K222098 · Manamed, Inc. · IMI · Mar 8, 2023 · Physical Medicine
Device Facts
Record ID
K222098
Device Name
ManaSport+
Applicant
Manamed, Inc.
Product Code
IMI · Physical Medicine
Decision Date
Mar 8, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Apply stationary use of ultrasound to: Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation. Apply continuous movement of ultrasound for: 1. Pain. 2. Pain relief, muscle spasms, and joint contractures. 3. Relief of pain, muscle spasms, and joint contractures that may be associated with: Adhesive capsulitis, Bursitis with slight calcification, Myositis, Soft tissue injuries, and Shortened tendons due to past injuries and scar tissues. 4. Relief of pain, muscle spasms, and joint contractures resulting from: Capsular tightness, and Capsular scarring. 5. Localized increase in blood flow. 6. Increased range of motion of contracted joint using heat and stretch techniques.
Device Story
ManaSport+ is a portable, rechargeable ultrasonic diathermy device for home or clinical use. It delivers therapeutic deep heat via piezoelectric ultrasound generation (1.5 MHz). The device features a wireless control option via Bluetooth app, though it remains fully functional independently. It includes safety features like automatic no-load trip, automatic shut-off, and user override. Healthcare providers or patients (under prescription) apply the device to the treatment site. The output (deep heat) is intended to relieve pain, muscle spasms, and joint contractures, and increase local circulation. The device is an adaptation of the ManaSport (K210284) with added home-use support and wireless connectivity.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), ultrasonic performance (IEC 60601-2-5), biocompatibility (ISO 10993-5, ISO 10993-10), home healthcare environment requirements (IEC 60601-1-11), and usability (IEC 60601-1-6).
Technological Characteristics
Piezoelectric ultrasound generator using PZT crystal; 1.5 MHz frequency; continuous wave (100% duty cycle); 120/240 VAC or 3.7 VDC battery power. ABS housing. Dimensions: 140x56x24 mm. Connectivity: Bluetooth wireless control. Standards: IEC 60601-1, 60601-1-2, 60601-2-5, 60601-1-11, 60601-1-6. Non-sterile.
Indications for Use
Indicated for patients requiring therapeutic deep heat for relief of pain, muscle spasms, and joint contractures, or localized increase in blood flow. Conditions include adhesive capsulitis, bursitis, myositis, soft tissue injuries, shortened tendons, capsular tightness, and scarring. Suitable for prescription use in home or clinical settings.
Regulatory Classification
Identification
An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
March 8, 2023
ManaMed. Inc. % Bill Ouangin Dai Official Correspondent JKH USA, LLC 14271 Jeffrey Rd. #246 Irvine, California 92620
Re: K222098
Trade/Device Name: ManaSport+ Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic diathermy Regulatory Class: Class II Product Code: IMI, PFW Dated: February 8, 2023 Received: February 9, 2023
#### Dear Bill Quanqin Dai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Lauren E. Woodard -S
for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22098
Device Name ManaSport+
Indications for Use (Describe)
Apply stationary use of ultrasound to:
Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
Apply continuous movement of ultrasound for:
1. Pain.
- 2. Pain relief, muscle spasms, and joint contractures.
- 3. Relief of pain, muscle spasms, and joint contractures that may be associated with:
- · Adhesive capsulitis,
- · Bursitis with slight calcification,
- · Myositis,
- · Soft tissue injuries, and
- · Shortened tendons due to past injuries and scar tissues.
- 4. Relief of pain, muscle spasms, and joint contractures resulting from:
- · Capsular tightness, and
- · Capsular scarring.
- 5. Localized increase in blood flow.
- 6. Increased range of motion of contracted joint using heat and stretch techniques.
Type of Use (Select one or both, as applicable)
| <span></span> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------|-----------------------------------------------------------|
|--------------------------------------------------------------------------|-----------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Submitter's Information
Submitter: ManaMed. Inc. Address: 5240 W. Charleston Blvd. Las Vegas, NV 89146 Date of Preparation: 03/08/2023
#### 2. Subject Device
Trade/Device Name: ManaSport+ 510(k): K222098 Common Name: Ultrasonic Diathermy Device For Use In Applying Therapeutic Deep Heat Review Panel: Physical Medicine Product Code: IMI, PFW Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic diathermy Device Class: II Use: Prescription
#### 3. Predicate device
Primary Predicate Device: ManaSport 510(k) Number: K210284 Clearance Date: April 26, 2022 Submitter: ManaMed, Inc.
