R2 STUDIO Q/RCT820 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified health care professionals. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
Device Story
R2 STUDIO Q/RCT820 is a dental CBCT, panoramic, and cephalometric X-ray imaging system. It utilizes a rotating C-arm with an integrated high-voltage X-ray generator and detector to capture 2D and 3D images of hard tissues (bone/teeth). The system supports dental model scanning and carpus imaging for orthodontic assessment. Operated by dentists or qualified professionals in clinical settings, the device uses digital sensors (direct/indirect conversion) to acquire X-ray signals. Data is reconstructed via PC software to provide 2D/3D diagnostic images on a monitor. These images assist clinicians in diagnosing dental and maxillofacial structures, supporting treatment planning. Benefits include improved diagnostic visualization of complex anatomical structures, aiding in precise orthodontic and surgical decision-making.
Clinical Evidence
Bench testing performed per IEC 61223-3-4 and IEC 61223-3-7 to assess imaging performance and patient dosage. Clinical evaluation report provided for the new FXDD-1724R detector, comparing sample images to the reference device (Jupi0606X) to support intended use. No formal clinical trial with primary endpoints reported; evidence relies on bench performance and clinical image assessment.
Indicated for radiographic examination of dento-maxillofacial, sinus, TMJ, and airway structures in adult and pediatric patients (≥5 years old, ≥21 kg, ≥113 cm). Includes optional model scanning and cephalometric imaging (including carpus images for orthodontic growth assessment).
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2022
MegaGen Implant CO., Ltd. % Yong Jae Kim Research Engineer 45, Secheon-ro, 7-gil Daegu, Dasa-eup, Dalesong-gun 42921 KOREA
Re: K222219
Trade/Device Name: R2 Studio Q/RCT820 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: July 18, 2022 Received: November 10, 2022
Dear Yong Jae Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
2022.12.06
07:13:02
-05'00'
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K222219
Device Name R2 STUDIO Q/RCT820
## Indications for Use (Describe)
R2 STUDIO Q/RCT820 is CBCT and panoramic x-ray imaging system with cephalometric.
Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option.
Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.
The device is to be operated and used by dentists or other legally qualified health care professionals.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 1. 510(k) Summary
The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
2. Date: July 18, 2022
## 3. Administrative Information
| APPLICANT | MegaGen Implant Co., Ltd. |
|--------------|-----------------------------------------------------------------------------------------------------|
| ADDRESS | 45, Secheon-ro, 7gil, Dasa-eup, Dalseong-gun, Daegu, Republic of Korea |
| Manufacturer | MegaGen Implant Co., Ltd.<br>45, Secheon-ro, 7gil, Dasa-eup, Dalseong-gun, Daegu, Republic of Korea |
| | TEL: +82-53-247-2261<br>FAX: +82-53-247-2254 |
Contact Person e-mail: ra3@imegagen.com
# 4. Device Information
#### Device Name
Trade/Proprietary Name: R2 STUDIO Q/RCT820 Common Name: Dental panoramic/tomography and cephalometric x-ray system
# Classification
Classification Name: Computed tomography x-ray system Regulation Number: 21 CFR 892.1750 Class: Il Product code: OAS Panel: Radiology
#### 5. Predicate device & Reference Device
| Parameter | Predicate Device | Reference Device |
|----------------------|----------------------------------|----------------------------------|
| Device Name | RCT800 | RCT700 |
| Manufacturer | RAY Co., Ltd | RAY Co., Ltd |
| 510(k) Number | K192737 | K213226 |
| Classification name | Computed tomography x-ray system | Computed tomography x-ray system |
| Regulation number | 892.1750 | 892.1750 |
| Primary product code | OAS | OAS |
{4}------------------------------------------------
## 7. Device Description
System purpose R2 STUDIO Q/RCT820 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle.
Panoramic image scanning function for attaining image of whole teeth, cephalometric scanning option for attaining cephalic image are included. and Model Scan option for attaining dental model CBCT image are included.
## 8. Indication for use
R2 STUDIO Q/RCT820 is CBCT and panoramic x-ray imaging system with cephalometric.
Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option.
Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.
The device is to be operated and used by dentists or other legally qualified heath care professionals.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height measurements approximately correspond to that of an average 5 vear old.
# 9. Patient population
The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure. There is no restriction for ethnic group, Gender, weight, health, or condition. We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.
# 10. Operating principle
The purpose of this unit is diagnose clinical structures of a tooth and head areas mainly by using the characteristics of permeability from X-ray. The principle of functions are as following. The machine is made of X-ray generator and arms in which transfers X-ray signals to a sensor in 2D. Also, an object that has a magnification is required in a distance. Moreover, the unit has to be adjustable depending on height of a patient and PC system to reconstruct an image. The arm parts are controlled for rotating and linear moving to synchronize between the sensor and X-ray generator to get the image of interests. The purpose of this mechanism is to provide the images in 2D or/and 3D as preferred to diagnose in a monitor. CBCT provides in 3D images as reconstructed and Panorama is to diagnose the structures in a panoramic view. Cephalometric allows for orthodontic treatment. These 3 functions could be in 1 system, Panorama with Cephalometric, or Panorama only system depending on the needs. To provide the features as mentioned above, digital transferring from permeated X-ray to absorbing to the sensor is essential and all the process are proceed in Detector. Detector transfers X-ray to light depending on the structure materials. Detector is separated into indirect method that the light is changed to digital signals on photodiode and direct method in which the light is directly transferred to digital signal. This unit is using both direct method depending on the interior structure materials.
