AI/ML, Software as a Medical Device, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K222176 · Mar 2, 2023
BoneView
Gleamer
Retrospective clinical radiographs from various manufacturers
Retrospective clinical datasets were used to conduct standalone performance testing (sensitivity/specificity) for the device across the expanded patient population (children/adolescents) and to validate performance against the adult population.
Training dataset of 44,649 radiographs, representing 151,096 images
—
Standalone performance testing on 2,000 radiographs (children/adolescents)
—
Fracture detection
Supervised Deep Learning
—
Sensitivity: 0.752, Specificity: 0.956 (Aided vs Unaided)
Training dataset of 44,649 radiographs, representing 151,096 images
—
MRMC retrospective reader study of 480 cases
24 (clinical readers)
Indications for Use
BoneView 1.1-US is intended to analyze radiographs using machine learning techniques to identify and highlight fractures during the review of radiographs of: [Table of anatomical areas and patient populations] BoneView 1.1-US is intended for use as a concurrent reading aid during the interpretation of radiographs. BoneView 1.1-US is for prescription use only.
Device Story
BoneView 1.1-US is a software-only device acting as a concurrent reading aid for clinicians (radiologists, primary care, emergency, urgent care, orthopedics). It receives 2D X-ray images from PACS or X-ray systems via DICOM. An AI algorithm processes images to identify potential fractures. Outputs are DICOM result files containing a summary table and annotated images. Annotations include bounding boxes: dotted-line for suspicious/subtle fractures (high-sensitivity operating point) and solid-line for definite fractures (high-specificity operating point). The device does not alter original images. It assists clinicians in diagnosis, potentially reducing missed fractures.
Clinical Evidence
No new clinical studies were conducted. The submission relies on the clinical performance study from the predicate device (K212365), which was a fully-crossed MRMC retrospective reader study with 24 readers evaluating 480 cases. Aided vs. unaided performance showed significant improvement: sensitivity increased from 0.648 to 0.752 (+10.4%) and specificity increased from 0.906 to 0.956 (+5%). Bench testing on 2,000 pediatric radiographs confirmed high sensitivity and specificity across all anatomical areas, comparable to adult performance.
Technological Characteristics
Standalone software; supervised deep learning algorithm. Deployed on-premise or cloud; integrates with PACS/X-ray systems via DICOM. No patient-contacting components; no electrical/EMC requirements. Software level of concern: Moderate.
Indications for Use
Indicated for fracture detection in radiographs of limbs (children/adolescents 2-21y; adults >21y) and pelvis, rib cage, and dorsolumbar spine (adults >21y).
Regulatory Classification
Identification
A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Special Controls
A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Gleamer % Antoine Tournier Head of Quality & Regulatory Affairs 5 Avenue du Général de Gaulle Saint Mandé, 94160 FRANCE
March 2, 2023
# Re: K222176
Trade/Device Name: BoneView 1.1-US Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software for fracture Regulatory Class: Class II Product Code: QBS Dated: January 31, 2023 Received: February 1, 2023
Dear Antoine Tournier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K222176
Device Name
BoneView 1.1-US
Indications for Use (Describe)
BoneView 1.1-US is intended to analyze radiographs using machine learning techniques to identify and highlight fractures during the review of radiographs of:
| Study Type (Anatomical Area of<br>Interest) | Compatible Radiographic View(s) | Patient population* |
|---------------------------------------------|---------------------------------|-------------------------------|
| Ankle | Frontal, Lateral, Oblique | Adults & Children/Adolescents |
| Foot | Frontal, Lateral, Oblique | Adults & Children/Adolescents |
| Knee | Frontal, Lateral | Adults & Children/Adolescents |
| Tibia/Fibula | Frontal, Lateral | Adults & Children/Adolescents |
| Wrist | Frontal, Lateral, Oblique | Adults & Children/Adolescents |
| Hand | Frontal, Oblique | Adults & Children/Adolescents |
| Elbow | Frontal, Lateral | Adults & Children/Adolescents |
| Forearm | Frontal, Lateral | Adults & Children/Adolescents |
| Humerus | Frontal, Lateral | Adults & Children/Adolescents |
| Shoulder | Frontal, Lateral, Axillary | Adults & Children/Adolescents |
| Clavicle | Frontal | Adults & Children/Adolescents |
| Pelvis | Frontal | Adults only |
| Hip | Frontal, Frog Leg Lateral | Adults only |
| Femur | Frontal, Lateral | Adults only |
| Ribs | Frontal Chest, Rib series | Adults only |
| Thoracic Spine | Frontal, Lateral | Adults only |
| Lumbosacral Spine | Frontal, Lateral | Adults only |
* Adults are patient aged above 21 years old and Children/Adolescents are patients aged from 2 to 21 years old.
