ARTICUL/EZE Ceramic Heads are zirconia-toughened alumina (ZTA) composite ceramic femoral head prostheses; used as a component in total hip arthroplasty systems. Device replaces diseased or damaged native femoral head; assembled intraoperatively with compatible femoral stem and acetabular component. Available in various outer diameters and offsets; utilizes 12/14 taper interface. Operated by orthopedic surgeons in clinical/hospital settings. Provides a bearing surface for hip joint articulation; restores joint function; reduces pain associated with hip pathologies. Performance verified via mechanical testing (compression, pull-off, torque, wear, impact) and biocompatibility assessment.
Clinical Evidence
No clinical tests were conducted. Substantial equivalence is supported by non-clinical bench testing, including static compression, axial pull-off, torque-off, wear testing per ISO 14242-1:2014, material wear after aging per ISO 6474-2:2019, impact strength per ISO 11491:2017, MRI safety evaluation, and biocompatibility testing per ISO 10993-1:2018.
Technological Characteristics
Material: Zirconia-toughened alumina (ZTA) ceramic compliant with ISO 6474-2:2019. Interface: 12/14 taper. Form factor: Femoral head prosthesis with various outer diameters and offsets. Sterilization: Standardized method (SAL, packaging, and shelf-life equivalent to predicates).
Indications for Use
Indicated for patients with severely painful and/or disabled hip joints due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia; avascular necrosis of the femoral head; acute traumatic fracture of the femoral head or neck; failed previous hip surgery (joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement); or certain cases of ankylosis.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 6, 2023
Depuy Ireland UC Paige Myers Regulatory Affairs Specialist Loughbeg, Ringaskiddy Cork Ireland
Re: K22296
Trade/Device Name: ARTICUL/EZE Ceramic Heads Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: December 2, 2022 Received: December 2, 2022
Dear Paige Myers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Limin Sun -S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
ARTICUL/EZE Ceramic Femoral Heads
Indications for Use (Describe)
#### INDICATIONS FOR USE
Total hip replacement is indicated in the following conditions:
1. A severely painful and/or disabled joint from osteoarthritis, rheumatoid arthritis, or congenital hip dysplasia.
- 2. Avascular necrosis of the femoral head.
3. Acute traumatic fracture of the femoral head or neck.
4. Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface
replacement arthroplasty, or total hip replacement.
5. Certain cases of ankylosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
(As required by 21 CFR 807.92 and 21 CFR 807.93)
| Submitter Information | |
|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Ireland UC |
| Address | Loughbeg, Ringaskiddy<br>Co. Cork Munster, IRELAND |
| Establishment Registration<br>Number | 3015516266 |
| Name of contact person | Paige Myers |
| E-Mail Address | pmyers1@its.jnj.com and<br>DePuySynthesJointsRegulatoryAffairs@its.jnj.com |
| Alternate Contact Person | Clare Hill |
| Alternate Contact E-Mail | Chill7@its.jnj.com |
| Alternate Contact Phone | +44 7795 389956 (UK time zone) |
| Date prepared | 6 January 2023 |
| Name of device | |
| Trade or proprietary name | ARTICUL/EZE Ceramic Heads |
| Common or usual name | Ceramic femoral ball prosthesis |
| Classification name | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous<br>uncemented prosthesis |
| Class | II |
| Classification panel | 87 Orthopedics |
| Regulation | 21 CFR 888.3353 |
| Product Code(s) | LZO |
| Legally marketed device(s) to<br>which equivalence is claimed | Primary Predicate Device: DEPUY CERAMIC FEMORAL HEADS:<br>K031803<br>Secondary Predicate Device: ARTICUL/EZE FEMORAL HEADS:<br>K980513<br>Reference Device: ICONACY I-Hip K151307 |
| Reason for 510(k) submission | The purpose of this 510K submission is to obtain market clearance for<br>ARTICUL/EZE Ceramic Heads |
| Device description | The ARTICUL/EZE Ceramic Heads are a zirconia toughened alumina<br>composite ceramic femoral head designed to be used as one component of a |
| system of prostheses in hip arthroplasty. The femoral heads are available in a<br>wide range of outer diameter sizes and offsets. | |
| Intended use of the device | Total hip arthroplasty |
| Indications for use | Total hip replacement is indicated in the following conditions:<br>1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis,<br>rheumatoid arthritis, or congenital hip dysplasia.<br>2. Avascular necrosis of the femoral head.<br>3. Acute traumatic fracture of the femoral head or neck.<br>4. Failed previous hip surgery including joint reconstruction, internal fixation,<br>arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.<br>5. Certain cases of ankylosis. |
| Performance Testing | The ARTICUL/EZE Ceramic Heads were tested to demonstrate the substantial<br>equivalence to the identified predicate devices. Testing and analyses included:<br>Static Compression (Burst) testing per ISO 7206-10:2018 Axial Pull-off testing per ISO 7206-10:2018 Torque-off testing per ISO 7206-13:2016 Wear testing per ISO 14242-1:2014 Material Wear after Aging testing per ISO 6474-2:2019 Impact strength per ISO 11491:2017 MRI Safety Evaluation testing per ISO 14630:2012 Biocompatibility testing per 10993-1:2018 Bacterial endotoxin per ANSI/AAMI ST 72:2019 |
