K222288 · Terrats Medical SL · NHA · Oct 27, 2022 · Dental
Device Facts
Record ID
K222288
Device Name
DESS Dental Smart Solutions
Applicant
Terrats Medical SL
Product Code
NHA · Dental
Decision Date
Oct 27, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with Ti Base abutments or Pre-milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.
Device Story
DESS Dental Smart Solutions are dental implant abutments used to support prosthetic restorations (crowns, bridges, overdentures). The system includes healing abutments, temporary abutments, Ti Base/C-Base/CrCo Base abutments for CAD/CAM zirconia superstructures, pre-milled blanks, multi-unit abutments (straight/angled), and DESSLoc overdenture attachments. Components are manufactured from titanium alloy (Ti-6Al-4V) or Co-Cr-Mo alloy, with some featuring zirconium nitride (ZrN) coatings or SelectGrip sandblasted surfaces for improved bonding. The device is used by clinicians in dental clinics. Custom abutments are fabricated via CAD/CAM at validated milling centers based on clinician prescriptions. The abutments connect to endosseous implants via specific interfaces; screws (some with DLC coating) secure the abutment to the implant or the prosthesis to the abutment. The device provides a range of prosthetic solutions compatible with various OEM implant systems, enabling clinicians to restore dental function and aesthetics for patients.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation (ISO 17665-1, ISO 14937), biocompatibility (ISO 10993-5, ISO 10993-12), and reverse engineering analysis of OEM implant bodies, abutments, and screws to confirm compatibility. No clinical data were included.
Technological Characteristics
Materials: Ti-6Al-4V (ASTM F136), Co-Cr-Mo (ASTM F1537), Y-TZP (ISO 13356). Surface treatments: SelectGrip (sandblasted), DLC coating (polycrystalline tungsten carbide/carbon/chromium), ZrN coating. Interfaces: Internal and external connections. Dimensions: 2.52-6.5 mm platform diameters. Sterilization: Non-sterile, validated per ISO standards. Connectivity: Mechanical interface to dental implants.
Indications for Use
Indicated for patients requiring prosthetic restoration support in the maxillary or mandibular arch using endosseous dental implants. Compatible with various OEM implant systems as specified in the device labeling.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
10/27/2022
Terrats Medical SL % Melissa Burbage Senior Regulatory Specialist PaxMed International. LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K222288
Trade/Device Name: DESS Dental Smart Solutions Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 29, 2022 Received: July 29, 2022
Dear Melissa Burbage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22288
#### Device Name
DESS Dental Smart Solutions
#### Indications for Use (Describe)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with Ti Base abutments or Pre-milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform Name |
|---------------------------------------------|------------------------------------|-----------------------|
| Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 |
| | 3.0 | 3.0 |
| | 3.6 | 3.6 |
| Astra Tech EV | 4.2 | 4.2 |
| | 4.8 | 4.8 |
| | 5.4 | 5.4 |
| | 3.0 | 3.0 |
| Astra Tech OsseoSpeed™ | 3.5/4.0 | 3.5/4.0 |
| | 4.5/5.0 | 4.5/5.0 |
| | 3.0, 3.4, 3.8 | 3.0 |
| | 3.8, 4.6 | 3.5 |
| BioHorizons | 4.6, 5.8 | 4.5 |
| | 5.8 | 5.7 |
| | 3.25 | 3.4 |
| Biomet 3i Certain® | 4.0 | 4.1 |
| | 5.0 | 5.0 |
| | 3.25 | 3.4 |
| Biomet 3i OSSEOTITE® | 3.75, 4.0 | 4.1 |
| | 5.0 | 5.0 |
| | 3.8 | 3.8 |
| Camlog | 4.3 | 4.3 |
| | 5.0 | 5.0 |
| Dentium SuperLine | 3.6, 4.0, 4.5, 5.0, 6.0, 7.0 | 3.3 |
| | 3.4 | 3.4 |
| FRIADENT XiVE® | 3.8 | 3.8 |
| | 4.5 | 4.5 |
| | 5.5 | 5.5 |
| MegaGen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 |
| Neodent Grand Morse | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | Grand Morse (GM) |
| | 3.0 | 3.0 |
| NobelActive®, NobelParallel Conical | 3.5 | NP |
| | 4.3, 5.0 | RP |
| | 3.5 | NP |
| NobelReplace® Trilobe | 4.3 | RP |
| | 5.0 | WP |
| | 6.0 | 6.0 |
| | 3.3 | NP |
| Nobel Brånemark System® | 3.75, 4.0 | RP |
| | 5.0 | WP |
| Osstem TS | 3.5 | Mini |
| | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular |
| Straumann BLX | 3.5, 3.75, 4.0, 4.5 | RB |
| | 5.0, 5.5, 6.5 | WB |
| Straumann® Bone Level | 3.3 | NC |
| | 4.1/4.8 | RC |
| | 3.3 | NNC, RN |
| Straumann® Tissue Level | 4.1 | RN |
| | 4.8 | RN, WN |
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform Name |
| Zimmer Screw Vent®/ Tapered Screw-<br>Vent® | 3.3, 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| | 6.0 | 5.7 |
