16ch Breast Coil / 16ch Breast Coils, 1.5T 16ch Breast Coil, 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil, 3T 16ch Breast Coil
K222407 · Neocoil, LLC · MOS · Oct 14, 2022 · Radiology
Device Facts
Record ID
K222407
Device Name
16ch Breast Coil / 16ch Breast Coils, 1.5T 16ch Breast Coil, 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil, 3T 16ch Breast Coil
Applicant
Neocoil, LLC
Product Code
MOS · Radiology
Decision Date
Oct 14, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 3T/3.0T MRI system software. The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and/or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures. When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and/or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.
Device Story
Phased array, receive-only RF coils for MRI imaging of breast, axilla, and chest wall; available in 1.5T and 3T configurations. Device consists of medial, lateral, and biopsy arrays, baseplate, and accessories (biopsy grids/drapes). Operates by receiving magnetic resonance signals from hydrogen nuclei in anatomy while blocking high-frequency fields from the MRI scanner; signals are amplified and transferred to the MRI scanner for image reconstruction. Used in clinical settings by trained medical professionals. Output is tomographic images (axial, sagittal, coronal oblique) used for diagnostic or interventional planning. Benefits include high-resolution imaging for clinical decision-making.
Clinical Evidence
No clinical performance testing performed. Substantial equivalence established via non-clinical bench testing, including biocompatibility, electrical safety (IEC 60601-1, 60601-2-33, 60601-1-2), usability, surface temperature, blocking network analysis, B0/B1 field distortion, and NEMA MS 6-2008 image quality metrics (SNR, uniformity).
Technological Characteristics
Phased array, receive-only RF coils. Materials assessed for biocompatibility. Energy source: MRI scanner (no internal power). Dimensions/form factor: contoured for breast anatomy. Connectivity: system cable to MRI scanner. Tuning: 1.5T or 3T field strengths. Includes antenna fuses, preamplifiers, and decoupling circuitry.
Indications for Use
Indicated for use by trained medical professionals to provide diagnostic or interventional planning images of the breast, axilla, and chest wall anatomy in patients undergoing MRI procedures using compatible 1.5T or 3T MRI systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
October 14, 2022
NeoCoil, LLC % Katie Gonzalez Ouality Systems & Regulatory Manager N27 W23910A Paul Road PEWAUKEE WI 53072
Re: K222407
Trade/Device Name: 16ch Breast Coil / 16ch Breast Coils, 1.5T 16ch Breast Coil, 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil, 3T 16ch Breast Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: September 9, 2022 Received: September 15, 2022
Dear Katie Gonzalez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K222407
#### Device Name
16ch Breast Coils: 1.5T 16ch Breast Coil / 3.0T 16ch Breast Coil, 1.5T 16ch Breast Coil / 3T 16ch Breast Coil
Indications for Use (Describe)
The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 373.0T MRI system software.
The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures.
When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:
Image /page/3/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape on the left, with a white curved line inside that forms an arrow pointing to the right. To the right of the shape is the word "NeoCoil" in blue. The font is sans-serif and the letters are bold.
# Special 510(k) Summary
K222407
Applicant
NeoCoil, LLC N27 W23910A Paul Rd Pewaukee, WI 53072 USA
Contact
Katie Gonzalez Quality Systems and Regulatory Manager 262-522-6124 (office) 262-347-1251 (fax) Katie.Gonzalez@neocoil.com
Preparation Date
August 3, 2022
## Name of Device
- . Trade/Proprietary name(s):
### 16ch Breast Coils
| Submission | Device Trade/Proprietary Name |
|----------------|-------------------------------|
| Existing model | 3.0T 16ch Breast Coil |
| Existing model | 1.5T 16ch Breast Coil |
| New model | 3T 16ch Breast Coil |
| New model | 1.5T 16ch Breast Coil |
- Magnetic Resonance Specialty Coil ● Common name:
- . Classification name: 21 CFR 892.1000, Magnetic resonance diagnostic device, Product Code MOS
### Predicate Device
3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017, manufactured by NeoCoil, LLC
### Reference Device
1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018, manufactured by NeoCoil, LLC
### Device Description
The NeoCoil 3.0T 16ch Breast Coil is a phased array coil for imaging structures of the breast, axilla and chest wall. The 3.0T 16ch Breast Coil is a three part receive-only coil designed to provide high resolution imaging. The.3.0T 16ch Breast Coil includes a coil support structure, patient support structure, biopsy components, accessories and comfort pads.