{4}------------------------------------------------
{5}------------------------------------------------
| SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE | | | | | |
|---------------------------------------------------------------------------------------------|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Characteristics | Subject Device:<br>ARTICUL/EZE Ceramic<br>Heads | Primary Predicate Device:<br>DEPUY CERAMIC<br>FEMORAL HEADS<br>(K031803) | Secondary Predicate Device:<br>ARTICUL/EZE COCR<br>FEMORAL HEADS<br>(K980513) | Reference Device:<br>ICONACY I-Hip (K151307) | |
| Intended Use | Same as primary predicate<br>device | Intended for use in total hip<br>arthroplasty | Same as primary predicate<br>device | Same as primary predicate<br>device | |
| Indications for<br>Use | Same as primary predicate<br>device | Total hip replacement is<br>indicated in the following<br>conditions:<br>1. A severely painful and/or<br>disabled joint from<br>osteoarthritis, traumatic<br>arthritis, rheumatoid arthritis,<br>or congenital hip dysplasia.<br>2. Avascular necrosis of the<br>femoral head.<br>3. Acute traumatic fracture of<br>the femoral head or neck.<br>4. Failed previous hip surgery<br>including joint reconstruction,<br>internal fixation, arthrodesis,<br>hemiarthroplasty, surface<br>replacement arthroplasty, or<br>total hip replacement.<br>5. Certain cases of ankylosis. | Total hip replacement is<br>indicated in the following<br>conditions:<br>1. A severely painful and/or<br>disabled joint from<br>osteoarthritis, traumatic<br>arthritis, or rheumatoid arthritis<br>2. Avascular necrosis of the<br>femoral head.<br>3. Acute traumatic fracture of<br>the femoral head or neck.<br>4. Failed previous hip surgery<br>including joint reconstruction,<br>internal fixation, arthrodesis,<br>hemiarthroplasty, surface<br>replacement arthroplasty, or<br>total hip replacement.<br>5. Certain cases of ankylosis. | 1. A severely painful and/or<br>disabled hip joint as a result of<br>osteoarthritis, traumatic<br>arthritis, rheumatoid arthritis,<br>or congenital hip dysplasia.<br>2. Avascular necrosis of the<br>femoral head.<br>3. Acute traumatic fracture of<br>the femoral head or neck.<br>4. Failed previous hip surgery<br>including joint reconstruction,<br>internal fixation, arthrodesis,<br>hemiarthroplasty, surface<br>replacement arthroplasty, or<br>total hip replacement.<br>5. Certain cases of ankylosis<br>6. Nonunions, correction of<br>functional deformity, and<br>trochanteric fractures of the<br>proximal femur with head<br>involvement that are<br>unmanageable using other<br>techniques. | |
{6}------------------------------------------------
| | Cerasurf-w, a Zirconia-toughened<br>Alumina Ceramic (ZTA)<br>(ISO 6474-2:2019) | CeramTec BIOLOX® delta, a<br>Zirconia-toughened Alumina<br>Ceramic (ZTA)(ISO 6474-2:2019) | CoCrMo (ASTM F1537-11) | Ceramic heads: Cerasurf-p, a<br>Zirconia-toughened Alumina<br>Ceramic (ZTA)(ISO 6474-2:2019) |
|-------------------------|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Material | | | | |
| Outer Diameter<br>Sizes | Same as secondary predicate<br>device | 28mm, 32mm, 36mm | 22.225mm, 28mm, 32mm,<br>36mm, 40mm | 28mm, 32mm, 36mm |
| Taper | Same as primary predicate<br>device | 12/14 ARTICUL/EZE taper | Same as primary predicate<br>device | 12/14 taper |
| Offsets | Same as secondary predicate<br>device | 28mm: +1.5, +5, +8.5<br>32mm: +1, +5, +9<br>36mm: +1.5, +5, +8.5, +12 | 22.225mm: +4, +7<br>28mm: +1.5, +5, +8.5,<br>32mm: +1, +5, +9<br>36mm: -2, +1.5, +5, +8.5, +12<br>40mm: -2, +1.5, +5, +8.5, +12 | 28mm: -4, +0, +4<br>32mm: -4, +0, +4, +7<br>36mm: -4, +0, +4, +8 |
The subject device has the same intended use and inte primary and secondary predicate devices. They are all femoral heads that are intended to replace diseased or broken femoral head and neck bone in combination with a femoral stem prosthesis. They all replace the native femoral head and are assembled intraoperatively with a compatible femoral stem to acetabular component. The subject devices and both predicate devices are intended for total hip arthroplasty and are compatible with 12/14 ARTICULEZE Tapers. The subject device have icentral sizes and offsets. The subject devices have the same method of sterlization. Sterility Asurance Level (SAL), packaging materials, and sheff-life. The ARTICUL/EZE Ceramic Heads are similar to the predicate DEPUY CERAMIC FEMORAL HEADS (K031803) and the predicate ARTICUL/EZE COCR FEMORAL HEADS (K980513) in design. Both the subject device, the primary predicate, and the reference device ICONACY I-Hin (K151307) are made of a zirconia toughened alumina (ZTA) ceramic that is compliant to ISO 6474-2:2019.
The reference device was used to demonstrate biological safety and equivalence for the manufacturing process and raw material of the subject device.
Results of performance testing and analyses demonstrate that the ARTICUL/EZE Ceramic Heads perform as well as the predicate devices.
## SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
No clinical tests were conducted to demonstrate substantial equivalence.
### CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The subject ARTICUL/EZE Ceramic Heads are substantially equivalent to the predicate DEPUY CERAMIC FEMORAL HEADS (K031803) and ARTICUL/EZE COCR FEMORAL HEADS (K980513)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.