BoneView 1.1-US is intended for use as a concurrent reading aid during the interpretation of radiographs. BoneView 1.1-US is for prescription use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "GLEAMER" in a simple, sans-serif font. To the left of the word is a red circular logo with a white center. The logo appears to be a stylized letter "G" or a symbol representing connection or collaboration. The overall design is clean and modern, suggesting a tech-related or innovative company.
### Date prepared: January 31th, 2023
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92 the 510(k) Summary for BoneView 1.1-US is provided below.
## 1. Submitter
| Submitter | GLEAMER SAS<br>5, avenue du Général de Gaulle<br>94160 Saint-Mandé - FRANCE | |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------|--|
| Primary Contact Person | Antoine Tournier<br>Head of Quality & Regulatory Affairs<br>Tel: 0033 6 15 81 23 45<br>Email: antoine.tournier@gleamer.ai | |
| Secondary Contact Perso | Christian Allouche<br>CEO<br>Tel: 0033 6 58 53 70 46<br>Email: christian@gleamer.ai | |
# 2. Device
| Trade Name | BoneView 1.1-US |
|------------------|--------------------------------------------------------------------------|
| 510(k) reference | K222176 |
| Common Name | Radiological computer assisted detection/diagnosis software for fracture |
| Regulation | 21 CFR 892.2090 |
| Product Code | QBS |
| Classification | Class II |
# 3. Predicate Device
| Predicate Device | Gleamer BoneView |
|------------------|------------------|
| 510(k) reference | K212365 |
{4}------------------------------------------------
# 4. Device Description
BoneView 1.1-US is a software-only device intended to assist clinicians in the interpretation of:
- . limbs radiographs of children/adolescents and
- . limbs, pelvis, rib cage, and dorsolumbar vertebra radiographs of adults.
BoneView 1.1-US can be deployed on-premise or on cloud and be connected to several computing platforms and X-ray imaging platforms such as X-ray radiographic systems, or PACS. More precisely, BoneView 1.1-US can be deployed:
- . In the cloud with a PACS as the DICOM Source
- On premise with a PACS as the DICOM Source
- On premise with an X-ray system as the DICOM Source
After the acquisition of the radiographs on the patient and their storage in the DICOM Source, the radiographs are automatically received by BoneView 1.1-US from the user's DICOM Source through an intermediate DICOM node (for example, a specific Gateway, or a dedicated API). The DICOM Source can be the user's image storage system (for example, the Picture Archiving and Communication System, or PACS), or other radiological equipment (for example X-ray systems).
Once received by BoneView 1.1-US, the radiographs are automatically processed by the AI algorithm to identify regions of interest. Based on the processing result, BoneView 1.1-US generates result files in DICOM format. These result files consist of a summary table and result images (annotations on a copy of the original images or annotations to be toggled on/off). BoneView 1.1-US does not alter the original images, nor does it change the order of original images or delete any image from the DICOM Source.
Once available, the result files are sent by BoneView 1.1-US to the DICOM Destination through the same intermediate DICOM node. Similar to the DICOM Source, the DICOM Destination can be the user's image storage system (for example, the Picture Archiving and Communication System, or PACS), or other radiological equipment (for example X-ray systems). The DICOM Source and the DICOM Destination are not necessarily identical.
The DICOM Destination can be used to visualize the result files provided by BoneView 1.1-US or to transfer the results to another DICOM host for visualization. The users are then as a concurrent reading aid to provide their diagnosis.
The general layout of images processed by BoneView is comprising:
(1) The "summary table" – it is a first image that is derived from the detected regions of interest in the following result images and that displays the results of the overall study along with the Gleamer – BoneView logo. This summary can be configured to be present or not.
(2) The result images – they are provided for all the images that were processed by BoneView and contain:
{5}------------------------------------------------
- . Around the Regions of Interest (if any), a rectangle with a solid or dotted line depending on the confidence of the algorithm (see below)
- . Around the entire image, a white frame showing that the images were processed by BoneView
- Below the image:
- o The Gleamer BoneView logo
- o The number of Regions of interest that are displayed in the result image
- (if any) The caution message if it was identified that the image was not part of the o indication for use of BoneView
The training of BoneView was performed on a training dataset of 44,649 radiographs, representing 151,096 images (52.4% of males, with age: range [0 – 109]; mean 42.4 +/- 24.6) for all anatomical areas of interest in the Indications for Use and from various manufacturers. BoneView has been designed to solve the problem of missed fractures including subtle fractures, and thus detects fractures with a high sensitivity. In this regard, the display of findings is triggered by a "high-sensitivity operating point" (DOUBT FRACT) that will enable the display of a dotted-line bounding box around the region of interest. Additionally, the users need to be confident that when BoneView identifies a fracture, it is actually a fracture. In this regard, an additional information is introduced to the user with a "high-specificity operating point" (FRACT).
