K222453 · Guangdong Comytens Medical Technology Co., Ltd. · GXY · Nov 4, 2022 · Neurology
Device Facts
Record ID
K222453
Device Name
Self-adhesive Electrode
Applicant
Guangdong Comytens Medical Technology Co., Ltd.
Product Code
GXY · Neurology
Decision Date
Nov 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Self-Adhesive Electrode is intended for as a reusable, conductive interface between the patient's skin and the marketed electrical stimulators (i.e.TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF or IFC(Interferential current),or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current, for OTC (Over-The-Counter) or Prescription use.
Device Story
Self-adhesive electrodes function as conductive interface between patient skin and electrical stimulators (TENS, EMS, IFC, PGF). Device consists of multi-layer structure: fabric/foam backing, conductive film (carbon/aluminum), and biocompatible conductive hydrogel. Connects to stimulators via lead wires, snap buttons, or magnetic buttons. Used in clinical or home settings; applied by patient or clinician. Device transmits electrical current from stimulator to patient skin; adhesive hydrogel secures electrode without external fasteners. Reusable design allows multiple applications on single patient. Benefits include non-invasive delivery of electrical stimulation for therapeutic purposes.
Clinical Evidence
No clinical data. Bench testing only, including electrical impedance, adhesion force, mechanical connector testing, and accelerated aging per ASTM F1980. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation).
Technological Characteristics
Multi-layer construction: fabric/foam backing, conductive film (silver-coated carbon/carbon/aluminum), and biocompatible conductive hydrogel. Connectivity: lead wire (2mm/2.5mm socket), snap button, or magnetic button. Impedance <200 ohms. Non-sterile. Complies with IEC 60601-1 (basic safety) and ISO 10993 (biocompatibility).
Indications for Use
Indicated for use as a reusable, conductive interface between patient skin and electrical stimulators (TENS, EMS, IFC, PGF) for general adult populations. Suitable for both OTC and prescription use.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 4, 2022
Guangdong Comytens Medical Technology Co., Ltd Mr. Amos Zou RA Manager Building A-102, No. 24, Jiangjunmao Industrial Zone Wulian, Longgang District Shenzhen, Guangdong 518116 China
Re: K222453
Trade/Device Name: Self-adhesive Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: August 12, 2022 Received: August 15, 2022
Dear Mr. Amos Zou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222453
Device Name Self-adhesive Electrode
### Indications for Use (Describe)
Self-Adhesive Electrode is intended for as a reusable, conductive interface between the patient's skin and the marketed electrical stimulators (i.e.TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF or IFC(Interferential current),or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current, for OTC (Over-The-Counter) or Prescription use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☑</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 16px;">☑</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K)SUMMARY
#### 1. Submitter of 510(K):
| Sponsor | |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|
| Company Name: | Guangdong Comytens Medical Technology Co.,Ltd |
| Address: | Building A-102, No. 24, Jiangjunmao Industrial Zone, Wulian, Longgang District ,<br>Shenzhen, Guangdong, China 518116 |
| Contact person: | Deng Jiuzhen |
| TEL: | +86-13418878095 |
| FAX: | / |
| E-mail: | info@comytens.com 276148005@qq.com |
| Date of Prepared: | Nov, 3 2022 |
## Application Correspondent:
| Company Name: | Guangdong Comytens Medical Technology Co.,Ltd |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|
| Address: | Building A-102, No. 24, Jiangjunmao Industrial Zone, Wulian, Longgang District ,<br>Shenzhen, Guangdong, China 518116 |
| Contact person: | Amos Zou |
| TEL: | +86-15015249549 |
| E-mail: | Amos.zou@139.copm |
| Date of Prepared: | Nov, 3 2022 |
#### 2. Proposed Device and code:
| Device Trade Name: | Self-Adhesive Electrode |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Model Series | Self-Adhesive Electrodes with lead wire;<br>Self-Adhesive Electrodes with Snap/Magnetic connector;<br>Self-Adhesive Electrode Pads |
| Regulation Medical Specialty | Neurology |
{4}------------------------------------------------
| Product Code: | GXY |
|-------------------|-----------------|
| Regulation number | 21 CFR 882.1320 |
| Device Class | II |
| 510(K) Number | K222453 |
#### Predicate Device: 3.
| 510(K) | Trade Name | Submitter | Product Code |
|---------|-------------------|-----------------------|--------------|
| K180865 | ZMI Self-adhesive | ZMI Electronics, Ltd. | GXY |
#### Description of Proposed Device: 4.
Self-Adhesive Electrodes are composed of conductive carbon film and adhesive Electrodes composed of Aluminum foil film,Adhesive Electrodes manufactured by Comytens are multi-layer reusable, flexible structures composed of laminated materials commonly used in this application:
| No. | Description | Material Composition |
|-----|-------------------------------|----------------------------------------------------------------------------|
| 1 | Top layer | Fabric/Foam/ Tan fabric |
| 2 | Middle layer | Conductive film (Silver coated Carbon Film/Carbon film/Aluminum foil film) |
| 3 | Bottom layer | Biocompatible self-adhesive conductive hydrogel |
| 4 | Connection | Leadwire/snap button/magnetic button |
| 5 | double sides<br>adhesive tape | which is used for attaching the non-woven fabrics and conducting film |
| 6 | Release liner | plastic film is a protective layer for the hydrogel. |
The electrodes are designed for single-patient/multiple application use. Because of the adhesive nature of the biocompatible conductive hydrogel, no securing materials are required to secure the device to the patient's skin.
