K222316 · Shenzhen Goodwind Technology Development Co., Ltd. · OHT · Sep 28, 2022 · General, Plastic Surgery
Device Facts
Record ID
K222316
Device Name
IPL Hair Removal Device
Applicant
Shenzhen Goodwind Technology Development Co., Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Sep 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.
Device Story
Personal, light-based hair reduction device; uses Intense Pulsed Light (IPL) technology to reduce hair growth. Input: electrical power from external adapter; user-initiated via finger switch. Mechanism: Xenon lamp emits light pulses; skin sensor ensures proper contact before emission; ice compress function provides skin cooling. Output: light pulses applied directly to skin. Used by consumers in home settings. Healthcare providers do not operate; output is visual/physical hair reduction. Benefits: non-invasive hair reduction with minimal pain.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-5 and ISO 10993-10. Electrical safety testing per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-83. Photobiological eye safety testing per IEC 62471. Software verification and validation performed for moderate level of concern.
Technological Characteristics
Xenon arc flashlamp energy source; 510-1100nm wavelength; 2.4-4.4J/cm² energy density; 3.6-4.1cm² spot size; 7.2-10.8ms pulse duration. Powered by external adapter. Features skin contact sensor and ice cooling function. Software-controlled. Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-83, and IEC 62471.
Indications for Use
Indicated for removal of unwanted body hair in over-the-counter, home-use settings.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
IPL Home Use Hair Removal Device, D-1198 (K221001)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
September 28, 2022
Shenzhen Goodwind Technology Development CO., LTD % Rain Yip Registration engineer Feiying Drug & Medical Consulting Technical Service Group Contact Address
Re: K222316
Trade/Device Name: IPL Hair Removal Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: July 12, 2022 Received: August 2, 2022
Dear Rain Yip:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222316
Device Name
IPL Hair Removal Device, Models: ZHF-IPL-II, ZHF-IPL-III.
Indications for Use (Describe)
The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## "510(k) Summary" as required by 21 CFR Part 807.92.
#### Date: 2022-07-12
#### I. Submitter
Shenzhen Goodwind Technology Development CO., LTD
4th&5Th Floor,2nd Part Of Building A1, Yinlong Industrial Park, Gate 292, Longgang Section Of Shenshan Road, Longdong Community, Longgang District, Shezhen Guangdong, CN Post code: 518116
Tel.: +86 138 0228 3843
Ziming Yang
Technical director
Tel: +86 138 0228 3843
Email: admin@goodwind.com.cn
#### II. Device
Name of Device: IPL Hair Removal Device
Model(s): ZHF-IPL-II, ZHF-IPL-III
Common or Usual Name: Light Based Over-The-Counter For Hair Removal
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: OHT
Regulation Number: 21 CFR 878.4810
## III. Predicate Device
Predicate device:
510(k) number: K221001
Manufacturer: Shenzhen Beauty Every Moment Intelligent Electric Co.,Ltd.
Trade name: IPL Home Use Hair Removal Device, D-1198
Product code: OHT
Clearance date: May 12, 2022
## IV. Device Description
The IPL Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
{4}------------------------------------------------
The IPL Hair Removal Device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the treated skin, the device will not emit the light pulse. Also, the subject device is equipped with ice compress function to cool down the skin temperature and give a user a better experience, but this function has no other therapeutic use.
The IPL Hair Removal Device includes ZHF-IPL-II and ZHF-IPL-III two models. Their intended use, performance, structure design and operation are basically identical, with the differences mainly in product appearance and its spot size of the treatment window, output energy, and display form, but these differences do not affect or change the intended use of the device. The model difference is embodied in #1) the spot size of the ZHF-IPL-II is 3.6cm2 and the corresponding output energy is 2.6~4.4J/cm²; the spot size of the ZHF-IPL-III is 4.1cm² and the corresponding output energy is 2.4~3.7J/cm²; #2) the display form of ZHF-IPL-II is LCD display screen and indicator light, and the display form of ZHF-IPL-III is indicator bar and light.
#### V. Indications for Use
The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.
#### VI. Comparison of Technological Characteristics With the Predicate Device
The subject device has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the subject device may be found substantially equivalent to its predicate device.
The subject device is compared with the following Predicate Device in terms of intended use, design, material, specifications, and performance:
| Comparison<br>Elements | Subject Device | Predicate device K221001 |
|------------------------|-------------------------------------------------------|--------------------------------------------|
| K Number | K222316 | K221001 |
| Trade name | IPL Hair Removal Device/ZHF-<br>IPL-II, ZHF-IPL-III | IPL Home Use Hair Removal<br>Device/D-1198 |
| Wavelength range | 510-1100nm | 530-1100nm |
| Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp |
| Energy density | ZHF-IPL-II: 2.6~4.4J/cm²<br>ZHF-IPL-III: 2.4~3.7J/cm² | 2.0~4.3J/cm² |
| Spot size | ZHF-IPL-II: 3.6cm²<br>ZHF-IPL-III: 4.1cm² | 2.7cm² |
| Pulse duration | 7.2~10.8ms | 5~12ms |
| Pulsing control | Finger switch | Finger switch |
| Delivery device | Direct illumination to tissue | Direct illumination to tissue |
| Indication<br>for | The IPL Hair Removal Device is | IPL Home Use Hair Removal |
{5}------------------------------------------------
| Comparison<br>Elements | Subject Device | Predicate device K221001 |
|------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| use/Intended use | an over-the-counter device intended for removal of unwanted body hair. | Device is an over-the-counter device intended for removal of unwanted body and/or facial hair. |
| Location for use | OTC | OTC |
#### VII.Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### 1) Biocompatibility Safety
The materials of the patient-directly contacting components of the subject device is performed the biocompatibility evaluation in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document issued on Sep. 4, 2020", as recognized by FDA. The battery of testing was performed to, and passed, including:
- > ISO 10993-5 Biological Evaluation of Medical Devices –Par t 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10 Biological Evaluation of Medical Devices –Par t 10: Tests for Irritation and Skin Sensitization
#### 2) Electrical Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
- > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
- > IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- > IEC 60601-2-83 Medical Electrical Equipment Part 2-83: Particular Requirements For The Basic Safety And Essential Performance Of Home Light Therapy Equipment
## 3) Eye Safety
- > IEC 62471 Photobiological safety of lamps and lamp systems
## 4) Software Verification and Validation
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. Summary
Based on the above performance as documented in this application, the subject device IPL Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device.
#### VIII. Conclusions
{6}------------------------------------------------
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device IPL Hair Removal Device is to be concluded substantial equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.