K222441 · Bk Medical Aps · IYN · Dec 7, 2022 · Radiology
Device Facts
Record ID
K222441
Device Name
Ultrasound System 2300
Applicant
Bk Medical Aps
Product Code
IYN · Radiology
Decision Date
Dec 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The system is a diagnostic ultrasound imaging system used by qualified and trained healthcare professionals for ultrasound imaging, human body fluid flow analysis and puncture and biopsy guidance.
Device Story
The Ultrasound System 2300 is a mobile, software-controlled diagnostic ultrasound system used in professional healthcare environments (hospitals, clinics) by trained professionals. It acquires, processes, and displays ultrasound data via various multi-frequency transducers (linear, phased, convex, endocavity, laparoscopic). The system features a mobile console with a 19-inch monitor and either a conventional keyboard or glass touch interface. It supports multiple imaging modes including 2D (B-mode), M-mode, PWD, CFM, Power Doppler, Contrast Imaging, CW Doppler, and Strain Elastography. The device provides puncture and biopsy guidance and includes an 'LAP Ablation Navigation' option using electromagnetic needle tracking. Output is displayed on the monitor for real-time clinical assessment, aiding in diagnosis, fluid flow analysis, and interventional guidance. The system benefits patients by providing non-invasive diagnostic imaging and precise guidance for surgical and biopsy procedures.
Clinical Evidence
No clinical data was required to support substantial equivalence. Evidence consists of bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in accordance with recognized standards (e.g., IEC 60601-1, IEC 60601-2-37, IEC 62359, ISO 10993-1).
Technological Characteristics
Mobile ultrasound console with 19" optical bonded glass monitor. Features 4 transducer ports and 196 TX/RX channels. Connectivity includes DICOM. Software-controlled with multiple configurations (bk3000, bk5000, bkActiv). Electromagnetic needle tracking uses a tracking control unit, portable EM field generator, and clip-on sensors. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359, and ISO 10993-1.
Indications for Use
Indicated for fetal (obstetrics), abdominal, pediatric, intra-operative, intra-operative neuro (neurosurgery), laparoscopic, small organ (small parts), adult cephalic (trans-cranial), neonatal cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional and superficial), cardiac adult, trans-esophageal (cardiac), and peripheral vessel (peripheral vascular) applications. Contraindicated for ophthalmic use or any use causing the acoustic beam to pass through the eye. Cardiac adult application is not intended for direct use on the heart.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
December 7, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
BK Medical Aps % Sandra Theodoridis SR Regulatory Affairs Specialist Mileparken 34 Herlev. 2730 DENMARK
Re: K222441
Trade/Device Name: Ultrasound System 2300 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: October 28, 2022 Received: October 28, 2022
Dear Sandra Theodoridis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K222441
Device Name Ultrasound System 2300
Indications for Use (Describe)
Intended Use:
The system is a diagnostic ultrasound imaging system used by qualified and trained healthcare professionals for ultrasound imaging, human body fluid flow analysis and puncture and biopsy guidance.
Indications to Use:
The clinical applications and exam types include:
Fetal (including Obstetrics), Abdominal, Pediatric, Intra-operative Neuro (also known as Neurosurgery), Laparoscopic, Small Organ (also known as Small Parts), Adult Cephalic is also known as Adult Trans-cranial), Neonatal Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional and Superficial), Cardiac Adult, Transesophageal (Cardiac) and Peripheral Vessel (also known as Peripheral Vascular).
Modes of Operation:
- 2D (B-Mode) including Tissue Harmonic Imaging
- M-Mode
- PWD Mode
- CFM Mode (C, VFI)
- Power Doppler
- Contrast Imaging
- CW Doppler
- Strain Elastography
Environment:
The Ultrasound System 2300 is intended for use in the professional healthcare environment (e.g. hospitals, physician offices)
Contraindications:
The Ultrasound System 2300 is not intended for ophthalmic use or any use causing the acoustic beam to pass through the eye.
The Cardiac Adult application is not intended for direct use on the heart.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the alphanumeric string "K222441" in a bold, sans-serif font. The string is horizontally oriented and centered. The characters are uniformly sized and spaced, creating a clear and legible sequence.