| Zimmer Spline | 3.25, | 3.25 |
| | 3.75, 4.0 | 3.75/4.0 |
| | 5.0 | 5.0 |
| Zimmer SwissPlus | 3.7 | 3.8 |
| | 4.8 | 4.8 |
Compatible Implant Systems
{3}------------------------------------------------
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of2
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
# 510(k) Summary
# K22288
# Terrats Medical SL DESS® Dental Smart Solutions
# October 27, 2022
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Terrats Medical SL<br>Carrer Mogoda, 75-99<br>08210 Barberà del Vallès<br>Barcelona, Spain |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone | +34 935 646 006 |
| Official Contact | Roger Terrats, CEO |
| Representative/Consultant | Melissa Burbage, BS<br>Floyd G. Larson, MS, MBA<br>Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 |
| Telephone | +1 858-792-1235 |
| Fax | +1 858-792-1236 |
| Email | mburbage@paxmed.com<br>flarson@paxmed.com<br>kthomas@paxmed.com |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | DESS Dental Smart Solutions |
|------------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Division | DHT1B: Division of Dental Devices |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K170588, DESS Dental Smart Solutions, Terrats Medical SL
Reference Devices
K173908, DESS Dental Smart Solutions, Terrats Medical SL K191986, DESS Dental Smart Solutions, Terrats Medical SL K203464, DESS Dental Smart Solutions, Terrats Medical SL K212577, DESS Dental Smart Solutions, Terrats Medical SL
{5}------------------------------------------------
K212628, DESS Dental Smart Solutions, Terrats Medical SL
K160965, SuperLine, Dentium Co., LTD.
K082639, Dental Tapered SwissPlus Implant, Zimmer Dental, Inc.
K013494, Spline Twist Implant, Sulzer Dental, Inc.
K012055, Spline Twist Implant, Sulzer Dental, Inc.
K142082, Zimmer 3.1mmD Dental Implant System, Zimmer Dental, Inc.
#### INDICATIONS FOR USE STATEMENT
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with Ti Base abutments or Pre-milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform Name |
|-------------------------------------|------------------------------------|-----------------------|
| Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 |
| Ankylos C/X | 3.0 | 3.0 |
| Astra Tech EV | 3.6 | 3.6 |
| Astra Tech EV | 4.2 | 4.2 |
| Astra Tech EV | 4.8 | 4.8 |
| Astra Tech EV | 5.4 | 5.4 |
| Astra Tech OsseoSpeed™ | 3.0 | 3.0 |
| Astra Tech OsseoSpeed™ | 3.5/4.0 | 3.5/4.0 |
| Astra Tech OsseoSpeed™ | 4.5/5.0 | 4.5/5.0 |
| BioHorizons | 3.0, 3.4, 3.8 | 3.0 |
| BioHorizons | 3.8, 4.6 | 3.5 |
| BioHorizons | 4.6, 5.8 | 4.5 |
| BioHorizons | 5.8 | 5.7 |
| Biomet 3i Certain® | 3.25 | 3.4 |
| Biomet 3i Certain® | 4.0 | 4.1 |
| Biomet 3i Certain® | 5.0 | 5.0 |
| Biomet 3i OSSEOTITE® | 3.25 | 3.4 |
| Biomet 3i OSSEOTITE® | 3.75, 4.0 | 4.1 |
| Biomet 3i OSSEOTITE® | 5.0 | 5.0 |
| Camlog | 3.8 | 3.8 |
| Camlog | 4.3 | 4.3 |
| Camlog | 5.0 | 5.0 |
| Dentium SuperLine | 3.6, 4.0, 4.5, 5.0, 6.0, 7.0 | 3.3 |
| Dentium SuperLine | 3.4 | 3.4 |
| FRIADENT XiVE® | 3.8 | 3.8 |
| FRIADENT XiVE® | 4.5 | 4.5 |
| FRIADENT XiVE® | 5.5 | 5.5 |
| MegaGen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 |
| Neodent Grand Morse | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | Grand Morse (GM) |
| NobelActive®, NobelParallel Conical | 3.0 | 3.0 |
| NobelActive®, NobelParallel Conical | 3.5 | NP |
| NobelActive®, NobelParallel Conical | 4.3, 5.0 | RP |
| NobelReplace® Trilobe | 3.5 | NP |
| NobelReplace® Trilobe | 4.3 | RP |
Compatible Implant Systems
{6}------------------------------------------------
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform Name |
|--------------------------------------------|---------------------------|-----------------------|
| | 5.0 | WP |
| | 6.0 | 6.0 |
| Nobel Brånemark System® | 3.3 | NP |
| Nobel Brånemark System® | 3.75, 4.0 | RP |
| | 5.0 | WP |
| Osstem TS | 3.5 | Mini |
| Osstem TS | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular |
| Straumann BLX | 3.5, 3.75, 4.0, 4.5 | RB |
| Straumann BLX | 5.0, 5.5, 6.5 | WB |
| Straumann® Bone Level | 3.3 | NC |
| Straumann® Bone Level | 4.1/4.8 | RC |
| Straumann® Tissue Level | 3.3 | NNC, RN |
| Straumann® Tissue Level | 4.1 | RN |
| Straumann® Tissue Level | 4.8 | RN, WN |
| Zimmer Eztetic | 3.1 | 2.9 |
| Zimmer Screw Vent®/ Tapered<br>Screw-Vent® | 3.3, 3.7, 4.1 | 3.5 |
| Zimmer Screw Vent®/ Tapered<br>Screw-Vent® | 4.7 | 4.5 |
| Zimmer Screw Vent®/ Tapered<br>Screw-Vent® | 6.0 | 5.7 |
| Zimmer Spline | 3.25, | 3.25 |
| Zimmer Spline | 3.75, 4.0 | 3.75/4.0 |
| Zimmer Spline | 5.0 | 5.0 |
| Zimmer SwissPlus | 3.7 | 3.8 |
