K222115 · MicroVention, Inc. · DQY · Jan 10, 2023 · Cardiovascular
Device Facts
Record ID
K222115
Device Name
ISAAC Neurovascular Navigation Catheter
Applicant
MicroVention, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Jan 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The ISAAC™ Neurovascular Navigation Catheter is indicated for use in facilitating advancement of catheters through the neuro and peripheral vasculature and introduction of diagnostic agents. The ISAAC™ Neurovascular Navigation Catheter is not intended for use in the coronary vasculature.
Device Story
ISAAC Neurovascular Navigation Catheter is a braid-reinforced, variable-stiffness, single-lumen catheter with a pre-shaped distal segment. It features a hydrophilic coating on the distal curve, a radiopaque tungsten-coiled distal section, and a proximal Luer hub. Used by clinicians in neuro/peripheral vascular procedures to navigate vasculature and deliver diagnostic agents. Device is inserted percutaneously; clinician manually advances catheter through vasculature. Radiopaque markers allow visualization under X-ray fluoroscopy to guide placement. Benefits include improved trackability and support during navigation. Device is sterile, single-use, and non-pyrogenic.
Clinical Evidence
No clinical testing performed. Evidence consists of benchtop performance testing (physical attributes, tip flexibility, simulated use, kink resistance, burst/leakage, tensile strength, coating durability) and an acute GLP-compliant porcine animal study. Animal study compared tracking performance, support, and safety against the predicate; results showed comparable performance with no significant device-associated complications (e.g., dissection, perforation, thrombus).
Indicated for facilitating catheter advancement through neuro and peripheral vasculature and introducing diagnostic agents. Not for use in coronary vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
January 10, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MicroVention, Inc. Sona Manickam Manager Regulatory Affairs 35 Enterprise Aliso Viejo, California 92656
Re: K222115
Trade/Device Name: ISAAC Neurovascular Navigation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, QJP Dated: December 9, 2022 Received: December 9, 2022
Dear Sona Manickam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222115
Device Name
ISAACTM Neurovascular Navigation Catheter
Indications for Use (Describe)
The ISAAC™ Neurovascular Navigation Catheter is indicated for use in facilitating advancement of catheters through the neuro and peripheral vasculature and introduction of diagnostic agents. The ISAAC™ Neurovascular Navigation Catheter is not intended for use in the coronary vasculature.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="white-space: nowrap;"> <span style="font-size: 1.2em;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="white-space: nowrap;"> <span style="font-size: 1.2em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo consists of a circular graphic on the left, followed by the text "MicroVention" in a larger, sans-serif font. Below "MicroVention" is the word "TERUMO" in a smaller, bolder font. The overall design is clean and corporate.
| 510(k) Owner | MicroVention, Inc. |
|-----------------------|-------------------------------------------|
| | 35 Enterprise |
| | Aliso Viejo, CA 92656 |
| | Establishment Registration No: 3013556777 |
| Contact Person | Sona Manickam |
| | Manager Regulatory Affairs |
| | Telephone: (712)-582-8133 |
| | Email: Sona.Manickam@microvention.com |
| Date Summary Prepared | January 09, 2023 |
| Trade Name | ISAACT™ Neurovascular Navigation Catheter |
| Common Name | Catheter, Percutaneous |
| Classification | Class II, DQY, QJP |
| Regulation | 21 CFR 870.1250 |
| Predicate Device | Chaperon Guiding Catheter (K082385) |
| Reference Devices | Neuron Select Catheter (K083125) |
| | Neuron MAX System (K111380) |
## K222115 510(k) Summary
#### Device Description
The ISAAC™ Neurovascular Navigation Catheter) is a braid-reinforced variable stiffness catheter with a pre-shaped distal segment. The distal end of the catheter is coated with a hydrophilic coating around the curve of the pre-shaped section of the Simmons (SIM) configuration. It is a single lumen catheter with a radiopaque distal coiled section and a Luer hub on the proximal end. The ISAAC Catheter is sterile, non-pyrogenic and intended for single use only.
#### Indications for Use
The ISAAC™ Neurovascular Navigation Catheter is indicated for use in facilitating advancement of catheters through the neuro and peripheral vasculature and introduction of diagnostic agents. The ISAAC™ Neurovascular Navigation Catheter is not intended for use in the coronary vasculature.
#### Comparison of Indications for Use and Technological Characteristics
The subject device, ISAAC Catheter, and the predicate device, the inner catheter of the Chaperon Guiding Catheter (K082385) are substantially equivalent in that these devices have similar intended use, principles of operation, and fundamental design.