Image /page/4/Picture/1 description: The image is the logo for BK Medical, a GE Healthcare company. The logo features the letters "bk" in a bold, green sans-serif font stacked on top of the word "medical" in a smaller, green sans-serif font. Below "medical" is the text "a GE Healthcare company" in a smaller, lighter font. To the right of the "bk" is a stylized green graphic that resembles sound waves or a signal emanating from a source.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| 1. Submitter: | BK Medical ApS<br>Mileparken 34<br>Herlev 2730<br>Denmark |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | BK Medical ApS<br>Mileparken 34<br>Herlev 2730<br>Denmark |
| Primary Contact Person: | Sandra Theodoridis<br>Senior Regulatory Affairs Specialist<br>BK Medical<br>Tel: (978) 578 9353<br>E-mail: stheodoridis@bkmedical.com |
| Date Prepared: | August 11, 2022 |
#### 儿 Device Names / Common Names / Classification Names:
| Trade Names: | Ultrasound System 2300 |
|-----------------------|------------------------------------------|
| Common Name: | Ultrasound System |
| Classification Name: | Ultrasonic pulsed doppler imaging system |
| Product Code: | IYN (primary), IYO, ITX (secondary) |
| Class: | II |
| Regulation Number: | 21 CFR §892.1550, §892.1560, §892.1570 |
| Classification Panel: | Radiology |
{5}------------------------------------------------
#### lll. Identification of Predicate or Legally Marketed Devices:
- . Primary predicate device: Ultrasound Scanner System bk2300 as cleared under 510(k) premarket notification No K180737.
| Trade Name: | Ultrasound System 2300 | |
|-----------------------|------------------------------------------|--|
| Common Name: | Ultrasound System | |
| Classification Name: | Ultrasonic pulsed doppler imaging system | |
| Product Code: | IYN (primary), IYO, ITX (secondary) | |
| Class: | II | |
| Regulation Number: | 21 CFR §892.1550, §892.1560, §892.1570 | |
| Classification Panel: | Radiology | |
- Reference predicate device: GE Logiq E10 cleared under 510(k) premarket notification No ● K211488.
| Trade Name: | GE Logiq E10 |
|-----------------------|------------------------------------------|
| Common Name: | GE Logiq E10 |
| Classification Name: | Ultrasonic pulsed doppler imaging system |
| Product Code: | IYN (primary), IYO, ITX (secondary) |
| Class: | II |
| Regulation Number: | 21 CFR §892.1550 |
| Classification Panel: | Radiology |
#### IV. Device Description
The Ultrasound System 2300 is a multi-purpose mobile, software-controlled diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms which are offered in different configurations/ models intended for urology, general imaging, surgical and anesthesiology applications.
The system consists of a mobile console (engine) that provides digital acquisition, processing and display capabilities. The user interface includes a conventional keyboard or a glass touchpad, a 19" Clinical Display Monitor (CDM). In addition, a variety of system accessories are available such as baskets, foot switch, printer start-up kit, remote control, and extra holders.
{6}------------------------------------------------
The Ultrasound System 2300 is available in the following marketing configurations:
- 1. bk3000 available with a conventional keyboard configuration. The bk3000 is primarily intended for applications such as urology and general imaging
- 2. bk5000 available with a conventional keyboard configuration. The bk5000 is primarily intended for surgery applications.
- 3. bkActiv is a configuration available with a glass user interface (UI). bkActiv is primarily intended for surgical and anesthesiology applications.
All configurations run on the previously cleared SW platform and HW platform (engine) (K180737). The various configurations of the Ultrasound System 2300 are intended to be used for different applications as described above with various transducers and options.