| Zimmer SwissPlus | 4.8 | 4.8 |
# SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K170588, K173908, K191986, K203464, K212577, and K212628 to:
- include new OEM platform compatibilities (that have not been previously cleared) for previously ● cleared DESS designs,
- include previously-cleared OEM platform compatibilities for previously-cleared DESS abutment ● designs for which the specific combinations of compatibility and design were not cleared previously,
- add angulation to previously cleared Ti Base Abutments, Pre-Milled Blank, Ti, and Multi Unit Abutment,
- include a new abutment design, ZRN Multi Unit Abutment,
- include additional gingival height variations for the Ti Base Interface and DESSLoc Abutment, ●
- include two attachments: Interface Attachment for UniAbutment and Bar Attachment. ●
This submission includes five (5) abutment designs [Healing Abutments, Temporary Abutments, Multi Unit Abutments (straight and angled), ZRN Multi Unit Abutments (straight and angled), and DESSLoc Abutments], one (1) abutment blank (Pre-Milled Blank, Ti), three (3) base designs (Ti Base, C-Base, CrCo Base), two (2) attachments, and twenty (20) screws.
This submission includes one (1) new abutment design (ZRN Multi Unit abutment), based on the previously cleared Multi Unit abutment, but with an added zirconium nitride (Zr\V) coating. The identical coating is used on the DESSLoc Abutments that have been cleared in K170588, and K212628.
This submission includes two attachments. The subject device Interface Attachment for UniAbutment attaches directly to the UniAbutments and is similar to the Interface Attachment for UniAbutment in
{7}------------------------------------------------
K170588. The subject device DESSLoc for Bar is designed to attached directly to the bar. The top portion of this attachment has the same design and material as the DESSLoc abutment cleared in K170588. K191986, and K212628.
The subject device includes the addition of new and previously cleared OEM compatibilities to previously cleared abutment design (Healing Abutments, Temporary Abutments, Ti Base, C-Base, Pre-Milled Blank Ti, Multi Unit abutments, and DESSLoc abutment). New compatibility is introduced for four (4) implant systems for which Terrats Medical has no prior clearance, for a total of eleven (11) new platforms. The direct correlation between each subject device design and the corresponding compatible implant platforms is shown in Summary of Subject Device Components and Implant Platforms for Compatible Implant Systems.
The subject device DESS Dental Smart Solutions provides a range of prosthetic solutions for dental implant restoration. DESS abutments, bases and blanks are offered in a variety of connection types to enable compatibility with currently marketed dental implants. Abutments are provided with the appropriate abutment screw (if applicable) for attachment to the corresponding implant, and the appropriate prosthetic screw (if applicable) for attachment of a screw-retained prosthesis. All abutments and screws are provided non-sterile.
#### Healing Abutment
Healing Abutments are designed to cover the implant connection during the period between implant placement and final abutment placement. Healing Abutments are provided in multiple gingival heights to aid in contouring the gingiva during healing. All healing abutments are marked to identify their gingival height and compatible implant platform. Healing abutments are made of titanium alloy (Ti-6Al-4V). Healing abutments that are the subject of this submission are identical to those cleared under K 170588. K191986, and K212628 but are provided for one (1) additional platform compatibility.
#### Temporary Abutment
Temporary Abutments are designed with horizontal grooves and a vertical flat portion to aid in the fabrication of temporary restorations. They can be reduced in height to adapt to individual patient occlusion. Temporary Abutments are not intended for angle correction. In addition, Temporary Abutments have a SelectGrip® surface to aid in bonding retention. SelectGrip is a sandblasted surface treatment process that significantly improves bonding strength between the treated abutment surface and the cemented prosthesis. Temporary Abutments are available for single-unit and multiple-unit restorations, the former with engaging connections to the implants and the latter with non-engaging connections. Temporary abutments are made of titanium alloy (Ti-6Al-4V). Healing abutments that are the subject of this submission are identical to those cleared under K170588 and K191986 but are provided for one (1) additional platform compatibility.