These two operating points are implemented in the User Interface as follow:
- Dotted-line Bounding Box: suspicious area / subtle fracture (when the level of confidence of the Al . algorithm associated with the finding is above "high-sensitivity operating point" and below "highspecificity operating point") displayed as a dotted bounding box around the area of interest
- . Solid-line Bounding Box: definite or unequivocal fractures (when the level of confidence of the Al algorithm associated with the finding is above "high-specificity operating point") displayed as a solid bounding box around the area of interest
BoneView can provide 4 levels of results:
- FRACT: BoneView identified at least one solid-line bounding box on the result images,
- . DOUBT FRACT: BoneView did not identify any solid-line bounding box on the result images but it identified at least one dotted-line bounding box in the result images,
- NO FRACT: BoneView did not identify any bounding box at all in the result images,
- . NOT AVAILABLE: BoneView identified that the original images are out of its Indications for Use
# 5. Intended use/Indications for use
BoneView 1.1-US is intended to analyze radiographs using machine learning techniques to identify and highlight fractures during the review of radiographs of:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Gleamer. The logo consists of a red circular icon with a white design inside, followed by the word "GLEAMER" in a dark blue sans-serif font. The logo is simple and modern, with a focus on the company name.
| Study Type (Anatomical Area of Interest) | Compatible Radiographic View(s) | Patient population* |
|------------------------------------------|---------------------------------|-------------------------------|
| Ankle | Frontal, Lateral, Oblique | Adults & Children/Adolescents |
| Foot | Frontal, Lateral, Oblique | Adults & Children/Adolescents |
| Knee | Frontal, Lateral | Adults & Children/Adolescents |
| Tibia/Fibula | Frontal, Lateral | Adults & Children/Adolescents |
| Wrist | Frontal, Lateral, Oblique | Adults & Children/Adolescents |
| Hand | Frontal, Oblique | Adults & Children/Adolescents |
| Elbow | Frontal, Lateral | Adults & Children/Adolescents |
| Forearm | Frontal, Lateral | Adults & Children/Adolescents |
| Humerus | Frontal, Lateral | Adults & Children/Adolescents |
| Shoulder | Frontal, Lateral, Axillary | Adults & Children/Adolescents |
| Clavicle | Frontal | Adults & Children/Adolescents |
| Pelvis | Frontal | Adults only |
| Hip | Frontal, Frog Leg Lateral | Adults only |
| Femur | Frontal, Lateral | Adults only |
| Ribs | Frontal Chest, Rib series | Adults only |
| Thoracic Spine | Frontal, Lateral | Adults only |
| Lumbosacral Spine | Frontal, Lateral | Adults only |
* Adults are patient aged above 21 years old and Children/Adolescents are patients aged from 2 to 21 years old.
BoneView 1.1-US is intended for use as a concurrent reading aid during the interpretation of radiographs. BoneView 1.1-US is for prescription use only.
# 6. Substantial equivalence
| Features and Characteristics | Subject Device<br>Gleamer<br>BoneView 1.1-US | Predicate Device<br>Gleamer<br>BoneView 1.0-US |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Information | | |
| Regulation Number/Name | 21 CFR 892.2090 / Radiological Computer<br>Assisted Detection and Diagnosis Software for<br>Fracture | Same |
| Product Code | QBS | Same |
| Features and Characteristics | Subject Device<br>Gleamer<br>BoneView 1.1-US | Predicate Device<br>Gleamer<br>BoneView 1.0-US |
| Regulation Description | A radiological computer assisted detection and<br>diagnostic software for suspected fracture is an<br>image processing device intended to aid in the<br>detection, localization, and/or characterization<br>of fracture on acquired medical images (e.g.<br>radiography, MR, CT). The device detects,<br>identifies, and/or characterizes fracture based<br>on features or information extracted from<br>images, and may provide information about the<br>presence, location, and/or characteristics of the<br>fracture to the user. Primary diagnostic and<br>patient management decisions are made by the<br>clinical user. | Same |
| Intended Use | The device is intended to aid in the detection,<br>localization, and characterization of fractures on<br>acquired medical images (per 21 CFR 892.2090<br>Radiological Computer Assisted Detection and<br>Diagnosis Software For Fracture). | Same |
| Image Modality | 2D Xray Images | Same |
| Clinical Finding and Clinical<br>Output | Fracture<br>To inform the primary diagnostic and patient<br>management decisions that are made by the<br>clinical user. | Same |
| Mode of action | Image processing software using machine<br>learning to aid in identifying and highlighting<br>fractures during the review of radiographs. | Same |
| Features and Characteristics | Subject Device<br>Gleamer<br>BoneView 1.1-US | Predicate Device<br>Gleamer<br>BoneView 1.0-US |
| Patient population and<br>Anatomic Areas of Interest | Adults (greater than 21 years of age) and<br>Children/Adolescents (between 2 years of age<br>and 21 years of age):<br>Ankle Foot Knee Tibia/Fibula Wrist Hand Elbow Forearm Humerus Shoulder Clavicle Adults (greater than 21 years of age) only: Pelvis Hip Femur Ribs Thoracic Spine Lumbosacral Spine | Adults (greater than 21 years<br>of age) only:<br>Ankle Foot Knee Tibia/Fibula Femur Wrist Hand Elbow Forearm Humerus Shoulder Clavicle Pelvis Hip Ribs Thoracic Spine Lumbosacral Spine |
| Intended Users | The intended users of BoneView are clinicians<br>with the authority to diagnose fractures in<br>various settings including primary care (e. g.,<br>family practice, internal medicine), emergency<br>medicine, urgent care, and specialty care (e. g.<br>orthopedics), as well as radiologists who review<br>radiographs across settings. | Same |
