K222459 · Centrix, Inc. · PHR · Oct 27, 2023 · Dental
Device Facts
Record ID
K222459
Device Name
Centrix FluoroSilver Silver Diamine Fluoride 38%
Applicant
Centrix, Inc.
Product Code
PHR · Dental
Decision Date
Oct 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3260
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Treatment of Dental Hypersensitivity for use in adults over the age of 21.
Device Story
Centrix FluoroSilver is a professional tooth desensitizer containing 38.3% to 43.2% w/v aqueous silver diamine fluoride (SDF). Used by dentists in a clinical setting, the solution is applied to cleaned, dried tooth surfaces using micro-fine brushes. The primary mechanism of action involves the formation of a precipitate upon contact with the tooth, which physically occludes open dentin tubules. This occlusion reduces dentin permeability, thereby alleviating pain associated with dentinal hypersensitivity. The device is supplied in squeezable dropper bottles and may be reapplied at one-week intervals if necessary. The treatment benefits patients by providing a non-invasive method to manage tooth sensitivity.
Clinical Evidence
Bench testing only. Studies included ISO 10993 biological evaluations (cytotoxicity, skin sensitization, irritation, systemic toxicity), SEM analysis of dentin tubule occlusion, fluoride release testing, and dentin permeability reduction. Permeability testing on human molar samples showed a 97.5% reduction in permeability for both subject and predicate devices. Chemical identity was confirmed via HPLC-Mass Spectrometry.
Technological Characteristics
Aqueous solution containing silver diamine fluoride (38.3-43.2% w/v), water, ammonia, and colorant. pH 8.5-9.5. Density 1.265-1.276 g/L. Principle of operation: chemical precipitation for physical tubule occlusion. Supplied in 5 ml squeezable dropper bottles with disposable micro-fine brush applicators. Shelf life: 24 months.
Indications for Use
Indicated for the treatment of dental hypersensitivity in adults over the age of 21.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
October 27, 2023
Centrix Inc % Joseph Azary Chief Advisor & Compliance Officer Aztech Regulatory & Quality LLC 543 Long Hill Avenue Shelton, Connecticut 06484
Re: K222459
Trade/Device Name: Centrix FluoroSilver Silver Diamine Fluoride 38% Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity varnish Regulatory Class: Class II Product Code: PHR Dated: August 12, 2022 Received: August 15, 2022
Dear Joseph Azary:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bobak Shirmohammadi -2
For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222459
### Device Name
Centrix FluoroSilver Silver Diamine Fluoride 38%
Indications for Use (Describe) Treatment of Dental Hypersensitivity for use in adults over the age of 21.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K222459 510(k) Summary CENTRIX FLUOROSILVER SILVER DIAMINE FLUORIDE 38%
#### SUBMITTER/510(K) HOLDER 1.
Centrix Inc 770 River Road Shelton, CT 06484
Contact Name: Joseph Azary (Aztech Regulatory & Quality LLC) Email: jazary@erols.com Telephone: (203) 242-6670
Date Prepared: October 26, 2023
#### DEVICE NAME 2.
| Proprietary Name: | Centrix FluoroSilver Silver Diamine Fluoride 38% |
|----------------------------|--------------------------------------------------|
| Common/Usual Name: | Tooth Desensitizer |
| Classification Name: | Tooth Desensitizer |
| Classification Regulation: | 21 CFR 872.3260 |
| Product code: | PHR |
| Classification: | Class 2 |
| Medical Specialty (Panel): | Division of Dental and ENT devices |
#### 3. PREDICATE DEVICES
- . Advantage Arrest (Elevate Oral Care aka ADP Silver Dental Arrest) K102973
{5}------------------------------------------------
#### 4. DEVICE DESCRIPTION
The Centrix FluoroSilver is a professional tooth desensitizer that is composed of Aqueous Silver Diamine Fluoride, 38.3% to 43.2% w/v. The solution is composed of water, ammonia, fluoride, silver and a colorant.
The device is provided in a single 5 ml squeezable dropper bottle or a multi-pack which includes three 5 ml squeezable dropper bottles. Each variation is packaged in a preformed trav and includes applicators (more information provided below) and the Instructions for Use (IFU).
The primary mode of action is to reduce dentinal hypersensitivity by occluding open dentin tubules by the formation of a precipitate.
The dentist must clean and dry the affected tooth surface and then dispense 1-2 drops of solution into a disposable dappen dish. The material is applied directly to the tooth surface using Centrix supplied applicators (Centrix Benda Micro Fine Brushes). The tooth must be allowed to air dry and then if needed one or two reapplications can be administered at intervals of one week.
The packaging will labeled with lot number, expiration date and UDI barcode. The shelf life of the device is 24 months.
The device is packaged with applicators (Centrix Benda Micro Fine Brushes) that are classified as class I 510(k) exempt medical devices under regulation 21 CFR 872.3140, Product Code KXR. A quantity of applicators will be provided in a zip lock bag.