#### Base Abutments
Subject device Bases, including Ti Base Interface, C-Base, are designed for custom abutment fabrication of a CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Bases is by prescription on the order of the clinician. All zirconia superstructures for use with the subject device Bases will be made at a Terrats Medical validated milling center under FDA quality system regulations, and the material will conform to ISO 13356 Implants for surgery - Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP).
{8}------------------------------------------------
510(k) Summary Page 5 of 11
The design parameters for the CAD/CAM zirconia superstructure to be used on Ti Base Interface compatible with all implants except those listed in the next paragraph are identical to those previously cleared in K170588, K191986, and K212628. They are:
Minimum wall thickness – 0.4 mm Minimum post height - 4.2 mm Minimum gingival height - 0.5 mm Maximum gingival height - 6.0 mm All zirconia superstructures are for straight abutments only.
For the CAD/CAM zirconia superstructure to be used on Ti Base Interface compatible with Biomet 3i Certain, NobelReplace Trilobe, and Zimmer Screw-Vent/Tapered Screw-Vent (except for 3.3 mm implants), the following design parameters may be used:
Minimum wall thickness - 0.4 mm Minimum post height - 4.2 mm Minimum gingival height - 0.5 mm Maximum gingival height - 6.0 mm Maximum angulation of the final abutment - 30°
The design parameters for the CAD/CAM zirconia superstructure to be used on C-Base are identical to those cleared in K203464. They are:
Minimum wall thickness – 0.4 mm Minimum post height - 4.7 mm Minimum gingival height - 0.5 mm Maximum gingival height - 6.0 mm All zirconia superstructures are for straight abutments only.
The design parameters for the CAD/CAM zirconia superstructure to be used on CrCo Base are identical to those cleared in K173908 and K191986. They are:
Minimum wall thickness - 0.4 mm Minimum post height for single-unit restorations - 4.5 mm Minimum gingival height - 0.5 mm Maximum gingival height – 6.0 mm All zirconia superstructures are for straight abutments only.
Pre-Milled Blank, Ti
Pre-milled Blank Abutments are available in engaging designs. They are made of titanium alloy (Ti-6Al-4V). Subject device Pre-milled (Blank) Abutments are identical to Pre-milled Blank Abutments cleared in K170588, K191986, K212628, and K212577.
The design parameters for the CAD/CAM fabrication of custom abutments from Pre-milled Blank Ti compatible with all implants except those listed in the next paragraph are identical to those cleared in K170588, K191986, and K212628. They are:
Minimum wall thickness – 0.45 mm Minimum post height - 4.0 mm Maximum gingival height - 6.0 mm Minimum gingival height - 0.3 mm Pre-Milled Blanks are for straight abutments only
{9}------------------------------------------------
The design parameters for the CAD/CAM Pre-milled Blanks that are compatible with Astra Tech EV (except for 3.0 mm implants), Astra Tech OssesSpeed, Biomet 3i Certain. Nobel Active/Nobel Parallel Conical (except for 3.0 mm implants), NobelReplace Trilobe, Nobel Branemark, Straumann Bone Level, and Zimmer Screw-Vent/Tapered Screw-Vent (except 3.3 mm implants) are identical to those cleared in K212577. They are:
Minimum wall thickness - 0.45 mm Minimum post height - 4.0 mm Maximum gingival height - 6.0 mm Minimum gingival height - 0.3 mm Maximum angulation of the final abutment - 30°
# Multi Unit Abutment
Multi Unit Abutment is designed for attachment of multi-unit screw-retained restorations and is provided in straight and angled designs. The design is the same as that of the Multi Unit Abutments cleared in K170588, K191986, and K212628 and the abutment is made from the same titanium alloy (Ti-6Al-4V). A new abutment line. Multi Unit Abutment ZRN, is being introduced. This subject device abutment has exactly the same design as the Multi Unit Abutment with an added coating of zirconium nitride (ZrN) to the top portion of the abutment. This coating is identical to the ZrN coating used on the DESSLoc Abutments that have been cleared in K170588. K191986, and K212628. Dedicated titanium alloy screws coated with Diamond-like carbon (DLC) are available to attach the abutment to the implant and a dedicated titanium alloy prosthetic screw is available to attach the abutment. The subject device Interface Attachment for UniAbutment attaches directly to the UniAbutments and is similar to the Interface Attachment for UniAbutment in K170588.
#### DESSLoc® Abutment
DESSLoc Abutment is designed for overdenture attachment. It is identical to the DESS LOC Abutment cleared in K170588, K191986, and K212628, except that additional gingival heights and new OEM Combabilities are provided. DESSLoc Abutments are made of titanium alloy (Ti-6Al-4V) and have a zirconium nitride (ZrN) coating.
