The Swan-Ganz IQ catheter (model AIQSGF8) is for use in patients who require hemodynamic monitoring. It is intended to be used in combination with clinical pressure monitoring, equipment to measure right heart and pulmonary artery pressures, and with a compatible cardiac output computer to measure intermittent cardiac output. Model AIOSGF8 also measures mixed venous oxygen saturation.
Device Story
Swan-Ganz IQ Pulmonary Arterial catheter; flow-directed catheter for hemodynamic monitoring in critically ill patients. Inputs: right heart/pulmonary artery pressures, mixed venous oxygen saturation, intermittent cardiac output (via thermodilution). Operation: catheter inserted into pulmonary artery; pressure transducers and bedside physiologic monitors process signals. Output: hemodynamic pressure data, cardiac output, oxygen saturation. Used in clinical settings (ICU/OR) by physicians/nurses. Provides real-time data on stroke volume, preload, afterload, and contractility to guide clinical management of circulatory imbalances. Benefits: enables precise diagnosis and management of complex cardiac/pulmonary conditions.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included packaging, shelf life, sterilization, biocompatibility, frequency response, pressure tubing pull testing, and lumen burst pressure testing per ISO 10555-1:2013 and IEC 60601:2015 Amd.1:2020.
Technological Characteristics
Flow-directed pulmonary arterial catheter. Features include an RV pressure monitoring port and thermistor. Materials and scientific technology are identical to predicate. Connectivity: interfaces with bedside physiologic monitors and cardiac output computers. Standards: ISO 10555-1:2013, IEC 60601:2015 Amd.1:2020. Sterilization: not specified.
Indications for Use
Indicated for patients requiring hemodynamic monitoring, including those with acute heart failure, severe hypovolemia, complex circulatory situations, medical emergencies, adult respiratory distress syndrome, gram-negative sepsis, drug intoxication, acute renal failure, hemorrhagic pancreatitis, high-risk intra/post-operative status, pulmonary or cardiac disease, fluid shifts, high-risk obstetrical conditions, toxemia, premature separation of placenta, or cardiac tamponade. Used for cardiac output determination, differential diagnosis of mitral regurgitation/ventricular septal rupture, blood sampling, and infusion of saline/dextrose solutions.
Regulatory Classification
Identification
A flow-directed catheter is a device that incorporates a gas-filled balloon to help direct the catheter to the desired position.
Predicate Devices
Edwards Swan-Ganz catheters (Models 777F8 and 780F75M) (K160084)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
December 9, 2022
Edwards Lifesiences, LLC Michelle Ducca Manager, Regulatory Affairs 1 Edwards Way Irvine, California 92614
Re: K222117
Trade/Device Name: Swan-Ganz IQ Pulmonary Arterial catheter Regulation Number: 21 CFR 870.1240 Regulation Name: Flow-Directed Catheter Regulatory Class: Class II Product Code: DYG Dated: November 4, 2022 Received: November 7, 2022
Dear Michelle Ducca:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K222117
#### Device Name
Swan-Ganz IQ Pulmonary Arterial Catheter
Indications for Use (Describe)
The primary indications for the Swan-Ganz IQ Pulmonary Arterial catheter includes:
- · Acute heart failure
- Severe hypovolemia
- Complex circulatory situations
- · Medical emergencies
- · Adult respiratory distress syndrome
- · Gram negative sepsis
- · Drug intoxication
- Acute renal failure
- · Hemorrhagic pancreatitis
- · Intra and post-operative management of high risk patients
- History of pulmonary or cardiac disease
- · Fluid shifts (e.g., extensive intra-abdominal operations)
- · Management of high-risk obstetrical patients
- · Diagnosed cardiac disease
- · Toxemia
- · Premature separation of placenta
- · Cardiac output determinations
- · Differential diagnosis of mitral regurgitation and ventricular septal rupture
- · Diagnosis of cardiac tamponade
Secondary indications include the following:
- · Blood Sampling
- · Infusion of saline and dextrose solutions
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Sponsor: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration<br>Number: | 2015691 |
| Contact<br>Person: | Michelle Ducca<br>Manager, Regulatory Affairs<br>One Edwards Way<br>Irvine, CA 92614<br>michelle_ducca@edwards.com<br>Telephone: (949) 250-4113 |
| Date: | July 15, 2022 |
| Trade Name: | Swan-Ganz IQ catheter (subject) |
| Common<br>Name: | Cardiovascular Diagnostic Catheters |
| Classification<br>Name: | Flow-directed catheter 21 CFR 870.1240 |
| Product<br>Code: | DYG, Class II |
| Predicate<br>Device(s): | The Edwards Swan-Ganz catheters (Models 777F8 and 780F75M),<br>manufactured by Edwards Lifesciences, K160084 cleared May 3, 2016, is<br>being utilized for substantial equivalence due to the same basic device<br>functionality, similar indications and intended use, labeling and principle of<br>operation as the subject device. |
| Device<br>Description: | The Edwards Swan-Ganz catheters are used to monitor the hemodynamic status<br>of critically ill and injured patients. The existing catheters give clinicians the<br>ability to measure right heart pressures, pulmonary artery occlusion pressure<br>("wedge"), sample mixed venous blood from the pulmonary artery, as well as<br>measure cardiac output through thermodilution when used with a bedside<br>physiologic monitor and pressure transducers.<br><br>The Edwards Swan-Ganz product line is comprised of Standard (Base) and<br>Advanced Technology monitoring catheters. The standard Swan-Ganz catheter |
# 510(k) Summary – Swan-Ganz IQ catheter model AIQSGF8
{5}------------------------------------------------
measures: right heart pressures, thermodilution cardiac output (room temperature and chilled) and provides a mechanism for pulmonary artery blood sampling for laboratory analysis.
