← Product Code [JXG](/productcode/JXG) · K222471

# IRRAflow Active Fluid Exchange System (AFES) (K222471)

_Irras USA · JXG · Dec 8, 2022 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222471

## Device Facts

- **Applicant:** Irras USA
- **Product Code:** [JXG](/productcode/JXG.md)
- **Decision Date:** Dec 8, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.5550
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The use of IRRAflow Active Fluid Exchange System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.

## Device Story

The IRRAflow Active Fluid Exchange System is an intracranial pressure (ICP) monitoring and drainage system. It utilizes a catheter inserted into the patient's head to monitor ICP via integrated transducers and drain cerebrospinal fluid (CSF). The system includes a control unit with a peristaltic pump and pinch valve mechanism that acts on a sterile, disposable tube set to regulate gravity-driven drainage. The control unit features a user interface for setting drainage parameters, managing alarms, and visualizing ICP data. It also supports bolus injections to clear catheter clogs. The system is used in clinical settings by healthcare providers. Output includes real-time ICP monitoring and controlled fluid drainage. The device benefits patients by providing automated, adjustable ICP management and reducing the risk of catheter occlusion through active flushing capabilities. Version 4.0 adds bedside monitoring integration, increased alarm volume, and enhanced data visualization.

## Clinical Evidence

Bench testing only. No clinical data presented. Verification and validation testing included inspection, BSM (bedside monitoring) algorithmic requirements, electrical requirements, mechanical performance, life cycle reliability, and packaging verification. All tests passed.

## Technological Characteristics

Class II device. Components: Control Unit (v4.0), sterile disposable tube set, catheter, drainage bag. Sensing: Integrated pressure transducers. Actuation: Stepper-motor controlled peristaltic pump and pinch valve for gravity-driven flow regulation. Connectivity: Bedside monitoring export, Bluetooth hardware (for future use). Power: External power supply, battery backup. Software: Version 4.0 with enhanced UI, data visualization, and auto-priming. Sterilization: Disposable components provided sterile.

## Regulatory Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

## Predicate Devices

- IRRAflow CNS System ([K200807](/device/K200807.md))
- IRRAflow CNS System ([K192289](/device/K192289.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

December 8, 2022

IRRAS USA, Inc. Jeanne Warner VP, Clinical & Regulatory Affairs 11975 El Camino Real, Suite 304 San Diego, California 92130

Re: K222471

Trade/Device Name: IRRAflow Active Fluid Exchange System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG, GWM Dated: November 4, 2022 Received: November 7, 2022

Dear Jeanne Warner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Digitally signed by Adam D. Pierce -S Pierce -> 07:36:05 -05'00'

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number K222471

Device Name IRRAflow Active Fluid Exchange System

Indications for Use (Describe)

The use of IRRAflow Active Fluid Exchange System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

Applicant:

IRRAS USA, Inc. 11975 El Camino Real, Suite 304 San Diego, CA 92130 US.A 858-220-3761 Phone Number: Fax number: 866-575-1002 FDA Registration#:

Primary Contact Personnel:

Jeanne S. Warner, RN MS VP, Clinical & Regulatory Affairs Phone Number: 269-270-2189 866-575-1002 Fax Number: Email: jeanne.warner@irras.com

Secondary Contact Personnel:

| Fernando Sanchez<br>Sr. Director, Quality |                                                                                  |
|-------------------------------------------|----------------------------------------------------------------------------------|
| Phone Number:                             | 858-212-6660                                                                     |
| Email:                                    | fernando.sanchez@irras.com                                                       |
| Date prepared:                            | December 8, 2022                                                                 |
| 510(k) Number:                            | K222471                                                                          |
| Trade name:                               | IRRAflow® Active Fluid Exchange System<br>(AFES, formerly branded as CNS System) |
| Common Name:                              | CSF Drainage System with ventricular catheter                                    |

{4}------------------------------------------------

| Primary Classification: |                                                  |
|-------------------------|--------------------------------------------------|
| Name:                   | Central Nervous System Fluid Shunt and Component |
| Product Code:           | JXG                                              |
| Regulation:             | 21 CFR 882.5550                                  |

Primary Predicate Device: IRRAflow® CNS System, K200807 Additional Predicate Device: IRRAflow® CNS System, K192289

# 1.0 DEVICE DESCRIPTION - IRRAflow® Active Fluid Exchange System

The IRRA/low" Active Fluid Exchange System (formerly branded as the CNS System) is an intracranial pressure (ICP) monitoring and drainage system, substantially equivalent to the predicate 510(k) K200807. Changes described herein were evaluated using design controls. The drainage flow of cerebrospinal fluid (CSF) into the IRRAflow Catheter is uni-directional and gravity-driven; there is no recirculation of the CSF. A parallel line from the saline infusion bag is used in case clearance at the tip of the catheter is required.