#### Screws
DESS Dental Smart Solutions screws are designed to attach the implant or the prosthesis to the abutment. There are twenty (20) new screws for the subject device components with similar designs to those of screws cleared in K170588, K173908, K191986, K203464, K212577, and K212628. Two (2) of these new screws are variations that are being used with the already cleared components. The subject device screw 19.349 has same connection and thread mating as screws cleared in K191986, and 19.102 has the same connection and thread mating as screws cleared in K170588. Screws are available with and without a DLC (Diamond-like Carbon) coating is a polycrystalline tungsten carbide/carbon and chromium coating that is identical to the DLC coating on screws cleared in K170588.
{10}------------------------------------------------
510(k) Summary Page 7 of 11
| Compatible Implant<br>Systems | DESS<br>Abutment<br>System | Healing Abutment | Temporary Abutment | Ti Base Interface,<br>Engaging and<br>non-engaging | C-Base, Engaging and<br>non-engaging | CrCo Base, Engaging and<br>non-engaging | Pre-milled Blank Ti,<br>Engaging | Multi Unit Abutment | Multi Unit ZRN | DESSLoc<br>Abutment | Screws |
|-------------------------------------------|----------------------------|----------------------------|-----------------------|----------------------------------------------------|-----------------------------------------|--------------------------------------------|------------------------------------------|------------------------------------------------------|------------------------------------------------------|-----------------------------|--------|
| Ankylos C/X | Internal<br>ANK | 2.52 | 2.52 | 2.52<br>(0°) | 2.52<br>(0°) | 2.52<br>(0°) | 2.52<br>(0°) | | | 2.52 | X |
| Astra Tech EV | Conic EVO | 3.0, 3.6,<br>4.2, 4.8, 5.4 | 3.6, 4.2, 4.8 | 3.0, 3.6,<br>4.2, 4.8, 5.4<br>(0°) | 3.6, 4.2, 4.8<br>(0°) | 3.6, 4.2, 4.8<br>(0°) | 3.0 (0°),<br>3.6, 4.2, 4.8,<br>5.4 (30°) | 3.0 (0°), 3.6,<br>4.2, 4.8, 5.4<br>(0°, 17°,<br>30°) | 3.0 (0°), 3.6,<br>4.2, 4.8, 5.4<br>(0°, 17°,<br>30°) | 3.6, 4.2,<br>4.8 | X |
| Astra Tech<br>OsseoSpeed | Internal Hex<br>Conic | 3.5/4.0,<br>4.5/5.0 | 3.5/4.0,<br>4.5/5.0 | 3.5/4.0,<br>4.5/5.0<br>(0°) | 3.5/4.0,<br>4.5/5.0<br>(0°) | 3.5/4.0,<br>4.5/5.0<br>(0°) | 3.0 (0°),<br>3.5/4.0,<br>4.5/5.0 (30°) | 3.5/4.0,<br>4.5/5.0<br>(0°, 17°,<br>30°) | 3.5/4.0,<br>4.5/5.0<br>(0°, 17°,<br>30°) | 3.0,<br>3.5/4.0,<br>4.5/5.0 | X |
| BioHorizons Internal | Internal Hex<br>BH | 3.5, 4.5, 5.7 | | 3.5, 4.5, 5.7<br>(0°) | 3.5, 4.5, 5.7<br>(0°) | 3.5, 4.5, 5.7<br>(0°) | 3.5, 4.5, 5.7<br>(0°) | 3.5, 4.5<br>(0°) | 3.5, 4.5<br>(0°) | 3.5, 4.5,<br>5.7 | X |
| Biomet 3i Certain | Internal Hex<br>"Click" | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0<br>(30°) | 3.4, 4.1, 5.0<br>(0°) | 3.4, 4.1, 5.0<br>(0°) | 3.4, 4.1, 5.0<br>(30°) | | | 3.4, 4.1,<br>5.0 | X |
| Biomet 3i<br>OSSEOTITE | External Hex<br>USA | 3.4, 4.1, 5.0 | 3.4. 4.1. 5.0 | 3.4, 4.1, 5.0<br>(0°) | 3.4, 4.1, 5.0<br>(0°) | 3.4, 4.1, 5.0<br>(0°) | 3.4, 4.1, 5.0<br>(0°) | 3.4, 4.1, 5.0<br>(0°) | 3.4, 4.1, 5.0<br>(0°) | 3.4, 4.1,<br>5.0 | |
| Camlog | Internal<br>CAM | 3.8. 4.3. 5.0 | 3.8. 4.3. 5.0 | 3.8, 4.3, 5.0<br>(0°) | | 3.8, 4.3, 5.0<br>(0°) | 3.8, 4.3, 5.0<br>(0°) | | | 3.8, 4.3,<br>5.0 | |
| Dentium SuperLine | DENT | | | | | | | 3.3<br>(0°) | 3.3<br>(0°) | | |
