CentriMag™ Blood Pump for use with CentriMag™ Acute Circulatory Support System
Applicant
Abbott (Formerly Thoratec Corporation)
Product Code
QNR · Cardiovascular
Decision Date
Dec 8, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4100
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K222038 · Dec 8, 2022
CentriMag™ Blood Pump for use with CentriMag™ Acute Circulatory Support System
Abbott (Formerly Thoratec Corporation)
ELSO Registry
A propensity-matched analysis of ELSO Registry data was used to compare the safety and effectiveness of the CentriMag System against a comparator group regarding seven ECMO-related clinical complications and outcomes.
Other devices used in the same manner (ECMO circuit)
Seven ECMO-related clinical complications and outcomes
Indications for Use
The CentriMag™ Blood Pump for use with CentriMag™ Acute Circulatory Support System (Motor, Monitor, Console, and Flow Probes) is indicated for controlling blood flow as part of an extracorporeal membrane oxygenation (ECMO) circuit. ECMO is intended to provide assisted extracorporeal circulation and physiologic gas exchange of the patients' blood for adult patients with acute respiratory failure and/or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
Device Story
CentriMag System provides temporary mechanical circulatory support for adult patients in acute hemodynamic compromise. System components include primary console, motor, blood pump, flow probe, and optional monitor. Centrifugal flow pump utilizes magnetically levitated (Full MagLev™) impeller; eliminates need for mechanical seals, bearings, or shafts; reduces friction, heat, and wear. Motor drives impeller via electromagnetic levitation. Console controls pump speed and resultant blood flow; monitors system operation. Used in clinical settings by healthcare providers to manage ECMO circuits. Provides assisted extracorporeal circulation and physiologic gas exchange. Benefits include temporary support for patients with failing treatment options and imminent risk of death.
Clinical Evidence
Included propensity-matched analysis from ELSO Registry comparing CentriMag System to a comparator group. Analysis focused on seven ECMO-related clinical complications and outcomes. Results showed no statistically significant or clinically meaningful differences between the CentriMag System and the comparator, supporting safety and effectiveness for long-term assisted extracorporeal circulation.
Technological Characteristics
Centrifugal flow pump with magnetically levitated impeller (Full MagLev™). Components: console, motor, pump, flow probe, monitor. Biocompatibility per ISO 10993-1. Sterilization via ethylene oxide (ANSI/AAMI/ISO 11135). Electrical safety/EMC testing performed. System is designed for extracorporeal circuit integration.
Indications for Use
Indicated for adult patients with acute respiratory failure and/or acute cardiopulmonary failure requiring extracorporeal membrane oxygenation (ECMO) support when other treatments have failed and clinical deterioration is expected or death is imminent.
Regulatory Classification
Identification
An extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary support (>6 hours) is a system of devices and accessories that provides assisted extracorporeal circulation and physiologic gas exchange of the patient's blood in patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, an oxygenator, blood pump, heat exchanger, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(1) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible;
(2) The devices and accessories in the circuit must be demonstrated to be biocompatible;
(3) Sterility and shelf-life testing must demonstrate the sterility of any patient-contacting devices and accessories in the circuit and the shelf life of these devices and accessories;
(4) Non-clinical performance evaluation of the devices and accessories in the circuit must demonstrate substantial equivalence of the performance characteristics on the bench, mechanical integrity, electromagnetic compatibility (where applicable), software, durability, and reliability;
(5) In vivo evaluation of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the clinical evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness if a specific indication (patient population and/or condition) is identified; and
(6) Labeling must include a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the devices and accessories in the circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.
