the<br>management and protection of<br>tendon injuries in which there has<br>been no substantial loss of<br>tendon tissue. | Substantially<br>Equivalent<br>(see<br>Discussion<br>below) |
| Color | D&C Violet #2 | D&C Violet #2 | Same |
| Physical form | Dart shaped Tack | Staple shaped Tack | Same as<br>reference<br>device<br>(K17027) |
| Size | 6mm, 7mm, 8mm | 6.5mm | Substantially |
## Table 2: Substantial Equivalence Comparison
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ANIKA. The logo consists of a blue circular graphic on the left, followed by the word "ANIKA" in blue capital letters. The graphic appears to be a stylized letter "A" within the circle.
| | | | Equivalent<br>(see<br>Discussion<br>below) |
|---------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Material | bioabsorbable Purasorb PLG 82/18 (L-<br>lactide/Glycolide Polymer) dyed with D&C<br>Violet No. 2 | bioabsorbable Purasorb PLG 82/18 (L-lactide/Glycolide<br>Polymer) dyed with D&C Violet<br>No. 2 | Same |
| Body Contact | Implant (>30 days) | Implant (>30 days) | Same |
| Sterilization | Gamma Irradiation | Gamma Irradiation | Same |
| Packaging | Tacks are pre-loaded into a caddy that is<br>inserted into sealed Tyvek and foil<br>pouches and inserted into a carton. | Tacks are pre-loaded into a<br>caddy that is inserted into sealed<br>Tyvek and foil pouches and<br>inserted into a carton. | Same |
#### Discussion
Indications for Use Statement:
The subject device and predicate device are used similarly to aid in the management of the tendon injuries, such as rotator cuff repair, by affixing a prosthetic material to tissue. The subject device can also affix biologic materials to tissues to aid in the management of the tendon injuries. The subject device has been mechanically tested for this additional indication and no concerns related to the safety and effectiveness were raised. Additionally, the reference device, TissueTak by Via Surgical Ltd. cleared under K203117 is also intended for fixation of prosthetic or biologic material to soft tissues.
Technological Characteristics:
The subject device ranges in size from 6-8mm in length to account for the variations in tendon thickness.
However, subject device's 6-8mm range reflects a nomenclature difference in comparison with predicate and reference devices, as Anika is using the overall length of the device in its naming convention. The longest "depth of purchase" of 6.5mm (for 8mm total length) is the same as the 6.5mm "depth of purchase" of the predicate device and reference device (K170278). Mechanical evaluations have demonstrated no significant difference in pull out or shear strength for varying shaft lengths. Therefore, the device length as described does not affect device safety and effectiveness compared to the predicate and reference devices.
In degradation testing results show that both the subject device and predicate device are identical in degradation time and are substantially equivalent. The differences in the shape and size of subject device versus the predicate does not raise any concerns in comparison to the predicate device as demonstrated by mechanical evaluation.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "ANIKA" in blue, with a blue circular logo to the left of the word. The logo appears to be a stylized letter "A" inside of a circle. The text is in a sans-serif font and is evenly spaced. The background is white.
## Performance Testing Summary
The following non-clinical tests and/or analysis were performed for the subject device to demonstrate the safety and effectiveness along with substantial equivalence to the predicate device:
- Mechanical retention strength, tensile & shear
- Degradation testing including mass loss, molar mass and mechanical evaluation.
- . Insertion and removal evaluation
- . Biocompatibility Testing
- . Bacterial endotoxin limit test
## Clinical Testing Summary
No clinical testing was submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence for the Anika Tissue Tack Fixation System.
## Conclusion
All similarities and differences described above were evaluated. Based on the justifications described in the subsequent sections of this submission, none of the differences raised significant concern regarding the safety or efficacy of the devices. Therefore, Anika Therapeutics has determined that the subject device, Anika Tissue Tack Fixation System is substantially equivalent to the predicate device and is safe and effective for the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.