| FRIADENT XiVE | Internal Hex<br>FD | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 | 3.4. 3.8,<br>4.5, 5.5<br>(0°) | 3.4. 3.8. 4.5<br>(0°) | 3.4, 3.8, 4.5<br>(0°) | 3.4, 3.8, 4.5<br>(0°) | | | 3.4, 3.8,<br>4.5 | |
| Megagen AnyRidge | Conic Anyr | | | 3.5<br>(0°) | 3.5<br>(0°) | | | | | | X |
| Neodent Grand<br>Morse | Neo GM | | | Grand<br>Morse<br>(0°) | Grand<br>Morse<br>(0°) | | | Grand Morse<br>(0°, 17°,<br>30°) | Grand<br>Morse<br>(0°, 17°,<br>30°) | | X |
| NobelActive®,<br>NobelParallel<br>Conical | Active Hex | NP, RP | NP, RP | 3.0, NP,<br>RP, WP<br>(0°) | NP, RP,<br>WP<br>(0°) | NP, RP<br>(0°) | 3.0 (0°), NP,<br>RP, WP<br>(30°) | NP, RP<br>(0°, 17°,<br>30°) | NP, RP<br>(0°, 17°,<br>30°) | NP, RP | X |
| NobelReplace<br>Trilobe | Tri-lobe | NP, RP,<br>WP | NP, RP, WP | NP, RP,<br>WP, 6.0<br>(30°) | NP, RP,<br>WP, 6.0<br>(0°) | NP, RP,<br>WP (0°) | NP, RP, WP,<br>6.0 (30°) | NP, RP, WP<br>(0°) | NP, RP, WP<br>(0°) | NP, RP | X |
| Nobel Branemark<br>System | External Hex<br>Universal | NP, RP,<br>WP | NP, RP, WP | NP, RP,<br>WP (0°) | NP, RP,<br>WP (0°) | NP, RP,<br>WP (0°) | NP, RP, WP,<br>6.0 (30°) | NP, RP, WP<br>(0°) | NP, RP, WP<br>(0°) | NP, RP,<br>WP | |
| Compatible Implant<br>Systems | DESS<br>Abutment<br>System | Healing<br>Abutment | Temporary<br>Abutment | Ti Base<br>Interface, Engaging and<br>non-engaging | C-Base,<br>Engaging and<br>non-engaging | CrCo Base,<br>Engaging and<br>non-engaging | Pre-milled<br>Blank Ti,<br>Engaging | Multi Unit<br>Abutment | Multi Unit ZRN | DESSLoc<br>Abutment | Screws |
| Osstem TS | Conic OSS | | Mini,<br>Regular | Mini,<br>Regular<br>(0°) | Mini,<br>Regular<br>(0°) | Mini,<br>Regular<br>(0°) | Mini,<br>Regular<br>(0°) | Mini,<br>Regular<br>(0°) | Mini,<br>Regular<br>(0°) | | |
| Straumann BLX | Conical<br>BLX | RB/WB,<br>WB | | RB/WB,<br>WB (0°) | | | RB/WB, WB<br>(0°) | RB/WB<br>(0°, 17°,<br>30°) | RB/WB<br>(0°, 17°,<br>30°) | | X |
| Straumann Bone<br>Level | Conical BL | NC, RC | NC, RC | NC, RC<br>(0°) | NC, RC<br>(0°) | NC, RC<br>(0°) | NC, RC<br>(30°) | NC, RC<br>(0°, 17°,<br>30°) | NC, RC<br>(0°, 17°,<br>30°) | NC, RC | X |
| Straumann Tissue<br>Level | Octagon | RN, WN | RN, WN | NNC, RN,<br>WN (0°) | NNC, RN,<br>WN (0°) | RN, WN<br>(0°) | NNC, RN,<br>WN (0°) | | | RN, WN | X |
| Zimmer Eztetic | Internal EZ | | | | | | | | | 2.9 | |
| Zimmer Screw Vent /<br>Tapered Screw Vent | Internal Hex<br>USA | 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7<br>(30°) | 3.5, 4.5, 5.7<br>(0°) | 3.5, 4.5, 5.7<br>(0°) | 3.5, 4.5, 5.7<br>(30°) | 3.5<br>(0°, 17°,<br>30°)<br>4.5<br>(0°, 17°,<br>30°) | 3.5, 4.5<br>(0°, 17°,<br>30°) | 3.5, 4.5,<br>5.7 | X |
| Zimmer Spline | External<br>SPL | | | | | | | | | 3.25,<br>3.75/4.0,<br>5.0 | |
| Zimmer SwissPlus | Internal<br>Swiss | | | | | | | | | 2.5, 3.0 | |
# Summary of Subject Device Components and Implant Platforms for Compatible Implant Systems
{11}------------------------------------------------
Bold indicates new components
{12}------------------------------------------------
## COMPATIBLITIES
All compatibilities are identical to those of the primary predicate K170588 and additional predicate K173908, K191986, K203464, K212577, K212628, except for the addition of compatibilities with four (4) new systems, eleven (11) platforms, for Dentium SuperLine, Zimmer Eztetic, Zimmer Spline, and Zimmer SwissPlus.
