K222335 · Luoyang Sunmed Devices Co., Ltd. · FXX · Oct 11, 2022 · General, Plastic Surgery
Device Facts
Record ID
K222335
Device Name
Surgical Mask
Applicant
Luoyang Sunmed Devices Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
Oct 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The product is intended to be worn to protect both the patients and healthcare personnel from transfer of microorganisms, body fluids and particulate materials. These masks are intended for adult use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile/sterile.
Device Story
Surgical mask; three-layer, flat-pleated design; earloop or tie-on styles. Composed of polypropylene spunbond nonwoven outer/inner layers and polypropylene meltblown nonwoven middle layer. Includes iron wire nose piece with polypropylene covering. Held in place by nylon/spandex ear loops or polypropylene nonwoven tie tapes. Used in clinical settings by healthcare personnel and patients for infection control. Acts as a physical barrier to reduce exposure to blood, body fluids, and particulates. Single-use, disposable; provided sterile or non-sterile. Performance meets ASTM F2100 Level 2 and Level 3 requirements.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM F2100-19, ASTM F1862 (fluid resistance), ASTM F2101 (BFE), ASTM F2299 (PFE), EN 14683 (Delta P), and 16 CFR 1610 (flammability). Biocompatibility confirmed via ISO 10993-5 and ISO 10993-10 (cytotoxicity, irritation, sensitization).
Technological Characteristics
Three-layer polypropylene nonwoven construction; iron wire/polypropylene nose piece; nylon/spandex ear loops or polypropylene tie tapes. ASTM F2100 Level 2/3 performance. Sterilization via Ethylene Oxide (EO) to SAL 10^-6. Dimensions: 145mmx95mm or 175mmx95mm.
Indications for Use
Indicated for adult use in infection control practices to protect patients and healthcare personnel from transfer of microorganisms, body fluids, and particulate materials, and to reduce potential exposure to blood and body fluids.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Surgical Face Masks (Sterile), Surgical Face Masks (Non-sterile) (K202843)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 11, 2022
Luoyang Sunmed Devices Co., Ltd. % Ms. Grace Liu Consultant Shenzhen Joyantech Consulting Co. Ltd 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518000 China
Re: K222335
Trade/Device Name: Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 2, 2022 Received: August 3, 2022
Dear Ms. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K222335
Device Name Surgical Mask
#### Indications for Use (Describe)
The product is intended to be worn to protect both the patients and healthcare personnel from transfer of microorganisms, body fluids and particulate materials. These masks are intended for adult use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile/sterile/sterile/sterile/sterile/sterile/sterile/sterile/sterile/ste
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K222335
## 1. Contact Details
### 1.1 Applicant information
| Applicant Name | Luoyang Sunmed Devices Co., Ltd. |
|----------------|-----------------------------------------------------------------------------------------|
| Address | No. 8 Huaxia Road, Hi-Tech Zone, Luoyang Area of China (Henan)<br>Pilot Free Trade Zone |
| Contact person | Tian Qian |
| Phone No. | +86-18637911566 |
| E-mail | qtian@suntech-power.com |
| Date Prepared | 2022-08-02 |
### 1.2 Submission Correspondent
| Image: Company Logo | Shenzhen Joyantech Consulting Co., Ltd |
|---------------------------|--------------------------------------------------------------------------------------------------------------|
| | 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan<br>District, Shenzhen, Guangdong Province, China |
| Image: Chinese Characters | |
| Phone No. | +86-755-86069197 |
| Contact person | Grace Liu; Field Fu; |
| Contact person's e-mail | grace@cefda.com; field@cefda.com |
| Website | http://www.cefda.com |
## 2. Device Information
| Trade name | Surgical Mask |
|---------------------|--------------------|
| Common name | Surgical Face Mask |
| Classification name | Mask, Surgical |
| Review Panel | General Hospital |
| Product code | FXX |
| Device Class | II |
| Regulation No. | 21 CFR 878.4040 |
### 3. Legally Marketed Predicate Device
Primary predicate device
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Trade Name | Surgical Face Masks (Sterile), Surgical Face Masks (Non-sterile) |
|---------------|------------------------------------------------------------------|
| 510(k) Number | K202843 |
| Product Code | FXX |
| Manufacturer | B.J.ZH.F.Panther Medical Equipment Co., Ltd. |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{4}------------------------------------------------
#### Additional predicate device
| Trade Name | Surgical Face Mask (Non-sterile) |
|---------------|-----------------------------------|
| 510(k) Number | K212398 |
| Product Code | FXX |
| Manufacturer | Hubei Kimsoul Industrial Co., Ltd |
#### 4. Device Description
The proposed devices are three-layer, flat pleated masks. It has two design styles, i.e. earloop and tie-on. Each mask is composed of a mask body, a nose piece, two ear loops or four tie tapes.
