# JR MEDIC  Nitrile Examination Gloves Powder Free (Orange, Purple) (K222349)

_Jr Engineering & Medical Technologies (M) Sdn. Bhd. · LZA · Oct 21, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222349

## Device Facts

- **Applicant:** Jr Engineering & Medical Technologies (M) Sdn. Bhd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Oct 21, 2022
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

JR MEDIC Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

## Device Story

JR MEDIC Nitrile Examination Gloves are disposable, non-sterile, powder-free patient examination gloves made from acrylonitrile-butadiene copolymer. Used in clinical settings by healthcare professionals to provide a barrier against contamination between the examiner and the patient. The device is worn on the hands; it functions as a physical barrier. No electronic, software, or algorithmic components are involved. The device is provided in orange and purple colors and has a 3-year shelf life. Benefits include protection against cross-contamination during medical examinations.

## Clinical Evidence

Bench testing only. No clinical data was required or provided. Performance was validated against ASTM D6319-2019 (dimensions, physical properties, powder residue), ASTM D5151-2019 (watertightness/holes), and ISO 10993 series (biocompatibility: irritation, sensitization, cytotoxicity, and systemic toxicity).

## Technological Characteristics

Material: Acrylonitrile-butadiene copolymer (nitrile). Form factor: Disposable, powder-free, non-sterile gloves. Dimensions and physical properties comply with ASTM D6319-2019. Biocompatibility tested per ISO 10993-5, 10, and 11. No electronic components, software, or energy sources.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Blue Nitrile Examination Gloves Powder Free ([K192333](/device/K192333.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

October 21, 2022

JR Engineering & Medical Technologies (M) SDN.BHD. % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE114 Aurora, Illinois 60504

Re: K222349

Trade/Device Name: JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 1, 2022 Received: August 3, 2022

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222349

Device Name

JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple)

Indications for Use (Describe)

Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable)                                                                           |                                                                                                                                            |
|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input checked="checked" type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR JR MEDIC NITRILE EXAMINATION GLOVES POWDER FREE (ORANGE, PURPLE)

# 510(k) SUMMARY As required by: 21CFR§807.92(c) K222349

| 510(k Owner's Name               | JR Engineering & Medical Technologies (M) Sdn. Bhd.                                                                                                        |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address                          | Lot 8 &10, Jalan Zurah 3 & Lot 1 & 3, Jalan Zurah<br>3A/1, Pusat Perindustrian 2, 44200 Rasa, Hulu<br>Selangor, Selangor<br>Darul Ehsan, Malaysia.         |
| Phone                            | +603-60572081                                                                                                                                              |
| Fax                              | +603-60572181                                                                                                                                              |
| E-mail                           | ganeshjrmt@gmail.com                                                                                                                                       |
| Contact Person                   | Mr. Ganesan Subramaniam                                                                                                                                    |
| Designation                      | Managing Director                                                                                                                                          |
| Contact Number                   | +6012 224 6677                                                                                                                                             |
| Contact Email                    | ganeshjrmt@gmail.com                                                                                                                                       |
| Official 510k Correspondent      | Manoj Zacharias, Consultant<br>Official Contact for JR Engineering & Medical<br>Technologies (M) Sdn. Bhd                                                  |
| Official 510k Correspondent Firm | Liberty Management Group Ltd<br>75 Executive Dr. STE114, Aurora, IL 60504 USA<br>(630) 270-2921 [voice] (815) 986-2632 [fax]<br>manoj@libertymanagement.us |
| Date Summary Prepared            | 19 OCTOBER 2022                                                                                                                                            |

#### APPLICANT INFORMATION ব

## B. DEVICE IDENTIFICATION

| Name of the device                | JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple) |
|-----------------------------------|------------------------------------------------------------------|
| Product proprietary or trade name | JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple) |
| Common or usual name              | Exam Gloves                                                      |
| Classification name               | Polymer Patient Examination Glove                                |
| Device Classification             | Class-1                                                          |
| Product Code                      | LZA                                                              |
| Regulation Number                 | 21 CFR 880.6250                                                  |
| Review Panel                      | General Hospital                                                 |

#### C. PREDICATE DEVICE

| Predicate Device | Blue Nitrile Examination Gloves Powder Free           |
|------------------|-------------------------------------------------------|
| Manufacturer     | JR Engineering & Medical Technologies (M)<br>SDN.BHD. |
| 510(k) Number    | K192333                                               |
| Regulatory Class | 1                                                     |
| Product code     | LZA                                                   |

{4}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR JR MEDIC NITRILE EXAMINATION GLOVES POWDER FREE (ORANGE, PURPLE)

## D. DESCRIPTION OF THEDEVICE:

JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple) are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are available in Orange & Purple color and are powder free and are provided non-sterile. These gloves have a shelf life for 3 years.