| Software and Technical Information | | |
| Machine Learning<br>Methodology | Supervised Deep Learning | Same |
| Image Source | DICOM Source (e.g., imaging device,<br>intermediate DICOM node, PACS system, etc.) | Same |
| Features and Characteristics | Subject Device<br>Gleamer<br>BoneView 1.1-US | Predicate Device<br>Gleamer<br>BoneView 1.0-US |
| Image Viewing | PACS system<br>Image annotations made on copy of original<br>image or image annotations toggled on/off | Same |
| Deployment Platform | Deployment on-premise or on cloud and<br>connection to several computing platforms and<br>X-ray imaging platforms such as X-ray<br>radiographic systems, or PACS | Same |
| Privacy | HIPAA Compliant | Same |
| Software Level of Concern | Moderate | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Gleamer. The logo consists of a red circular icon with a white design inside, followed by the word "GLEAMER" in a simple, sans-serif font. The text is a dark blue color and is spaced evenly.
{8}------------------------------------------------
# 6 GLEAMER
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Gleamer. The logo consists of a red circular icon with a white center, followed by the word "GLEAMER" in a simple, sans-serif font. The text is in a dark blue color. The logo is clean and modern in appearance.
# 7. Performance data
### 7.1. Biocompatibility Testing
As a standalone software, BoneView has no direct patient or user contacting components. Therefore, biocompatibility information is not required for this device.
### 7.2. Software Verification and Validation Testing
BoneView is a standalone software that is considered a moderate level of concern as per the guidance document from the FDA: "Guidance for the Content of Premarket Submissions for Software in Medical Devices". Indeed, a failure or latent design flaw of BoneView could directly result in minor injury to the patient or operator.
Consequently, software verification and validation testing were conducted and documented as per the requirements of the abovementioned FDA guidance document for a moderate level of concern device.
### 7.3. Electrical safety and Electromagnetic compatibility Testing
As a standalone software, BoneView is not subject to electromagnetic compatibility or electrical safety testing activities. Therefore, Electrical safety and Electromagnetic compatibility information is not required for this device.
### 7.4. Bench Testing
### 7.4.1. Testing for the children/adolescent population
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Gleamer. The logo consists of a red circular icon with a white design inside, followed by the word "GLEAMER" in a simple, sans-serif font. The text is in a dark blue or black color, providing a clear contrast against the white background.
In order to include the children and adolescents population in the indications for use of BoneView 1.1-US, Gleamer performed a standalone performance testing on a dataset of 2,000 radiographs (52.8% of males, with age: range [2 – 21]; mean 11.54 +/- 4.7) for all anatomical areas of interest in the Indications for Use for the children and adolescents population and from various manufacturers (Canon, Fujifilm, GE Healthcare, Konica Minolta, Philips, Primax, Samsung, Siemens). This dataset was independent of the data used for model training, tuning, and establishment of device operating points.
The overall goal of the conducted study was to compare the diagnostic performances of BoneView 1.1-US on the children/adolescents clinical performance study dataset to the diagnostic performances of BoneView on the adult clinical performance study dataset (included in the submission of the predicate device).
The results of the study demonstrated that BoneView 1.1-US detects fractures in radiographs with similar performances on the adult population and on the children/adolescents population:
Sensitivity (with 95% Clopper-Pearson Cl) and Specificity (with 95% Clopper-Pearson Cl) of BoneView 1.1-US at the examinationlevel at the high-sensitivity operating point on the children/adolescents clinical performance study dataset VS adult clinical performance study dataset
| Operating Point | Dataset | Sensitivity | Specificity |
|------------------------------------------------------|---------------------------------------------------------------|----------------------------|------------------------|
| High-sensitivity<br>operating point (DOUBT<br>FRACT) | Adult clinical<br>performance study<br>dataset | 0.928 [0.919 - 0.936] | 0.811 [0.8 - 0.821] |
| | Children/adolescents<br>clinical performance<br>study dataset | 0.909 [0.889 - 0.926] | 0.821 [0.796 - 0.844] |
| | 95% confidence interval<br>on the difference | -0.019 [-0.039 -<br>0.001] | 0.010 [-0.016 - 0.037] |
Sensitivity (with 95% Clopper-Pearson Cl) and Specificity (with 95% Clopper-Pearson Cl) of BoneView 1.1-US at the examinationlevel at the high-specificity operating point on the children/adolescents clinical performance study dataset VS adult clinical performance study dataset
| Operating Point | Dataset | Specificity | Sensitivity |
|---------------------------------------------|------------------------------------------------------------|-----------------------|--------------------------|
| High-specificity<br>operating point (FRACT) | Adult clinical<br>performance study<br>dataset | 0.932 [0.925 - 0.939] | 0.841 [0.829 - 0.853] |
| | Children/adolescents clinical performance<br>study dataset | 0.965 [0.952 - 0.976] | 0.792 [0.766 - 0.817] |
| | 95% confidence interval<br>on the difference | 0.033 [0.019 - 0.046] | -0.049 [-0.079 - -0.021] |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "GLEAMER" in a sans-serif font. To the left of the word is a red circular logo with a white space in the middle. The logo appears to be a stylized letter "G" or a circle with a gap in it. The text and logo are aligned horizontally.
