LOSPA TKR System Instrumentation, EXULT TKR System Instrumentation
Applicant
Corentec Co., Ltd.
Product Code
JWH · Orthopedic
Decision Date
Aug 26, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Indications for Use
Corentec knee instruments are accessory devices and are intended to be used to assist in the implantation of LOSPA Knee Systems (Primary, Modular), EXULT Knee Replacement System in their cleared Indications for Use as provided below: LOSPA Knee System, EXULT Knee Replacement System are intended for the treatment of diseases as follows: · Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis; · Post-traumatic loss of knee joint configuration and function; · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability; · Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. The LOSPA, EXULT Knee Replacement System are intended for cemented application only.
Device Story
Surgical instruments for LOSPA Modular Knee System and EXULT Knee System; used as line extension of existing reusable orthopedic knee instruments. Includes implant-specific trials, cutting guides, alignment instruments, and AP sizer external rotation instruments. Facilitates total knee arthroplasty (TKA) procedures by assisting in implant placement and bone preparation. Used in clinical/OR settings by orthopedic surgeons. Instruments are non-sterile; require sterilization before use. Output is physical guidance and positioning for implant components; assists surgeon in achieving proper joint alignment and stability.
Clinical Evidence
No clinical or animal testing were required. Substantial equivalence is supported by non-clinical bench testing, including dimension measurement and assembly testing.
Technological Characteristics
Reusable orthopedic surgical instruments for TKR. Materials are consistent with previously cleared Corentec instrumentation. Supplied non-sterile; intended for steam sterilization. Mechanical design includes guides, trials, and sizers for bone preparation and implant positioning. No software or energy source components.
Indications for Use
Indicated for patients with painful, disabling knee joint disease (osteoarthritis, traumatic arthritis, avascular necrosis, or rheumatoid arthritis), post-traumatic loss of joint function, or moderate varus/valgus/flexion deformity requiring stabilization. Also indicated for revision of failed osteotomy, arthrodesis, or previous arthroplasty. Intended for cemented application only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
August 26, 2022
Corentec Co., Ltd. Yoorim Bae RA Manager 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu Cheonan-si, Chungcheongnam-do 31056 Republic of Korea
Re: K22278
Trade/Device Name: LOSPA TKR System Instrumentation, EXULT TKR System Instrumentation Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH, OOG Dated: June 27, 2022 Received: July 29, 2022
Dear Yoorim Bae:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K222278
Device Name
LOSPA TKR System Instrumentation
Indications for Use (Describe)
LOSPA TKR System is intended for use in total knee arthroplasty surgery for the following indications:
· Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
· Post-traumatic loss of knee joint configuration and function;
• Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability:
· Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.
LOSPA TKR System is intended for cemented application only.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------|
| <span style="font-size:10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K222278
#### Device Name
EXULT TKR System Instrumentation
Indications for Use (Describe)
EXULT Knee System is intended for the treatment of diseases as follows:
· Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
· Post-traumatic loss of knee joint configuration and function;
• Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability:
· Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.
EXULT Knee System is intended for cemented application only.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 10pt;"> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 10pt;"> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for LOSPA, EXULT Knee Instruments changes.
Date: June 27, 2022
# ADMINISTRATIVE INFORMATION
| Manufacturer: | Corentec Co., Ltd. |
|----------------------------|-----------------------------------------------------------------------------------------------|
| | 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu |
| | Cheonan-si, Chungcheongnam-do, Rep. of Korea -31056 |
| | Telephone: +82-41-585-7114; Fax: +82-41-585-7113 |
| Official Contact: | Yoorim Bae<br>RA Manager |
| | 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu |
| | Cheonan-si, Chungcheongnam-do, Rep. of Korea -31056 |
| | Ph: +82-41-410-7116 (Direct) ; Fax: +82-41-585-7113 |
| | Email: yoorim.bae@corentec.com |
| Trade or Proprietary Name: | LOSPA TKR System Instrumentation,<br>EXULT TKR System Instrumentation |
| Common Name: | Orthopedic Surgical Instruments |
| Regulatory Class: | Class II |
| Regulation Number: | 21CFR888.3560 |
| Regulation Name: | Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-<br>Constrained Cemented Prosthesis |
| Product Code(s): | JWH, MBH, OOG |
## Legally Marketed Device(s) to Which Equivalence is Claimed
| 510(k) Number | Trade Name | Submitter | Product Code | |
|---------------|------------|-----------------------------------------------------|------------------|----------|
| Primary | K160157 | LOSPA Modular Knee System | JWH, MBH | |
| Reference | K192507 | LOSPA II Knee System | | JWH |
| | K110404 | LOSPA Total Knee System<br>(EAUM Total Knee System) | | JWH |
| | K130673 | LOSPA Total Knee Replacement System | Corentec Co.,Ltd | JWH |
| | K200395 | EXULT Knee Replacement System | | JWH |
| | K201851 | EXULT Knee Replacement System | | JWH, OOG |
| | K212034 | LOSPA TKR System | | JWH, MBH |
## Reason for 510(k) Submission
The purpose of this submission is to include additional instruments for LOSPA Modular Knee System and EXULT Knee System.
{5}------------------------------------------------
## DEVICE INFORMATION
# A. Device Description
The subject devices in this special 510(k) are the surgical instruments for the LOSPA Modular Knee System and EXULT Knee System, and submitted as a line extension of optional instruments for Corentec's FDA-cleared reusable orthopedic knee joint replacement instruments. The subject instruments are part of Corenter' 510(k)-cleared knee product lines and are required to facilitate total knee arthroplasty procedures.
# LOSPA Modular Knee System Instruments
Implant specific trials - the design of the connection part of the trials to the instrument has been revised due to customer's needs.
Guide, alignment instruments - the design of the cutting guides and alignments instruments have been revised due to customer's needs.
## EXULT Knee System Instruments
AP sizer external rotation instrument and cutting block instruments have been developed to facilitate to check in posterior access during the operation based on US surgeon's request.
# B. Intended Use
Corentec knee instruments are accessory devices and are intended to be used to assist in the implantation of LOSPA Knee Systems (Primary, Modular), EXULT Knee Replacement System in their cleared Indications for Use as provided below:
LOSPA Knee System, EXULT Knee Replacement System are intended for the treatment of diseases as follows:
· Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
· Post-traumatic loss of knee joint configuration and function;
· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;
· Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.
The LOSPA, EXULT Knee Replacement System are intended for cemented application only.
## C. Summary of Technological Characteristics
The subject instruments performed comparable to predicate devices and are as safe and effective as the predicate devices. Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy. The subject instruments have the following similarities to the predicate devices:
- Have the same intended use, .
- Have the same indications for use,
- . Use the same operating principles,
{6}------------------------------------------------
- Incorporate the same basic designs, ·
- Incorporate the same/similar materials, and ・
- . Are supplied non-sterile.
## D. Non-Clinical Testing
The following non-clinical laboratory testing and/or engineering analysis were performed to determine substantial equivalence:
- ・ Dimensions measurement testing
- . Assembly testing
## E. Clinical and Animal Testing
No clinical or animal testing were required.
# F. Substantial Equivalence Conclusion
Based on the similarities to the predicate devices and rationale to support substantial equivalence, the subject devices are substantially equivalent to the commercially available predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.