Surgical Face Mask

K222483 · Hubei Woozon Healthcare Co., Ltd. · FXX · Jan 4, 2023 · General, Plastic Surgery

Device Facts

Record IDK222483
Device NameSurgical Face Mask
ApplicantHubei Woozon Healthcare Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateJan 4, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Story

Single-use, three-layer, flat-pleated surgical face mask; features ear loops and nose piece. Constructed with inner/outer polypropylene layers and middle melt-blown polypropylene filter layer. Nose piece made of galvanized iron wire; ear loops made of polyester silk and polyurethane filament. Provided non-sterile. Used in clinical settings by healthcare personnel and patients to provide barrier protection against microorganisms, body fluids, and particulates. Device functions as physical barrier; performance verified via bench testing for filtration efficiency, fluid resistance, and breathability.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100, F1862 (synthetic blood penetration), F2101 (BFE), F2299 (particulate filtration), and 16 CFR 1610 (flammability). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.

Technological Characteristics

Three-layer construction: polypropylene inner/outer layers, melt-blown polypropylene filter. Galvanized iron wire nose piece. Polyester silk/polyurethane ear loops. Non-sterile. ASTM F2100 Level 1, 2, 3 compliant. Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization).

Indications for Use

Indicated for use by patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 4, 2023 Hubei Woozon Healthcare Co.,Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161 Lujiazui East Rd., Pudong Shanghai, Shanghai 200120 China Re: K222483 Trade/Device Name: Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 22, 2022 Received: December 5, 2022 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Brent Showalter -S Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K222483 Device Name Surgical Face Mask #### Indications for Use (Describe) The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| | | | | | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92. #### 1.0 Submitter's Information Name: Hubei Woozon Healthcare Co.,Ltd. Address: Nongfeng Road, Nonwoven Fabric Industrial Park, Wangzhou Avenue, Pengchang Town, Xiantao City, Hubei, China. Tel: +86-728-3280667 Contact: Li Chunlin Date of Preparation: Aug.8, 2022 #### Designated Submission Correspondent Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong, Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn ### 2.0 Device Information | Trade name: | Surgical Face Mask | |----------------------|--------------------| | Common name: | Surgical Face Mask | | Classification name: | Surgical Face Mask | | Model: | Ear-loop Type | #### 3.0 Classification | Production code: | FXX | |--------------------|------------------| | Regulation number: | 21CFR 878.4040 | | Classification: | Class II | | Panel: | Surgical Apparel | #### 4.0 Predicate Device Information | Manufacturer: | Anhui Tiankang Medical Technology Co., Ltd. | |----------------|---------------------------------------------| | Device: | Surgical Face Mask | | 510(k) number: | K212368 | {4}------------------------------------------------ # 5.0 Device Description The Surgical Face Mask is single use, three-layer, flat-pleated style with ear loops and nose piece. The Surgical Face Mask is manufactured with three layers, the inner and outer layers are made of polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Galvanized iron wire. The Surgical Face Mask will be provided in blue. The Surgical Face Mask is sold as non-sterile and are intended to be single use, disposable devices. ### 6.0 Indication for Use Statement The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | Table 1 General Comparison | | | | |--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Item | Subject Device | Predicate Device<br>K212368 | Remark | | Product Name | Surgical Face Mask | Surgical Face Mask | -- | | Product Code | FXX | FXX | Same | | Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | | Class | II | II | Same | | Intended Use&<br>Indications for use | The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use ininfection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use ininfection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | Same | | Design features | Ear Loops, | Ear loop, Tie-On. | Different | # 7.0 Comparison to the Predicate Device | Table 1 General Comparison | |----------------------------| |----------------------------| {5}------------------------------------------------ | Mask Styles | | | 3 layers | 3 layers | Analysis 1 | |------------------|--------------------------------------|--------------------------------------------|-------------------------------------------------------|-----------------------------------------------------|------------| | Material | Outer facing layer | Polypropylene | Spun-bond