The primary mode of action is to reduce dentinal hypersensitivity by occluding open dentin tubules by the formation of a precipitate.
| Parameter | Specification |
|-------------------------------|-------------------|
| Silver Diamine Fluoride (SDF) | 360 - 400 g/L |
| Fluoride | 50 - 60 g/L |
| pH | 8.5 - 9.5 |
| Density | 1.265 - 1.276 g/L |
### Specifications
#### 5. INDICATIONS FOR USE
Treatment of Dental Hypersensitivity for use in adults over the age of 21.
{6}------------------------------------------------
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The solution is applied to the tooth enamel to block dentinal tubules for the purpose of reducing tooth sensitivity. The solution precipitates upon application resulting in a physical blockage of dentinal tubules.
The subject device has equivalent indications for use, technical characteristics, mode of action, material composition (confirmed through chemical identity testing), SDF % and cytotoxicity results (comparative cytotoxicity testing) when compared to the selected predicate device.
Additionally, the subject device has a different colorant, uses different size bottles, and some minor differences in wording in labeling. These differences are minor and the overall similarities support the equivalence between the subject and predicate device.
| | Centrix<br>FluoroSilver<br>Silver Diamine<br>Fluoride 38% | Advantage Arrest | Discussion |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) Number | K222459 | K102973 | |
| Company | Centrix Inc | Elevate Oral Care<br>(named as ADP<br>Silver Dental Arrest<br>LLC in 510k) | |
| Classification Name | Cavity Varnish | Cavity Varnish | IDENTICAL |
| Common Name | Tooth Desensitizer | Tooth Desensitizer | IDENTICAL |
| Regulation Number | 872.3260 | 872.3260 | IDENTICAL |
| Product Code | PHR | PHR | IDENTICAL |
| Indications for Use | Treatment of Dental<br>Hypersensitivity for<br>use in adults over the<br>age of 21. | Treatment of dentinal<br>hypersensitivity.<br><br>For use in adult over<br>the age of 21. | IDENTICAL |
| Technical Method / Characteristics | Silver diamine<br>fluoride forms<br>precipitates with<br>calcium or phosphate<br>in the dentinal<br>tubules to block open<br>dentinal tubules. | Silver diamine<br>fluoride forms<br>precipitates with<br>calcium or phosphate<br>in the dentinal<br>tubules to block open<br>dentinal tubules. | IDENTICAL |
| Mode of Action | Tubule occlusion | Tubule occlusion | IDENTICAL |
| Material Composition | Silver diamine<br>fluoride<br>Purified Water | Silver diamine<br>fluoride<br>Purified Water | IDENTICAL |
## Comparison Table
{7}------------------------------------------------
| | Fast Green FCF Dye<br>HPLC – Mass Spec | FD&C Blue 1<br>HPLC Mass Spec | |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|------------------------------------------------------------------|
| Chemical<br>Composition Profile | | | Nearly Identical<br>- Compositional<br>Equivalency |
| Color | Green | Blue / Violet | IDENTICAL |
| % SDF | 38% | 38% | IDENTICAL |
| Application | Liquid | Liquid | IDENTICAL |
| Rx / OTC | Rx | Rx | IDENTICAL |
| Packaging | Squeezable Dropper<br>Bottle<br>5 ml bottle<br>Three 5ml bottle kit<br>Applicator(s) | Squeezable Dropper<br>Bottle<br>8ml bottle<br>Three 3ml bottle kit | IDENTICAL<br>Packaging type<br>Different Size<br>bottles |
| Cytotoxicity | Yielded positive<br>cytotoxicity when<br>tested to ISO 10993-5 | Yielded positive<br>cytotoxicity when<br>tested to ISO 10993-5 | IDENTICAL |
| Fluoride Release | 0.023 +/- 0.003<br>ppmF/mm² | 0.024 +/- 0.006<br>ppmF/mm² | IDENTICAL<br>No statistical<br>difference<br>between<br>devices. |
| SEM Analysis | Silver deposition on dentin and in tubules<br>between Centrix SDF and Elevate SDF<br>showed no discernable difference based on<br>SEM analyses. | | IDENTICAL<br>No discernable<br>difference<br>between<br>devices. |
| Dentin Permeability | Dentin Permeability. Human molar teeth<br>samples were evaluated to calculate the %<br>reduction of dentin permeability using the<br>ratio of average treated volume of water<br>passing through the dentin divided by the<br>average baseline rate of flow.<br>The results found that tooth samples treated<br>with the subject and predicate device reduced<br>the dentin permeability by an average 97.5%.<br>Dentin permeability is thought to be related to<br>dentin sensitivity. Treatments that reduce<br>dentin permeability have been found to stop<br>pain due to dentin sensitivity.<br><br>Centrix % Reduction in Dentin Permeability<br>97.5% +/- 4.3%<br><br>Elevate % Reduction in Dentin Permeability<br>97.5% +/- 2.5% | | IDENTICAL |
{8}------------------------------------------------
#### 7. PERFORMANCE TESTING
The following standards were either used as a reference and/or for testing of the subject device:
| Summary of Testing | | |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Standard | Description of Test | Results |
