The Hailie® sensor is intended for single-patient multiple use in the home environment as an electronic data capture accessory for monitoring and recording actuation, inspiratory flow and inhaler shake for prescribed inhaler usage. The Hailie® sensor may be used in the following applications: in clinical trials, where specialists, general practitioners, nurses, and educators need to know if a patient has used their prescribed medication, or assess inspiratory flow and inhaler technique; and in patient self-management. The Hailie® sensor is compatible only with the Teva HFA MDIs like ProAir, Albuerol Sulphate inhalers. The Hailie® sensor is not intended to indicate remaining quantity of medication in an inhaler and does not include a dose counting function. The Hailie® sensor is not intended to provide spirometry measurements.
Device Story
Clip-on electronic accessory for Teva HFA MDI inhalers; monitors medication adherence and technique. Inputs: optical sensor (inhaler presence), induction coil (actuation), motion sensor (shake), flow sensor (inspiratory flow). Device logs date/time of events via internal clock. User interface: status button, multi-color LED, piezo buzzer. Data transmitted via Bluetooth to mobile applications. Used by patients for self-management or by clinicians/educators in clinical practice/trials to assess technique and adherence. Output: usage history logs. Benefits: provides objective data on inhaler usage to improve patient management and adherence.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical testing including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software verification/validation, and performance testing (optical calibration, flow/motion detection accuracy, shelf life, and usability).
Technological Characteristics
Microprocessor-controlled clip-on sensor. Materials evaluated per ISO 10993-1 for limited contact (≤24h). Sensors: optical, induction coil, motion, flow. Connectivity: Bluetooth. Power: battery-powered. Safety standards: ANSI/AAMI ES60601-1, IEC 60601-1-11, IEC 60601-1-2. Software: embedded firmware for data logging and communication.
Indications for Use
Indicated for patients using prescribed Teva HFA MDI inhalers (ProAir HFA, Albuterol Sulphate HFA) to monitor and record inhaler actuation, inspiratory flow, and shake technique in home or clinical settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
February 8, 2023
Adherium (NZ) Ltd Tara Creaven-Capasso Vice President Quality, Regulatory and Clinical Affairs Level 11, 16 Kingston Street Auckland, 1010 New Zealand
Re: K222247
Trade/Device Name: Hailie Sensor NF0110 Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: February 3, 2023 Received: February 6, 2023
Dear Tara Creaven-Capasso:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
for James Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K222247
Device Name
Hailie® Sensor
Indications for Use (Describe)
The Hailie® sensor is intended for single-patient multiple use in the home environment as an electronic data capture accessory for monitoring and recording actuation, inspiratory flow and inhaler shake for prescribed inhaler usage.
The Hailie® sensor may be used in the following applications: in clinical trials, where specialists, general practitioners, nurses, and educators need to know if a patient has used their prescribed medication, or assess inspiratory flow and inhaler technique; and in patient self-management.
The Hailie® sensor is compatible only with the Teva HFA MDIs like ProAir, Albuerol Sulphate inhalers. The Hailie® sensor is not intended to indicate remaining quantity of medication in an inhaler and does not include a dose counting function. The Hailie® sensor is not intended to provide spirometry measurements.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page of
EF PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "adherium". The word "adherium" is written in lowercase letters in a salmon color. To the right of the word is a graphic of a pill capsule, with one half of the capsule in salmon and the other half in gray.