#### E. INTENDED USE OF THE DEVICE:

JR MEDIC Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

| Characteristics                                                                                        | Standards                                                        | Device Performance                                                                                                                                                                                               |                                                                                                                                                                                                          | Comparison<br>Analysis                                                           |                                                                                 |
|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
|                                                                                                        |                                                                  | Predicate<br>K192333                                                                                                                                                                                             | Subject<br>K222349                                                                                                                                                                                       |                                                                                  |                                                                                 |
| Name of device                                                                                         | -                                                                | Blue Nitrile Examination<br>Gloves Powder-free                                                                                                                                                                   | JR MEDIC Nitrile<br>Examination Gloves Powder<br>Free (Orange, Purple)                                                                                                                                   | Similar                                                                          |                                                                                 |
| Manufacturer(s)                                                                                        | -                                                                | JR Engineering & Medical<br>Technologies<br>(M) SDN.BHD. Malaysia.                                                                                                                                               | JR Engineering & Medical<br>Technologies<br>(M) SDN.BHD. Malaysia.                                                                                                                                       | Same                                                                             |                                                                                 |
| Intended use                                                                                           | -                                                                | JR MEDIC Blue Nitrile<br>Examination Gloves Powder-free<br>is disposable devices intended<br>for medical purpose that are wonon the examiner's hand to<br>prevent contamination between<br>patient and examiner. | Nitrile Examination Gloves<br>Powder free is a disposable<br>device intended for medical<br>purpose that are worn on the<br>examiner's hand to prevent<br>contamination between patient<br>and examiner. | Similar                                                                          |                                                                                 |
| Color                                                                                                  | -                                                                | Blue                                                                                                                                                                                                             | Orange, Purple                                                                                                                                                                                           | Different                                                                        |                                                                                 |
| Texture                                                                                                | -                                                                | Finger Texture                                                                                                                                                                                                   | Finger texture                                                                                                                                                                                           | Same                                                                             |                                                                                 |
| Size                                                                                                   | ASTM D6319-<br>2019                                              | Extra - Small, Small, Medium,<br>Large, Extra - Large                                                                                                                                                            | Extra - Small, Small, Medium,<br>Large, Extra - Large                                                                                                                                                    | Same                                                                             |                                                                                 |
| Single Use                                                                                             | Medical Glove<br>Guidance Manual<br>Labeling                     | Single Use                                                                                                                                                                                                       | Single Use                                                                                                                                                                                               | Same                                                                             |                                                                                 |
| Shelf Life                                                                                             | Medical Glove<br>Guidance Manual<br>Labeling                     | 3 Years                                                                                                                                                                                                          | 3 Years                                                                                                                                                                                                  | Same                                                                             |                                                                                 |
| Characteristics                                                                                        | Standards                                                        | Device Performance                                                                                                                                                                                               |                                                                                                                                                                                                          | Comparison                                                                       |                                                                                 |
|                                                                                                        |                                                                  | Predicate<br>K192333                                                                                                                                                                                             | Subject<br>K222349                                                                                                                                                                                       | Analysis                                                                         |                                                                                 |
| Color                                                                                                  |                                                                  | Blue                                                                                                                                                                                                             | Orange, Purple                                                                                                                                                                                           | Different                                                                        |                                                                                 |
| Dimensions                                                                                             | ASTM D6319-<br>2019                                              | Length min 220mm (XS, S)<br>min 230 mm (M, L, XL)<br>Width<br>70 +/-10 mm (XS)<br>80 +/-10 mm (S)<br>95+/-10 mm (M)<br>110+/-10 mm (L)<br>120+/-10 mm (XL)                                                       | Length min 220mm (XS, S)<br>min 230 mm (M, L, XL)<br>70 +/-10 mm (XS)<br>Width<br>80 +/-10 mm (S)<br>95+/-10 mm (M)<br>110+/-10 mm (L)<br>120+/-10 mm (XL)                                               | Same                                                                             |                                                                                 |
| Physical<br>Properties                                                                                 | ASTM D6319-2019                                                  | Before Aging Tensile<br>Strength min 14 Mpa<br>Ultimate Elongation Min<br>500%<br>After Aging Tensile<br>Strength min 14 Mpa<br>Ultimate Elongation Min<br>400%                                                  | Before Aging Tensile<br>Strength min 14 Mpa<br>Ultimate Elongation Min<br>500%<br>After Aging Tensile Strength<br>min 14 Mpa Ultimate<br>Elongation Min 400%                                             | Same                                                                             |                                                                                 |
