K222463 · Ultromics Limited · QUO · Nov 23, 2022 · Cardiovascular
Device Facts
Record ID
K222463
Device Name
EchoGo Heart Failure
Applicant
Ultromics Limited
Product Code
QUO · Cardiovascular
Decision Date
Nov 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2200
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K222463 · Nov 23, 2022
EchoGo Heart Failure
Ultromics Limited
Retrospective clinical data cohort
The retrospective clinical data was used to validate the device's performance, specifically to assess sensitivity, specificity, repeatability, and reproducibility against ground truth classifications.
Retrospective clinical performance study; Retrospective case-control study
Adult patients undergoing routine functional cardiovascular assessment using echocardiography; Sample Size: 1,285 patients (639 controls and 646 cases); Number of Sites: 8 clinical sites
Not applicable for this study
Sensitivity and specificity for HFpEF detection; repeatability and reproducibility of device output
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Heart failure with preserved ejection fraction (HFpEF)
Convolutional neural network
—
Sensitivity of 87.8% (95% CI: 85.0, 90.3%) and specificity of 82.0% (95% CI: 78.6, 85.0%)
—
—
Retrospective case-control study of 1,285 patients (639 controls and 646 cases) across 8 independent clinical sites in the USA.
—
Indications for Use
EchoGo Heart Failure 1.0 is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing routine functional cardiovascular assessment using echocardiography. When utilised by an interpreting clinician, this device provides information that may be useful in detecting heart failure with preserved ejection fraction (HFpEF). EchoGo Heart Failure 1.0 is indicated in adult populations over 25 years of age. Patient management decisions should not be made solely on the results of the EchoGo Heart Failure 1.0 analysis. EchoGo Heart Failure 1.0 takes as input an apical 4-chamber view of the heart that has been captured and assessed to have an ejection fraction ≥50%.
Device Story
Software-only device; processes DICOM echocardiogram files (apical 4-chamber view, EF ≥50%, no contrast, full cardiac cycle). Uses convolutional neural network (AI) to classify presence/absence of HFpEF. Operates on Ultromics or third-party infrastructure; fully automated; no user interface. Produces report for interpreting clinician. Used as adjunctive diagnostic aid in clinical pathways; results must be confirmed by other investigations. Benefits patient by providing automated decision support for HFpEF detection; assists clinicians in diagnostic decision-making.
Clinical Evidence
Retrospective case-control study (N=1,285; 639 controls, 646 cases) across 8 US sites. Sensitivity 87.8% (95% CI: 85.0-90.3%); specificity 82.0% (95% CI: 78.6-85.0%). p<0.001 for both. Repeatability 100%. Reproducibility: 86.7% positive agreement, 76.9% negative agreement. 7.3% non-diagnostic rate. Sub-group analysis for SGLT2i patients (n=105) showed 80.5% sensitivity and 87.3% specificity.
Indicated for adult patients >25 years undergoing routine functional cardiovascular assessment via echocardiography to aid in detecting heart failure with preserved ejection fraction (HFpEF). Requires apical 4-chamber view with ejection fraction ≥50%. Not for sole diagnostic use.
Regulatory Classification
Identification
The adjunctive cardiovascular status indicator is a prescription device based on sensor technology for the measurement of a physical parameter(s). This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Software description, verification, and validation based on comprehensive hazard analysis must be provided, including:
(i) Full characterization of technical parameters of the software, including any proprietary algorithm(s);
(ii) Description of the expected impact of all applicable sensor acquisition hardware characteristics on performance and any associated hardware specifications;
(iii) Specification of acceptable incoming sensor data quality control measures; and
(iv) Mitigation of impact of user error or failure of any subsystem components (signal detection and analysis, data display, and storage) on accuracy of patient reports.
(2) Scientific justification for the validity of the status indicator algorithm(s) must be provided. Verification of algorithm calculations and validation testing of the algorithm using a data set separate from the training data must demonstrate the validity of modeling.
