Xtrallux Alpha (XA136R-USA), Xtrallux Super Plus (XS276R-USA/XS276L-USA), Xtrallux Turbo Pro (XP316R-USA), Xtrallux Extreme RX (XR352R-USA)
K222364 · Xtrallux, LLC · OAP · Nov 16, 2022 · Physical Medicine
Device Facts
Record ID
K222364
Device Name
Xtrallux Alpha (XA136R-USA), Xtrallux Super Plus (XS276R-USA/XS276L-USA), Xtrallux Turbo Pro (XP316R-USA), Xtrallux Extreme RX (XR352R-USA)
Applicant
Xtrallux, LLC
Product Code
OAP · Physical Medicine
Decision Date
Nov 16, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Xtrallux Alpha, Super Plus, Turbo Pro, and Extreme RX are intended to treat Androgenetic Alopecia (AGA) and promote hair growth in males who have Norwood-Hamilton Classifications of hair loss and, females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
Device Story
Non-invasive low-level laser therapy (LLT) devices; utilize red visible light diode lasers at 650 nm wavelength; 5 mW maximum power per diode. Diodes configured within wearable cap; deliver non-thermal energy to scalp for photostimulation of hair follicles. Intended for over-the-counter use by patients at home. Device stimulates hair growth to treat Androgenetic Alopecia. No complex software or algorithmic processing involved.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-5:2009 and ISO 10993-10:2010. Electrical safety and EMC testing per IEC 60601-1:2005, IEC 60601-1-2:2014, IEC 60601-1-6:2010, and IEC 60601-1-11:2015. Laser safety testing per IEC 60825-1:2014. Usability study conducted to confirm consumer self-selection and operation.
Technological Characteristics
Low-level laser therapy (LLT) device; 650 nm wavelength red light diodes; 5 mW max power per diode. Form factor: wearable cap. Standards: ISO 10993-5/10 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60825-1 (laser safety).
Indications for Use
Indicated for males with Norwood-Hamilton Classifications of hair loss and females with Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal hair loss; both with Fitzpatrick Skin Types I to IV, to treat Androgenetic Alopecia and promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG" in a larger font size and "ADMINISTRATION" in a smaller font size below.
November 16, 2022
Xtrallux, LLC % Freeman Anike Principal Consultant Ram+ 2251 San Diego Avenue Suite B-257 San Diego, California 92110
# Re: K222364
Trade/Device Name: Xtrallux Alpha (XA136R-USA), Xtrallux Super Plus (XS276R-USA/XS276L-USA), Xtrallux Turbo Pro (XP316R-USA), Xtrallux Extreme RX (XR352R-USA) Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: October 20, 2022 Received: October 20, 2022
Dear Freeman Anike:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222364
#### Device Name
Xtrallux Alpha (XA136R-USA); Xtrallux Super Plus (XS276R-USA/XS276L-USA); Xtrallux Turbo Pro (XP316R-USA); Xtrallux Extreme RX (XR352R-USA)
#### Indications for Use (Describe)
The Xtrallux Alpha, Super Plus, Turbo Pro, and Extreme RX are intended to treat Androgenetic Alopecia (AGA) and promote hair growth in males who have Norwood-Hamilton Classifications of hair loss and, females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K222364
## DATE PREPARED
November 15, 2022
## MANUFACTURER AND 510(k) OWNER
Xtrallux 7260 NW 58th Street Miami, FL 33166 USA Telephone: Official Contact: Carlos Piña
# REPRESENTATIVE/CONSULTANT
Anike Freeman, MEng, PMP Allison C. Komiyama, PhD, RAC RQM+ +1 (412) 899-7422 Telephone: Email: afreeman@rqmplus.com, akomiyama@rqmplus.com
# DEVICE INFORMATION
| Proprietary Name: | Xtrallux Alpha, Xtrallux Super Plus, Xtrallux Turbo Pro,<br>Xtrallux Extreme RX |
|--------------------|---------------------------------------------------------------------------------|
| Common Name: | Laser, comb, hair |
| Regulation Number: | 21 CFR 890.5500 |
| Class: | II |
| Product Code: | OAP |
| Premarket Review: | General Surgery Devices (DHT4A) |
| Review Panel: | General & Plastic Surgery |
### PREDICATE DEVICE IDENTIFICATION
The Xtrallux devices are substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|--------------------------------------------------------------------------------------------------------------|-------------------|
| K192012 | CapillusX & CapillusX+ by Capillus | ✓ |
| K163170 | Capillus 82, Capillus 202, Capillus 272 Pro, 272 Office<br>Pro, Capillus 302, Capillus 312, and Capillus 352 | |
| K180885 | HairMax Laser by Lexington International, LLC | Reference |
The predicate device has not been subject to a design related recall.
{4}------------------------------------------------
# DEVICE DESCRIPTION
The Xtrallux devices are non-invasive, low level laser therapy (LLT) devices containing red, visible light diode lasers operating at 650 nm wavelength. The maximum power output of each diode is 5 mW. The diodes are configured within a cap, enclosed between the outer shell and inner liner. They deliver non-thermal energy to hair follicles via photostimulation of the scalp and are intended to stimulate hair growth. Xtrallux has developed four models:
- Xtrallux Alpha ●
- Xtrallux Super Plus ●
- Xtrallux Turbo Pro
- Xtrallux Extreme RX ●
# INDICATIONS FOR USE
The Xtrallux Alpha, Super Plus, Turbo Pro, and Extreme RX are intended to treat Androgenetic Alopecia (AGA) and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and, females who have Ludwig (Savin) Scale I-1 to I-4, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Xtrallux believes that the Xtrallux Devices are substantially equivalent to the predicate devices based on the information summarized here:
The subject device has the same intended use and similar design as the predicate devices cleared in K192012 and K163170. The subject device also has the same technological characteristics (laser wavelength, number of laser diodes, energy per laser diode, and total output mode) as the predicate devices.
# SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the Xtrallux devices. The following tests were performed to demonstrate safety based on current industry standards:
# Biocompatibility:
The subject device was evaluated for cytotoxicity, sensitization, and irritation in compliance to ISO10993-5:2009 and ISO 10993-10:2010.
# Electromagnetic Compatibility and Electrical Safety:
The subject device was tested in compliance to IEC 60601-1:2005, IEC 60601-1-2:2014, IEC 60601-1-6:2010, and IEC 60601-1-11:2015
{5}------------------------------------------------
# Performance and Laser Safety:
The subject device was tested in compliance to IEC 60825-1:2014.
# Usability:
Xtrallux conducted a usability study to confirm consumers could appropriately self-select and correctly operate the Xtrallux devices.
The results of these tests indicate that the Xtrallux devices are substantially equivalent to the predicate devices.
# CONCLUSION
Based on the non-clinical testing performed it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed Xtrallux devices are assessed to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.