polypropylene | Similar<br>Analysis 1 | | | | Middle layer | Melt-blown cloth<br>(polypropylene) | Melt blown<br>polypropylene filter | Same | | | | Inner Facing layer | Polypropylene | Spun-bond polypropylene | Similar | | | | Nose piece | Galvanized iron wire | PP coated steel wire | Different<br>Analysis 1 | | | | Ear loops | Plolyester silk<br>& Polyurethane filament | Ear loops: Nylon and<br>spandex;<br>Ties: PP nonwoven | Different<br>Analysis 1 | | | Color | | Blue | Blue | Same | | | Dimension | | Length: 175±4mm<br>Width: 95±4mm | Length: 175±5mm<br>Width: 95±2.85mm | Similar<br>Analysis 2 | | | OTC use | | | Yes | Yes | Same | | Single Use | | | Yes | Yes | Same | | Sterility | | | Non-sterile | Non-sterile | Same | | ASTM F2100 Level | | | Level 1<br>Level 2<br>Level 3 | Level 1<br>Level 2<br>Level 3 | Same | | Biocompatibility | Cytotoxicity<br>ISO10993-5 | | Non-cytotoxic under the<br>conditions of the study | Non-cytotoxic under the<br>conditions of the study | Same | | | Skin Irritation<br>ISO10993-10 | | Non-irritating under the<br>conditions of the study | Non-irritating under the<br>conditions of the study | Same | | | Skin<br>Sensitization<br>ISO10993-10 | | Non-sensitizer under the<br>conditions of the study | Non-sensitizer under the<br>conditions of the study | Same | Analysis 1: The two devices have some difference in design features and materials, product materials safety is proved by its biocompatibility, and the difference does not raise additional questions for safety and effectiveness of device. Analysis 2: The two devices share same dimensions otherwise the tolerance is different, the little deviation in tolerance does not raise additional questions for safety and effectiveness of device. # 8.0 Non-Clinical Test Conclusion The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Medical face masks - Premarket Notification [510(k)] Submission issued on March 5, 2004. {6}------------------------------------------------ ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at A Known Velocity); ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus; ASTM F2299, stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres; 16 CFR 1610, Standard for the Flammability of clothing textiles; | Item | Purpose | Acceptance Criteria | Results | |----------------------------------------------------------|-------------------------------------------------------|---------------------------------------------------------------------|-----------------------------------------------| | Synthetic<br>Blood<br>Penetration<br>ASTM<br>F1862 | Demonstrate<br>resistance<br>to liquid<br>penetration | 29 samples out of 32<br>pass (AQL 4%)<br>Level 1 pass at 80mmHg | Pass<br>32 out of 32 pass at 80 mmHg, 3 lots | | | | 29 samples out of 32<br>pass (AQL 4%)<br>Level 2 pass at<br>120mmHg | Pass<br>32 out of 32 pass at 120 mmHg, 3 lots | | | | 29 samples out of 32<br>pass (AQL 4%)<br>Level 3 pass at<br>160mmHg | Pass<br>32 out of 32 pass at 160 mmHg, 3 lots | | Particulate<br>Filtration<br>Efficiency<br>ASTM<br>F2299 | Demonstrate<br>particulate<br>filtration | Level 1 pass at $≥95%$ | Pass<br>Average 98.24% | | | | Level 2 pass at $≥98%$ | Pass<br>Average 98.75% | | | | Level 3 pass at $≥98%$ | Pass<br>Average 99.67% | | Bacterial<br>Filtration<br>Efficiency<br>ASTM<br>F2101 | Demonstrate<br>bacterial<br>filtration | Level 1 pass at $≥95%$ | Pass<br>Average 99.18% | | | | Level 2 pass at $≥98%$ | Pass<br>Average 99.20% | | | | Level 3 pass at $≥98%$ | Pass<br>Average 99.21% | | Differential<br>Pressure | Demonstrate<br>breathability | Level 1 pass at $≤ 5.0$ mmH2O/cm² | Pass<br>Average 3.75 mmH2O/cm² | #### Table 2 - Performance Testing {7}------------------------------------------------ | (Delta P)<br>EN 14683<br>Annex C | Level 2 pass at ≤ 6.0<br>mmH2O/cm² | Pass<br>Average 4.25 mmH2O/cm² | |----------------------------------|------------------------------------|--------------------------------| | Flammability<br>16 CFR 1610 | Demonstrate flame resistance | Class I<br>Pass | # Table 3 - Biocompatibility Testing | Item | Purpose | Subject Device | Results | |--------------------------------------|-------------------------------------------|------------------------------------------------------------------------|---------| | Cytotoxicity<br>ISO 10993-5 | Demonstrate cytotoxic<br>biocompatibility | Under the conditions of the<br>study, the device is<br>non-cytotoxic. | Pass | | Skin Irritation<br>ISO10993-10 | Demonstrate<br>non irritability | Under the conditions of the<br>study, the device is<br>non-irritating. | Pass | | Skin<br>Sensitization<br>ISO10993-10 | Demonstrate<br>non sensitization | Under the conditions of the<br>study, the device is<br>non-sensitizing | Pass | # 9.0 _Clinical Test Conclusion No clinical study is included in this submission. #### 10.0 Conclusion The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device in K212368.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...