K222385 · Anhui Tiankang Medical Technology Co., Ltd. · SCL · Dec 6, 2022 · General Hospital
Device Facts
Record ID
K222385
Device Name
Bifurcated Needle
Applicant
Anhui Tiankang Medical Technology Co., Ltd.
Product Code
SCL · General Hospital
Decision Date
Dec 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bifurcated Needle is intended for use in administering vaccines by the scarification method or administering epidermal allergens.
Device Story
The Bifurcated Needle is a manual, single-use device featuring two prongs designed to hold a small drop of vaccine or allergen. A healthcare provider performs approximately fifteen punctures into the patient's skin to deliver the substance. The device is constructed from SUS 304 stainless steel and includes a PVC needle tip protector. It is supplied sterile, having undergone ethylene oxide (EO) sterilization to achieve a sterility assurance level (SAL) of 10^-6. The device is intended for manual use by healthcare professionals in clinical settings. It serves as a delivery mechanism for vaccines or allergens, with the bifurcated design facilitating the scarification technique.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993), sterilization validation (ISO 11135), shelf-life/stability (ASTM F1980-16), package integrity (ASTM F88, F1929, F1886), and performance testing (ISO 7864, ISO 9626).
Technological Characteristics
Materials: SUS 304 stainless steel needle, PVC tip protector. Dimensions: 64mm length, 2mm width. Principle: Manual scarification. Sterilization: Ethylene Oxide (EO) per ISO 11135. Biocompatibility: ISO 10993 compliant. Standards: ISO 7864 (sterile hypodermic needles), ISO 9626 (stainless steel tubing).
Indications for Use
Indicated for administration of vaccines via scarification or administration of epidermal allergens in patients requiring such procedures.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2022
Anhui Tiankang Medical Technology Co., Ltd. Zhang Yong Management No.228,Weiyi Road, Economic Development Zone Tianchang, Anhui 239300 China
Re: K222385
Trade/Device Name: Bifurcated Needle Regulatory Class: Unclassified Product Code: LDH Dated: November 10, 2022 Received: November 10, 2022
Dear Zhang Yong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan M.
Stevens -
S3
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222385
Device Name Bifurcated Needle
Indications for Use (Describe)
The Bifurcated Needle is intended for use in administering vaccines by the scarification method or administering epidermal allergens.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K222385 510(k) Summary
### l Submitter
Device submitter: Anhui Tiankang Medical Technology Co., Ltd. No.228, Weiyi Road, Economic Development Zone, Tianchang City, Anhui, China.
Contact person: Name: Zhang yong Title: Management representative Phone: +86-13705505106 Fax: +86-550-7309158 E-mail: zy@tkmedical.com
Preparation Date: December 6, 2022
#### II Device
Trade Name of Device: Bifurcated Needle Common Name: System, Delivery, Allergen And Vaccine Regulation Name: None Requlatory Class: Unclassified Product code: LDH Review Panel: General Hospital
### III Predicate Devices
| Trade name: | BD Bifurcated Needle |
|-------------------------|-------------------------|
| Common name: | Bifurcated Needle |
| Classification: | Unclassified |
| Product Code: | LDH |
| Premarket Notification: | K020523 |
| Manufacturer: | BECTON, DICKINSON & CO. |
### IV Device description
The proposed Bifurcated Needle have two prongs. A small drop of smallpox vaccine was placed between the prongs and approximately fifteen punctures would be made into the skin. The needles were sterilized by EO to achieve a SAL of 10-6 and supplied sterile in packaging with a shelf life of five years.
{4}------------------------------------------------
## V Indications for use
The Bifurcated Needle is intended for use in administering vaccines by the scarification method or administering epidermal allergens.
