Al Smile Aligner is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Device Story
Al Smile Aligner consists of custom-made, clear, removable thermoplastic orthodontic appliances. Clinicians take patient impressions and submit them with a prescription to the manufacturer. An FDA-cleared dental software designs a series of 3D models representing the treatment plan; these models are reviewed and approved by a dental professional. Aligners are manufactured by thermoforming thermoplastic sheets over the 3D models. The device is used in a clinical setting under the supervision of a dentist or orthodontist. Patients wear the sequential aligners to apply continuous gentle force to teeth, gradually moving them into alignment. The system benefits patients by providing a discreet, removable method for correcting malocclusion.
Clinical Evidence
Bench testing only. Performance testing included physical, chemical, and mechanical property evaluations following ASTM D570, D638, D790, and D1525. Biocompatibility testing was conducted per ISO 10993, including Ames Assay, cytotoxicity, oral mucosa irritation, delayed-type hypersensitivity, and systemic toxicity.
Technological Characteristics
Customized, patient-specific thermoplastic aligners. Manufactured via thermoforming. Non-sterile. Biocompatibility per ISO 10993. Mechanical properties validated per ASTM D570, ASTM D638, ASTM D790, and ASTM D1525. Utilizes dental software for treatment planning and 3D model design.
Indications for Use
Indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients requiring orthodontic correction.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2022
Nanjing Jiahe Dental Technology Co., Ltd. % Charles Shen Director Manton Business and Technology Services 37 Winding Ridge Oakland. New Jersey 07436
Re: K222308
Trade/Device Name: Al Smile Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: November 22, 2022 Received: November 22, 2022
Dear Charles Shen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
## Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222308
Device Name Al Smile Aligner
Indications for Use (Describe)
"Al Smile Aligner" is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K22308
# Section 5: 510(k) Summary:
This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92
#### Submitter & Foreign Manufacture Identification 5.1
Nanjing Jiahe Dental Technology Co., Ltd. Room 201, Building 2, No.2 Shuanglong Street, Oinhuai District, Naujing, Jiangsu Province, China, Zipcode 210000 Tel: (086) 400-0880468 Submitter's FDA Registration Number: N/A
#### Contact Person 5.2
Charles Shen Manton Business and Technology Services 37 Winding Ridge, Oakland, NJ 07436 Tel: 608-217-9358 Email: cyshen@aol.com
#### Date of Summary: July 11, 2022 5.3
| 5.4<br>Device Name: | |
|------------------------|----------------------------------------------------|
| Proprietary Name: | Al Smile Aligner |
| Common Name: | Dental Plastic Aligner |
| Classification Name: | Orthodontic Plastic Brackets - Sequential Aligners |
| Device Classification: | II |
| Regulation Number: | 21 CFR 872.5470 |
| Panel: | Dental |
| Product Code: | NXC |
#### ર્સ્ટ Predicate Device Information:
K200908, "iSMILE", manufactured by "3D Diagnostix Inc." (1)
#### ર્સ્વ Device Description:
"Al Smile Aligner" consists of a custom-made series of thin, clear plastic removable orthodontic appliances (aligners) that apply gentle pressure to teeth, gradually moving them into alignment.
A dentist or orthodontist assesses the patient to determine if the patient is a good candidate. Impressions are taken by the dental clinician and submitted to Nanjing Jiahe Dental Technology Co., Ltd along with the physician's prescription.
{4}------------------------------------------------
An FDA cleared dental software used for teeth alignment was used to design a series of plastic aligners intended to gradually realign the patient's teeth in accordance to the doctor's prescription.
Once the treatment plan is reviewed and approved by a dental health professional, each 3D model from the treatment plan is manufactured. The aligner travs are then manufactured by thermoforming a dental thermoplastic sheet over each model.
The "Al Smile Aligners" have the same technological characteristics as the predicate device, in that all the devices are made from commercially available plastic that is thermoformed to create a customized, patient-specific aligners are then used for minor tooth movement by way of continuous gentle force.
#### 5.7 Indications for Use:
"Al Smile Aligner" is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
#### ર્સ્ક Summary of Device Testing:
As part of demonstrating substantial equivalence of "Al Smaile Aligner" to the predicate device that are subject to this 510(k) submission, we have completed a number of nonclinical performance tests. "Al Smile Aligner" meets all the requirements for overall design, biocompatibility, and performance results confirming that the design output meets the design inputs and specifications for the device.
Laboratory Testing
- Performance including physical, chemical, and mechanical properties ●
- . Validate the processes used for the design and manufacture of the customized aligners.
- Additionally, mechanical property and elasticity of the thermoformed device were ● studied following ASTM standards such as ASTM D570, ASTM D638, ASTM D790, and ASTM D1525.
Biocompatibility Testing
The device has been tested according to Good Laboratory Practices for its biocompatibility according to ISO 10993, as follows:
- Part 3 (Ames Assay) ●
- Part 5 (Cytotoxicity) ●
- Part 10 (Oral Mucosa Irritation)
- Part 10 (Maximization for Delayed-Type Hypersensitivity) ●
- Part 11 (Systemic Toxicity) ●
{5}------------------------------------------------
#### 5.9 Technological Comparison with Predicate Device
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
| Description | Subject Device | Predicate Device (K200908) | SE<br>Decision |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Brand Name | Al Smile Aligner | iSMILE | SE |
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | SE |
| Product Code | NXC | NXC | SE |
| Classification | Class 2 | Class 2 | SE |
| Indication for<br>Use | "Al Smile Aligner" is indicated for the<br>alignment of teeth during orthodontic<br>treatment of malocclusion. | iSMILE is indicated for the<br>alignment of teeth during<br>orthodontic treatment of<br>malocclusion. | SE |
| Mode of Action | Alignment of teeth by application of<br>continuous gentle force, by sequential<br>use of preformed plastic trays,<br>fabricated based on doctor's<br>prescription. | Alignment of teeth by application<br>of continuous gentle force, by<br>sequential use of preformed<br>plastic trays, fabricated based on<br>doctor's prescription. | SE |
| Materials | Thermoplastic | Thermoplastic | SE |
| Processing | Thermoforming | Thermoforming | SE |
| Dimension | Customized patient specific | Customized patient specific | SE |
| Anatomy<br>Location | Dental mucosal | Dental mucosal | SE |
| Single Use | Yes | Yes | SE |
| Prescription or<br>OTC | Rx | Rx | SE |
| Sterile | Non-sterile | Non-sterile | SE |
| Software Use | Yes | Yes | SE |
| Product Picture | Image: Product Picture | Image: Product Picture | SE |
| Biocompatibility | | | |
| Performance | Meet ISO 10993-1 | Meet ISO 10993-1 | SE |
| | Bench Testing and Process Validation | Bench Testing and Process Validation | SE |
{6}------------------------------------------------
Our device is essentially identical to the predicate device in terms of indications for use, design, material, and processing between our device and the predicate devices.
#### 5.10 Substantial Equivalence Conclusion
It has been shown in this 510(k) submission that "Al Smile Aligner" and its predicate devices have the identical indications for use, similar composition and biocompatibility, similar manufacturing process, and adequate performance.
The difference between the "Al Smile Aligner" and their predicate device do not raise any question regarding its equivalence.
"Al Smile Aligner", as designed and manufactured, is equivalent to its predicate device, and therefore is substantially equivalent as its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.