{5}------------------------------------------------
# 11. Comparison with predicate devices
The following table provides the summary of the technological characteristics of R2 STUDIO Q/RCT820 compared to the ence.
| Parameter | Proposed Device | Predicate Device | Reference Device | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Manufacturer | MegaGen Implant Co., Ltd. | RAY Co., Ltd. | RAY Co., Ltd. | |
| Device name | R2 STUDIO Q/RCT820 | RCT800 | RCT700 | |
| 510(k) Number | (Traditional 510(k)) | K192737 | K213226 | |
| | | (Traditional 510(k)) | (Traditional 510(k)) | |
| Common Name | Dental panoramic/tomography and<br>cephalometric x-ray system | Dental panoramic/tomography and<br>cephalometric x-ray system | Dental panoramic/tomography and<br>cephalometric x-ray system | |
| Indications for use | R2 STUDIO Q/RCT820 is CBCT and panoramic<br>x-ray imaging system with cephalometric.<br>Which is intended to radiographic<br>examination of the dento-maxillofacial, sinus,<br>TMJ, Airway for diagnostic support for adult<br>and pediatric patients. And a model scan is<br>included as an option.<br>Cephalometric image is also includes wrist to<br>obtain carpus images for growth and maturity<br>assessment for orthodontic treatment.<br>The device is to be operated and used by<br>dentists or other legally qualified heath care<br>professionals.<br>This device is not intended for use on patients<br>less than approximately 21 kg (46 lb) in weight<br>and 113 cm (44.5 in) in height; these height<br>and weight measurements approximately<br>correspond to that of an average 5 year old. | RCT800 is CBCT and panoramic x-ray imaging<br>system with cephalometric.<br>Which is intended to radiographic<br>examination of the dento-maxillofacial, sinus,<br>TMJ, Airway for diagnostic support for adult<br>and pediatric patients. And a model scan is<br>included as an option.<br>Cephalometric image is also includes wrist to<br>obtain carpus images for growth and maturity<br>assessment for orthodontic treatment.<br>The device is to be operated and used by<br>dentists or other legally qualified heath care<br>professionals. | RCT700 is CBCT and panoramic x-ray imaging<br>system with cephalometric. Which is intended<br>to radiographic examination of the dento-<br>maxillofacial, sinus, TMJ, Airway and ENT<br>structure for diagnostic support for adult and<br>pediatric patients. And a model scan is<br>included as an option. Cephalometric image is<br>also includes wrist to obtain carpus images for<br>growth and maturity assessment for<br>orthodontic treatment. The device is to be<br>operated and used by dentists or other legally<br>qualified heath care professionals. | |
| Mode of Operation | Continuous operation with intermittent,<br>stated permissible loading | Continuous operation with intermittent,<br>stated permissible loading | Continuous operation with intermittent,<br>stated permissible loading | |
| 3D technology | CBCT Cone beam | CBCT Cone beam | CBCT Cone beam | |
| | Computed Tomography | Computed Tomography | Computed Tomography | |
| Performance Specification | 1) CBCT Computed tomography | 1) CBCT Computed tomography | 1) CBCT Computed tomography | |
| | - Patient | - Patient | - Patient | |
| | - Dental Model Scan(Optional) | - Dental Model Scan(Optional) | - Dental Model Scan(Optional) | |
| | 2) Panoramic | 2) Panoramic | 2) Panoramic | |
| | 3) Cephalometric(optional) | 3) Cephalometric(optional) | 3) Cephalometric(optional) | |
| | - Scan type | - One shot type | - One shot type | |
| | | | - Scan type | |
| | | | - Scan type | |
| | | Base | Base | Base |
| | | CT + PANO | CT + PANO | CT + PANO |
| | | | | |
| | | Option(CEPH) | Option(CEPH) | Option(CEPH) |
| | | CT + PANO + SCAN CEPH | CT + PANO + SCAN CEPH | CT + PANO + SCAN CEPH |
| Functional Option | | | | |
| | | CT + PANO + One shot(One shot, Standard | CT + PANO + One shot(One shot, Standard | |
| | | Type) | Type) | |
| | | | | |
| | | | CT + PANO + One shot(One shot, Large Type) | CT + PANO + One shot(One shot, Large Type) |
| | CT | FXDD-1724R | FXDD-0606CA | FXDD-0606CA |
| | | | FXDD-1012CHA | Jupi0606X |
| | PANO | FXDD-1724R | FXDD-0606CA | FXDD-0606CA |
| Detector Type | | | FXDD-1012CHA | Jupi0606X |
| | Ceph (Scan) | XID-C24DC | XID-C24DC | XID-C24DC |
| | Ceph (One shot)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.