| Thickness                                                                                              | ASTM D6319-2019                                                  | Palm min 0.05 mm<br>Finger min 0.05 mm                                                                                                                                                                           | Palm min 0.05 mm<br>Finger min 0.05 mm                                                                                                                                                                   | Same                                                                             |                                                                                 |
| Powder Residue                                                                                         | ASTM D6319-2019                                                  | ≤2 mg/glove                                                                                                                                                                                                      | ≤2 mg/glove                                                                                                                                                                                              | Same                                                                             |                                                                                 |
| Watertight<br>(1000 ml)                                                                                | ASTM D5151-2019                                                  | Passes AQL-1.5                                                                                                                                                                                                   | Passes AQL-2.5                                                                                                                                                                                           | Similar                                                                          |                                                                                 |
| Material                                                                                               | ASTM D6319-2019                                                  | Nitrile                                                                                                                                                                                                          | Nitrile                                                                                                                                                                                                  | Same                                                                             |                                                                                 |
| Biocompatibility                                                                                       | Primary Skin<br>Irritation-ISO<br>10993 - 10:2010<br>(E)         | Under the condition of<br>study not an irritant                                                                                                                                                                  | Under the condition of<br>study not an irritant                                                                                                                                                          | Same                                                                             |                                                                                 |
|                                                                                                        | Dermal<br>Sensitization- ISO<br>10993-10:2010 -<br>(E)           | Under the conditions of the<br>study not a sensitizer                                                                                                                                                            | Under the conditions of the<br>study not a sensitizer                                                                                                                                                    | Same                                                                             |                                                                                 |
|                                                                                                        | In vitro cytotoxicity<br>ISO10993-5:2009<br>(E)                  | Under the conditions of the<br>study, cytotoxic                                                                                                                                                                  | Under the conditions of the<br>study, cytotoxic                                                                                                                                                          | Same                                                                             |                                                                                 |
|                                                                                                        | Acute Systemic<br>Toxicity Test ISO<br>10993-11:2017(E)          | Under the conditions of study the<br>device extracts do not pose a<br>systemic toxicity concern                                                                                                                  | Under the conditions of study<br>the device extracts do not pose a<br>systemic toxicity concern                                                                                                          | Same                                                                             |                                                                                 |
| Test Method                                                                                            | Purpose                                                          | Acceptance<br>Criteria                                                                                                                                                                                           | Result                                                                                                                                                                                                   |                                                                                  |                                                                                 |
|                                                                                                        |                                                                  |                                                                                                                                                                                                                  |                                                                                                                                                                                                          |                                                                                  |                                                                                 |
| ASTM D6319-2019<br>Standard Specification for<br>Nitrile Examination Gloves for<br>Medical Application | To determine the<br>length of the gloves                         | Min 220mm (XS, S)<br>Min 230 mm (M, L, XL)                                                                                                                                                                       | X-Small:- 236 mm<br>Small:- 236 mm<br>Medium:- 238 mm<br>Large:- 238 mm<br>X-Large:- 240 mm                                                                                                              |                                                                                  |                                                                                 |
| ASTM D6319-2019<br>Standard Specification for<br>Nitrile Examination Gloves for<br>Medical Application | To determine the<br>width of the gloves                          | X-Small:- 70+/-10 mm<br>Small:- 80+/-10 mm<br>Medium:- 95+/-10mm<br>Large:- 110+/-10 mm<br>X-Large:- 120+/-10 mm                                                                                                 | X-Small- 74 mm<br>Small:- 84 mm<br>Medium:- 94 mm<br>Large:- 104 mm<br>X-Large:- 114 mm                                                                                                                  |                                                                                  |                                                                                 |
| ASTM D6319-2019<br>Standard Specification for Nitrile<br>Examination Gloves for<br>Medical Application | To determine the<br>thickness of the<br>gloves                   | Palm 0.05 mm min<br>Finger 0.05 mm min<br>for all sizes                                                                                                                                                          | Size<br>X-Small<br>Small<br>Medium<br>Large<br>X-Large                                                                                                                                                   | Palm<br>0.18mm<br>0.18mm<br>0.18mm<br>0.18mm<br>0.18mm                           | Finger<br>0.20mm<br>0.20mm<br>0.20mm<br>0.20mm<br>0.20mm                        |
| ASTM D6319-2019<br>Standard Specification for<br>Examination Gloves for<br>Medical Application         | To determine the<br>physical properties -<br>Tensile strength    | Before Aging Tensile<br>Strength 14Mpa Min<br>for all sizes<br><br>After Aging<br>Tensile Strength<br>14Mpa Min for all<br>sizes                                                                                 | Size<br>X-Small<br>Small<br>Medium<br>Large<br>X-Large                                                                                                                                                   | Before<br>Aging<br>22.77 Mpa<br>22.80 Mpa<br>24.46 Mpa<br>24.51 Mpa<br>24.59 Mpa | After<br>Aging<br>20.50 Mpa<br>20.69 Mpa<br>21.28 Mpa<br>21.34 Mpa<br>21.36 Mpa |
|                                                                                                        | To determine the<br>physical properties -<br>Ultimate Elongation | Before Aging<br>Ultimate Elongation<br>500% Min for all<br>sizes<br><br>Before Aging<br>Ultimate Elongation<br>400% Min for all<br>sizes                                                                         | Size<br>X-Small<br>Small<br>Medium<br>Large<br>X-Large                                                                                                                                                   | Before<br>Aging<br>885%<br>886%<br>888%<br>891%<br>892%                          | After<br>Aging<br>760%<br>764%<br>767%<br>769%<br>772%                          |