(3) Usability assessment must be provided to demonstrate that risk of misinterpretation of the status indicator is appropriately mitigated.
(4) Clinical data must be provided in support of the intended use and include the following:
(i) Output measure(s) must be compared to an acceptable reference method to demonstrate that the output measure(s) represent(s) the predictive measure(s) that the device provides in an accurate and reproducible manner;
(ii) The data set must be representative of the intended use population for the device. Any selection criteria or limitations of the samples must be fully described and justified;
(iii) Agreement of the measure(s) with the reference measure(s) must be assessed across the full measurement range; and
(iv) Data must be provided within the clinical validation study or using equivalent datasets to demonstrate the consistency of the output and be representative of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment.
(5) Labeling must include the following:
(i) The type of sensor data used, including specification of compatible sensors for data acquisition;
(ii) A description of what the device measures and outputs to the user;
(iii) Warnings identifying sensor reading acquisition factors that may impact measurement results;
(iv) Guidance for interpretation of the measurements, including warning(s) specifying adjunctive use of the measurements;
(v) Key assumptions made in the calculation and determination of measurements;
(vi) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance; and
(vii) A detailed description of the patients studied in the clinical validation (
*e.g.,* age, gender, race/ethnicity, clinical stability) as well as procedural details of the clinical study.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 23, 2022
Ultromics Limited Jaco Jacobs Chief Regulatory and Compliance Officer 4630 Kingsgate Cascade Way, Oxford Business Park Oxford, OX4 2SU United Kingdom
Re: K222463
Trade/Device Name: EchoGo Heart Failure Regulation Number: 21 CFR 870.2200 Regulation Name: Adjunctive Cardiovascular Status Indicator Regulatory Class: Class II Product Code: OUO Dated: October 24, 2022 Received: October 25, 2022
Dear Jaco Jacobs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222463
Device Name EchoGo Heart Failure
### Indications for Use (Describe)
EchoGo Heart Failure 1.0 is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing routine functional cardiovascular assessment using echocardiography. When utilised by an interpreting clinician, this device provides information that may be useful in detecting heart failure with preserved ejection fraction (HFpEF).
EchoGo Heart Failure 1.0 is indicated in adult populations over 25 years of age. Patient management decisions should not be made solely on the results of the EchoGo Heart Failure 1.0 analysis.
EchoGo Heart Failure 1.0 takes as input an apical 4-chamber view of the heart that has been captured and assessed to have an ejection fraction ≥50%.
| | Type of Use (Select one or both, as applicable) | |
|--|-------------------------------------------------|--|
| | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular graphic on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular graphic is a gradient of blue and green, creating a swirling effect.
#### 1 Submitter
| Company | Ultromics Limited<br>4630 Kingsgate Cascade Way, Oxford Business Park South, Oxford,<br>Oxfordshire, United Kingdom, OX4 2SU |
|---------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact | Dr. Jaco Jacobs |
| Date Prepared | 18th November 2022 |
#### 2 Subject Device
| Product Trade Name | EchoGo Heart Failure |
|--------------------|--------------------------------------------------------------|
| Model Number | 1.0 |
| 510(k) | K222463 |
| Manufacturer | Ultromics Limited |
| Medical Speciality | Cardiology |
| Regulation | 21 CFR 870.2200 – Adjunctive Cardiovascular Status Indicator |
| Product Code | QUO – Adjunctive Heart Failure Status Indicator |
| Regulatory Class | II |
EchoGo Heart Failure is the product trade name and 1.0 is the model number. For the avoidance of doubt, in this submission we combine the product trade name and model number and refer to the subject device as EchoGo Heart Failure 1.0.
#### 3 Predicate Device
| Predicate Device | T3 Platform Software |
|------------------|----------------------|
| 510(k) | K213230 |
| Manufacturer | Etiometry Inc. |
#### Device Description ব
EchoGo Heart Failure 1.0 is a software-only medical device manufactured by Ultromics Limited and granted breakthrough status by the FDA under Q212613.