As evidenced by the data, the subject, predicate, and reference devices:
- have similar intended use,
- . use the same principles of operation,
- . incorporate the same basic design,
- . use similar construction and materials,
- . are Ethylene Oxide (EtO)- sterilized and packaged using the same processes.
The following table provides a comparison of the key characteristics of the ISAAC Catheter to the predicate and reference devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a stylized, circular graphic on the left, followed by the text "MicroVention" in a blue, sans-serif font. Below "MicroVention" is the word "TERUMO," also in blue and in a bolder, sans-serif font. The overall design is clean and corporate, suggesting a company in the medical or technology sector.
## Device Comparison Table
| Device<br>Characteristics | Proposed Device<br>ISAAC Neurovascular<br>Navigation Catheter (K222115) | Predicate Device<br>Chaperon Guiding Catheter<br>(K082385) | Reference Device<br>Neuron Select Catheter<br>(K083125) |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The ISAAC Neurovascular<br>Navigation Catheter is<br>indicated for use in<br>facilitating advancement of<br>catheters through the neuro<br>and peripheral vasculature<br>and introduction of<br>diagnostic agents. The ISAAC<br>Neurovascular Navigation<br>Catheter is not intended for<br>use in the coronary<br>vasculature. | Chaperon Guiding Catheter is<br>intended for general<br>intravascular use, including the<br>neuro and peripheral<br>vasculature. Chaperon Guiding<br>Catheter can be used to<br>facilitate introduction of<br>diagnostic or therapeutic<br>devices. Chaperon Guiding<br>Catheter is not intended for use<br>in coronary arteries. | The Neuron Intracranial<br>Access System is indicated<br>for the introduction of<br>interventional devices into<br>the peripheral, coronary<br>and neuro vasculature. |
| Device Class | Class II DQY, QJP<br>21 CFR 870.1250 | Class II DQY<br>21 CFR 870.1250 | Class II DQY<br>21 CFR 870.1250 |
| Catheter Body | Braid: Stainless Steel<br>Catheter:<br>Polytetrafluoroethylene<br>(PTFE)<br>Coiling: Tungsten<br>Outer: Polymer | Braid: Stainless Steel<br>Catheter: PTFE | Braid: Stainless Steel<br>Catheter: PTFE |
| Marker | Radiopaque coil | Radiopaque marker | Radiopaque jacket |
| Hub | Nylon | Nylon | Nylon |
| Strain Relief | Polyurethane | Polyurethane | Polyurethane |
| Inner Diameter | 6F: 1.02 mm (0.040") | 5F: 1.22 mm (0.048") | 6F: 1.02 mm (0.040") |
| Outer Diameter | 6F: 2.11 mm (0.083") | 5F: 1.70 mm (0.067") | 6.5F: 2.16 mm (0.085") |
| Effective Length | SIM2-3D: 150 cm<br>SIM3-3D: 150 cm<br>TIP45: 140 cm | 117 cm | 125 cm |
| Coating | Hydrophilic | Hydrophilic | Hydrophilic |
| Tip<br>Configuration | Variable Tip Shape & Size | Variable Tip Shape & Size | Variable Tip Shape & Size |
| Guidewire<br>Compatibility | 0.035" and 0.038" | 0.035" and 0.038" | 0.035" and 0.038" |
| Accessories | Not applicable | Not applicable | Not applicable |
| Method of<br>Supply | Sterile and single use | Sterile and single use | Sterile and single use |
| Sterilization<br>Method | Sterilized using 100%<br>Ethylene Oxide | Sterilized using 100%<br>Ethylene Oxide | Sterilized using 100%<br>Ethylene Oxide |
| Device<br>Characteristics | Proposed Device | Predicate Device | Reference Device |
| Packaging<br>Configuration | ISAAC Neurovascular<br>Navigation Catheter<br>(K222115) | Chaperon Guiding Catheter<br>(K082385) | Neuron Select Catheter<br>(K083125) |
| | Placed on mounting card,<br>Tyvek pouch, shipping<br>carton | Placed on mounting card,<br>Tyvek pouch, shipping<br>carton | Placed on mounting card,<br>Tyvek pouch, shipping<br>carton |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a stylized graphic of concentric circles on the left. To the right of the graphic is the text "MicroVention" in a blue sans-serif font, with the trademark symbol to the upper right. Below "MicroVention" is the word "TERUMO" in a bolder, blue sans-serif font.