Predicate Device: sam 2.0 Long Duration Ultrasound System 510(k) Number: K191568 Clearance Date: March 6, 2020 Submitter: ZetrOZ Systems, LLC
#### 4. Description of Subject Device
The ManaSport+ is an adaptation of the previously cleared ManaSport device, which was cleared under K210284. A wireless control option was added to the ManaSport+ and the subject device was evaluated for home use. As a portable and rechargeable prescriptive device in the home or clinical/hospital setting, the subject device is intended to apply ultrasound to the patient's treatment site for selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation. Ultrasound can be used to apply therapeutic deep heat for the treatment of selected medical conditions such as soft tissue injuries, shortened tendons due to past injuries and scar tissues, relief of pain, muscle spasms and joint contractures.
## 5. Indications for Use
Apply stationary use of ultrasound to:
Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
{4}------------------------------------------------
Apply continuous movement of ultrasound for:
- 1. Pain.
- 2. Pain relief, muscle spasms, and joint contractures.
- Relief of pain, muscle spasms, and joint contractures that may be associated with: 3.
- Adhesive capsulitis, ●
- Bursitis with slight calcification, ●
- Myositis, ●
- Soft tissue injuries, and .
- Shortened tendons due to past injuries and scar tissues. ●
- 4. Relief of pain, muscle spasms, and joint contractures resulting from:
- Capsular tightness, and ●
- Capsular scarring. .
- 5. Localized increase in blood flow.
- 6. Increased range of motion of contracted joint using heat and stretch techniques.
#### 6. Summary of Substantial Equivalence
The following Table 1 summarizes the comparison between the subject device and the predicate device, indicating the intended use and technical characteristics of the subject device are substantially equivalent to those of the predicate device.
| Parameter &<br>Predicate Device(s) | Subject Device | Primary Predicate Device | Predicate Device | Equivalence |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 510(k)/PMA Number | K222098 | K210284 | K191568 | N/A |
| Submitter<br>/Manufacturer | ManaMed, Inc | ManaMed, Inc. | ZetrOZ Systems, LLC | N/A |
| Device Name/Model | ManaSport+ | ManaSport | sam 2.0 Long Duration Ultrasound<br>System | N/A |
| Product Code | IMI, PFW | IMI, PFW | PFW | Identical |
| Parameter & Predicate<br>Device(s) | Subject Device | Primary Predicate Device | Predicate Device | Equivalence |
| 510(k)/PMA Number | K222098 | K210284 | K191568 | N/A |
| Submitter<br>/Manufacturer | ManaMed, Inc | ManaMed, Inc. | ZetrOZ Systems, LLC | N/A |
| Device Name/Model | ManaSport+ | ManaSport | sam 2.0 Long Duration Ultrasound<br>System | N/A |
| Product Code | IMI, PFW | IMI, PFW | PFW | Identical |
| Indications for Use | Apply stationary use of ultrasound<br>to:<br>Generate deep heat within body<br>tissues for the treatment of selected<br>medical conditions such as the<br>relief of pain, the relief of muscle<br>spasms, the treatment of joint<br>contractures, and the local increase<br>in circulation.<br>Apply continuous movement of<br>ultrasound for:<br>Pain.<br>1.<br>2.<br>Pain relief, muscle<br>spasms, and joint contractures.<br>3.<br>Relief of pain, muscle<br>spasms, and joint contractures that<br>may be associated with:<br>Adhesive capsulitis. | Apply stationary use of ultrasound<br>to:<br>Generate deep heat within body<br>tissues for the treatment of selected<br>medical conditions such as the<br>relief of pain, the relief of muscle<br>spasms, the treatment of joint<br>contractures, and the local increase<br>in circulation.<br>Apply continuous movement of<br>ultrasound for:<br>1.<br>Pain.<br>2.<br>Pain relief, muscle<br>spasms, and joint contractures.<br>Relief of pain, muscle<br>3.<br>spasms, and joint contractures that<br>may be associated with:<br>Adhesive capsulitis. | The sam 2.0 Long Duration Ultrasound<br>Device is intended for home use to<br>apply ultrasonic energy to generate deep<br>heat within body tissues for the<br>treatment of selected medical conditions<br>such as the relief of pain, the relief of<br>muscle spasms, the treatment of joint<br>contractures, and the local increase in<br>circulation. | Identical |