The IRRA/low Tube Set has a cassette that clicks on to the IRRA/low Control Unit and aligns the tubing against a peristaltic pump and pinch valve. A drainage bag is attached to the Control Unit, using the Laser Leveler for defining the height of the bag relative to the catheter's tip position in the patient's head. This positioning is used for controlling the speed of drainage. The tubing and catheter can be disconnected and connected by standard Luer-Lock connectors. Settings can be changed via the user interface on the Control Unit.

The default mode provides drainage and measuring ICP, allowing bolus injections when indicated. The bolus injections allow the catheter to be flushed when it becomes clogged. CSF or intracranial fluid samples can be taken from the aspiration port.

# 2.0 INDICATIONS FOR USE

The use of the IRRAflow Active Fluid Exchange System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.

{5}------------------------------------------------

# 3.0 SUBSTANTIAL EQUIVALENCE

The IRRAflow® AFES (formerly branded as CNS System) intended use, technological characteristics and principles of operation are the substantially equivalent to the predicate device described in K200807. Changes present in the Control Unit 4.0 that would have potential to affect the technological characteristics or principles of operation of the IRRAflow® System were evaluated by design controls. Earlier predicate devices described in K192289 and K200807 are shown for comparison in Table 2 below.

Comparison of these Monitoring Systems will show that the only difference in the device presented in this Special 510(k) are the enhanced and new features present in the Control Unit, designated as version 4.0. The currently cleared and marketed Control Unit is designated as version 3.0. Data presented definitively support the conclusion that Control Unit version 4.0 continues to be as safe and effective as the predicate device.

{6}------------------------------------------------

Therefore, this Special 510(k) captures detailed descriptions of changes to the Control Unit which is part of the overall IRRAflow® AFES, classified by FDA as a Class II medical device. IRRAS is submitting this Special 510(k) as substantially equivalent to the IRRA/low® CNS System device described in K192289 and K200807.