| Compatible Implant<br>System | Implant<br>510(k) | Implant 510(k) Name | Manufacturer |
|-------------------------------------------|-------------------|---------------------------------------------------------------------------|--------------------------------------------------------|
| Ankylos C/X | K140347 | Ankylos C/X Implant System | DENTSPLY International, Inc. |
| Astra Tech EV | K111287 | Astra Tech Implant System | Astra Tech AB |
| | K120414 | OsseoSpeed Plus | |
| Astra Tech OsseoSpeed | K101732 | OsseoSpeed™ | Astra Tech AB |
| BioHorizons Internal | K042429 | The Prodigy System Dental Implants | BioHorizons Implants System, Inc. |
| | K071638 | BioHorizons Tapered Internal Implant System | |
| Biomet 3i Certain® | K063341 | 3i OSSEOTITE Certain® Dental Implants | Implant Innovations, Inc. |
| Biomet 3i OSSEOTITE | K063286 | OSSEOTITE® Dental Implants | Implant Innovations, Inc. |
| Camlog | K083496 | Camlog Implant System | Altatech GmbH |
| Dentium SuperLine | K160965 | SuperLine | Dentium Co., Ltd |
| Friadent XiVE | K073075 | Friadent Implant Systems | DENTSPLY International, Inc. |
| MegaGen AnyRidge | K110955 | AnyRidge Internal Implant System | Megagen Co., Ltd. |
| Neodent Grand Morse | K163194 | Neodent Implant System - GM Line | JJGC Industria E Comercio De<br>Materials Dentarios SA |
| NobelActive,<br>NobelParallel Conical | K142260 | NobelActive® | Nobel Biocare AB |
| | K102436 | NobelActive 3.0 | |
| | K073142 | NobelReplace Hexagonal Implants | |
| Nobel-Replace Tri-Lobe | K050705 | TiUnite Implants | Nobel Biocare AB |
| | K050406 | NobelSpeedy Implants | |
| Nobel Brånemark | K022562 | Various Brånemark System Implants-<br>Immediate Function Indication | |
| OSSTEM TS | K161604 | Osstem Implant System | OSSTEM IMPLANT Co., Ltd |
| Straumann BLX | K173961 | Straumann BLX Implant System | Institut Straumann AG |
| Straumann Bone Level | K140878 | Straumann® Bone Level Tapered Implants | Straumann USA, LLC |
| Straumann Tissue Level | K130222 | Straumann® Dental Implant System SLActive and<br>Roxolid Product Families | Straumann USA, LLC |
| Zimmer Eztetic | K142082 | Zimmer 3.1mm Dental Implant | Zimmer Dental, Inc. |
| Zimmer Screw-Vent /<br>Tapered Screw-Vent | K011028 | Screw-Vent Dental Implant System | Sulzer Dental, Inc. |
| | K112160 | Tapered Screw-Vent® X Implant | Zimmer Dental, Inc. |
| Zimmer Spline | K012055 | 3.25mm Spline Twist Implant | Sulzer Dental, Inc. |
| | K013494 | 3.75mm and 5.0mm Spline Twist Implant | |
| Zimmer SwissPlus | K082639 | Zimmer Tapered SwissPlus Implants | Zimmer Dental, Inc. |
| | | Compatible Implant Systems |
|--|--|----------------------------|
| | | |
# MATERIAL COMPOSITION
All subject device abutments and screws are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) or from Co-Cr-Mo alloy (CrCo) conforming to ASTM F1537 Standard Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539).
{13}------------------------------------------------
Zirconia superstructures for Ti Base, C-Base, and CrCo Base are made of Y-TZP conforming to ISO 13356 Implants for surgery – Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP). The cement recommended in labeling for bonding of superstructures is Multi Link Hybrid Abutment Cement from Ivoclar Vivadent AG, cleared under K130436. All of these materials are equivalent to those cleared in K170588, K173908, K191986, and K203464.
All subject device components are manufactured from the same materials, are treated with the same surface treatments (SelectGrip surface, DLC coating and anodization), and are manufactured in the same facilities using the same manufacturing processes as corresponding devices cleared in K170588, K173908. K191986. K203464. K212577. and K212628.
# PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence included: sterilization validation according to ISO 17665-1 and ISO 14937, and biocompatibility according to ISO 10993-5 and ISO 10993-12, leveraged from K170588, K173908, K191986, K203464, K212577, and K212628; and reverse engineering analysis of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility for new OEM connections. No clinical data were included in this submission.
# EQUIVALENCE TO MARKETED DEVICES
All subject device abutments are identical in design, materials and technological characteristics to corresponding abutments of the primary predicate K170588 and additional predicates K173908, K191986. K203464. K212577. K212628. The subject device Multi Unit Abutment ZRN are substantially equivalent to Multi Unit abutments cleared in primary predicate K170588 and additional predicates K 191986, and K212628, except that it has a zirconium nitride (ZrN) coating added to the top portion of the abutment. This coating is used on the DESSLoc Abutments that have been cleared in primary predicate K170588 and additional predicates K191986 and K212628.