The mask body is manufactured with three layers, the inner layer and the outer layer are made of polypropylene spunbond nonwoven fabric, and the middle layer is made of polypropylene meltblown nonwoven fabric.
The proposed device is held in place over the user's mouth and nose by two elastic ear loops or four tie tapes welded to the mask body. The elastic ear loops are knitted elastic loops (made of nylon and spandex), and the tie tapes are made of polypropylene nonwoven fabric.
The nose piece is enclosed between the layers of face mask to allow the user to fit the face mask around their nose, which is an iron wire with polypropylene covering.
The proposed devices can simultaneously meet the requirements for the performance class of level 2 and level 3 specified in ASTM F2100.
Both of the two design styles are available non-sterile. They are intended to be single use, disposable devices.
### 5. Intended Use/Indication for Use
The product is intended to be worn to protect both the patients and healthcare personnel from transfer of microorganisms, body fluids and particulate materials. These masks are intended for adult use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile/sterile.
### 6. Technological Characteristics Comparison
| Comparison item | Proposed Device<br>(K222335) | First Predicate Device<br>(K202843) | Second Predicate Device<br>(K212398) | Comment | |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Manufacturer | Luoyang Sunmed<br>Devices Co., Ltd. | B.J.ZH.F.Panther Medical<br>Equipment Co., Ltd. | HUBEI KIMSOUL<br>INDUSTRIAL CO., LTD | None | |
| Product name | Surgical Mask | Surgical Face Masks<br>(Sterile), Surgical Face<br>Masks (Non-sterile) | Surgical Face Mask<br>(Non-sterile) | None | |
| Product Code | FXX | FXX | FXX | Same | |
| Regulation Number | 21 CFR § 878.4040 | 21 CFR § 878.4040 | 21 CFR § 878.4040 | Same | |
| Classification | Class II | Class II | Class II | Same | |
| OTC use | Yes | Yes | Yes | Same | |
| Indications for Use | The product is intended to<br>be worn to protect both<br>the patients and<br>healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids and particulate<br>materials. These masks<br>are intended for adult use<br>in infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single use,<br>disposable device,<br>provided<br>non-sterile/sterile. | The Surgical Face Mask is<br>intended for single use by<br>operating room personnel<br>and other general<br>healthcare workers to<br>protect both patients and<br>healthcare workers against<br>transfer of microorganisms,<br>blood and body fluids, and<br>particulate materials. | The masks are intended<br>to be worn to protect both<br>the patient and<br>healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids and particulate<br>material. These masks<br>are intended for adult use<br>in infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single use,<br>disposable device(s),<br>provided non-sterile. | Similar | |
| Mask style | Flat-pleated, 3 layers | Flat-pleated, 3 layers | Flat-pleated, 3 layers | Same | |
| Design feature | Ear loop/Tie-on | Ear loop/Tie-on | Ear-loop | Same | |
| Single use | Yes | Yes | Yes | Same | |
| Color | Blue | Blue | White | Same | |