For the electrical connection, Comytens provides lead wire type and snap button/Magnetic button type:
- Leadwire assembly:
The electrode is connected to the electrical stimulator by lead wire, at least 40 mm long wire with a standard connector: 0.08"/0.1"(2mm/2.5mm) diameter female socket connector with insulating outer. By design, the insulated outer jacket prevents the connection to earth or hazardous voltages. The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR part 898 by testing under ANSI/AAMI ES60601-1, subclause 8.5.2.3.
- Snap button/Magnetic button assembly:
With 2.5~5mm diameter male socket. The snap button and the magnetic button provide different types of physical contact with the source of stimulation current via spring tension or magnetic force among male and femaleconnectors. Either method is similar to the lead wire connection providing the friction to create the physicalcontact.
#### 5. Intended for Use
Self-Adhesive Electrode is intended for as a reusable, conductive interface between the patient's skin and the marketed electrical stimulators (i.e.TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF or IFC(Interferential current), or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current, for OTC (Over-The-Counter) or Prescription use.
{5}------------------------------------------------
#### 6. Technical and Performance
The following table compares the device to the predicate device with basic technological characteristics.
Comparison to predicate device:
| Element of<br>Comparison | Subject Device | Predicate Device | Comm<br>ent |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| 510(k) | K222453 | K180865 | N/A |
| Company | Guangdong Comytens Medical<br>Technology Co.,Ltd | ZMI Electronics, Ltd. | N/A |
| Device Name | Self-Adhesive Electrodes | ZMI Self-adhesive electrodes | N/A |
| Model name | ● Self-Adhesive Electrodes with leadwire;<br>● Self-Adhesive Electrodes with Snap<br>/Magnetic connector;<br>● Self-Adhesive Electrode Pads | SAE Type | N/A |
| Shape | Square(40*40mm) | Round, Rectangular, Square, Oval, and<br>Butterfly | similar<br>1# |
| Color | White, Red, Black, and Tan | White, Red, Black, and Tan | Same |
| Regulation<br>Number | 882.1320 | 882.1320 | Same |
| Product Code | GXY | GXY | Same |
| Classification<br>Name | Cutaneous electrode | Cutaneous electrode | Same |
| Intended Use | Self-Adhesive Electrode is intended for as<br>a reusable, conductive adhesive interface<br>between the patient's skin and the<br>marketed electrical stimulators (i.e.TENS<br>(Transcutaneous Electrical Nerve<br>Stimulation), EMS (Electrical Muscular<br>Stimulation), IF or IFC(Interferential<br>current), RUSS(Russian stimulation),<br>MIC(Microcurrent) or PGF (Pulsed<br>Galvanic Stimulation) for transmitting<br>electrical current, for OTC (Over-The-<br>Counter) or Prescription use. | ZMI Self-adhesive electrodes are<br>intended for as a reusable, conductive<br>adhesive interface between the patient's<br>skin and the marketed electrical<br>stimulators (i.e. TENS (Transcutaneous<br>Electrical Nerve Stimulation), EMS<br>(Electrical Muscular Stimulation), IF<br>(Interferential) or PGF (Pulsed Galvanic<br>Stimulation) for transmitting electrical<br>current, for OTC (Over-The-Counter) or<br>Prescription use. | Same |
| OTC or<br>Prescription | OTC and Prescription | OTC and Prescription | Same |
| Target<br>Population | General [Adult] | General [Adult] | Same |
| Design<br>Feature | Six layers:<br>1. Insulation backing material:<br>Fabric/Foam/Tan fabric<br>2. Conductive film: Aluminum foil film | Three layers:<br>1. Insulation backing material:<br>Fabric/Foam/Tan fabric<br>2.Conductive film: Aluminum foil | similar<br>2# |
| | /Carbon film/Carbon film coated with<br>silver/<br>3.Conductive hydrogel<br>4.Connection<br>5.double sides adhesive tape<br>6.Release liner | film/Carbon film/Carbon film coated with<br>silver/<br>3.Conductive hydrogel | |
| Electrical<br>Connection | Leadwire<br>Snap button<br>Magnetic button | Leadwire<br>Snap button<br>Magnetic button | Same |
| Lead Wire<br>connector | .080"/0.1"(2mm/2.5mm) female socket<br>or Pin connector | .080" (2mm) female socket connector | similar<br>3# |
| Non-sterile | Non-sterile | Non-sterile | Same |
| Reusable | Reusable | Reusable | Same |
| Packaging | Re-sealable bag packed | Re-sealable bag packed | Same |
| Adhesive<br>Type | Self-adhesive | Self-adhesive | Same |
| Biocompatibi<br>lity | Complies with ISO10993 | Complies with ISO10993 | Same |
| A.C.<br>Impedance | <200 ohms | <200 ohms | Same |
| Force<br>required to<br>remove<br>wire from<br>electrode | More than 6 pounds of force | More than 6 pounds of force | Same |
| Single<br>Patient Use | Yes | Yes | Same |
{6}------------------------------------------------
Comparison in Detail(s):
1#: The shape and size are related to the treatment area may cause higher current density, which leads to discomfort and burns. The current densities for any shaped electrodes exceeding 2mA r.m.s/cm2 may require the special attention of the user. The color is mainly for identification and marketing purpose, and does not raise any new concerns of safety or effectiveness.