The primary difference between the system configurations (also refer to Table 1) are:
- bk5000 is the premier product offering with all features and probes available
- Bk3000 is a basic product offering with only a subset of features
- bkActiv is a configuration available with a glass user interface (UI). bkActiv is ● primarily intended for surgical and anesthesiology applications
| Catalog/<br>Reference<br>(REF) | Model | Model Description |
|--------------------------------|---------|------------------------------------------------------------------------------------------------------------------------------------------|
| 2300 | 2300-01 | BK3000 ULTRASOUND SYSTEM W/O<br>BATTERY<br>This configuration is primarily<br>intended for Urology and General<br>imaging applications. |
| Catalog/<br>Reference<br>(REF) | Model | Model Description |
| 2300 | | BK3000 ULTRASOUND SYSTEM<br>W/BATTERY<br>This configuration is primarily<br>intended for Urology and General<br>imaging applications. |
| 2300 | | BK5000 ULTRASOUND SYSTEM W/O<br>BATTERY<br>This configuration is primarily<br>intended for surgical applications. |
| 2300 | | BK5000 ULTRASOUND SYSTEM<br>W/BATTERY<br>This configuration is primarily<br>intended for surgical applications. |
| Catalog/<br>Reference<br>(REF) | Model | Model Description |
| 2300 | 2300-56 | BKACTIV ULTRASOUND SYSTEM W/O<br>BATTERY<br>This configuration is primarily<br>intended for surgical and<br>anesthesiology applications. |
| 2300 | 2300-66 | BKACTIV ULTRASOUND SYSTEM W/<br>BATTERY<br>This configuration is primarily<br>intended for surgical and<br>anesthesiology applications. |
Table 1: Ultrasound Scanner System bk2300 available configurations
{7}------------------------------------------------
{8}------------------------------------------------
The various configurations of the Ultrasound System 2300 are intended to be used with various multi-frequency transducers (see Table 2). The indicated uses are different and specific for each transducer listed.
- . Linear Array
- Phased Linear Array ●
- . Convex / Curved Array
The interaction with patients is dependent upon the transducer type which may include:
- Surface ●
- Inter-operative ●
- Laparoscopic ●
- Endocavity
{9}------------------------------------------------
| Transducer | bk3000 | bk5000 | bkActiv |
|-----------------------------------------------------------|--------|--------|---------|
| 5Cle (9085) CURVED ARRAY TRANSDUCER | X | X | X |
| 6C2 (9040) CURVED ARRAY TRANSDUCER | X | X | X |
| 6C2s (9023) SMALL CURVED ARRAY TRANSDUCER | X | X | X |
| 9C2 (9002) CURVED ARRAY TRANSDUCER | X | X | X |
| 14L3 (9051) LINEAR ARRAY TRANSDUCER | X | X | X |
| 13L4w (9011) WIDE LINEAR ARRAY TRANSDUCER | X | X | X |
| 10L2w (9022) WIDE LINEAR ARRAY TRANSDUCER | X | X | |
| 18L5 (9070) SMALL GIGH-FREQUENCY LINEAR ARRAY TRANSDUCER | X | X | X |
| 18L5s (9081) SMALL GIGH-FREQUENCY LINEAR ARRAY TRANSDUCER | X | X | X |
| 8L2 (9032) LINEAR ARRAY TRANSDUCER | X | X | X |
| E13C2 (9029) ENDFIRE ENDOCAVITY TRANSDUCER | X | X | |
| E14C4t (9018) TRIPLANE ENDOCAVITY TRANSDUCER | X | X | |
| X14CL4b (9048) BIPLANE ENDOCAVITY TRANSDUCER | X | X | |
| E10C4 (9019) ENDOCAVITY TRANSDUCER | X | X | |
| 20R3 (9052) ANORECTAL TRANSDUCER | X | X | |
| N13C5 (9062) CURVED ARRAY TRANSDUCER | X | X | X |
| 5P1 (9077) PHASED ARRAY TRANSDUCER | X | X | |
| X18L5S (9009) HOCKEY STICK TRANSDUCER | X | X | X |
| N11C5S (9063) BURR HOLE TRANSDUCER | X | X | X |
| I14C5I (9015) INTRAOPERATIVE I-SHAPE TRANSDUCER | X | X | X |
| I14C5T (9016) INTRAOPERATIVE T-SHAPE TRANSDUCER | X | X | X |
| I12C5B (9024) INTRAOPERATIVE BIPLANE TRANSDUCER | X | X | X |
| 112C5 (9034) MINI-TRANSDUCER | X | X | X |
| I12C4f (9066) LAPAROSCOPIC TRANSDUCER | X | X | X |
| X12C4 (9026) DROP-IN TRANSDUCER | X | X | X |
| X14L4 (9038) 3D ENDOCAVITY TRANSDUCER | X | X | |
| Rob12C4 (9096) ROBOTIC TRANSDUCER | X | X | X |
| N20P6 (9007) MINIMALLY INVASIVE TRANSDUCER | X | X | X |
| 113C3f (9076) | X | X | X |
| ADVANCED LAPAROSCOPIC TRANSDUCER | | | |
| 113C3fx (9078) | | | X |
| ADVANCED LAPAROSCOPIC TRANSDUCER WITH TRACKING | | | |
### Table 2: Transducers used with Ultrasound System 2300 configurations
#### > Indications / Intended Use:
### Intended Use:
The system is a diagnostic ultrasound imaging system used by qualified and trained healthcare professionals for ultrasound imaging, human body fluid flow analysis and puncture and biopsy guidance.