In addition to the equivalence of performances with the performances on the adult population, the results of the standalone testing demonstrated that BoneView detects fractures in radiographs with high sensitivity and high specificity:
Specificity (with 95% Clopper-Pearson Cl) and Sensitivity (with 95% Clopper-Pearson Cl) of BoneView at the examination-level at the high-sensitivity operating point and high-specificity operating point on the children/adolescents clinical performance study dataset
| | High-sensitivity operating point | | High-specificity operating point | |
|----------------------------------------------------|-----------------------------------------|-----------------------------------------|-----------------------------------------|-----------------------------------------|
| Standalone<br>Performance | Specificity – 95%<br>Clopper-Pearson CI | Sensitivity – 95%<br>Clopper-Pearson CI | Specificity – 95%<br>Clopper-Pearson CI | Sensitivity – 95%<br>Clopper-Pearson CI |
| Global<br>n(positive)= 1,000<br>n(negative)= 1,000 | 0.821 [0.796 -<br>0.844] | 0.909 [0.889 -<br>0.926] | 0.965 [0.952 -<br>0.976] | 0.792 [0.766 -<br>0.817] |
Specificity (with 95% Clopper-Pearson Cl) and Sensitivity (with 95% Clopper-Pearson Cl) of BoneView at the examination-level for the subgroup analysis of anatomical areas of interest at the high-sensitivity operating point and high-specificity operating point on the children/adolescents clinical performance study dataset
| | High-sensitivity operating point<br>DOUBT FRACT | | High-specificity operating point<br>FRACT | |
|-------------------------------------------------|-------------------------------------------------|-----------------------------------------|-------------------------------------------|-----------------------------------------|
| Anatomical<br>Areas of<br>Interest | Specificity – 95%<br>Clopper-Pearson CI | Sensitivity – 95%<br>Clopper-Pearson CI | Specificity – 95%<br>Clopper-Pearson CI | Sensitivity – 95%<br>Clopper-Pearson CI |
| Ankle<br>n(positive)= 88<br>n(negative)= 157 | TP=75 FP=38<br>0.758 [0.683 - 0.823] | TN=119 FN=13<br>0.852 [0.761 - 0.919] | TP=57 FP=11<br>0.93 [0.878 - 0.965] | TN=146 FN=31<br>0.648 [0.539 - 0.747] |
| Clavicle<br>n(positive)= 113<br>n(negative)= 45 | TP=110 FP=9<br>0.8 [0.654 - 0.904] | TN=36 FN=3<br>0.973 [0.924 - 0.994] | TP=108 FP=1<br>0.978 [0.882 - 0.999] | TN=44 FN=5<br>0.956 [0.9 - 0.985] |
| Elbow<br>n(positive)= 96<br>n(negative)= 120 | TP=87 FP=32<br>0.733 [0.645 - 0.81] | TN=88 FN=9<br>0.906 [0.829 - 0.956] | TP=60 FP=2<br>0.983 [0.941 - 0.998] | TN=118 FN=36<br>0.625 [0.52 - 0.722] |
| Foot<br>n(positive)= 151<br>n(negative)= 173 | TP=129 FP=47<br>0.728 [0.656 - 0.793] | TN=126 FN=22<br>0.854 [0.788 - 0.906] | TP=113 FP=12<br>0.931 [0.882 - 0.964] | TN=161 FN=38<br>0.748 [0.671 - 0.815] |
| Forearm | TP=59 FP=5 TN=35 FN=6 | | TP=53 FP=1 TN=39 FN=12 | |
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for Gleamer. The logo consists of a red circular icon with a white design inside, followed by the word "GLEAMER" in a simple, sans-serif font. The text is in a dark blue color, contrasting with the red icon.