| Attributes                                                                                                                                                            | PREDICATE<br>IRRAflow® CNS<br>System (K192289)<br>(JXG), (GWM)                                                                                                                                                                                                                            | PREDICATE<br>IRRAflow® CNS<br>System (K200807)<br>(JXG), (GWM)                                                                                                                                                                                                                                     | IRRAflow® AFES System<br>Special 510(k)<br>K222471<br>(JXG), (GWM)<br>Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Product<br>Code                                                                                                                                               | JXG                                                                                                                                                                                                                                                                                       | JXG                                                                                                                                                                                                                                                                                                | JXG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Primary Regulation<br>Number:                                                                                                                                         | 21 CFR 882.5550                                                                                                                                                                                                                                                                           | 21 CFR 882.5550                                                                                                                                                                                                                                                                                    | 21 CFR 882.5550                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Secondary<br>Product Code                                                                                                                                             | GWM                                                                                                                                                                                                                                                                                       | GWM                                                                                                                                                                                                                                                                                                | GWM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Secondary<br>Regulation Number:                                                                                                                                       | 21 CFR 882.1620                                                                                                                                                                                                                                                                           | 21 CFR 882.1620                                                                                                                                                                                                                                                                                    | 21 CFR 882.1620                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Indications for Use                                                                                                                                                   | The use of IRRAflow®<br>CNS is indicated when<br>intracranial pressure<br>monitoring is required<br>and for externally<br>draining intracranial fluid<br>as a means of reducing<br>intracranial pressure in<br>patients where an external<br>drainage and monitoring<br>system is needed. | Same<br>The use of IRRAflow®<br>CNS is indicated when<br>intracranial pressure<br>monitoring is required and<br>for externally draining<br>intracranial fluid  as a<br>means of reducing<br>intracranial pressure in<br>patients where an external<br>drainage and monitoring<br>system is needed. | Same<br>The use of IRRAflow® AFES<br>is indicated when intracranial<br>pressure monitoring is<br>required and for externally<br>draining intracranial fluid as a<br>means of reducing intracranial<br>pressure in patients where an<br>external drainage and<br>monitoring<br>system is needed. **                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Attributes                                                                                                                                                            | PREDICATE<br>IRRAflow® CNS System<br>(K192289)<br>(JXG), (GWM)                                                                                                                                                                                                                            | PREDICATE<br>IRRAflow® CNS System<br>(K200807)<br>(JXG), (GWM)                                                                                                                                                                                                                                     | IRRAflow® AFES System<br>Special 510(k) K222471<br>(JXG), (GWM)<br>Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Injection/ CSF<br>Sampling<br>Ports                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Unidirectional<br>Flow of Drained Fluid                                                                                                                               | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Fluid Injection<br>Capability                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Attaches to separate,<br>commercially<br>available EVD<br>Catheter                                                                                                    | The IRRAflow® system<br>attaches to IRRAflow®<br>Catheter which is an EVD<br>Catheter part of the complete<br>system.                                                                                                                                                                     | Same, the IRRAflow®<br>system attaches to<br>IRRAflow® Catheter which<br>is an EVD Catheter part of<br>the complete system.                                                                                                                                                                        | Same, the IRRAflow®<br>system attaches to<br>IRRAflow® Catheter which<br>is an EVD Catheter part of<br>the complete system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Sterile Disposable<br>tubing set                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| CSF Drainage Bag                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Gravity drainage of<br>CSF                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Method to control<br>gravity drainage of<br>CSF                                                                                                                       | Automated adjustment based<br>on user settings via a<br>stepper-motor controlled,<br>tube pinching mechanism to<br>either compress or release<br>the compliant drainage<br>tubing contained within the<br>sterile, disposable Cartridge.                                                  | Same, automated adjustment<br>based on user settings via a<br>stepper-motor controlled,<br>tube pinching mechanism to<br>either compress or release<br>the compliant drainage<br>tubing contained within the<br>sterile, disposable Cartridge.                                                     | Same, automated<br>adjustment based on user<br>settings via a stepper-motor<br>controlled, tube pinching<br>mechanism to either<br>compress or release the<br>compliant drainage tubing<br>contained within the sterile,<br>disposable Cartridge.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Pressure Transducer<br>for ICP Measurement                                                                                                                            | The IRRAflow®<br>system integrates transducers<br>into its design for<br>measurement and visual<br>display of ICP                                                                                                                                                                         | Same, the IRRAflow®<br>system integrates transducers<br>into its design for<br>measurement and visual<br>display of ICP                                                                                                                                                                            | Same, the IRRAflow®<br>system integrates<br>transducers into its design<br>for measurement and<br>visual display of ICP.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Attributes                                                                                                                                                            | PREDICATE<br>IRRAflow® CNS System<br>(K192289)<br>(JXG), (GWM)                                                                                                                                                                                                                            | PREDICATE<br>IRRAflow® CNS System (K200807)<br>(JXG), (GWM)                                                                                                                                                                                                                                        | IRRAflow® AFES System<br>Special 510(k) (GWM, JXG)<br>Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Software-based,<br>Powered Console for<br>User Interface, User<br>Settings and Alarm<br>Adjustments, Data<br>Storage and Display,<br>and Alarms for ICP<br>monitoring | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                | Substantially equivalent to<br>predicate, with addition of:<br>• Bedside monitoring ability<br>(BSMA)to export ICP data<br>stream compatible with<br>marketed bedside monitors,<br>with cables<br>• Increased volume for<br>system alarms<br>• Power system changes to<br>extend shelf life and improve<br>power shut down sequencing<br>• Allowance for 'off the<br>shelf' external power<br>supply<br>• Manufacturability<br>related changes to<br>board layout<br>• Bluetooth hardware for<br>future application<br>development<br>• User software enhancements,<br>including: case-specific<br>irrigation and drainage<br>settings, additional data<br>visualization onscreen<br>(trending, axis and unit<br>adjustments), auto priming<br>cassette functionality,<br>pressure accuracy<br>improvement<br>• Drainage collection system<br>mounting that allows<br>enhanced fluid drainage<br>collection |
| Measured Pressure<br>Range                                                                                                                                            | -80 mmHg to +100 mmHg                                                                                                                                                                                                                                                                     | Same, -80 mmHg to<br>+100 mmHg                                                                                                                                                                                                                                                                     | Substantially Equivalent<br>-100 to +300 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Displayed ICP                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Battery Back-up                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