EchoGo Heart Failure 1.0 takes as input a DICOM file containing an echocardiogram and reports a classification decision suggestive of the presence or absence of heart failure with preserved ejection fraction (HFpEF). The output of this device is based on an artificial intelligence (Al) model developed using a convolutional neural network that produces the classification result. The model takes as input a 2D echocardiogram in which an apical 4-chamber view of the heart has been captured and assessed to have an ejection fraction ≥50% (this would normally be computed using a medical device for the assessment of cardiac function of the left
Image /page/3/Picture/13 description: The image shows the text 'K222463' in a clear, sans-serif font. The text is presented in a vibrant blue color, which contrasts with the white background. The numbers and letters are evenly spaced and easily readable.
{4}------------------------------------------------
ventricle, for example K213275). The echocardiogram should be acquired without contrast and contain at least one full cardiac cycle.
Independent training, validation and test datasets were used for training and performance assessment of the device. EchoGo Heart Failure 1.0 is fully automated and does not comprise a user interface.
EchoGo Heart Failure 1.0 produces a report containing the result of the classification, and this report is intended to be used by an interpreting clinician as an aid to diagnosis for HFpEF. The results are intended as an additional input to standard diagnostic pathways and should only be used by an interpreting clinician. The device is a diagnostic aid and thus according to common medical sense and the principles of differential diagnosis any diagnostic finding derived from usage of this product must be confirmed by additional diagnostic investigations, if in doubt. The ultimate diagnostic decision remains the responsibility of the interpreting clinician using patient presentation, medical history, and the results of available diagnostic tests, one of which may be EchoGo Heart Failure 1.0.
EchoGo Heart Failure 1.0 is a prescription only device.
## 5 Context
#### 5.1 Intended Use
Providing adjunctive information on a patient's cardiovascular condition (diagnostic aid for Heart Failure with Preserved Ejection Fraction (HFpEF)).
### 5.2 Intended User
The clinician interpreting the report produced by EchoGo Heart Failure 1.0 and making a diagnostic decision.
#### 5.3 Indications for Use
EchoGo Heart Failure 1.0 is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing routine functional cardiovascular assessment using echocardiography. When utilised by an interpreting clinician, this device provides information that may be useful in detecting heart failure with preserved ejection fraction (HFpEF).
EchoGo Heart Failure 1.0 is indicated in adult populations over 25 years of age. Patient management decisions should not be made solely on the results of the EchoGo Heart Failure 1.0 analysis.
EchoGo Heart Failure 1.0 takes as input an apical 4-chamber view of the heart that has been captured and assessed to have an ejection fraction ≥50%.
## 5.4 Patient Population
Patients undergoing routine functional cardiovascular assessment using diagnostic echocardiography or those suspected of heart failure.
{5}------------------------------------------------
## 6 Comparison of Intended Use
Both the subject and predicate devices are software-only devices as the subject device is an "automated machine learning based decision support system" and the predicate device "features the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Risk Analytics Engine software module version 8.0″.
Both devices are indicated as adjunctive devices that provide information that may be useful for healthcare professionals in forming a diagnostic decision. The subject device is a "diagnostic aid for patients undergoing routine functional cardiovascular assessment" and the predicate device is intended to be used "to utilize this information to aid in clinical decisions".
Both devices are intended to provide information on the cardiovascular state of patients. The subject device "provides information that may be useful in detecting heart failure with preserved ejection fraction (HFpEF)", whereas the predicate device "presents partial quantitative information about the patient's cardiovascular condition".
Both the subject and predicate devices are adjunctive devices and patient management decisions should not be made solely on device recommendations. Specifically, for the subject device "Patient management decisions should not be made solely on the results of the EchoGo Heart Failure 1.0 analysis"; the predicate states that "The primary data should be reviewed as part of standard patient evaluations and no decisions should be solely based on the indices".
