K222066 · Waldemar Link GmbH & Co. KG · LPH · Apr 14, 2023 · Orthopedic
Device Facts
Record ID
K222066
Device Name
LINK MobileLink Acetabular Cup System
Applicant
Waldemar Link GmbH & Co. KG
Product Code
LPH · Orthopedic
Decision Date
Apr 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
General indications: The MobileLink Acetabular Cup System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. Indications: 1) Primary and secondary osteoarthritis 2) Rheumatoid arthritis 3) Correction of functional deformities 4) Avascular necrosis 5) Femoral neck fractures 6) Revision after implant loosening dependent on bone mass and quality The MobileLink Dual Mobility Insert is additionally indicated for: 7) Dislocation risks The MobileLink Acetabular Shells are intended for cementless fixation.
Device Story
The LINK MobileLink Acetabular Cup System is a hip joint prosthesis component used in total hip replacement surgery. This submission introduces additively manufactured TrabecuLink Acetabular Shells. These shells are implanted into the acetabulum to provide a cementless, porous-coated surface for biologic fixation. The device is used by orthopedic surgeons in a clinical/OR setting. The shells feature a 3D porous structure integrated during the additive manufacturing process. They are designed to mate with previously cleared UHMWPE liners, Dual Mobility CoCr inserts, shell/insert adapters, and bone screws. The surgeon selects the appropriate shell size and configuration (cluster-hole or multi-hole) based on patient anatomy and bone quality. The device replaces the damaged hip joint surface, aiming to restore mobility and function while reducing pain associated with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through non-clinical performance testing, including acetabular deformation testing, fatigue testing, assessment of modular connection disassembly (push-out, lever-out, torque-out), range of motion assessment, characterization of the porous surface, biocompatibility evaluation, and particle analysis.
Technological Characteristics
Materials: Ti6Al4V alloy per ISO 5832-3. Construction: Additively manufactured with integrated 3D porous surface (TrabecuLink). Fixation: Cementless. Design: Cluster-hole and multi-hole patterns. Compatibility: Mates with previously cleared MobileLink UHMWPE liners, Dual Mobility CoCr inserts, adapters, and bone screws.
Indications for Use
Indicated for patients with mobility-limiting diseases, fractures, or defects (e.g., osteoarthritis, rheumatoid arthritis, functional deformities, avascular necrosis, femoral neck fractures, revision after implant loosening) requiring hip joint replacement. Dual Mobility Insert indicated for dislocation risks.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
April 14, 2023
Waldemar Link GmbH & Co KG % Terry Powell Regulatory Affairs LinkBio Corp. 59 King Street Dover, New Jersey 07801
Re: K222066
Trade/Device Name: LINK MobileLink Acetabular Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, OQG Dated: March 20, 2023 Received: March 21, 2023
Dear Terry Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Limin Sun -S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known) K222066
Device Name
LINK MobileLink Acetabular Cup System
#### Indications for Use (Describe)
General indications:
Mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. Indications:
1) Primary and secondary osteoarthritis
- 2) Rheumatoid arthritis
- 3) Correction of functional deformities
- 4) Avascular necrosis
- 5) Femoral neck fractures
- 6) Revision after implant loosening dependent on bone mass and quality
- The MobileLink Dual Mobility Insert is additionally indicated
- for: 7) Dislocation risks
The MobileLink Acetabular Shells are intended for cementless fixation.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
| 510(k)<br>Submitter: | Waldemar Link GmbH & Co. KG<br>Barkhausenweg 10<br>22339 Hamburg, Germany<br>Phone: +49-40-539950<br>Facility Registration: 3003386935 |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Waldemar Link GmbH & Co. KG<br>Stefanie Fuchs (Regulatory Affairs)<br>Oststraße 4-10<br>22844 Norderstedt, Germany<br>Phone: +49-40 53995-530<br>Fax: +49-40 53995-174<br>E-Mail: st.fuchs@linkhh.de |