Adherium (NZ) Limited Level 2, 63 Albert Street, Auckland 1010, New Zealand PO Box 106-612, Auckland 1143, New Zealand Phone +64 9 307 2771 contact@hailie.com www.hailie.com
15 Jul 2022
## 510(k) SUMMARY
## I. General Information
| 510(k) Sponsor | Adherium (NZ) Ltd |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Address | Level 2, 63 Albert Street, Auckland 1010, New Zealand<br>PO Box 106-612, Auckland 1143, New Zealand<br>Phone +64 9 307 2771 |
| Correspondence Person | Tara Creaven-Capasso, FRAPS, RAC, RQAP-GLP, MTPORA |
| Title | Vice President Quality, Regulatory and Clinical Affairs |
| Contact Information | Regulatory@adherium.com |
| Date Prepared | 15 July 2022 |
## II. Subject Device
| Proprietary Name | Hailie® |
|---------------------|-----------------------------------|
| Common Name | Hailie® Sensor |
| Model Number | NF0110 |
| Classification Name | Nebulizer Anesthesiology Devices, |
| Regulation Number | 21 CFR 868.5630 |
| Product Code | CAF |
| Regulatory Class | II |
## III. Predicate Device
| Predicate device | Hailie® Sensor NF0109 (for use with Symbicort® MDI),<br>manufactured by Adherium (NZ) Limited. Cleared under K211233 |
|------------------|----------------------------------------------------------------------------------------------------------------------|
| Reference device | Smartinhaler™ NF0091 for use with (ProAir®) also manufactured<br>by Adherium (NZ) Ltd. Cleared under K180407 |
## IV. Device Description
The subject device Hailie® Sensor (Hailie® sensor) NF0110 model, is a modification to the Hailie® sensor NF0109 model (the predicate device) and is used to provide medication actuation monitoring, shake and airflow recording functions, for use as an accessory to the inhaler specified on the device label. The Hailie® sensor is indicated for use only with the following TEVA MDIs: ProAir® HFA & Albuterol Sulphate HFA.
{4}------------------------------------------------
The Hailie sensor NF0110 is a clip-on device that attaches externally around the housing of the inhaler. The optical sensor and induction coil sensor are used to detect the inhaler presence and monitor actuation. Motion and flow sensors are used to record inhaler usage technique parameters. The Haile sensor contains an electronic clock and calendar that are used to log the date and time of inhaler usage events.
The user interface consists of a single Status Button and a multi-color LED indicator to check device status and initiate communications functions. The Hailie sensor has a Bluetooth interface to wirelessly exchange medication usage data with a paired communications device and compatible mobile software applications.
## V. Indications for Use
The Hailie sensor is intended for single patient multiple use in the home environment as an electronic data capture accessory for monitoring and recording actuation, inspiratory flow, and inhaler shake for prescribed inhaler usage.
The Hailie sensor may be used in the following applications: in clinical practice or clinical trials, where specialists, general practitioners, nurses, and educators need to know if a patient has used their prescribed medication or assess inspiratory flow and inhaler usage technique; and in patient selfmanagement.
The Hailie sensor is compatible only with the following TEVA MDI inhalers: ProAir® HFA & Albuterol Sulphate HFA. The Hailie sensor is not intended to indicate remaining quantity of medication in an inhaler and does not include a dose counting function. The Hailie sensor is not intended to provide spirometry measurements.
## VI. Comparison of Technological Characteristics with the Predicate Device
Technological characteristics of the Hailie sensor are equivalent to the predicate and reference devices listed above. They are both microprocessor-controlled electronic devices that clip on to an inhaler, using a combination of sensors to detect inhaler use which is logged to compile a usage history.
The subject device NF0110 has substantially equivalent technological characteristics to the predicate device Hailie Sensor model NF0109, including intended use, indication for use, materials, design, device functionality, features and operating principles. Nonclinical testing results support the substantial equivalence claim.
The subject device NF0110 is substantially equivalent to the reference device, Adherium's Smartinhaler model NF0091, cleared under K180407, manufactured by Adherium (NZ) Ltd. which uses same ProAir®, TEVA MDI medication.
These design changes were verified by non-clinical testing to establish equivalent performance to the predicate device.
## VII. Performance Data
Non-clinical testing and evaluation of the Hailie sensor has been carried out to cover biocompatibility evaluation, electrical safety and electromagnetic compatibility testing, software verification and validation testing, performance testing, and usability evaluation.