| | High-sensitivity operating point<br>DOUBT FRACT | | High-specificity operating point<br>FRACT | |
|--------------------------------------------|-------------------------------------------------|-----------------------------------------|-------------------------------------------|-----------------------------------------|
| Anatomical<br>Areas of<br>Interest | Specificity - 95%<br>Clopper-Pearson Cl | Sensitivity - 95%<br>Clopper-Pearson Cl | Specificity - 95%<br>Clopper-Pearson Cl | Sensitivity - 95%<br>Clopper-Pearson Cl |
| n(positive)= 65<br>n(negative)= 40 | 0.875 [0.732 -<br>0.958] | 0.908 [0.81 -<br>0.965] | 0.975 [0.868 -<br>0.999] | 0.815 [0.7 - 0.901] |
| Hand | TP=174 FP=18 | TN=142 FN=14 | TP=154 FP=6 | TN=154 FN=34 |
| n(positive)=<br>188<br>n(negative)=<br>160 | 0.887 [0.828 -<br>0.932] | 0.926 [0.878 -<br>0.959] | 0.963 [0.92 -<br>0.986] | 0.819 [0.757 -<br>0.871] |
| Humerus | TP=24 FP=4 | TN=8 FN=0 | TP=22 FP=1 | TN=11 FN=2 |
| n(positive)= 24<br>n(negative)= 12 | 0.667 [0.349 -<br>0.901] | 1.0 [0.858 - 1.0] | 0.917 [0.615 -<br>0.998] | 0.917 [0.73 - 0.99] |
| Knee | TP=36 FP=12 | TN=155 FN=7 | TP=20 FP=4 | TN=163 FN=23 |
| n(positive)= 43<br>n(negative)=<br>167 | 0.928 [0.878 -<br>0.962] | 0.837 [0.693 -<br>0.932] | 0.976 [0.94 -<br>0.993] | 0.465 [0.312 -<br>0.623] |
| Shoulder | TP=80 FP=21 | TN=82 FN=5 | TP=79 FP=2 | TN=101 FN=6 |
| n(positive)= 85<br>n(negative)=<br>103 | 0.796 [0.705 -<br>0.869] | 0.941 [0.868 -<br>0.981] | 0.981 [0.932 -<br>0.998] | 0.929 [0.853 -<br>0.974] |
| Tibia/Fibula | TP=50 FP=7 | TN=33 FN=8 | TP=43 FP=1 | TN=39 FN=15 |
| n(positive)= 58<br>n(negative)= 40 | 0.825 [0.672 -<br>0.927] | 0.862 [0.746 -<br>0.939] | 0.975 [0.868 -<br>0.999] | 0.741 [0.61 -<br>0.847] |
| Wrist | TP=136 FP=20 | TN=70 FN=5 | TP=127 FP=4 | TN=86 FN=14 |
| n(positive)=<br>141<br>n(negative)= 90 | 0.778 [0.678 -<br>0.859] | 0.965 [0.919 -<br>0.988] | 0.956 [0.89 -<br>0.988] | 0.901 [0.839 -<br>0.945] |
Additionally, the performance of BoneView 1.1-US on the children and adolescents population was validated for potential confounders including weight-bearing and non-weight bearing bone fractures and different X-ray system manufacturers.
### 7.4.2. Testing for adult population
BoneView 1.1-US is using the same Al algorithm than the predicate device: BoneView 1.0-US (K212365). Thus, the bench testing (standalone testing) on the adult population described in the 510(k) submission of the predicate device are still valid and applicable to BoneView 1.1-US and are provided here for reference.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Gleamer. The logo consists of a red circular icon with a white center, followed by the word "GLEAMER" in a simple, sans-serif font. The letters are a dark blue color.
Gleamer performed a standalone performance testing on a dataset of 8,918 radiographs (47.2% of males, with age: range [21 – 113]; mean 52.5 +/- 19.8) for all anatomical areas of interest in the Indications for Use and from various manufacturers (Agfa, Fujifilm, GE Healthcare, Kodak, Konica Minolta, Philips, Primax, Samsung, Siemens). This dataset was independent of the data used for model training, tuning, and establishment of device operating points.