#### Indications for Use:
The clinical applications and exam types include:
{10}------------------------------------------------
- Fetal (including obstetrics) ●
- Abdominal ●
- Pediatric
- Intra-operative
- Intra-operative Neuro (also known as Neurosurgery) ●
- Laparoscopic ●
- Small Organ (also known as Small Parts) ●
- Adult Cephalic (Cephalic is also known as Adult Trans-cranial) ●
- Neonatal Cephalic
- Trans-rectal ●
- Trans-vaginal ●
- Musculo-skeletal (Conventional and Superficial) ●
- Cardiac Adult ●
- Trans-esophageal (Cardiac) ●
- Peripheral vessel (also known as Peripheral Vascular)
### Modes of operation:
- B-Mode (including Tissue Harmonic Imaging) -
- -M-Mode
- PWD Mode -
- CFM Mode (C, VFI) -
- -Power Doppler
- Contrast Imaging -
- CW Doppler -
- -Strain Elastography
### Environment:
The Ultrasound System 2300 is intended for use in the professional healthcare environment (e.g. hospitals, physician offices).
### Contraindications
The Ultrasound System 2300 is not intended for ophthalmic use or any use causing the acoustic beam to pass through the eye.
The Cardiac Adult application is not intended for direct use on the heart.
{11}------------------------------------------------
#### VL Comparison of Technological Characteristics with the Predicate Device
| | Ultrasound System 2300 | Ultrasound Scanner<br>System bk2300 | Ultrasound System LOGIQ<br>E10 | |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Proposed device<br>(TBD) | Primary predicate<br>(K180737) | (Volume Navigation/GPS<br>tracking option)<br>Reference predicate | Comment on Comparison |
| | | | (K211488) | |
| Manufacturer | BK Medical ApS | BK Medical ApS | GE Healthcare | N/A |
| Common Name | Ultrasound system | Ultrasound system | Ultrasound system | Equivalent |
| Name | bk3000 (2300-01, 2300- | bk3000 (2300-01, 2300- | LOGIQ E10 | bkActiv 2300-56 & 66 are |
| (Configuration | 11) | 11) | | configurations of the cleared |
| models) | bk5000 (2300-51, 2300- | bk5000 (2300-51, 2300- | | system K180737 |
| | 61) | 61) | | |
| | bkActiv 2300-56, 2300-66 | | | |
| Mode of Operation | B, M, PW, CFM, P, THI, | B, M, PW, CFM, P, THI, | B, M, PW Doppler, CW | Identical - Primary predicate |
| | CI, SE, CW | CI, SE, CW | Doppler, Color Doppler, | Equivalent - reference |
| | | | Color M Doppler, Power | predicate. |
| | Combination modes : | Combination modes: | Doppler, Harmonic | |
| | 2D+M, 2D+PW, | 2D+M, 2D+PW, | Imaging, Coded Pulse, | |
| | 2D+C+PW, 2D+P+PW, | 2D+C+PW, 2D+P+PW, | 3D/4D Imaging mode, | |
| | 2D+2D, 2D+2D (Biplane | 2D+2D, 2D+2D (Biplane | Elastography, Shear Wave | |
| | Imaging), 2D+(2D+C), | Imaging), 2D+(2D+C), | Elastography (Attenuation | |
| | 2D+(2D+P), 2D+THI, | 2D+(2D+P), 2D+THI, | Imaging and Combined | |
| | 2D+SE, 2D+CI | 2D+SE, 2D+CI | modes: B/M, B/Color, B/ | |
| | | | PWD, B/Color/PWD, | |
| | | | B/Power/PWD) | |
| Intended Use | Intended Use: | Intended Use: | The LOGIQ E10 is a | Identical - Primary predicate |
| | The system is a diagnostic | The system is a diagnostic | general-purpose diagnostic | Equivalent - reference |
| | ultrasound imaging system | ultrasound imaging system | ultrasound system intended | predicate. |
| | used by qualified and…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.