| Specifications and<br>dimensions | 1. 145mm×95mm<br>Nose piece:<br>85mm×3.0mm<br>Ear loop: 155mm×3.5mm<br>Tie tape: N.A.<br>2. 175mm×95mm:<br>Nose piece:<br>100mm×3.0mm<br>Ear loop: 190mm×3.5mm<br>Tie tape: 890mm×8.5mm | 1. 14.5cm×9cm:<br>Nose clip:<br>85mm×2.9mm<br>Ear loop: 180mm×3mm<br>Ties: 910mm×10mm<br>2. 17.5cm×9.5cm:<br>Nose clip:<br>100mm×2.9mm<br>Ear loop: 180mm×3mm<br>Ties: 910mm×10mm | 1. 145mm × 95mm<br>Nose piece:<br>95mm×2.8mm<br>Ear loop: 160mm×3.5mm<br>Tie tape: N.A.<br>2. 175mm×95mm<br>Nose piece:<br>105mm×2.8mm<br>Ear loop: 165mm×3.5mm<br>Tie tape: N.A. | Similar | |
| Sterility | Non-sterile/Sterile | Non-sterile/Sterile | Non-Sterile | Same | |
| Sterilization<br>method | EO (SAL: 10-6) | EO (SAL: 10-6) | N.A. | Same | |
| ASTM F2100 Level | Level 2, Level 3 | Level 2 | Level 3 | Different | |
| Labeling | Complied with 21<br>CFR<br>part 801 | Complied with 21 CFR part<br>801 | Complied with 21 CFR<br>part 801 | Same | |
| Materials | | | | | |
| Outer layer | Polypropylene spunbond<br>nonwoven | Polypropylene spunbond<br>nonwoven | Polypropylene spunbond<br>nonwoven | Same | |
| Middle layer | Polypropylene meltblown<br>nonwoven | Polypropylene meltblown<br>nonwoven | Polypropylene meltblown<br>nonwoven | Same | |
| Inner layer | Polypropylene spunbond nonwoven | Polypropylene spunbond nonwoven | Polypropylene spunbond nonwoven | Same | |
| Nose piece | Iron wire with polypropylene covering | Medical polypropylene and Q235 | Polyethylene (PE) | Different | |
| Ear loop | Nylon and spandex | Nylon and spandex | Polyester and Polyurethane | Same | |
| Tie tape | Polypropylene spunbond nonwoven | Polypropylene spunbond nonwoven | N.A. | Same | |
| Performances | | | | | |
| / | Sterile | Non-sterile | Surgical Face Masks (Sterile), Surgical Face Masks (Non-sterile) | Surgical Face Mask (Non-sterile) | / |
| Fluid Resistance (ASTM F1862) | Pass at 120mmHg and 160mmHg | Pass at 120mmHg and 160mmHg | Pass at 120mmHg | Pass at 160 mmHg | |
| Particulate Filtration Efficiency (ASTM F2299) | Average: 98.38% | Average: >99.99% | Average: 98.98% | Average: 99.6% | |
| Bacterial Filtration Efficiency (ASTM F2101) | Average: 99.6% | Average: >99.9% | Average: 98.92% | Average: ≥99.9% | Different |
| Differential Pressure (Delta P) (EN 14683) | Average: 2.7 mmH2O/cm² | Average: 5.5 mH2O/cm² | Average: 4.4 mmH2O/cm² | Average: 3.7 mmH2O/cm² | |
| Flammability (16CFR 1610) | Class 1 | Class 1 | Class 1 | Class 1 | |
| Biocompatibility | ISO 10993-5 and ISO 10993-10; Under the conditions of the study, the proposed device extract was determined to be non-cytotoxic, non-sensitizing, and non-irritating. | ISO 10993-5 and ISO 10993-10; Under the conditions of the study, the proposed device extract was determined to be non-cytotoxic, non-sensitizing, and non-irritating. | ISO 10993-5 and ISO 10993-10; Under the conditions of the study, the proposed device extract was determined to be non-cytotoxic, non-sensitizing, and non-irritating. | ISO 10993-5 and ISO 10993-10; Under the conditions of the study, the proposed device extract was determined to be non-cytotoxic, non-sensitizing, and non-irritating | Same |
Table 1 Technological Characteristics Comparison Table
{5}------------------------------------------------
{6}------------------------------------------------
The proposed device has the similar indication for use as the two predicate devices as well as comparable technical and biocompatibility properties and characteristics, and the differences don't raise any additional questions for safety and effectiveness, therefore, the proposed device is substantially equivalent to the predicate device.