The results of the standalone testing demonstrated that BoneView detects fractures in radiographs with high sensitivity and high specificity:
Specificity (with 95% Clopper-Pearson Cl) and Sensitivity (with 95% Clopper-Pearson Cl) of BoneView at the examination-level at the high-sensitivity operating point and high-specificity operating point on the merged datasets
| | High-sensitivity operating point | | High-specificity operating point | |
|----------------------------------------------------|-----------------------------------------|-----------------------------------------|-----------------------------------------|-----------------------------------------|
| Standalone<br>Performance | Specificity – 95%<br>Clopper-Pearson CI | Sensitivity – 95%<br>Clopper-Pearson CI | Specificity – 95%<br>Clopper-Pearson CI | Sensitivity – 95%<br>Clopper-Pearson CI |
| Global<br>n(positive)= 3,886<br>n(negative)= 5,032 | 0.811 [0.8 - 0.821] | 0.928 [0.919 - 0.936] | 0.932 [0.925 - 0.939] | 0.841 [0.829 - 0.853] |
Specificity (with 95% Clopper-Pearson Cl) and Sensitivity (with 95% Clopper-Pearson Cl) of BoneView at the examination-level for the subgroup analysis of anatomical areas of interest at the high-sensitivity operating point and high-specificity operating noint on the merged datasets
| | and high-specificity operating point on the merged datasets<br>High-sensitivity operating point<br>DOUBT FRACT | | High-specificity operating point<br>FRACT | |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------------|--------------------------------------------|--------------------------------------------|
| Anatomical Areas<br>of Interest | Specificity - 95%<br>Clopper-Pearson<br>CI | Sensitivity - 95%<br>Clopper-Pearson<br>CI | Specificity - 95%<br>Clopper-Pearson<br>CI | Sensitivity - 95%<br>Clopper-Pearson<br>CI |
| Ankle<br>n(positive)= 378<br>n(negative)= 805 | 0.784 [0.754 -<br>0.812] | 0.95 [0.923 -<br>0.969] | 0.897 [0.874 -<br>0.917] | 0.899 [0.865 -<br>0.928] |
| Clavicle<br>n(positive)= 147<br>n(negative)= 255 | 0.757 [0.699 -<br>0.808] | 0.905 [0.845 -<br>0.947] | 0.929 [0.891 -<br>0.958] | 0.83 [0.759 -<br>0.887] |
| Elbow<br>n(positive)= 145<br>n(negative)= 227 | 0.718 [0.655 -<br>0.776] | 0.924 [0.868 -<br>0.962] | 0.899 [0.852 -<br>0.935] | 0.531 [0.446 -<br>0.614] |
| Femur<br>n(positive)= 63<br>n(negative)= 161 | 0.733 [0.658 -<br>0.799] | 0.937 [0.845 -<br>0.982] | 0.944 [0.897 -<br>0.974] | 0.825 [0.709 -<br>0.909] |
| Foot<br>n(positive)= 985<br>n(negative)=<br>1,097 | 0.793 [0.768 -<br>0.817] | 0.934 [0.917 -<br>0.949] | 0.924 [0.907 -<br>0.939] | 0.874 [0.852 -<br>0.894] |
| | High-sensitivity operating point<br>DOUBT FRACT | | High-specificity operating point<br>FRACT | |
| Anatomical Areas<br>of Interest | Specificity - 95%<br>Clopper-Pearson<br>CI | Sensitivity - 95%<br>Clopper-Pearson<br>CI | Specificity - 95%<br>Clopper-Pearson<br>CI | Sensitivity - 95%<br>Clopper-Pearson<br>CI |
| Forearm<br>n(positive)= 94<br>n(negative)= 102 | 0.676 [0.577 -<br>0.766] | 0.989 [0.942 - 1.0] | 0.912 [0.839 -<br>0.959] | 0.851 [0.763 -<br>0.916] |
| Hand<br>n(positive)= 1,168<br>n(negative)=<br>1,003 | 0.809 [0.783 -<br>0.832] | 0.966 [0.954 -<br>0.975] | 0.917 [0.898 -<br>0.934] | 0.915 [0.898 -<br>0.931] |
| Hip<br>n(positive)= 145<br>n(negative)= 235 | 0.77 [0.711 -<br>0.822] | 0.938 [0.885 -<br>0.971] | 0.953 [0.918 -<br>0.976] | 0.793 [0.718 -<br>0.856] |
| Humerus<br>n(positive)= 114<br>n(negative)= 175 | 0.731 [0.659 -<br>0.796] | 0.904 [0.834 -<br>0.951] | 0.92 [0.869 -<br>0.956] | 0.833 [0.752 -<br>0.897] |
| Knee<br>n(positive)= 128<br>n(negative)=<br>1,045 | 0.889 [0.868 -<br>0.907] | 0.891 [0.823 -<br>0.939] | 0.975 [0.964 -<br>0.984] | 0.797 [0.717 -<br>0.863] |
| Lumbosacral<br>Spine<br>n(positive)= 125<br>n(negative)= 209 | 0.737 [0.672 -<br>0.795] | 0.776 [0.693 -<br>0.846] | 0.947 [0.908 -<br>0.973] | 0.6 [0.509 - 0.687] |
| Pelvis<br>n(positive)= 230<br>n(negative)= 479 | 0.745 [0.704 -<br>0.784] | 0.887 [0.839 -<br>0.925] | 0.939 [0.914 -<br>0.959] | 0.743 [0.682 -<br>0.799] |
| Ribs<br>n(positive)= 252<br>n(negative)= 95 | 0.684 [0.581 -<br>0.776] | 0.753 [0.7 - 0.802] | 0.926 [0.854 -<br>0.97] | 0.488 [0.425 -<br>0.552] |
| Shoulder<br>n(positive)= 255<br>n(negative)= 586 | 0.782 [0.746 -<br>0.814] | 0.929 [0.891 -<br>0.958] | 0.947 [0.926 -<br>0.964] | 0.851 [0.801 -<br>0.892] |
| Thoracic Spine<br>n(positive)= 74<br>n(negative)= 105 | 0.676 [0.578 -<br>0.764] | 0.878 [0.782 -<br>0.943] | 0.905 [0.832 -<br>0.953] | 0.689 [0.571 -<br>0.792] |
| Tibia/Fibula<br>n(positive)= 72<br>n(negative)= 184 | 0.712 [0.641 -<br>0.776] | 0.972 [0.903 -<br>0.997] | 0.815 [0.751 -<br>0.869] | 0.931 [0.845 -<br>0.977] |
| | High-sensitivity operating point<br>DOUBT FRACT | | High-specificity operating point<br>FRACT | |
| Anatomical Areas<br>of Interest | Specificity - 95%<br>Clopper-Pearson<br>Cl | Sensitivity - 95%<br>Clopper-Pearson<br>Cl | Specificity - 95%<br>Clopper-Pearson<br>Cl | Sensitivity - 95%<br>Clopper-Pearson<br>Cl |
| Wrist<br>n(positive)= 573<br>n(negative)= 502 | 0.771 [0.732 -<br>0.807] | 0.97 [0.953 -<br>0.983] | 0.892 [0.862 -<br>0.918] | 0.934 [0.91 -<br>0.953] |
{14}------------------------------------------------
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Gleamer. The logo consists of a red circular icon to the left of the word "GLEAMER" in a dark blue sans-serif font. The icon appears to be a stylized letter "G" or a circular shape with a break in the upper right quadrant.