Table 1. Comparison between the subject device and the predicate device
{5}------------------------------------------------
| Bursitis with slight calcification, Myositis, Soft tissue injuries, and Shortened tendons due to past injuries and scar tissues. 4. Relief of pain, muscle spasms, and joint contractures resulting from: Capsular tightness, and Capsular scarring. 5. Localized increase in blood flow. | Bursitis with slight calcification, Myositis, Soft tissue injuries, and Shortened tendons due to past injuries and scar tissues. 4. Relief of pain, muscle spasms, and joint contractures resulting from: Capsular tightness, and Capsular scarring. 5. Localized increase in blood flow. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{6}------------------------------------------------
| | 6. Increased range of motion of contracted joint using heat and stretch techniques. | 6. Increased range of motion of contracted joint using heat and stretch techniques. | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Users | Health Care Professionals and lay users (under prescription) | Not Publicly Available | Health Care Professionals and lay users (under prescription) | Identical | |
| Prescription or OTC | Prescription | Not Publicly Available | Prescription | Identical | |
| Power Source(s) | 120/240 Vac with 5Vdc Input and rechargeable 3.7Vdc battery | Not Publicly Available | 120/240 VAC with 5V DC Input Power Jack and Lithium Battery Powered | Identical | |
| Number of Output | 1 | Not Publicly Available | 1-2 | Identical | |
| Software/Firmware/Microprocessor Control? | Yes | Not Publicly Available | Yes | Identical | |
| Automatic No-Load Trip? | Yes | Not Publicly Available | Not Publicly Available | Identical | |
| Automatic Shut Off? | Yes | Not Publicly Available | Not Publicly Available | Identical | |
| User Override Control? | Yes | Not Publicly Available | Not Publicly Available | Identical | |
| Indicator Display: | On/Off Status? | Yes | Not Publicly Available | Not Publicly Available | Identical |
| | Low Battery? | Yes | Not Publicly Available | Not Publicly Available | Identical |
| | Voltage / Current Level? | Yes | Not Publicly Available | Not Publicly Available | Identical |
| | Timer Range (minutes) | 0-20 | Not Publicly Available | 0-240 | Identical or similar. This does not affect the safety or effectiveness. |
| | Compliance with Voluntary Standards? | Yes | Not Publicly Available | Yes | Identical |
| | Biocompatibility? | Yes | Not Publicly Available | Yes | Identical |
| Sterility | | Non-Sterile | Not Publicly Available | Non-sterile | Identical |
| Housing construction material | | ABS | Not Publicly Available | ABS | Identical |
| Console/Generator Dimensions (mm) [L x W x H] | | 140×56×24 | Not Publicly Available | 61×70.9×18.8 | Identical or similar. This does not affect the safety or effectiveness. |
| Console/Generator Weight (kg) | | 0.13 | Not Publicly Available | 0.01 | Identical or similar. This does not affect the safety or effectiveness. |
| Treatment head dimensions (mm) [L x W x H] | | 30 x 30 x 12 | Not Publicly Available | 38.1 x 33 x 11.4 | Identical or similar. This does not affect the safety or effectiveness. |
| Treatment Head Weight (kg) | | 0.031 | Not Publicly Available | 0.10 | Identical or similar. This does not affect the safety or effectiveness. |
| Functions and design | | Ultrasound | Not Publicly Available | Ultrasound | Identical |