Both the subject and predicate devices have indications for adult populations. The subject device "EchoGo Heart Failure 1.0 is indicated in adult populations over 25 years of age" and the predicate is indicated for use on "adult, paediatric, and neonatal patients".
For purposes of substantial equivalence, the term intended use means the general purpose of the device or its function and encompasses the indications for use. The term indications for use describes the disease the device is intended to serve as a diagnostic aid including a description of the patient population for which the device is intended. Both devices are intended as adjunctive aids for cardiovascular states or conditions and both devices can be used on adult populations. It follows that both devices have intended use of providing adjunctive information on the cardiovascular condition of a patient.
Any minor differences between the intended use and/or indications for use do not raise any new concerns with regards to safety and effectiveness. The regulation product code of the predicate device, PPW, under the associated regulation includes special controls that were applied to the subject device. The general controls and special controls of the predicate device are sufficient to ensure the substantial equivalence of the subject device and adequately control any differences.
#### 7 Comparison of Technological Characteristics
A full comparison of the technological characteristics of the subject and predicate devices follows.
The following technological differences exist between the subject and predicate devices:
{6}------------------------------------------------
- . The subject device takes as input a DICOM file containing an echocardiogram and the predicate device uses as input numeric physiological information from medical devices to which it is connected. Both devices therefore receive as input data that is the output of another medical device.
- The output of the subject device is based on an artificial intelligence (AI) model developed using a . convolutional neural network that produces a classification result. The output of the predicate device is derived from mathematical manipulations of physiologic data and measurements received from connected devices.
- . The subject device reports a classification as suggestive or not suggestive of the presence of heart failure with preserved ejection fraction (HFpEF), while the predicate device calculates indices that provides information on the respiratory or cardiovascular status of patients (IVCO2 and ACD, HLA and IDO2, respectively). Both devices are adjunctive cardiovascular status indicators.
- The predicate device outputs or displays amalgamated information from connected devices not output by ● the subject device, for example:
- Airway flow, volume, and pressure .
- . Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean)
- Bi-spectral index (BIS, signal quality index, suppression ratio) .
- . Cardiac Index
- Cardiac output ●
- Central venous pressure .
- Cerebral perfusion pressure ●
- End-tidal CO2 ●
- Heart rate ●
- Heart rate variability ●
- Intracranial pressure ●
- Left atrium pressure ●
- Oxygen saturation (intravascular, regional, SpO2) ●
- Premature ventricular counted beats ●
- Pulmonary artery pressure (systolic, diastolic, and mean) .
- Pulse pressure variation ●
- Pulse Rate ●
- Respiratory rate ●
- Right atrium pressure ●
- Temperature (rectal, oesophageal, tympanic, blood, core, nasopharyngeal, skin) ●
- Umbilical arterial pressure (systolic, diastolic, and mean) .
The following table summarises technological characteristics.
