Spinal Inner Fixation System

K222031 · Changzhou Geasure Medical Apparatus and Instruments Co., Ltd. · NKB · Oct 28, 2022 · Orthopedic

Device Facts

Record IDK222031
Device NameSpinal Inner Fixation System
ApplicantChangzhou Geasure Medical Apparatus and Instruments Co., Ltd.
Product CodeNKB · Orthopedic
Decision DateOct 28, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Spinal Inner Fixation System is intended for posterior, non-cervical, pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization of the spine as an adjunct to fusion with bone graft.

Device Story

Spinal Inner Fixation System; surgical fixation device for spine. Components: pedicle screws, set screws, rods, cross-connecting devices. Used for stabilization/fusion of two or more spinal segments; adjunct to bone graft fusion. Implants are single-use; surgical instruments are reusable. Used in OR by surgeons. Provides mechanical stabilization of spine; intended for removal after stabilization function complete. Benefits patient by providing structural support during spinal fusion process.

Clinical Evidence

Bench testing only. Performance evaluated per ASTM F1717-18 standards, including static compression bending, dynamic compression bending, and static torsion. No clinical data provided.

Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V) ELI. Surface treatment: Anodized. Provided non-sterile; requires manual cleaning and steam sterilization. Mechanical fixation system consisting of pedicle screws, set screws, rods, and cross-connectors. Standards: ASTM F1717-18.

Indications for Use

Indicated for skeletally mature patients requiring posterior, non-cervical, pedicle fixation for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudarthrosis, or failed previous fusion, as an adjunct to fusion with bone graft.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ October 28, 2022 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Changzhou Geasure Medical Apparatus and Instruments Co., Ltd % Ms. Xiaoqing Xue Registration Engineer Sinow Medical AS Vestre Fantoftasen 44, 5072, Bergen, Norway ### Re: K222031 Trade/Device Name: Spinal Inner Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: October 6, 2022 Received: October 7, 2022 Dear Ms. Xue: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K222031 Device Name Spinal Inner Fixation System #### Indications for Use (Describe) Spinal Inner Fixation System is intended for posterior, non-cervical, pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization of the spine as an adjunct to fusion with bone graft. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(K) Summary # K222031 | Preparation Date: | Oct. 06, 2022 | | |-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | Changzhou Geasure Medical Apparatus and Instruments Co., Ltd<br>No. 12, Jinfeng Road, West Taihu Science and Technology<br>Industrial Park, Changzhou, Jiangsu, P.R. China<br>Jing Huang, Management Representative<br>Email: huangjing@geasure.com<br>Phone:+86 13656146897 | | | Designated<br>Submission<br>Correspondent | Company: Sinow Medical AS | | | | Address: Vestre Fantoftåsen 44, 5072, Bergen, Norway | | | | Contact Person: Xiaoqing Xue | | | | Telephone: +86 15161196032<br>Email: xue@bergemed.com | | | Subject Device | Trade name | Spinal Inner Fixation System | | | Common name | Pedicle Screw Spinal System | | | Regulatory Class | II | | | Regulation<br>Number | 21 CFR 888.3070 | | | Product Codes | NKB | | | Classification<br>Panel | Orthopedic | | | Classification<br>Name | Thoracolumbosacral Pedicle Screw system | | | Manufacturer | Changzhou Dingjian Medical Appliance<br>Co., Ltd. | | | Trade name | Spinal Inner Fixation System | | | 510(K) Number | K143013 | | | Regulatory Class | II | | Primary Predicate<br>Device | Regulation<br>Number | 21 CFR 888.3070,888.3050 | | | Product Codes | MNH, MNI, KWP | | | Classification<br>Panel | Orthopedic | | | Classification<br>Name | Thoracolumbosacral Pedicle Screw<br>System, Spinal Interlaminal Fixation<br>Orthosis | | | Intended Use /<br>Indications for Use | Spinal Inner Fixation System is intended for posterior,<br>non-cervical, pedicle fixation for the following indications:<br>degenerative disc disease (defined as back pain of discogenic | | | origin with degeneration of the disc confirmed by history and | | | | radiographic studies); spondylolisthesis; trauma (i.e., fracture or | | | | dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis | | | | and/or lordosis); tumor; pseudarthrosis; and/or failed previous | | | | fusion. The device is to be used in skeletally mature patients, and for | | | | | | | | stabilization and immobilization of the spine as an adjunct to fusion | | | | with bone graft. | | | Device Description | The Spinal Inner Fixation System is a medical device system for | | | | surgical fixation of the spine. It is made up of pedicle screws, set | | | | screw, rods and cross connecting devices. The implanted parts | | | | confer stabilization and fusion of two or more spinal segments | | | | and are to be removed once their stabilizing function is no longer | | | | required. The implants are single-use. The surgical instruments | | | | may be reprocessed and re-used. | | | | The Spinal Inner Fixation System includes model GB1Z-I and | | | | GB1Z-II according to the rod diameter, and each model | | | | comprises of different length rods, various length and diameter | | | | screws, set screws and accessories. | | | | Both models are designed for internal posterior thoracolumbar | | | | fixation of the spine. Patient diagnosis and individual conditions | | | | should be taken into consideration when selecting the surgical | | | | option. | | | | Surgical instruments are provided with the device system. | | | Materials | Titanium alloy (Ti-6Al-4V) ELI | | | | The devices are colored using an anodizing manufacturing | | | Coatings/Colorants | process. No coatings or colorants are added to the device with | | | | this process. | | | Sterilization<br>Method | The devices are provided non-sterile; validated manual cleaning | | | | and steam sterilization instructions are provided for the end user | | | | before implantation. | | | Performance -<br>Bench | The ASTM F1717-18 standards were utilized to complete the<br>following bench performance testing on the subject and predicate | | | | device: | | | | • static compression bending | | | | • dynamic compression bending, | | | | • static torsion | | | | The Spinal Inner Fixation System is substantially equivalent to | | | | the predicate device when evaluating intended use and | | | Substantial<br>Equivalence | technological characteristics. | | | | • The subject device has the identical intended use as the | | | | predicate device. | | | | The subject device and predicate devices are substantially<br>equivalent with only minor differences regarding:<br>Mechanism of action Rods Set screw Cross connector Bench performance: static compression, dynamic<br>compression and static torsion These differences do not raise new questions of safety and<br>effectiveness. | | | Conclusion | Non-clinical data demonstrates the Spinal Inner Fixation System<br>device is substantially equivalent to the predicate device. | | {4}------------------------------------------------ {5}------------------------------------------------
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...