K222052 · CIVCO Medical Instruments Co., Inc. · ITX · Nov 22, 2022 · Radiology
Device Facts
Record ID
K222052
Device Name
VitroPRO
Applicant
CIVCO Medical Instruments Co., Inc.
Product Code
ITX · Radiology
Decision Date
Nov 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
This device provides physicians with a tool for performing needle/instrument guided procedures with the use of diagnostic ultrasound endocavity transducers. - Transvaginal - Diagnostic ultrasound needle / instrument guided procedures such as tissue biopsy, fluid aspiration, catheter placement, treatment, and oocyte retrieval. - Transrectal Diagnostic ultrasound needle / instrument guided procedures such as tissue biopsy, fluid aspiration, catheter placement, treatment.
Device Story
Disposable endocavity needle guide; attaches to ultrasound transducer to direct needles/instruments along fixed path. Used in transvaginal/transrectal procedures (biopsy, fluid aspiration, catheter placement, oocyte retrieval). Operated by physicians in clinical settings. Device ensures needle path accuracy and stability during ultrasound-guided interventions. Benefits include precise instrument placement and reduced risk of procedure-related complications. No electronic components; purely mechanical guidance tool.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10), cover/probe damage testing, retention force testing (min 8N), needle drag force testing (<1.5N), needle path verification, and simulated usability testing.
Technological Characteristics
Materials: ABS thermoplastic, stainless steel cannula. Principle: Mechanical needle/instrument guidance for ultrasound transducers. Sterilization: Ethylene Oxide. Shelf-life: 3 years. No electronic or software components.
Indications for Use
Indicated for physicians performing needle/instrument guided procedures using diagnostic ultrasound endocavity transducers, including transvaginal (biopsy, fluid aspiration, catheter placement, treatment, oocyte retrieval) and transrectal (biopsy, fluid aspiration, catheter placement, treatment) procedures.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
November 22, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
CIVCO Medical Instruments Co., Inc. % Jim Leong Regulatory Affairs Manager 102 First Street South KALONA IA 52247
Re: K222052
Trade/Device Name: VitroPRO Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: ITX, MOE Dated: October 14, 2022 Received: October 24, 2022
Dear Jim Leong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Yanna S. Kang -S
Yanna Kang, Ph. D. Assistant Director Mammography and Ultrasound Team DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K222052 Device Name VitroPRO
#### Indications for Use (Describe)
This device provides physicians with a tool for performing needle/instrument guided procedures with the use of diagnostic ultrasound endocavity transducers.
- · Transvaginal Diagnostic ultrasound needle / instrument guided procedures such as tissue biopsy fluid aspiration, catheter placement, treatment, and oocyte retrieval.
- · Transrectal Diagnostic ultrasound needle / instrument guided procedures such as tissue biopsy, fluid aspiration, catheter placement, treatment.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------|
|---------------------------------------------------------------------------|--------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being prepared in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) numberis: K222052 .
- 1. Submitter's Identifications:
| Establishment: | CIVCO Medical Instruments Co., Inc. |
|--------------------------------------------|--------------------------------------------|
| Address: | 102 First Street South<br>Kalona, IA 52247 |
| Registration Number: | 1937223 |
| Operations Manufacturer<br>Owner/Operator: | CIVCO Medical Instruments Co., Inc. |
| Owner/Operator Number: | 1937223 |
| Contact Person: | Jim Leong |
| Phone: | 319-248-6502 |
| e-mail: | James.Leong@civco.com |
# 2. Date 510(k) Summary Prepared: July 5, 2022
# 3. Name of the Subject Device and Classification Information:
| Trade/Device Name | VitroPRO / Disposable Endocavity Needle Guide |
|---------------------|-------------------------------------------------------------------------|
| Regulation Number | 21 CFR 892.1570<br>21 CFR 884.6100 |
| Classification Name | Diagnostic ultrasonic transducer Assisted reproduction needles Class II |
| Regulatory Class | ITX & MQE |
| Product Code | |
# 4. Information for the Predicate Device:
| Tradename/Device Name | Disposable Endocavity Needle / Biopsy Guide |
|-----------------------|---------------------------------------------|
| Manufacturer | CIVCO Medical Instruments Co., Inc. |
| 510(k) Number | K970514 |
| Regulation Number | 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | II |
| Product Code | ITX |
{4}------------------------------------------------
# 5. Information for Reference Device:
| Tradename/Device Name | Embryon® Ultrasound Needle Guides |
|-----------------------|-----------------------------------|
| Manufacturer | Rocket Medical PLC |
| 510(k) Number | K032015 |
| Regulation Number | 884.6100 |
| Regulation Number | 892.1560 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | II |
| Product Code | MQE & IYO |
### 6. Device Description:
The disposable endocavity needle guides are devices used to direct needles or instruments along a fixed path to a target location with an ultrasound traducer. They are provided in a variety of sizes to fit different equipment and situations. They can be utilized for various types of procedures from tissue biopsy and fluid aspirations to instrument placements and oocyte harvesting.
