The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.
Device Story
TruWave Disposable Pressure Transducer is a sterile, single-use device for monitoring intravascular, intracranial, and intrauterine pressures. It features a straight, flow-through design with an integrated pressure-sensitive silicon chip. Fluid flow deflects a piezo-electric diaphragm on the chip, changing circuit resistance and generating a voltage signal proportional to pressure. This signal is transmitted via a disposable cable to a reusable cable connected to a patient monitor. The device includes an integrated stopcock and flush device (3mL or 30mL) and can be mounted on an IV pole or patient arm. Healthcare providers (physicians, nurses) use the monitor output to assess patient hemodynamic status and guide clinical decision-making. The device benefits patients by providing continuous, real-time pressure data for critical care management.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), EMC/EMI (IEC 60601-1-2), electrical safety (ANSI/AAMI BP22, IEC 60601-2-34), MR safety (ASTM F2503), and mechanical performance. Usability/human factors were evaluated by healthcare providers. All acceptance criteria were met.
Indicated for patients requiring intravascular, intracranial, or intrauterine pressure monitoring.
Regulatory Classification
Identification
A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 22, 2022
Edwards Lifesciences LLC Carmen Chen Manager, Regulatory Affairs One Edwards Way Irvine, California 92614
Re: K222216
Trade/Device Name: TruWave Disposable Pressure Transducer Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter Tip Pressure Transducer Regulatory Class: Class II Product Code: DXO Dated: November 22, 2022 Received: November 22, 2022
Dear Carmen Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K222216
Device Name TruWave Disposable Pressure Transducer
Indications for Use (Describe)
The Pressure Monitoring Kit with TruWave Disposable Pressure Transducer is for use on patients requiring intravascular, intracranial, or intrauterine pressure monitoring.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 5. 510(K) SUMMARY
| 510(k) Submitter | Edwards Lifesciences, LLC<br>One Edwards Way<br>Irvine, CA, USA 92614 | | |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | <b>Primary Contact</b><br>Carmen G Chen<br>Manager, Regulatory Affairs<br>Edwards Lifesciences<br>One Edwards Way<br>Irvine, CA 92614<br>Telephone: (949) 250 - 5469<br>Fax: (949) 809 - 2954<br>Email:<br>Carmen.Chen@edwards.com | <b>Secondary Contact</b><br>Karen O'Leary<br>Sr. Director, Regulatory Affairs<br>Edwards Lifesciences<br>One Edwards Way<br>Irvine, CA 92614<br>Telephone: (949) 610-9179<br>Fax: (949) 809 - 2954<br>Email:<br>Karen.OLeary@edwards.com | |
| | Date Prepared | December 21, 2022 | |
| Trade Name | TruWave™ | | |
| Common Name | Disposable Pressure Transducer | | |
| Regulation<br>Number/<br>Regulation<br>Name | 21 CFR 870.2870 / Transducer, Pressure, Catheter Tip | | |
| Product Code | DXO | | |
| Regulation Class | Class II | | |
| Predicate Device | K183413- TruWave Disposable Pressure Transducer (cleared 01 May 2019) | | |
| Device<br>Description | The TruWave Disposable Pressure Transducer is a sterile, single-use<br>device that is used to monitor intravascular, intracranial, and intrauterine<br>pressures. The pressure transducer has a straight, flow-through design<br>across the pressure sensor, and is available with or without the integral<br>flush device. The pressure sensor is a pressure sensitive silicon chip with<br>two electrodes for excitation voltage and two electrodes for signal output.<br>A transparent fluid path with an integral stopcock at one end and an<br>integral flush device (either 3mL or 30mL) at the other end encloses the<br>sensor. The enclosure of the TruWave Disposable Pressure Transducer has<br>a pathway for air to enter the housing and acts as a vent. A disposable<br>cable (available in 10-inch/25 cm and 48-inch/120 cm lengths) attached to<br>the pressure transducer housing interfaces with an Edwards Lifesciences<br>reusable cable that is specifically wired for the monitor being used. The<br>TruWave Disposable Pressure Transducer can be mounted on the patient's | | |
| | arm using an arm strap or it may be mounted on an IV pole in a holder. | | |
| | The TruWave Disposable Pressure Transducer may be a component in various pressure monitoring kits or systems. | | |
| Indications for<br>Use/Intended<br>Use | The Pressure Monitoring Kit with TruWave Disposable Pressure<br>Transducer is for use on patients requiring intravascular, intracranial, or<br>intrauterine pressure monitoring. | | |