If the subject device is manufactured from the identical raw materials manufacturing processes as a predicate device with the same type and duration of tissue contact, and any changes in geometry are not expected to impact the biological response, this is typically sufficient to establish substantially equivalent biocompatibility.
2#: The subject device and predicate device are similar in construction. They both contain three basic components: a nonconductive top layer, a patient contacting layer, a carbondispersion pad middle layer, and; while the Self-Adhesive Electrode also contains three additional layer:conductive carbon fiber lead wire (or snap for Magnetic Style Electrode),double sides adhesive tape which is used for attaching the non-woven fabrics and conducting film,and plastic film is a protective layer for the hydrogel. Which the differences will not affectiveness of the Self-Adhesive Electrode.
3#: The snap button and the magnetic button provide different types of physical contact with the source of stimulation current via spring tension or magnetic force among male and female connectors. Either method is similar to the lead wire connection providing the friction to create the physical contact.
{7}------------------------------------------------
## 7. Performance Testing:
## Test Summary:
To establish substantial equivalence to the identified predicate devices, we performed the following tests on the subject device, Adhesive Electrodes TS serial,and the testing results provide evidence that the device complies with the applicable standards requirement and it is substantially equivalent to the predicate devices. Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.
## 7.1 Non-Clinical Data:
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrate that the proposed device complies with the following standards:
## 7.2 Biocompatibility testing
The biocompatibility evaluation for the Self-Adhesive Electrode conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is contacting for duration of less than 24 hours. And the testing included the following tests:
- Cytotoxicity
- Skin Sensitization
- . Skin Irritation
## 7.3 Safety and EMC
To verify the basic safety and essential performance of the subject device, weperformed the test noted below: - IEC60601-1:2005+AMD1:2012+AMD2:2020; Medical electrical equipment - Part 1: General requirementsfor basic safety and essential performance
The submitted sample(s) complied with the requirements of IEC 60601-1:2005+AMD1:2012+AMD2:2020, Cl.8.5.2.3
## 7.4 Bench Testing
Bench tests were conducted on Self-Adhesive Electrode to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
1. Electrical & Impedance Tests
The Self-Adhesive Electrode was tested to assure the current is equally distributed. The Self-Adhesive Electrode was also tested for Impedance to assure proper electrical connection to the skin.
2. Adhesion tests
The Self-Adhesive Electrode was tested for initial adhesion force after using multiple times. 3. Connector & Mechanical tests
The Self-Adhesive Electrode has a Snap or magnetic surface with the purpose to clamp the Self-Adhesive Electrode. This connection was tested to assure multiple use and also to assure the TENS stay fixed to the Self-Adhesive Electrode during use.
4. Packaging testsReal-Time Shelf Life Test Report
The Self-Adhesive Electrode was tested for storage conditions:
· The Self-Adhesive Electrode should not be stored at temperatures lower than 41°E(+5°C) or higher than 80.6°F (+27°C). 5. Accelerated Aging
The Self-Adhesive Electrode was tested for accelerate aging in worst storage condition to assure within the expiration date the electrode function as expected.
- ASTM F1980 (Reapproved 2016), Standard guide for accelerated aging of sterile barrier systems for medical devices. (Sterility)
## 7.5 Clinical data:
No clinical testing was performed
## 7.6 Summarv
Based on the non-clinical and clinical performance as documented in the device devices were found to have a safety and effectiveness profile that is similar to the predicate device.
## 8. Conclusions:
The proposed device has the same intended use and similar characteristics as the predicate device, Meanwhile, performance testing, bench testing, and safety report documentation supplied in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Guangdong Comytens Medical Technology Co.,Ltd. maintains that the Self-Adhesive Electrode is substantially equivalent to the
{8}------------------------------------------------
predicate device in indications for use, patient population, and environment for use, technology characteristics,
specifications / performance and compliance with internation
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.