{4}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:
Image /page/4/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue. The "N" and "e" are in a lighter shade of blue, while the "o" is in a darker shade of blue. The "Coil" part of the word is in black.
The NeoCoil 3.0T 16ch Breast Coil is tuned to receive RF frequency corresponding to the proton precession in a 3 tesla magnetic field, which is governed by the Larmor equation.
The 16ch Breast Coils are intended for use in a manner that is identical to the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018. The 16ch Breast Coil consists of the following arrays available in 1.5T, 3.0T and 3T field strengths:
- Medial Array, Breast Coil
- . Lateral Array Left, Breast Coil
- Lateral Array Right, Breast Coil ●
- . Baseplate Assembly, Breast Coil
- Biopsy Array Left, Breast Coil ●
- Biopsy Array Right, Breast Coil ●
The coils receive magnetic resonance signals generated in hydrogen nuclei (protons) in the Head. Neck and Brachial Plexus anatomy while blocking the high-frequency magnetic field applied by the MRI scanner at specified timings. The received signals are amplified before being transferred to the MRI scanner through the coil's system cable. The amplified signals are processed into tomographic images of the breast, axilla and chest wall anatomy by the MRI scanner. Images are typically generated as axial, sagittal, coronal oblique slices. Accessories associated with the 16ch Breast Coils include biopsy grids and biopsy drapes.
The 1.5T Breast Coil, the 3T 16ch Breast Coil, and the 3.0T 16ch Breast Coil are tuned to receive RF frequency corresponding to the proton precession in a 1.5 tesla and 3.0 tesla magnetic field (respectively), which is governed by the Larmor equation.
The 16ch Breast Coil is intended for use in a manner that is identical to the predicate device described in this submission.
Proposed labeling is documented in the Instructions for Use manual for the 16ch Breast Coil (NC149IFU-en).
#### Intended Use, including indications
The NeoCoil 16ch Breast Coils are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended for use by trained medical professionals, in combination with and controlled by compatible 1.5T or 3T/3.0T MRI system software.
The NeoCoil 16ch Breast Coils can be used with compatible ancillary components, accessories, and/or devices to provide access to breast anatomy for diagnostic or interventional planning/procedures.
When used as intended, the 16ch Breast Coil provides information used by the MRI system to produce diagnostic and/or interventional planning images of the breast, axilla, and chest wall. The images produced are interpreted by medical professionals as part of clinical decision making.
#### Technological Characteristics
16ch Breast coils are similar in design, material, chemical composition and energy source to the legally marketed predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
{5}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:
Image /page/5/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with a white crescent moon shape inside, and a black arrow pointing to the right. To the right of the logo is the word "NeoCoil" in blue and black font. The word "Neo" is in blue, and the word "Coil" is in black.
At a high level, the 16ch Breast Coils included as part of this submission, and the predicate device, 3.0T 16ch Breast Coil (K173377) are based on the following same technological elements:
- Prescription use;
- . Coil designs are receive-only phased array coils;
- . Preamplification methodology
- Decoupling methodology
- Patient contacting materials and chemical composition are known materials that have ● been assessed for compliance with recognized biocompatibility standards;
- Energy source for the coils is the MRI scanner; ●
- Energy is not intended to be supplied by the coils; ●
- . Mechanical designs are contoured for the breast anatomy;
- Facilitates imaging for diagnostic and interventional planning of the breast, axilla and chest wall anatomy:
- Channel count
- Field strength when compared to the predicate device and reference device ●
The following technological differences exist between the subject and predicate and reference devices:
- MRI specific system cabling
- MRI specific interface circuitry .
- MRI specific RF loop and preamplifier tuning ●
- Addition of fuses on antenna elements ●
Testing performed per defined standards demonstrates that the safety and/or effectiveness of the 16ch Breast Coils compared to the predicate device is not adversely affected as a result of the differences.