### 7. Summary of Non-clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications as
{7}------------------------------------------------
similar to the predicate device. The tests were conducted according to the following standards, and the results conducted on the representative models, the sterile earloop masks and non-sterile earloop masks with the largest size, demonstrated that the proposed device complies with the following standards (see Table 2 and 3):
- > ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- > ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: Tests for in Vitro cytotoxicity
- > ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin sensitization
- > ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks
- > ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- > ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
- > EN 14683:2019+AC:2019 Medical Face Masks Requirements and Test Methods
- > ASTM F2299/F2299M-03(R2017) Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
- > 16 CFR 1610 Standard for the Flammability of Clothing Textiles
- > ISO 11135:2014 Sterilization of health-care products Ethylene oxide Requirements for the development, validation and routine control of a sterilization process for medical devices
- > ISO 11737-2:2019 Sterilization of medical devices Microbiological methods Part 2 Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- > ISO 10993-7:2008 Biological evaluation of medical devices Part 7 Ethylene oxide sterilization residuals
- > ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Device
- > ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- > ASTM F1886M-2016 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
{8}------------------------------------------------
| | | Table 2 Summary of Performance Testing | | |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------|
| | Purpose | Acceptance<br>Criteria<br>per ASTM<br>F2100-19<br>(AQL=4.0%) | Results<br>(Statistics of three lots, 32 per lot) | |
| Test | | | Sterile erarloop<br>mask, 175*95mm | Non-sterile erarloop<br>mask, 175*95mm |
| Fluid<br>Resistance<br>(ASTM F1862) | Verify the fluid<br>resistance of the<br>proposed device can<br>meet the requirements<br>for Level 2 and 3<br>specified in ASTM<br>F2100-19. | Level 2:<br>Pass at 120<br>mmHg | 96 out of 96 pass at<br>120 mmHg | 96 out of 96 pass at<br>120 mmHg |
| | | Level 3:<br>Pass at 160<br>mmHg | 96 out of 96 pass at<br>160 mmHg | 94 out of 96 pass at<br>160 mmHg¹ |
| Bacterial<br>filtration<br>efficiency (BFE)<br>(ASTM F2101) | Verify the bacterial<br>filtration efficiency of<br>the proposed device<br>can meet the<br>requirements for Level<br>2 and 3 specified in<br>ASTM F2100-19. | Level 2:<br>≥98%<br><br>Level 3:<br>≥98% | 99.4%~99.7%<br>(Average: 99.6%) | >99.9%<br>(Average: >99.9%) |
| Particulate<br>filtration<br>efficiency (PFE)<br>(ASTM F2299) | Verify the particulate<br>filtration efficiency of<br>the proposed device<br>can meet the<br>requirements for Level<br>2 and 3 specified in<br>ASTM F2100-19. | Level 2:<br>≥98%<br><br>Level 3:<br>≥98% | 97.64%~99.11% ²<br>(Average: 98.38%) | 99.88%~>99.9977%<br>(Average: >99.99%) |
| Differential<br>pressure<br>(Delta-P)<br>(EN 14683) | Verify the differential<br>pressure of the<br>proposed device can<br>meet the requirements<br>for Level 2 and 3<br>specified in ASTM<br>F2100-19. | Level 2:<br><6.0<br>mmH₂O/cm²<br><br>Level 3:<br><6.0<br>mmH₂O/cm² | (2.4~3.4)<br>mmH₂O/cm²<br>(Average: 2.7<br>mmH₂O/cm²) | (4.3~5.9)<br>mmH₂O/cm²<br>(Average: 5.5<br>mmH₂O/cm²) |
| Flammability<br>(16 CFR 1610) | Verify the flammability<br>of the proposed device<br>can meet the<br>requirements for Level<br>2 and 3 specified in<br>ASTM F2100-19. | Level 2:<br>Class 1<br><br>Level 3:<br>Class 1 | Class 1 | Class 1 |
Table 2 Summary of Performance Testing
Note:
1 Two samples (one sample per lot in two lots among the three lots) don't meet the acceptance criteria.
2 Only one data (97.64%) doesn't meet the acceptance criteria.
{9}------------------------------------------------
| Test | Purpose | Acceptance Criteria | Result |
|-------------------------------------------|----------------------------------------------------------------|-------------------------------------------------------------------------------------------|--------|
| In vitro<br>Cytotoxicity<br>(ISO 10993-5) | Verify that the proposed device<br>extract is non-cytotoxic. | The extract is non-cytotoxic under<br>the research conditions. | Pass |
| Skin Irritation<br>(ISO 10993-10) | Verify that the proposed device<br>extract is non-irritating. | The polar and non-polar extracts are<br>non-irritating under the research<br>conditions. | Pass |
| Skin Sensitization<br>(ISO 10993-10) | Verify that the proposed device<br>extract is non-sensitizing. | The polar and non-polar extracts are<br>non-sensitizing under the research<br>conditions. | Pass |
### Table 3 Summary of Biocompatibility Testing
### 8. Clinical testing
Clinical testing was not performed for the proposed device.
### 9. Conclusion
The nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed device (K202843 and K212398).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.