#### F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

{5}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR JR MEDIC NITRILE EXAMINATION GLOVES POWDER FREE (ORANGE, PURPLE)

{6}------------------------------------------------

SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR JR MEDIC NITRILE EXAMINATION GLOVES POWDER FREE (ORANGE, PURPLE)

There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices meet the ASTM standard D6319-2019.

## G.NON-CLINICAL TESTING SUMMARY OF JR MEDIC NITRILE EXAMINATION GLOVES POWDER FREE

#### PERFORMANCE DATA – ORANGE COLOR

{7}------------------------------------------------

# JR ENGINEERING & MEDICAL TECHNOLOGIES (M) SDN.BHD SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR JR MEDIC NITRILE EXAMINATION GLOVES POWDER FREE (ORANGE, PURPLE)

| Test Method                                                                                            | Purpose                                                         | Acceptance Criteria                                                                                                           | Result                                                                                                                                                                               |
|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D6319-2019<br>Standard Specification for<br>Nitrile Examination Gloves for<br>Medical Application | To determine the<br>length of the gloves                        | Min 220mm (XS, S)<br>Min 230 mm (M, L, XL)                                                                                    | X-Small:- 245 mm<br>Small:- 245 mm<br>Medium:- 246 mm<br>Large:- 248 mm<br>X-Large:- 248 mm                                                                                          |
| ASTM D6319-2019<br>Standard Specification for<br>Nitrile Examination Gloves for<br>Medical Application | To determine the<br>width of the gloves                         | X-Small:- 70+/-10 mm<br>Small:- 80+/-10 mm<br>Medium:- 95+/-10mm<br>Large:- 110+/-10 mm<br>X-Large:- 120+/-10 mm              | X-Small:- 76 mm<br>Small:- 85 mm<br>Medium:- 95 mm<br>Large:- 104 mm<br>X-Large:- 115 mm                                                                                             |
| ASTM D6319-2019<br>Standard Specification for<br>Nitrile Examination Gloves<br>for Medical Application | To determine the<br>thickness of the<br>gloves                  | Palm 0.05 mm min<br>Finger 0.05 mm min<br>for all sizes                                                                       | Size Palm Finger<br>X-Small 0.18mm 0.21mm<br>Small 0.18mm 0.21mm<br>Medium 0.18mm 0.21mm<br>Large 0.18mm 0.21mm<br>X-Large 0.18mm 0.21mm                                             |
| ASTM D6319-2019<br>Standard Specification for<br>Nitrile Examination Gloves<br>for Medical Application | To determine the<br>physical properties-<br>Tensile strength    | Before Aging<br>Tensile Strength 14Mpa<br>Min for all sizes<br>After Aging<br>Tensile Strength 14Mpa<br>Min for all sizes     | Size Before Aging After Ageing<br>X-Small 22.78 Mpa 20.48 Mpa<br>Small 22.82 Mpa 20.67 Mpa<br>Medium 24.47 Mpa 21.30 Mpa<br>Large 24.52 Mpa 21.35 Mpa<br>X-Large 24.58 Mpa 21.36 Mpa |
|                                                                                                        | To determine the<br>Physical properties-<br>Ultimate Elongation | Before Aging<br>Ultimate Elongation<br>500% Min for all sizes<br>After Aging<br>Ultimate Elongation<br>400% Min for all sizes | Size Before Aging After Aging<br>X-Small 884% 759%<br>Small 885% 763%<br>Medium 887% 766%<br>Large 892% 768%<br>X-Large 893% 772%                                                    |
| ASTM D5151-2019<br>Standard Test Method for<br>Detection of Holes in Medical<br>Gloves                 | To determine the holes<br>in the gloves                         | AQL 2.5                                                                                                                       | Gloves Passes AQL 2.5                                                                                                                                                                |
| ASTM D6124-06<br>(Reapproved 2017) Standard<br>Test Method for Residual<br>Powder on Medical Gloves    | To determine the<br>residual powder in<br>the gloves            | 2 Mg/Glove Max                                                                                                                | Size Residual Powder Content<br>X-Small 0.16 mg/glove<br>Small 0.16 mg/glove<br>Medium 0.16 mg/glove<br>Large 0.16 mg/glove<br>X-Large 0.16 mg/glove                                 |