### Testing
Through NeoCoil's design controls process, a risk assessment leveraging FDA's quidance document decision trees for when to submit a new 510(k) for changes to an existing device, a summary of non-clinical performance data using well-established methodologies is included, referenced, or relied on to demonstrate that the 16ch Breast Coils are safe and effective and performs in a manner that demonstrates substantial equivalence to the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
### Performance testing - Bench:
A Test Report Summary for non-clinical Bench Testing performed, including testing to FDArecognized consensus standards identified as relevant in FDA guidance document Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016, is outlined below:
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions /<br>Conclusions |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>Assessment | Assess potential<br>biological risks | Evaluation of<br>data; historical<br>use, biologic<br>testing, where<br>warranted | Acceptable<br>level of risk | Pass | No identified significant<br>risks. |
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions /<br>Conclusions |
| Electrical<br>Safety | Basic electrical<br>safety/essential<br>performance, 60601-1 | Test Lab | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements for basic<br>electrical safety and<br>essential performance<br>met. |
| Electrical<br>Safety | Particular electrical<br>requirements: MR<br>equipment, 60601-2-<br>33 | Test Lab | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements of the<br>particular standard<br>were met. |
| Electrical<br>Safety | Collateral electrical<br>safety/essential<br>performance, 60601-<br>1-2 | Test Lab /<br>Bench Testing | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements of the<br>collateral standard were<br>met. |
| Usability<br>Assessment | Devices meet<br>customer, end user<br>and patient needs | Actual.<br>simulated or<br>retrospective<br>evaluation of<br>the device<br>and/or data | Pre-defined<br>requirements | Pass | The devices met the<br>needs of the customer,<br>end user and patient. |
| Entrapment,<br>Trapping Zone<br>and Cable<br>Looping<br>(assessment w/<br>scanner) | Assess the device for<br>pinch points,<br>entrapment, cable<br>looping - interfacing<br>with MRI scanner | Evaluation of<br>coil-to-scanner<br>entrapment,<br>trapping and<br>cable looping<br>not covered by<br>test lab<br>assessments. | Requirements<br>based on pre-<br>defined<br>requirements<br>in 60601-1<br>and customer<br>requirements | Pass | Requirements were<br>met. |
| Surface<br>Temperature | Surface temperatures<br>do not exceed limits | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | Surface temperatures<br>were within IEC limits. |
| Unplugged<br>Surface<br>Temperature | Devices remain safe in<br>first fault condition | MRI scanner<br>test | Acceptable<br>level of risk | Pass | Surface temperatures<br>were within IEC limits<br>when the coil is left<br>unplugged in the MRI<br>scanner. |
| Blocking<br>Network<br>Analysis | Ensures devices are<br>designed with<br>adequate active and<br>passive transmit<br>decoupling | Theoretical<br>calculations | Adequate<br>transmit<br>decoupling | Pass | Blocking network<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| B1 Field<br>Distortion | Measure amount of<br>distortion produced<br>due to presence of an<br>RF coil in the scanner | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | B1 field inhomogeneity<br>meets performance<br>requirements and<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions /<br>Conclusions |
| B0 Filed<br>Distortion | Measure amount of<br>distortion produced<br>due to presence of an<br>RF coil in the scanner | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | B0 field inhomogeneity<br>meets performance<br>requirements and<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| NEMA MS 6-<br>2008 | Evaluate single-<br>channel non-volume<br>special purpose<br>radiofrequency (RF)<br>coils for use with<br>magnetic resonance<br>(MR) imaging (MRI)<br>systems | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | SNR and Image<br>Uniformity are<br>consistent with the<br>requirements for<br>indications for use. |
{6}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 Fax: (262) 347-1251
Image /page/6/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue. The word "Neo" is in a slightly larger font than the word "Coil".
{7}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 (262) 347-1251 Fax:
Image /page/7/Picture/1 description: The image contains the logo for NeoCoil. To the left of the company name is a black circular logo with an arrow pointing to the right. The company name is to the right of the logo, with "Neo" in blue and "Coil" in black.
## Performance testing - Clinical:
Due to the substantial equivalency of the devices and the evaluation of performance testing against defined standards, no additional clinical performance testing has been performed above that present for the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
No adverse events have been reported against the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
### Conclusion
This submission demonstrates by means of nonclinical testing that the 16ch Breast Coils are substantially equivalent to and perform as well as the predicate device and reference device described in this submission 3.0T 16ch Breast Coil, K173377, cleared on 11/28/2017 and 1.5T 16ch Breast Coil, K182958, cleared on 11/16/2018.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.