| Frequency (MHz) | | 1.5 | Not Publicly Available | 3 | Identical or small. The frequency of the subject device is identical to that of the primary predicate device, and smaller than that of the predicate device. Therefore, the frequency does not affect the safety or effectiveness. |
| | Leakage current | <0.3mA | Not Publicly Available | 0.3mA | Identical |
| | Crystal material | PZT | Not Publicly Available | Lead Zirconate-Titanate | Identical |
| | Technology of Ultrasound generation | Piezoelectric | Not Publicly Available | Piezoelectric | Identical |
| | Output Mode | Continuous Wave - 100% duty cycle | Not Publicly Available | Continuous Wave - 100% duty cycle | Identical |
| | | | | | |
| Ratio of temporal<br>maximum output<br>power to the output<br>power | 1:1 | Not Publicly Available | Not Publicly Available | Identical | |
| Deviation | $\pm$ 20% | Not Publicly Available | $\pm$ 20% | Identical | |
| | | | | | |
| Maximum Value of the<br>Output Power (W<br>$\pm$ 20%) | 0.60 | Not Publicly Available | Single Applicator: 0.65W $\pm$ 20%<br>Dual Applicator:<br>1.3W $\pm$ 20% | Identical or similar. Therefore, the<br>maximum power does not affect the<br>safety or effectiveness. | |
| Temporal average power<br>(W $\pm$ 20%) | 0.60 | Not Publicly Available | Not Publicly Available | Identical or similar. Therefore, the<br>temporal average power does not affect<br>the safety or effectiveness. | |
| Maximum value of the<br>effective intensity (W/cm²<br>$\pm$<br>20%) | 0.16 | Not Publicly Available | 0.264 | Identical or similar. Therefore, the<br>maximum effective intensity does not<br>affect the safety or effectiveness. | |
| Beam Maximum<br>Intensity and Accuracy<br>(W/cm² $\pm$ 20%) | 0.16 | Not Publicly Available | 0.132 | Identical or similar. Therefore, the<br>bean maximum intensity does not affect<br>the safety or effectiveness. | |
| Duty factor | 100% | Not Publicly Available | 100% | Identical | |
| Beam Type | Collimated | Not Publicly Available | Divergent | Identical or similar. This does not affect<br>the safety or effectiveness. | |
| Applicator size | Area: 3.9 or 5 cm² | Not Publicly Available | Up to two circular Applicators One<br>Applicator: 5 cm2<br>Two Applicators : 10 cm2 | Identical | |
| Maximum patient contact<br>surface temperature of the<br>treatment head under<br>simulated or actual use<br>conditions for all<br>operating conditions<br>(continually operated for<br>maximum<br>treatment time) (°C) | Meets IEC 60601-2-5,<br>section 201.11 Protection against<br>excessive temperature and other<br>HAZARDS. | Not Publicly Available | 44 °C | Identical | |
| Peak Temperature Rise<br>vs. Time and Tissue<br>Depth to Maximum<br>Treatment Time (for fixed<br>Treatment Head<br>Placement) (°C) | 7.6°C at 1 cm<br>3.3°C at 2 cm<br>1.4°C at 3 cm<br>Max treatment time: 20 min | Not Publicly Available | 8°C at 1 cm 6°C at 3 cm 3°C at 5 cm<br>Max treatment time: 4 hours | Identical or similar. This value of the<br>subject device is identical to that of the<br>primary predicate device, and smaller<br>than that of the predicate device.<br>Therefore, it does not affect<br>the safety or effectiveness. | |
| Beam Non-Uniformity<br>Ratio | 4 $\pm$ 20% | Not Publicly Available | <5 $\pm$ 20% | Identical or similar. Therefore, the beam<br>non-uniformity ratio does not<br>affect the safety or effectiveness. | |
| Effective Radiating Area<br>(cm2) | 3.9 or 5 $\pm$ 20% | Not Publicly Available | One: 6 cm2 Two:12 cm2<br>$\pm$ 20% | Identical or smaller. Therefore, the<br>effective radiating area does not<br>affect the safety or effectiveness. | |
| Electrical Safety<br>Standards Compliance | IEC 60601-1<br>IEC 60601-2-5<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-1-6<br>Wireless Coexistence Cybersecurity | Not Publicly Available | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | Identical or similar with IEC 60601- 1-<br>11 and IEC 60601-1-6 added…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.