The SelectGrip® surface on Ti Base Interface is identical to the SelectGrip surface on equivalent abutments cleared in primary predicate K170588. The ZrN coating on DESSLoc and Multi Unit ZRN Abutment is identical to that on DESS LOC Abutments cleared in primary predicate K170588 and additional predicate K191986 and K212628.
The cement recommended in labeling for bonding of superstructures is Multi-Link cement from Ivoclar Vivadent, cleared under K130436. This is the same cement recommended in labeling for the primary predicate K170588 and the additional predicates K173908, K191986, and K212628.
All screws are identical in design, materials and technological characteristics to those cleared in primary predicate K170588 and additional predicates K173908, K203464, K212577, K212628 except for threads that accommodate the new compatibilities. Diamond-like carbon (DLC) coatings that are available on certain screws are identical to those on screws cleared in primary predicate K170588 and additional predicates K173908, K191986, K203464, K212577, K212628.
#### CONCLUSION
The subject device, the primary predicate device, and the additional predicate devices have the same intended use, have similar technological characteristics, and are made of the same materials. The subject device, the primary predicate, and additional predicate devices encompass the same range of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{14}------------------------------------------------
| Comparison | Subject Device | Primary Predicate Device | Additional Predicate Devices |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DESS Dental Smart Solutions<br>Terrats Medical SL | K170588<br>DESS Dental Smart Solutions<br>Terrats Medical SL | K173908, K191986, K203464,<br>K212577, K212628<br>DESS Dental Smart Solutions<br>Terrats Medical SL |
| Indications | DESS Dental Smart Solutions<br>abutments are intended to be used in<br>conjunction with endosseous dental<br>implants in the maxillary or<br>mandibular arch to provide support<br>for prosthetic restorations.<br><br>All digitally designed custom<br>abutments for use with DESS Bases<br>or Blanks are to be sent to a Terrats<br>Medical validated milling center for<br>manufacture.<br><br>For complete Indications for Use<br>statement on OEM Compatibility<br>see Section 4. | DESS Dental Smart Solutions<br>abutments are intended to be used<br>in conjunction with endosseous<br>dental implants in the maxillary or<br>mandibular arch to provide<br>support for prosthetic restorations.<br><br>All digitally designed custom<br>abutments for use with DESS<br>Bases or Blanks are to be sent to a<br>Terrats Medical validated milling<br>center for manufacture.<br><br>For complete Indications for Use<br>statement on OEM Compatibility<br>see 510(k) Summary for K170588<br>in Section 12. | DESS Dental Smart Solutions<br>abutments are intended to be used<br>in conjunction with endosseous<br>dental implants in the maxillary or<br>mandibular arch to provide<br>support for prosthetic restorations.<br><br>All digitally designed custom<br>abutments for use with DESS<br>Bases or Blanks are to be sent to a<br>Terrats Medical validated milling<br>center for manufacture.<br><br>For complete Indications for Use<br>statement on OEM Compatibility<br>see 510(k) Summaries for<br>K173908, K191986, K203464,<br>K212577, K212628 in Section 12. |
| Design | | | |
| Designs | Healing, Temporary Abutment, Ti<br>Base, C-base, CrCo Base, Pre-<br>milled Blank, Multi Unit Abutment,<br>Multi Unit ZRN Abutment,<br>DESSLoc | Healing, Temporary, Straight,<br>Multi-unit, Locator-type,<br>CAD/CAM Bases, CAD/CAM<br>Blanks | Healing, Temporary, Straight,<br>Uniabutment, Multi-unit,<br>DESSLoc, CAD/CAM Bases,<br>CAD/CAM Blanks,Ti Bases,<br>CrCo Base, Pre-milled (Blank) |
| Prosthesis Attachment | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained |
| Restoration | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit |
| Abutment/Implant<br>Platform<br>Diameter, mm | 2.52 - 6.5 | 2.52 - 6.5 | 2.3 - 6.0 |
| Prosthetic Platform<br>Diameter, mm | 4.5-6.5 | 4.5 | 4.0 - 6.5 |
| Abutment Angle | 0°, 17°, 30° | Straight (0°) | 0°, 17°, 30° |
| Abutment/ Implant<br>Interface | Internal, External | Internal, External | Internal, External |
| Material | | | |
| Abutments | Ti-6Al-4V ELI<br>Co-Cr-Mo Alloy | Ti-6Al-4V ELI | Ti-6Al-4V ELI<br>Co-Cr-Mo Alloy |
| Screws | Ti-6Al-4V ELI<br>DLC coating | Ti-6Al-4V ELI<br>DLC coating | Ti-6Al-4V ELI<br>DLC coating |
#### Table of Substantial Equivalence
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.