The subject Swan-Ganz IO catheter combines from the existing Advanced catheter design (model 777F8) and the Paceport catheter design (model 780F75M). Using the 777F8 model as the base design, the new catheter modifies the design to remove the thermal filament functionality and replaces it with an additional port which will allow for pressure monitoring at the right ventricular level. The RV port is utilized just like the Swan-Ganz Paceport Oximetry catheter model 780F75M.
Just like the existing advanced catheters, the Swan-Ganz IQ catheter provides the same functionality as the standard and Advanced Technology Swan-Ganz catheters by providing the ability to continuously monitor the patient's balance between oxygen delivery and consumption as well as the ability to help investigate the root cause of an imbalance through analysis of the components of stroke volume (preload, afterload, and contractility).
## Indications Indications for Use - Swan-Ganz IQ catheter: for Use: The primary indications for the Swan-Ganz IQ Pulmonary Arterial catheter includes: • Acute heart failure • Severe hypovolemia · Complex circulatory situations • Medical emergencies · Adult respiratory distress syndrome · Gram negative sepsis • Drug intoxication • Acute renal failure · Hemorrhagic pancreatitis • Intra and post-operative management of high risk patients • History of pulmonary or cardiac disease · Fluid shifts (e.g., extensive intra-abdominal operations) · Management of high-risk obstetrical patients • Diagnosed cardiac disease
- Toxemia
- Premature separation of placenta
- Cardiac output determinations
- · Differential diagnosis of mitral regurgitation and ventricular septal rupture
- Diagnosis of cardiac tamponade
Secondary indications include the following:
- Blood Sampling
{6}------------------------------------------------
- · Infusion of saline and dextrose solutions
#### Intended Use: Intended Use- Swan-Ganz IQ catheter:
Device:
The Swan-Ganz IQ catheter (model AIQSGF8) is for use in patients who require hemodynamic monitoring. It is intended to be used in combination with clinical pressure monitoring, equipment to measure right heart and pulmonary artery pressures, and with a compatible cardiac output computer to measure intermittent cardiac output. Model AIOSGF8 also measures mixed venous oxygen saturation.
#### Comparison The following technological similarities exist between the subject and predicate to Predicate devices:
The scientific technology and materials of the subject device are unchanged from the legally marketed predicate devices. Both the subject and predicate devices provide the same hemodynamic pressure monitoring and provides intermittent cardiac output, mixed venous oxygen saturation and stroke volume. The modified Swan-Ganz catheter device is identical to the predicate devices in terms of indications for use and has the same overall intended use without the thermodilution functions.
The following technological differences exist between the subject and predicate devices:
- . Thermal filament removed: The currently cleared predicate catheter model 777F8 (K160084 cleared May 3, 2016) has been modified to remove the components relating to the thermal filament including white heater extension tubing.
- . Additional port: An additional port has been implemented to replace the thermal filament with extension tubing (violet) to allow for pressure monitoring from the right ventricular level. The same RV port for pressure monitoring is already available in the existing predicate device model 780F75M (K160084 cleared May 3, 2016).
- Addition of y-connector as tether point on thermistor extension tubing . (yellow) along with short white extension tubing for EEPROM connector.
These differences are not a fundamental change in scientific technology and have no impact to the indications for use or overall intended use.
The existing Swan-Ganz catheter product line includes IFU's for the Pacing, Base and Advanced models. An additional separate IFU for the Swan-Ganz IQ catheter model will be added to the Swan-Ganz product line, which includes the same Indication for Use and the same overall Intended Use as the predicate
{7}------------------------------------------------
devices (K160084), an updated image, and removal of thermal filament functionality and associated bolus thermodilution parameters (CCO, EDV, RVEF) in the Instructions for Use.
Performance The Swan-Ganz IQ catheter has successfully passed functional and Data: performance testing including packaging, shelf life, sterilization, biocompatibility and bench testing, which included frequency response testing, pressure tubing pull testing and lumen burst pressure testing per ISO 10555-1:2013 and IEC 60601:2015 Amd.1:2020.
- The subject Swan-Ganz IQ catheter has successfully passed all functional and Conclusions performance testing including verification and validation, sterilization, shelf life, biocompatibility and bench testing. Completion of all performance verification and validation activities demonstrated that the subject device meets its predetermined design and performance specifications. Verification activities performed confirmed that the differences in the features and design did not adversely affect the safety and effectiveness of the subject device. The testing performed demonstrates that the subject Swan-Ganz IQ (model AIQSGF8) is substantially equivalent to the predicate Swan-Ganz catheter predicate devices models 777F8 and 780F75M (K160084, cleared May 3, 2016).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.