Predicate Devices
FDA final order for Extracorporeal Circuit and Accessories for Long-Term Respiratory/Cardiopulmonary Failure (21 CFR 870.4100, section (b) Class II (special controls))
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
December 8, 2022
Abbott (formerly Thoratec Corporation) Meera Mehta Project Manager, Regulatory Affairs 6035 Stoneridge Drive Pleasanton, California 94588
# Re: K222038
Trade/Device Name: CentriMag™ Blood Pump for use with CentriMag™ Acute Circulatory Support System Regulation Number: 21 CFR 870.4100 Regulation Name: Extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure Regulatory Class: Class II Product Code: QNR Dated: November 21, 2022 Received: November 21, 2022
Dear Meera Mehta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222038
### Device Name
CentriMag™ Blood Pump for use with CentriMag™ Acute Circulatory Support System
### Indications for Use (Describe)
The CentriMag™ Blood Pump for use with CentriMag™ Acute Circulatory Support System (Motor, Monitor, Console, and Flow Probes) is indicated for controlling blood flow as part of an extracorporeal membrane oxygenation (ECMO) circuit. ECMO is intended to provide assisted extracorporeal circulation and physiologic gas exchange of the patients' blood for adult patients with acute respiratory failure and/or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M0 2a2 2 0 0 1 2-2h12a2 2 0 0 1 2 2v12a2 2 0 0 1-2 2H2a2 2 0 0 1-2-2V2zm1.5 0a.5.5 0 0 0-.5.5v12a.5.5 0 0 0 .5.5h12a.5.5 0 0 0 .5-.5V2.5a.5.5 0 0 0-.5-.5H1.5z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Abbott logo, which is a stylized blue letter 'a' with a rounded design. Below the logo, the word "Abbott" is written in bold, black font. Underneath "Abbott", the text "CentriMag Pre-connected Pack" is written in a smaller, regular font.
# 6. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92.
#### SUMBITTER l.
II.
| Date Prepared: | July 8, 2022 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Submitter's Name & Address: | Abbott<br>6035 Stoneridge Drive<br>Pleasanton, CA 94588 |
| Establishment Registration No. | 2916596 |
| Contact Person: | Meera Mehta<br>6035 Stoneridge Dr,<br>Pleasanton, CA-94588<br>Project Manager, Regulatory Affairs<br>Ph: 925-315-6483 |
| DEVICE | |
| Trade or Proprietary Name: | CentriMag™ Blood Pump for use with<br>CentriMag™ Acute Circulatory Support<br>System |
| Common or Usual Name: | CentriMag System |
| Classification Name: | Extracorporeal circuit and accessories for<br>long-term respiratory/cardiopulmonary<br>failure |
| Regulation Number: | 21 CFR 870.4100 |
Class II (special controls)
QNR
Product Code:
Regulatory Class:
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Abbott logo, which consists of a stylized blue "A" shape above the word "Abbott" in bold, black font. Below the word "Abbott" is the text "CentriMag Pre-connected Pack" in a smaller, regular font. The logo is clean and modern, with a focus on the company name and a product line.
#### III. PREDICATE DEVICE
Predicate device: FDA final order for Extracorporeal Circuit and Accessories for Long-Term Respiratory/Cardiopulmonary Failure (21 CFR 870.4100, section (b) Class II (special controls)).
Secondary/Reference device: Medtronic BPX-80 Centrifugal Pump (K973011)
#### IV. DEVICE DESCRIPTION
The CentriMag™ Blood pump for use with CentriMag™ Acute Circulatory Support System (hereafter referred to as the CentriMag System) is designed to provide assisted extracorporeal circulation and physiologic gas exchange of patients' blood for adult patients with acute respiratory and/or acute cardiopulmonary failure.
The CentriMag System was designed to provide temporary mechanical circulatory support. The CentriMag System provides circulatory assistance for patients in acute hemodynamic compromise, a population whose treatment options are limited.
- CentriMag 2nd Generation Primary Console
- CentriMag Motor
- . CentriMag Blood Pump
- Flow Probe
- . Mag Monitor (optional)
The CentriMag System features a centrifygal flow pump with inflow and outflow ports that are at right angles to one another, and a magnetically levitated impeller (Full MagLev™ technology).
The CentriMag Motor is a reusable, non-sterile component of the CentriMag Acute Circulatory Support System. The CentriMag Motor holds the blood pump and drives the impeller inside the blood pump. When the pump is inserted into the motor and activated, the internal impeller is electromagnetically levitated and centered, eliminating the need for shafts, seals, and bearings in the pump. Utilizing magnetic levitation technology (Full MagLev™ technology) to suspend and spin the impeller eliminates bearing and seal friction, resulting in minimal heat generation and wear of the pump components. the console is used to control pump speed, the resultant blood flow, and monitor the operation of the system. A cable connects the console to the motor, allowing flexibility in the pump motor and pump positioning.
Note: The CentriMag Drainage (Venous) and Return (Arterial) Cannula Kits (K200306) are not part of this 510(k) premarket notification seeking clearance per 21 CFR 870.4100 Extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure, section (b) Class II (special controls).