| ISO 10993-5 (2009)<br>Cytotoxicity | Biological Evaluation of Medical<br>Devices – Part 5: Tests for in vitro<br>cytotoxicity | Cytotoxicity testing was conducted on both<br>the subject (Centrix) and predicate (Elevate)<br>devices. The testing found both the subject<br>device and predicate device samples were<br>"unsuitable for analysis" due to the high<br>level of precipitate generated when the test<br>articles were combined with culture media.<br>The subject device performed identically to<br>the predicate device. |
| ISO 10993-10 (2021) | Biological Evaluation of Medical<br>Devices - Part 10: Tests for Skin<br>Sensitization | The testing concluded that the subject<br>device is NOT considered to be a contact<br>skin sensitizer. |
| ISO 10993-23 (2021) | Biological Evaluation of Medical<br>Devices – Part 23: Tests for Irritation | The testing concluded that the subject<br>device is considered to be a Non-Irritant to<br>oral mucosa. |
| ISO 10993-11 (2017) | Biological Evaluation of Medical<br>Devices – Part 11: Tests for Systemic<br>Toxicity | Acute Systemic Toxicity testing was<br>conducted on both the subject device<br>(Centrix) and Predicate Device (Elevate).<br>The testing concluded that there were no<br>signs of gross toxicity, adverse clinical<br>effects, or abnormal behavior. No gross<br>abnormalities were noted after a 3 day<br>observation and necropsy. Both the subject<br>and predicate devices met the requirements<br>of the tests and performed identically. |
| ISO 10993-1 (2018) | Biological Evaluation of Medical<br>Devices - Part 1: Evaluation and<br>Testing within a Risk Management<br>Process | The remaining biological endpoints of<br>Acute, Subacute / Subchronic, Chronic<br>Systemic Toxicity, Genotoxicity and<br>Carcinogenicity were evaluated and<br>documented in a Toxicological Risk<br>Assessment report. The evaluation<br>concluded that information gathered in<br>published literature on the chemical<br>ingredients, intended use, predicate device<br>information that the risk of adverse effects<br>are low with respect to biological endpoints<br>of Acute, Subacute / Subchronic, Chronic<br>Systemic Toxicity, Genotoxicity and<br>Carcinogenicity. |
| N/A | SEM Analysis of Subject and<br>Predicate Devices | The SEM report summarizes comparison<br>between subject device (Centrix SDF) and<br>predicate device (Elevate SDF) on human<br>tooth samples. Both the subject and<br>predicate devices treated teeth samples have<br>significant deposits of silver on the cut |
| | | surface of the tooth and within the dentinal<br>tubules. There is no discernable difference<br>in the silver deposited on the tooth surface<br>and within the dentinal tubules as a result of<br>exposure to subject and predicate devices. |
| N/A | Fluoride Release Testing. Evaluation<br>of fluoride release in SDF treated<br>tooth cross sections comparing<br>Centrix FluoroSilver Silver Diamine<br>Fluoride 38% and Predicate Device<br>(Elevate Advantage Arrest SDF) | The study confirmed the release of fluoride<br>at the same rate with no statistical<br>difference between the groups. The<br>maximum fluoride concentrations were<br>equivalent.<br>Subject 0.023 +/- 0.003 ppmF/mm²<br>Predicate 0.024 +/- 0.006 ppmF/mm² |
| N/A | Dentin Permeability. Human molar<br>teeth samples were evaluated to<br>calculate the % reduction of dentin<br>permeability using the ratio of<br>average treated volume of water<br>passing through the dentin divided by<br>the average baseline rate of flow. | The results found that tooth samples treated<br>with SDF reduced the dentin permeability<br>by 97.5%. Dentin permeability is thought to<br>be related to dentin sensitivity. Treatments<br>that reduce dentin permeability have been<br>found to stop pain due to dentin sensitivity |
| N/A | Chemical Identity Testing: Mass<br>Spectrometry (MS) performed on<br>subject and predicate device solutions | MS analysis found a strong match for<br>chemical composition between the subject<br>and predicate device to demonstrate<br>compositional equivalency. |
Summary of Testing
{9}------------------------------------------------
# 8. SAFETY AND EFFICACY
The subject device is found to be equivalent to the predicate device based on comparative evaluation and testing. The overall safety and efficacy of FluoroSilver Silver Diamine Fluoride 38% usage for treating dentinal sensitivity has been demonstrated through testing conducted on the subject device, comparative testing between subject and predicate device and published literature.
### 9. CONCLUSION
Information presented supports substantial equivalence of the Centrix FluoroSilver Silver Diamine Fluoride 38% the predicate device based on similarities in intended use, design, principles of operation, performance specifications, and testing.
Centrix Inc believes that based on the indications for use, technological characteristics, and comparison to predicate devices the Centrix FluoroSilver Silver Diamine Fluoride 38% has been shown to be substantially equivalent to the predicate and have demonstrated that the device is as safe and as effective as the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.