# 7. Intended Use / Indications for Use:
This device provides physicians with a tool for performing needle/instrument guided procedures with the use of diagnostic ultrasound endocavity transducers.
- . Transvaginal - Diagnostic ultrasound needle / instrument guided procedures such as tissue biopsy, fluid aspiration, catheter placement, treatment, and oocyte retrieval.
- Transrectal Diagnostic ultrasound needle / instrument guided procedures such as tissue biopsy, ● fluid aspiration, catheter placement, treatment.
| Item | Subject Device<br>CIVCO Disposable<br>Endocavity Needle Guide<br>K21XXXX | Predicate Device<br>CIVCO Disposable<br>Endocavity Ultrasound<br>Needle / Biopsy Guide<br>K970514 |
|--------------------------|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Material | ABS Thermoplastic | ABS Thermoplastic |
| Cannula: Stainless Steel | Cannula: Stainless<br>Steel | |
# 8. Comparison to Legally Marketed Device
{5}------------------------------------------------
| Item | Subject Device<br>CIVCO Disposable<br>Endocavity Needle Guide<br>K21XXXX | Predicate Device<br>CIVCO Disposable<br>Endocavity Ultrasound<br>Needle / Biopsy Guide<br>K970514 |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | All materials have met the<br>requirements of ISO 10993-1<br>for biocompatibility.<br>Meets ISO 10993-1<br>biocompatibility<br>requirements for limited<br>contact duration:<br>• surface devices of<br>breached or<br>compromised surface<br>• Externally<br>communicating tissue<br>bone/dentin<br>Demonstrated to be non-toxic,<br>non-sensitizing, non-irritating,<br>(not labeled non-pyrogenic) | All materials have met the<br>requirements of ISO 10993-1<br>for biocompatibility.<br>Meets ISO 10993-1<br>biocompatibility<br>requirements for limited<br>contact duration:<br>• surface devices of<br>breached or<br>compromised surface<br>• External<br>communicating<br>indirect blood<br>path/tissue contact<br>Demonstrated to be non-toxic,<br>non-sensitizing, non-irritating,<br>non-hemolytic, and non-<br>pyrogenic |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Shelf-life | 3 years | 3 years |
| Accessory | None provided. Intended<br>user to provide cover and<br>guides which are IVF use<br>cleared. | Ultrasound gel packet and<br>covers. |
- 9. Comparison of Indications to the Legally Marketed Device:
The proposed endocavity needle guides have the same intended use. The language used has been updated for clarity and to add an additional indication for use for oocyte retrieval. The differences in the indications for use statement would not affect the safety or effectiveness of the device because the safety and efficacy related to the indicated procedures is dictated by the use of the underlying ultrasound equipment. Any questions related to safety and effectiveness of the endocavity guides have been addressed using the same testing performed by the legally marketed device, including updated biocompatibility evaluations.
{6}------------------------------------------------
## 10. Summary of Non-Clinical Tests Performed:
- . Biocompatibility:
The disposable endocavity needle guides devices met ISO 10993-1 biocompatibility requirements for limited contact duration for surface devices of breached or compromised surface and external communicating tissue/bone/dentin contact:
- Cytotoxicity ISO 10993-5 —
- -Sensitization - ISO 10993-10
- Irritation ISO 10993-10 -
- Cover breach and probe damage testing ●
Water leak testing was performed to demonstrate material attachment of needle guide to over a cover did not cause damage to cover or probe.
- Retention and movement testing Force testing was performed on needle guide attachment to ensure a minimum force of 8N would not cause the guide to dislodge.
- Needle drag testing
Force testing was performed by passing a cannula through the needle guide to ensure binding would not occur and force was less than a 1.5N threshold.
- Needle path verification testing . Needle guides were tested on test fixtures to ensure needle path falls within the design tolerances specified for the design.
- Simulated Usability Testing Simulated use evaluations were performed by customers to ensure the design of the needle guide conforms to the user needs and intended use.
# 11. Clinical Test Performed:
Clinical tests were not required to demonstrate substantial equivalence.
#### 12. Conclusions:
The disposable endocavity needle guide devices have intended use and its technological characteristics do not raise any different questions of safety or effectiveness, as compared to the legally marketed device. The addition of oocyte harvesting to the indications for use have been tested to show the devices are safe for such usage. Therefore, the endocavity needle guides are substantially equivalent to the legally marketed disposable endocavity needle guide devices marketed by CIVCO.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.