## VI Comparison of technological characteristics with the predicate devices
The Bifurcated Needle has the same intended use, technology, design and biocompatibility is either identical or substantially equivalent to existing legally marketed predicate devices. The comparison between the subject device and the predicate devices are listed in below tables:
| Device<br>feature | Subject Device<br>K222385 | Predicate Device<br>K020523 | Comment |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Product | Bifurcated Needle | BD Bifurcated Needle | / |
| Indications<br>for use | The Bifurcated Needle is<br>intended for use in<br>administering vaccines by<br>the scarification method or<br>administering epidermal<br>allergens. | The BD Bifurcated Needle is<br>intended for use in<br>administering vaccines by<br>the scarification method or<br>administering epidermal<br>allergens. | Same |
| Product code | LDH | LDH | Same |
| Principle of<br>operation | For Manual Use Only, For<br>Single Use Only | For Manual Use Only, For<br>Single Use Only | Same |
| Models | BFN001 | 301754; 301755; 301756;<br>301757 | Difference<br>comment 1 |
| Needle Length | $64mm\pm0.3mm$ Needle | 60.325mm | |
| Needle<br>dimension | $2mm\pm0.05mm$ | 1.905mm | |
| Materials | Bifurcated needle-(SUS<br>304)<br>Needle tip protector-(PVC<br>MT) | Stainless steel | Difference<br>comment 2 |
| Sterilization | EO Sterilization | Sterilization | Difference<br>comment 3 |
#### Table 6-1 Substantial equivalence discussion for Bifurcated Needle
{5}------------------------------------------------
| Device<br>feature | Subject Device<br>K222385 | Predicate Device<br>K020523 | Comment |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Biocompatibility | Contact level: blood path,<br>indirect, limited contact (<24<br>hours).<br>Conforms to the<br>requirement of ISO 10993<br>series Standards | Contact level: blood path,<br>indirect, limited contact (<24<br>hours).<br>Conforms to the<br>requirement of ISO 10993<br>series Standards | Same |
| Labeling | Complied with 21 CFR part<br>801 | Complied with 21 CFR part<br>801 | Same |
# Difference comment 1
The subject device is similar to the predicate device. There are minor differences in needle length and dimension. These differences do not affect intended use. In addition, differences were addressed through ISO 7864 and ISO 9626. Therefore, the differences on models and the specification of needle length and needle width will not raise different question of safety and effectiveness.
# Difference comment 2
The material of the needle of the subject device is similar to the predicated device. They are both stainless steels. The differences on materials do not raise new questions about safety and effectiveness.
# Difference comment 3
The sterilization of the predicate device was unknown while the subject device was sterilized by EO to achieve a SAL of 10-6 and it has been validated to ISO11135. Both devices are sterile.
# Discussion:
From clinical aspect: The subject device has the same indication for use as the predicate.
From technological aspect: The subject device has the same principle of operation, and configuration.
From product performance. The main difference is that specification of models and l the specification of needle length and needle width. However, all of them will be selected by health care provider per injection requirement and this difference does not affect indication for use. Additionally, the performance of needle has been evaluated and the test results met the requirements of ISO 7886-1 and ISO 7864. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.
From material perspective: The biocompatibility of the subject device was evaluated through tests according to the requirements of ISO10993 series standards. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.
{6}------------------------------------------------
From sterilization: The subject device was sterilized by EO to achieve a SAL of 10-6 and it has been validated to ISO11135. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.
### VII Performance data
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility testing
Biocompatibility of the Bifurcated Needle was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communication device – Blood path indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended:
| In Vitro Cytotoxicity Test | ISO 10993-5: 2009 |
|--------------------------------|--------------------|
| Skin Sensitization Test | ISO 10993-10: 2010 |
| In Vitro Hemolysis Test | ISO 10993-4: 2017 |
| Intracutaneous Reactivity Test | ISO 10993-10: 2010 |
| Acute Systemic Toxicity Test | ISO 10993-11: 2017 |
| Pyrogen Test | ISO 10993-11: 2017 |
#### Sterilization and shelf life testing
The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters.
The testing is performed according to the following standards:
| EO residue | ISO 10993-7:2008 |
|-------------|------------------|
| ECH residue | ISO 10993-7:2008 |
The shelf life of 5 year is determined based on stability studies which include ageing test according to FDA recognized standard ASTM F1980-16.
Package integrity testing was conducted on the final, packaged, and sterile devices after environmental conditioning and simulated transportation. All packaging deemed acceptable for protection of product and sterility maintenance.
The testing is performed according to the following standards:
| Seal strength | ASTM F88/F88M-15 |
|------------------------------------|-------------------------|
| Blue Dye Penetration | ASTM F 1929-2015 |
| Seal Integrity (Visual Inspection) | ASTM F 1886/ F 1886M-16 |
### Performance testing
Performance testing is performed according to the following standards:
- A ISO 7864: 2016, Sterile hypodermic syringes for single use
- A ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices-Requirements and test methods.
{7}------------------------------------------------
# VIII Clinical Test Conclusion
No clinical study is included in this submission.
# IX Conclusion
The Bifurcated Needle IS substantially equivalent to its predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.