Additionally, the performance of BoneView was validated for potential confounders including weightbearing and non-weight bearing bone fractures and different X-ray system manufacturers.
### 7.5. Animal Studies
No animal studies were conducted in support of the 510(k) submission of BoneView.
### 7.6. Clinical Studies
No clinical studies were conducted in support of the 510(k) submission of BoneView 1.1-US.
BoneView 1.1-US is based on the same Al algorithm than the predicate device: BoneView 1.0-US (K212365). Thus, the clinical performance described in the 510(k) submission of the predicate device are still valid and applicable to BoneView 1.1-US, for both the adult and children adolescent population. The results are provided here for reference.
Gleamer conducted a fully-crossed multiple reader, multiple case (MRMC) retrospective reader study to determine the impact of BoneView on reader performance in diagnosing fractures. The primary objective of the study was to determine whether the diagnostic accuracy of readers aided by BoneView is superior to the diagnostic accuracy of readers unaided by BoneView as determined by the Specificity/Sensitivity pair (primary endpoint).
The clinical validation study design was the following:
- . 24 clinical readers each evaluated a dataset of 480 cases (31.9% of males, with age: range [21 – 93]; mean 59.2 +/- 16.4) in BoneView's Indications for Use and from various manufacturers (GE Healthcare, Kodak, Konica Minolta, Philips, Samsung) under both Aided and Unaided conditions.
- This dataset was independent of the data used for model training, and establishment of device operating points.
- . Each case had been previously evaluated by a panel of three U.S. board-certified radiologists who assigned a ground truth label indicating the presence of a fracture and its location.
- Cases are from all the anatomical areas of interest included in BoneView's Indications for Use.
- The MRMC study consisted of two independent reading sessions separated by a washout period of at least one month in order to avoid memory bias.
- . For each case, each reader was required to provide a determination of the presence of a fracture and provide its location.
GLEAMER
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Gleamer. The logo consists of a red circular icon with a white design inside, followed by the word "GLEAMER" in a dark blue sans-serif font. The icon is positioned to the left of the text, creating a visual balance.
The results of the study found that the diagnostic accuracy of readers in the intended use population is superior when aided by BoneView than when unaided by BoneView, as measured at the task of fracture detection using the Specificity/Sensitivity pair.
In particular, the study results demonstrated:
- . Reader specificity improved significantly from 0.906 (95% bootstrap Cl: 0.898-0.913) to 0.956 (95% bootstrap CI: 0.951-0.960): +5% increase of the Specificity
- Reader sensitivity improved significantly from 0.648 (95% bootstrap Cl: 0.640-0.656) to 0.752 (95% . bootstrap CI: 0.745-0.759): +10.4% increase of the Sensitivity
Additionally, subgroup analysis was carried out by anatomical areas of interest, listed in the Indications for Use. The subgroup analysis found that the Sensitivity and Specificity were higher for Aided reads versus Unaided reads for all of the anatomical areas of interest.
# 8. Conclusion
BoneView 1.1-US and BoneView 1.0-US predicate device have the same intended use and technological characteristics. Only the indications for use are different with the inclusion of children and adolescents in the intended patient population.
Performance testing was conducted to validate the performance of BoneView 1.1-US on the new patient population. The results of the testing show that the device performs as intended and the differences in indications for use including the new patient population of children and adolescents does not raise different questions of safety or effectiveness as compared with the predicate device.
Therefore, BoneView 1.1-US subject device and BoneView 1.0-US predicate device (K212365) are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.