# Table 2: Substantial Equivalence (SE)Comparison

{7}------------------------------------------------

{8}------------------------------------------------

# K222471, Page 6 of 8

** Equivalent-change from CNS System to Active Fluid Exchange System is a branding change only

{9}------------------------------------------------

A review of Table 2 shows that there is no change to the overall IRRA/low® system. Changes made only to the Control Unit software required additional testing, including verification and validation of new and enhanced functionality. Changes present in the Control Unit 4.0 that would have potential to affect the technological characteristics or principles of operation of the IRRAflow® System were evaluated by design controls. Any new safety and effectiveness issues related to the Control Unit modifications are addressed in the FMEA and Risk Analysis.

Because changes were made only to the Control Unit software (with minor hardware changes to accommodate hanger for drainage bag and Scale), with the required additional testing including verification and validation new and enhanced functionality, the continued safe and effective use of Control Unit 4.0 was confirmed. Changes present in the Control Unit 4.0 that would have potential to affect the technological characteristics or principles of operation of the IRRAflow® System were evaluated by design controls. Any new safety and effectiveness issues related to the Control Unit modifications are addressed in the Control Unit FMEA. No changes were made to the EVD catheter or the remaining components of the IRRA/low® system.

### 4.0 SUMMARY OF DESIGN CONTROLS

Procedural design controls were followed, and all design elements were satisfied.

### 5.0 SUMMARY OF VERIFICATION & VALIDATION

Additional verification and validation performance testing was done on the IRRAflow AFES Control Unit 4.0. Any new safety and effectiveness issues related to software and other described changes were addressed in the FMEA and Risk Analysis.

The IRRAflow® AFES is substantially equivalent to the system described in 510(k) K200807. Enhanced and new Control Unit features are described herein. These changes do not adversely affect the safety and effectiveness of the IRRA/low® Control Unit. Any new risks to the overall system were address in the FMEA and Risk Analysis. The risk was evaluated separately, as well as, in conjunction with the system components of the IRRAflow® AFES. Since any new concerns or risks identified were mitigated by the actual proper use and function of Control Unit 4.0, it is appropriate to conclude that the IRRAflow® AFES is Substantially Equivalent to the predicate device (K200807).

A summary of the testing done at IRRAS is presented in the Table below. In Table 4 below, IRRAS has identified the various well-established and recognized Standards for evaluation of the performance of the Control Unit 4.0 separately. Results are provided in a summary format. Complete test reports are available upon request.

{10}------------------------------------------------

| Test                                 | Test Method Summary                                                                                                                                                                                            | Results    |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Verification by Inspection           | No physical testing is performed for the<br>associated requirements. Coverage was<br>demonstrated by analysis and presented in<br>a final trace matrix.                                                        | Acceptable |
| BSM Requirements Verification        | These testes are algorithmic in nature, with<br>demonstration of passing results one one<br>sample, with repeated tests as defined in<br>the associated protocol, are valid for full<br>requirement coverage.  | Pass       |
| Electrical Requirements Verification |                                                                                                                                                                                                                |            |
| Mechanical Performance Verification  | As defined in the associated protocol, the<br>requirement pertains to the hardware<br>component itself (i.e. pole clamp, drainage<br>bag holder clip, etc.) and not on its interface<br>with the control unit. | Pass       |
| Life Cycle Verification Test         | All testing was completed on capital<br>equipment, with multiple test repeats (as<br>required) were used to demonstrate lifecycle<br>reliability.                                                              | Pass       |
| Packaging Verification               | No risk to user since device was found to<br>work based on device functionality checked<br>upon arrival after simulated transit.                                                                               | Pass       |

#### Conclusions

The IRRAflow® AFES (formerly branded as the CNS system) is Substantially Equivalent to the predicate device (K200807). The IRRAflow® AFES has the same Indications for Use. Changes present in the Control Unit 4.0 that would have potential to affect the technological characteristics or principles of operation of the IRRAflow® System were evaluated by design controls.

---

**Source:** [https://fda-staging.innolitics.com/device/K222471](https://fda-staging.innolitics.com/device/K222471)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