| Comparison to<br>Predicate Device | The subject TruWave Disposable Pressure Transducer device of this<br>Traditional 510(k) is identical to the predicate device cleared in K183413<br>in terms of indications for use/intended use except for the proposed design,<br>material, and labeling changes to the device. See table below. Differences<br>in technological characteristics do not raise any new concerns of safety and<br>effectiveness. Verification and validation testing for the subject device<br>demonstrate safety and effectiveness. The TruWave Disposable Pressure<br>Transducer has shown to be substantially equivalent to the predicate device<br>for its intended use in a hospital setting or other appropriate clinical<br>environment. | | |
| Device Testing | Biocompatibility testing was performed in accordance with ISO 10993-1:<br>2018 – Biological evaluation of medical devices – Part 1: Evaluation and<br>testing within a risk management process, and FDA guidance document,<br>Use of International Standard ISO 10993-1, “Biological evaluation of<br>medical devices – Part 1: Evaluation and testing within a risk management<br>process”, issued on September 4, 2020. | | |
| | Electromagnetic compatibility (EMC) and Electromagnetic Immunity<br>(EMI) testing were conducted and comply with the IEC 60601-1, IEC<br>60601-2-37 standards for safety and the IEC 60601-1-2 standard for EMC. | | |
| | Electrical safety testing (including defibrillator challenge, dielectric<br>strength, liquid ingress, and leakage current) was performed in accordance<br>with ANSI/AAMI BP22:1994/(R)2016 and IEC 60601-2-34: 2011 | | |
| | Testing (including accuracy, excitation/signal impedance, light sensitivity,<br>symmetry, and overpressure) was conducted per ANSI/AAMI<br>BP22:1994/(R)2016 and IEC 60601-2-34: 2011. | | |
| | Magnetic resonance safety testing was performed based on ASTM F2503-<br>20 Standard Practice for Marking Medical Devices and Other Items for<br>Safety in the Magnetic Resonance Environment. | | |
| | Mechanical testing was performed in accordance with ANSI/AAMI<br>BP22:1994/(R) 2016 and Edwards' design requirements. | | |
| | The usability/human factors of the TruWave Disposable Pressure<br>Transducer were evaluated by healthcare provider users. | | |
| | All device acceptance criteria were met. Results of non-clinical testing | | |
| Conclusion | show that the subject device TruWave Disposable Pressure Transducer | | |
| | meets its intended use and demonstrate that the device is as safe, as | | |
| | effective, and performs as well as the predicate device. The differences | | |
| | between the subject device and predicate device do not raise new issues of | | |
| | safety and/or effectiveness. Therefore, the subject device TruWave | | |
| | Disposable Pressure Transducer is substantially equivalent to the predicate | | |
| | device TruWave Disposable Pressure Transducer (K183413). | | |
| Function/Parameter | Subject Device<br>(K222216) | Predicate Device<br>(K183413) | Comparison |
| Device Name | TruWave™ Disposable<br>Pressure Transducer | TruWave™ Disposable<br>Pressure Transducer | Same |
| 510(k) Number | K222216 | K183413<br>(cleared on 01 April 2019) | Not Applicable |
| Manufacturer | Edwards Lifesciences, LLC | Edwards Lifesciences, LLC | Same |
| Device Classification | Class II | Class II | Same |
| Regulation Number | 21 CFR §870.2870- Catheter<br>tip pressure transducer | 21 CFR §870.2870- Catheter<br>tip pressure transducer | Same |
| Product Code | DXO | DXO | Same |
| Intended Use | The Pressure Monitoring Kit<br>with TruWave DPT is used<br>on patients requiring<br>intravascular, intracranial, or<br>intrauterine pressure<br>monitoring. | The Pressure Monitoring Kit<br>with TruWave DPT is used<br>on patients requiring<br>intravascular, intracranial, or<br>intrauterine pressure<br>monitoring. | Same |
| Indications for Use | The Pressure Monitoring Kit<br>with TruWave DPT is for<br>use on patients requiring<br>intravascular, intracranial, or<br>intrauterine pressure<br>monitoring. | The Pressure Monitoring Kit<br>with TruWave DPT is for<br>use on patients requiring<br>intravascular, intracranial, or<br>intrauterine pressure<br>monitoring. | Same |
| Operating Principle | The TruWave DPT sensor<br>consists of a silicon chip<br>which when fluid flows<br>through the fluid path allows<br>a piezo-electric diaphragm of<br>the chip to be deflected,<br>which changes resistance of<br>the circuit, which<br>correspondingly causes a<br>change in voltage. This<br>voltage change is transmitted<br>through the cable to a patient<br>monitor. | The TruWave DPT sensor<br>consists of a silicon chip<br>which when fluid flows<br>through the fluid path allows<br>a piezo-electric diaphragm of<br>the chip to be deflected,<br>which changes resistance of<br>the circuit, which<br>correspondingly causes a<br>change in voltage. This<br>voltage change is transmitted<br>through the cable to a patient<br>monitor. | Same |
| Design | Straight, flow-through design<br>across the pressure sensor<br>Housing with curved side<br>grips and arm mount strap<br>slots…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.