{5}------------------------------------------------
#### Biocompatibility Evaluation
The biocompatibility evaluation for the Hailie sensor was conducted in accordance with the FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (16 Jun 2016), to meet reguirements from the following standards: ISO 10993-1:2009 (biocompatibility), ISO 10993-5:2009 (cytotoxicity), ISO 10993-10:2010 (sensitization and intracutaneous irritation), and ISO 10993-12:2012 (sample preparation for biocompatibility testing). The materials used in the Hailie sensor were evaluated according to requirements for a surface device contacting intact skin for limited duration ≤ 24 hours.
## Electrical Safety and Electromagnetic Compatibility (EMC) Testing
Electrical safety and EMC testing was conducted by external laboratories on the Hailie sensor. The device complies with the following standards: ANSI/AAMI ES60601-1:2005 +A1:2012, C1:2009, A2:2020 (general safety), IEC 60601-1-11:2020 (home-use safety), and IEC 60601-1-2:2020 (electromagnetic compatibility). General safety testing was conducted according to applicable requirements for a home use, batterypowered device. EMC testing was conducted according to applicable requirements for an internally powered, non-patient coupled, Bluetooth radio device. Information was provided according to FDA guidance Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices (11 Jul 2016), and Radio-Frequency Wireless Technology in Medical Devices (14 Aug 2013).
## Software Verification and Validation Testing
Software verification and validation testing were conducted to ensure correct functionality for the Haile sensor software release, for all software modules. Documentation was provided as recommended by FDA guidance Content of Premarket Submissions for Software Contained in Medical Devices (11 May 2005), Off-The-Shelf Software Use in Medical Devices (27 Sep 2019), and Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (2 Oct 2014).
## Performance Testing
Performance testing was conducted to establish correct functionality and compatibility of the Haile sensor with the with the following TEVA MDI inhalers: ProAir® HFA and Albuterol Sulphate HFA:
- Optical inhaler presence detection determined optical calibration limits and confirmed accurate . detection of the installed TEVA MDI inhalers.
- Airflow detection sensor confirmed performance of flow detection during inhaler usage. .
- Airflow verification testing confirmed the effect of the Hailie sensor on the airflow path. .
- Motion detection sensor confirmed performance of shake detection before inhaler use. .
- General performance testing confirmed acceptable performance over the specified shelf life and specified Bluetooth communications range.
- . User interface testing - confirmed visibility of device LED display and audibility of device Piezo buzzer.
## Usability Evaluation
Usability evaluation for the Hailie sensor was carried out to evaluate impact on critical tasks indicated by the usability risk analysis for the updated design, and established validity of the results obtained from testing carried out in accordance with the FDA guidance Applying Human Factors and Usability Engineering to Medical Devices (3 Feb 2016).
## Clinical Testing
Clinical testing was not required for a determination of substantial equivalence of the Hailie sensor. The product functionality has been adequately assessed by non-clinical testing as above.
## VIII. Conclusions
Finished device testing carried out for the Hailie® sensor indicates it meets design, safety, and performance requirements. Performance verification demonstrates that device functions, Inhaler detection, communication, Shelf life, Battery life and user interface are substantially equivalent to the predicate device. Inhaler usage parameters are measured with sufficient accuracy for monitoring inhaler
{6}------------------------------------------------
use equivalent to reference device, and the change of the sensor does not adversely affect normal use of the inhaler. Software verification demonstrates that the device should perform as intended in the specified use conditions, and equivalently to the predicate for common software functions. The device meets standard requirements for biocompatibility, electrical safety, and electromagnetic compatibility. The usability evaluation indicates there are no issues for successful use with the compatible inhaler TEVA ProAir® HFA and Albuterol Sulphate HFA.
This information indicates that the subject device Hailie Sensor model NF0110 is substantially equivalent to the predicate device Hailie Sensor model NF0109. Non-clinical testing results are acceptable and support the substantial equivalence claim.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.