#### V. INDICATION FOR USE
The CentriMag™ Blood Pump for use with CentriMag™ Acute Circulatory Support System (Motor, Monitor, Console, and Flow Probes) is indicated for controlling blood flow as part of an extracorporeal membrane oxygenation (ECMO) circuit. ECMO is intended to provide assisted extracorporeal circulation and physiologic gas exchange of the patients' blood for adult patients with acute respiratory failure and/or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Abbott, a healthcare company. The logo features a stylized letter "A" in blue, positioned above the company name "Abbott" in bold, black font. Below the company name is the text "CentriMag Pre-connected P" in a smaller font size.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.
Substantial equivalence of the CentriMag Blood Pump for use with CentriMag Acute Circulatory Support System was demonstrated by meeting the special controls in FDA Final Order, 21 CFR 870.4100 Extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure, section (b) Class II (special controls) and through bench testing, mechanical testing, electrical safety and electromagnetic compatibility testing, usability, and software testing.
- . Technological Characteristics: Geometry and design parameters of the subject device are consistent with devices intended for use in extracorporeal life support procedures. The subject device is designed to be compatible with other extracorporeal circuit devices and accessories.
- . Biocompatibility: The subject device (CentriMag blood pump) is demonstrated to be biocompatible in accordance with FDA biocompatibility guidance Use of International Standard ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
- . Sterility and Shelf-life: The CentriMag blood pump has been validated over the stated shelf life of the device to ensure a minimum sterility assurance level (SAL) of 10th as described in ANSI/AAMI/ISO 11135, Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices.
- . Non-clinical Performance: Substantial equivalence has been demonstrated by performance testing (bench), mechanical integrity, durability, and reliability.
- . In vivo Evaluation: In vivo evaluation demonstrates the subject device's performance over an intended duration of use.
- . Labeling: The labeling summarizes non-clinical performance of the CentriMag System and provides adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.
#### VII. PERFORMANCE DATA
The CentriMag Blood Pump for use with CentriMag Acute Circulatory Support System performance characteristics were demonstrated through bench testing, mechanical testing, electrical safety and electromagnetic compatibility testing, usability, and software testing to support the determination of substantial equivalence.
| Component | Test |
|----------------------------------------------|------------------------------|
| System (console,<br>monitor, pump,<br>motor) | 30-day reliability |
| | In Vivo Animal Testing |
| | Water Ingress Resistance |
| | Electrical Safety & EMC |
| Motor | Thermal Operating conditions |
| | Fluid Intrusion |
| Console &<br>Monitor | Battery life |
| | Battery run time |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Abbott CentriMag Pre-connected Pack. The logo features a stylized letter "a" in blue, followed by the word "Abbott" in bold, black font. Below "Abbott" is the text "CentriMag Pre-connected Pack" in a smaller, regular font.
| Blood Pump | Rough Handling (Shock/Drop/Vibration) |
|------------|------------------------------------------|
| | Environmental Test |
| | Software validation |
| | HQ performance |
| | In Vitro Hemolysis, Platelet count, WBC, |
| | RBC, hematocrit and thrombus |
| | Air handling |
| | Priming volume |
| | Leak and burst pressure testing |
| | Sterilization |
| | Transportation distribution / simulation |
| | Shelf life |
| | Biocompatibility |
#### VIII. CLINICAL SUMMARY
This submission included a pre-specified statistical analysis plan from the ELSO Registry. The propensity matched analysis demonstrates the safety and effectiveness of the CentriMag Blood Pump for use with CentriMag Acute Circulatory Support System as part of an ECMO circuit to provide longterm assisted extracorporeal circulation and physiologic gas exchange of the blood for adult patients with acute respiratory failure and/or acute cardiopulmonary failure. This pre-specified analysis focused on seven ECMO-related clinical complications and outcomes.
The results of the propensity analysis did not identify any statistically significant or clinically meaningful differences in the seven ECMO-related clinical outcomes between the CentriMag System and the comparator group. This analysis demonstrated the CentriMag Blood pump for use with CentriMag System as part of an ECMO circuit is substantially equivalent to other devices used in the same manner.
#### IX. CONCLUSION
The CentriMag Blood Pump along with CentriMag Acute Circulatory Support System met the special controls in 21 CFR 870.4100 Extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure, section (b) Class II (special controls) and are substantially equivalent to the predicate device. The information and data in this 510(k) premarket notification demonstrate that the CentriMag Blood Pump for use with CentriMag Acute Circulatory Support System are substantially equivalent to the predicate to provide assisted extracorporeal circulation and physiologic gas exchange of the patients' blood for adult patients with acute respiratory failure and/or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.