| Alternate<br>Contact: | Link Bio Corp.<br>Terry Sheridan Powell (Regulatory Affairs)<br>69 King Street<br>Dover, NJ 07801<br>e-mail: t.powell@linkbio.com |
| Date Prepared: | April 14, 2023 |
| Trade Name: | LINK MobileLink Acetabular Cup System |
| Common Name: | Total hip replacement system |
| Classification<br>Name: | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented<br>prosthesis. 21 CFR §888.3358, product code LPH, OQG |
| Classification<br>and Panel: | Class II, Orthopedic / 87 |
## Predicate
Devices:
| Subect Device<br>Components | Predicate Devices | | 510(k)<br>Number |
|-----------------------------|-------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| TrabecuLink<br>Shells | 1. | MobileLink Acetabular Cup System<br>Pro Codes: LPH, LZO, MEH, OQG<br>Waldemar Link GmbH & Co KG<br>[Primary Predicate] | K182321,<br>K192559,<br>K200607 |
| | 2. | G7 OsseoTi Acetabular Shell<br>Pro Codes: LPH, JDI, KWZ, LZO, OQG, OQH, OQI, PBI<br>Biomet, Inc. (Zimmer Biomet) | K140669 |
Reason for Submission
New Device System Components – TrabecuLink Shells
Device This 510k adds MobileLink TrabecuLink Acetabular Shells to the MobileLink Description: Acetabular Cup System previously cleared in K182321, K192559, K200607.
> These shells are additively manufactured from Ti6Al4V alloy per ISO 5832-3 and feature a porous structure (TrabecuLink) on the bone interfacing surface for biologic fixation. The shells come in cluster-hole and in multi-hole patterns.
{4}------------------------------------------------
mate with the same UHMWPE liners (in conventional PE and highly-crosslinked vitamin E PE) as previously cleared in K182321, and with the same Dual Mobility Inserts and Shell/Insert Adapters previously cleared in K200607. They can be used with the same bone screws cleared in K192559. Intended Use: General indications: The MobileLink Acetabular Cup System is indicated for patients with mobilitylimiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. Indications: 1) Primary and secondary osteoarthritis Rheumatoid arthritis 2) 3) Correction of functional deformities 4) Avascular necrosis Femoral neck fractures 5) 6) Revision after implant loosening dependent on bone mass and quality The MobileLink Dual Mobility Insert is additionally indicated for: 7) Dislocation risks The MobileLink Acetabular Shells are intended for cementless fixation. Comparison to The subject MobileLink TrabecuLink Acetabular Shells differ from the predicate Predicate Device: MobileLink Acetabular Shells (K182321) in that the subject shells are additive manufactured and feature a 3-D porous surface (TrabecuLink) that is integrated with the shell substrate during the additive manufacturing process. This feature (additively manufactured shell with integrated porous surface) is equivalent to the predicate Zimmer/Biomet G7 Acetabular Cup System with OsseoTi Shells (K140669). The subject MobileLink TrabecuLink Acetabular Shells are similar to the predicate MobileLink Acetabular Shells (K182321) in that they are made of Ti6Al4V alloy, and come in the same sizes and design variants (cluster hole or multi-hole), have the same interior geometry and mate with all of the same previously cleared MobileLink polyethylene inserts, Dual Mobility CoCr Inserts, Adapters (face changers) and bone screws. Performance Non-clinical performance testing and analysis were provided, including: Testing: -Acetabular deformation testing -Acetabular fatigue testing -Assessment of disassembly (Push-out, lever-out, torque out for all acetabular modular connections in MobileLink system) Assessment of range of motion --Characterization of the TrabecuLink porous surface -Biocompatibility evaluation -Particle analysis The results of non-clinical performance testing demonstrate that the device is as
safe and effective as the predicate device, and therefore Substantially Equivalent.
{5}------------------------------------------------
- Clinical Testing: Clinical performance testing was not required to demonstrate the substantial equivalence of this device.
- Conclusion: The subject MobileLink TrabecuLink Acetabular Shells are substantially equivalent to the predicate devices identified in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.