### Performance Testing Summary
The following performance testing was conducted to support the ISAAC Catheter and demonstrate that it performs as intended:
| Test | Test Method Summary | Results |
|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Physical Attributes | The usable length, proximal and distal outer diameters, distal length, and<br>inner diameters were measured. | Pass |
| Tip Flexibility | The device shall have less force to bend at the distal tip than the<br>comparator reference device. | Pass |
| Simulated Use | The performance of the catheter was rated during simulated-use testing<br>in benchtop vessel model. | Pass |
| Radio Detectability | The catheter must be visible under X-ray fluoroscopy. | Pass |
| Kink Resistance | Kink resistance was measured by subjecting the device to bending in<br>simulated tortuous anatomy. | Pass |
| Static Burst | The distal tip of the catheter was occluded, and fluid was injected into the<br>lumen at increasing pressure until the catheter burst. | Pass |
| Liquid Leakage | The distal tip of the catheter was occluded, and fluid was injected.<br>Pressure was maintained for 30 seconds, and the device was inspected<br>for leakage per ISO 10555-1. | Pass |
| Leakage and Damage<br>Under High Static<br>Pressure Conditions | Dyed fluid was injected into the lumen until the rated burst pressure was<br>reached and the pressure was held for 10 seconds while the distal tip is<br>occluded. | Pass |
| Air Leakage | A vacuum was applied to the catheter and observed for air bubbles, per<br>ISO 80369-7 and ISO 10555-1. | Pass |
| Dynamic Burst | The catheter was injected with fluid at a set pressure and inspected for<br>damage. | Pass |
| Force at Break<br>(Hub/Distal) | The distal and proximal sections of the catheter are secured into a tensile<br>test machine. The machine pulled until the catheter broke, and the pull<br>force was recorded. | Pass |
| Particulate Testing | The catheter underwent simulated-use testing in a benchtop model and<br>was evaluated for particulates. | Pass |
| Surface Contamination | Device must be free from visible surface defects. | Pass |
| Corrosion Resistance | The catheter was tested to evaluate corrosion resistance per ISO 10555-1<br>and ISO 11070. | Pass |
| Coating Durability and<br>Lubricity | Device was secured to a tensile machine and put into heated water bath<br>for hydration. The force to slide through the clamp was recorded as<br>lubricity through 20 cycles. Durability was recorded through 100 cycles. | Pass |
| Catheter Flexural<br>Fatigue | The tensile strength and pressure characteristics were measured per ISO<br>10555-1. | Pass |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a stylized, circular graphic on the left, composed of curved lines suggesting movement or flow. To the right of the graphic, the word "MicroVention" is written in a blue sans-serif font, with the "®" symbol next to it. Below "MicroVention", the word "TERUMO" is written in a bolder, blue sans-serif font.
| Test | Test Method Summary | Results |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Hub and Luer<br>Connector | The Luer connector was tested to dimensional and performance<br>requirements per ISO 80369-7. | Pass |
| Stiffness | The catheter stiffness profile was compared to the reference device. | Pass |
| Torque Strength | The device was evaluated for torque strength by measuring the number of<br>catheter rotations until failure after tracking through a tortuous<br>anatomical model. | Pass |
#### Biocompatibility Testing Summary
The biological safety of the ISAAC Catheter was verified in accordance with ISO 10993-1 categorized as an externally communicating device directly contacting blood for a limited duration (≤ 24 hours). All ISAAC Catheter materials have a history of use in medical devices. The table below illustrates specific tests performed.
| Test Method Summary | | | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Test | Extract(s) & Test Systems | Extraction Conditions | Results |
| Cytotoxicity<br>(ISO MEM Elution Test) | L-929 mouse fibroblast cells<br>prepared using MEM<br>Maintenance Growth Media<br>with 5% FBS | 6.0 cm2/mL (exposed surface area<br>to extraction medium volume<br>ratio), extracted at 37 °C for 24 hrs. | Non-cytotoxic |
| Irritation Reactivity<br>(ISO Intracutaneous<br>Reactivity Test) | Normal saline and sesame<br>seed oil (SSO) (tested<br>separately)<br>New Zealand White Rabbits | 6.0 cm2/mL (exposed surface area<br>to extraction medium volume<br>ratio), extracted at 50 ± 2°C for 72<br>± 2hrs. | Non-irritant |
| Maximization<br>(ISO Guinea Pig Maximization<br>Test) | Normal saline and SSO (tested<br>separately) | 6.0 cm2/mL (exposed surface area<br>to extraction medium volume<br>ratio), extracted at<br>50 ± 2°C for 72 ± 2hrs. | Non- sensitizing |
| Systemic Toxicity<br>(ISO Acute Systemic<br>Toxicity Test) | Normal Saline and SSO (tested<br>separately)<br>Albino outbred strain (ND4)<br>mice (20 male, young<br>adults) | 6.0 cm2/mL (exposed surface area<br>to extraction medium volume<br>ratio), extracted at 50 ± 2°C for 72<br>± 2hrs. | Non-acute<br>systemically toxic |
| Pyrogenicity<br>(ISO/USP Material Mediated<br>Pyrogenicity Test) | Normal saline and SSO (tested<br>separately)<br>New Zealand White Rabbits | 6.0 cm2/mL (exposed surface area<br>to extraction medium volume<br>ratio), extracted at 50 ± 2°C for 72<br>± 2hrs. | Non- pyrogenic |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a circular design on the left, with the text "MicroVention" in a combination of blue and gray. Below "MicroVention" is the word "TERUMO" in blue, indicating a partnership or affiliation between the two entities.