## PERFORMANCE DATA-PURPLE COLOR

{8}------------------------------------------------

#### SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR JR MEDIC NITRILE EXAMINATION GLOVES POWDER FREE (ORANGE, PURPLE)

#### BIOCOMPATIBILITY DATA

| Test Method                                                                                                                                 | Purpose                                                                                                                                                                           | Acceptance<br>Criteria                                                                                | Result                                                                                                                                              |
|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-10<br>Biological Evaluation of<br>Medical Devices Test for<br>Irritation and Skin Sensitization.<br>Test done for irritation.     | To determine the potential of<br>the material under test to<br>produce dermal irritation in<br>Rabbits                                                                            | Under the condition<br>of<br>study<br>not<br>an<br>irritant                                           | Under the condition of<br>study not an irritant                                                                                                     |
| ISO 10993-10<br>Biological Evaluation of Medical<br>Devices Test for Irritation and<br>Skin Sensitization. Test done Skin<br>sensitization. | To<br>the<br>skin<br>determine<br>sensitization potential of the<br>material both in terms of<br>induction and elicitation in<br>Guinea Pig.                                      | Under the conditions<br>of the study not a<br>sensitizer                                              | Under the conditions of<br>the study not a sensitizer                                                                                               |
| ISO 10993-5:2009<br>Biological evaluation of medical<br>devices - part 5, tests for in vitro<br>cytotoxicity.                               | To evaluate the in vitro<br>cytotoxic potential of the test<br>item (both inner and outer<br>surface) Extracts in L-929<br>fibroblasts<br>cells<br>mouse<br>using elution method. | Under the conditions<br>of study non cytotoxic                                                        | Under the conditions of<br>the study, cytotoxic.<br>Additional testing was<br>performed to determine<br>if this was a systemic<br>toxicity concern. |
| ISO 10993-11:2017<br>Biological evaluation of medical<br>devices - part 11, tests for<br>systemic toxicity.                                 | To<br>determine<br>the<br>acute<br>systemic toxicity potential of<br>the test item extracts (both<br>inside and outer surfaces) in<br>Swiss Albino mice.                          | Under the conditions<br>of study the device<br>extracts do not pose a<br>systemic toxicity<br>concern | Under the conditions of<br>study the device<br>extracts do not pose a<br>systemic toxicity<br>concern                                               |

# H. Clinical Testing Summary

Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

# I. CONCLUSION

The conclusions drawn from the non-clinical test demonstrate that the subject device in 510(K) submission, JR MEDIC Nitrile Examination Gloves Powder Free (Orange, Purple) is as safe, as effective, and performs as well as or better than the legally marketed predicate device K192333.

---

**Source:** [https://fda-staging.innolitics.com/device/K222349](https://fda-staging.innolitics.com/device/K222349)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).