| Characteristic | Subject Device<br>EchoGo Heart Failure 1.0 | Predicate Device<br>T3 Platform Software |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 21 CFR 870.2200 | 21 CFR 870.2200 |
| Generic Device<br>Type | Adjunctive cardiovascular status<br>indicator | Adjunctive cardiovascular status<br>indicator |
| Characteristic | Subject Device | Predicate Device |
| | EchoGo Heart Failure 1.0 | T3 Platform Software |
| SaMD | Yes | Yes |
| Intended Use | Providing adjunctive information on a<br>patient's cardiovascular condition<br>(diagnostic aid for Heart Failure with<br>Preserved Ejection Fraction (HFpEF)). | Providing adjunctive information on a<br>patient's cardiovascular and respiratory<br>condition. |
| Indications for Use | EchoGo Heart Failure 1.0 is an<br>automated machine learning-based<br>decision support system, indicated as a<br>diagnostic aid for patients undergoing<br>routine functional cardiovascular<br>assessment using echocardiography.<br>When utilised by an interpreting<br>clinician, this device provides<br>information that may be useful in<br>detecting heart failure with preserved<br>ejection fraction (HFpEF).<br>EchoGo Heart Failure 1.0 is indicated in<br>adult populations over 25 years of age.<br>Patient management decisions should<br>not be made solely on the results of the<br>EchoGo Heart Failure 1.0 analysis.<br>EchoGo Heart Failure 1.0 takes as input<br>an apical 4-chamber view of the heart<br>that has been captured and assessed to<br>have an ejection fraction ≥50%. | The T3 Data Aggregation & Visualization<br>software module is intended for the<br>recording and display of multiple<br>physiological parameters of the adult,<br>paediatric, and neonatal patients from<br>supported bedside devices. The<br>software module is not intended for<br>alarm notification or waveform display,<br>nor is it intended to control any of the<br>independent bedside devices to which it<br>is connected. The software module is<br>intended to be used by healthcare<br>professionals for the following purposes:<br>• To remotely consult regarding a<br>patient's status, and<br>• To remotely review other standard<br>or critical near real-time patient data<br>in order to utilize this information to<br>aid in clinical decisions and deliver<br>patient care in a timely manner.<br>The T3 Data Aggregation & Visualization<br>software module can display numeric<br>physiologic data captured by other<br>medical devices:<br>• Airway flow, volume, and pressure<br>• Arterial blood pressure (invasive and<br>non-invasive, systolic, diastolic, and<br>mean)<br>• Bi-spectral index (BIS, signal quality<br>index, suppression ratio)<br>• Cardiac Index • Cardiac output<br>• Central venous pressure<br>• Cerebral perfusion pressure<br>• End-tidal CO2<br>• Heart rate<br>• Heart rate variability |
| Characteristic | Subject Device | Predicate Device |
| | EchoGo Heart Failure 1.0 | T3 Platform Software |
| | | • Intracranial pressure<br>• Left atrium pressure<br>• Oxygen saturation (intravascular,<br>regional, SpO2)<br>• Premature ventricular counted beats<br>• Pulmonary artery pressure<br>(systolic, diastolic, and mean)<br>• Pulse pressure variation<br>• Pulse Rate<br>• Respiratory rate<br>• Right atrium pressure<br>• Temperature (rectal, oesophageal,<br>tympanic, blood, core,<br>nasopharyngeal, skin)<br>• Umbilical arterial pressure (systolic, diastolic, and mean)<br>The T3 Data Aggregation & Visualization<br>software module can display laboratory<br>measurements including arterial and<br>venous blood gases, complete blood<br>count, and lactic acid. T3 Data<br>Aggregation & Visualization software<br>module can display information<br>captured by the T3 Risk Analytics Engine<br>software module.<br>The T3 Risk Analytics Engine software<br>module calculates four indices: the IDO2<br>Index for inadequate delivery of oxygen,<br>the IVCO2 Index for inadequate<br>ventilation of carbon dioxide, the ACD<br>Index for acidaemia, and the HLA Index<br>for hyperlactatemia.<br>The IDO2 Index is indicated for use by<br>health care professionals with post-<br>surgical patients 0 to 12 years of age and<br>weighing 2 kg or more under intensive<br>care. The IDO2 Index is derived by<br>mathematical manipulations of the<br>physiologic data and laboratory<br>measurements received by the T3 Data<br>Aggregation & Visualization software<br>module. When the IDO2 Index is |
| Characteristic | Subject Device | Predicate Device |
| | EchoGo Heart Failure 1.0 | T3 Platform Software |
| | | increasing, it means that there is an |
| | | increasing risk of inadequate oxygen |
| | | delivery and attention should be |
| | | brought to the patient. |
| | | The IDO2 Index presents partial |
| | | quantitative information about the |
| | | patient's cardiovascular condition, and |
| | | no therapy or drugs can be administered |
| | | based solely on the interpretation |
| | | statements. The IVCO2 Index is indicated |
| | | for use by health care professionals with |
| | | invasively ventilated patients 0 to 12 |
| | | years of age and weighing 2 kg or more |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.