| Test Method Summary | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|--|
| Test | Extract(s) & Test Systems | Extraction Conditions | Results | |
| Hemocompatibility In-Vitro<br>Blood Loop Assay<br>(ISO In-Vitro Blood Loop<br>Assay) | A loop system circulated with<br>freshly drawn sheep blood | Direct exposure at 37±2°C for 4<br>hours ± 30 minutes. | Thrombogenic<br>risk potential<br>similar to the<br>predicate | |
| Hemocompatibility Hemolysis<br>Assay<br>(ISO ASTM Hemolysis Assay) | Phosphate Buffered Saline<br>(PBS)<br>New Zealand White Rabbits | 6.0 cm²/mL (exposed surface area<br>to extraction medium volume<br>ratio), extracted at 50 ± 2°C for 72<br>± 2hrs. | Non- hemolytic | |
| Hemocompatibility<br>Complement Activation Assay<br>(ISO Hemocompatibility:<br>Complement Activation Assay<br>(C3a and SC5b-9)) | Normal Human Serum (NHS) | 6.0 cm²/mL (exposed surface area<br>to extraction medium volume<br>ratio) extracted at 37±2°C for 60<br>min ± 1 minute. | Non-activator of<br>complement<br>system | |
| Hemocompatibility Partial<br>Thromboplastin<br>(ISO Hemocompatibility:<br>Partial Thromboplastin<br>Time (PTT) Assay) | Human Plasma | 6.0 cm²/mL (exposed surface area<br>to extraction medium volume<br>ratio) extracted at 37±2°C for 60<br>min ± 1 minute. | No effect on the<br>PTT | |
| Hemocompatibility<br>Heparinized Blood Platelet<br>and Leukocyte<br>(Hemocompatibility:<br>Heparinized Blood Platelet<br>and Leukocyte Count Assay) | Human Blood | 6.0 cm²/mL (exposed surface area<br>to extraction medium volume<br>ratio) extracted at 37±2°C for 60<br>min ± 1 minute. | Pass | |
#### Animal Study
Acute animal testing was conducted in accordance with FDA Good Laboratory Practice (GLP) Regulation (21 CFR Part 58) comparing the ISAAC Catheter to the predicate Chaperon Guiding Catheter and testing with the BOBBY Guiding Catheter. The testing was intended to assess preclinical safety and efficacy for ISAAC Catheter (test article) and the inner catheter of the Chaperon Guiding Catheter (control article) in a porcine model. The porcine model was chosen since the vessel sizes of the pig model allow for insertion and navigation of standard-sized devices used in humans; porcine vessel diameters are comparable with human vasculature. The test article performed comparably to the control article with regards to tracking performance, support, and safety in this acute in vivo model. No significant device-associated complications (dissection, perforation, embolic debris, thrombus formation, hemorrhage, ischemia, necrosis, fibrin deposition, internal elastic lamina (IEL) rupture, external elastic lamina (EEL) rupture, mineralization, neointimal maturation, medial injury/fibrosis and adventitial injury/fibrosis) were noted in any of the test article and control article treated vessels. The tracking results demonstrated that the ISAAC Catheter and the Chaperon Guiding Catheter performed equally. The results of the present study did not raise any safety issues with either the test ISAAC Catheter or control Chaperon Guiding Catheter. The devices are deemed equivalent.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a stylized, circular graphic element on the left, composed of curved lines in shades of gray and blue, suggesting movement or connectivity. To the right of this graphic is the word "MicroVention" in a blue, sans-serif font, with the "®" symbol next to it. Below "MicroVention" is the word "TERUMO" in a bolder, blue, sans-serif font.
Clinical Testing: No clinical testing was deemed necessary to support the substantial equivalence of t ISAAC Catheter.
#### Conclusion
MicroVention concludes through a review of the benchtop testing, non-clinical animal study assessments, the comparison of the device classifications for use, operating principles, technological characteristics, sterility, and biocompatibility testing that the ISAAC Neurovascular Navigation Catheter is substantially equivalent to the predicate Chaperon Guiding Catheter (K082385). Any differences between the subject device and the predicate device do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.