The IdentiTi Porous Ti Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the IdentiTi Porous Ti System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The IdentiTi Porous Ti Interbody System is intended for use on patients who have had at least six months of nonoperative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with IdentiTi LIF interbody spacers to provide integrated fixation. IdentiTi LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. IdentiTi ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.
Device Story
Thoracolumbar intervertebral body fusion systems; inserted via anterior or posterior surgical approaches. Implants feature internal graft apertures for bone graft placement; endplate teeth mitigate expulsion risk. Available in PEEK, tantalum, titanium alloy, or porous titanium. NanoTec variants feature 20-40 nm hydroxyapatite surface treatment. Used by surgeons in clinical settings to treat degenerative spinal conditions. Implants provide structural support to facilitate fusion; integrated fixation (AMP Anti-Migration Plate or anterior plates) used for specific lordotic angles (>20°). Benefits include stabilization of spinal segments and promotion of fusion in patients failing conservative therapy.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on non-clinical bench testing, including ASTM F2077 (axial compression, compression-shear, torsion), ASTM F2267 (subsidence), ASTM Draft F-04.25.02.02 (push-out), gravimetric analysis, and particulate analysis.
Technological Characteristics
Materials: PEEK Optima LT1 (ASTM F2026), Tantalum (ASTM F560), Ti-6Al-4V ELI (ASTM F136), CPTi Grade 2 (ASTM F67). Surface: Porous titanium microstructure or 20-40 nm hydroxyapatite (NanoTec) coating. Form factor: Various lengths, widths, heights, lordotic options. Fixation: Standalone or with supplemental fixation (AMP plate/anterior plate).
Indications for Use
Indicated for spinal fusion (T1-S1) in skeletally mature patients with symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, or thoracic disc herniation (with myelopathy/radiculopathy). Also indicated as adjunct for multilevel degenerative scoliosis and sagittal deformity. Requires 6 months non-operative treatment. Used with autograft/allograft bone and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
NuVasive Thoracolumbar Interbody Systems (K203201)
IdentiTi™ and Transcend™ Interbody Systems (K220782)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
October 7, 2022
Alphatec Spine Neha Mohindroo Regulatory Affairs Associate 1950 Camino Vide Roble Carlsbad, California 92008
# Re: K222028
Trade/Device Name: IdentiTi™ Porous Ti Interbody System, IdentiTi™ NanoTec™ Interbody System, Transcend™ PEEK Interbody System, Transcend™ NanoTecTM Interbody System, IdentiTiTM ALIF Standalone Interbody System, IdentiTiTM NanoTecTM ALIF Standalone Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX,OVD, PHM Dated: July 8, 2022 Received: July 11, 2022
# Dear Ms. Mohindroo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
# K222028
Device Name IdentiTi™ Porous Ti Interbody System
## Indications for Use (Describe)
The IdentiTi Porous Ti Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondy lolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the IdentiTi Porous Ti System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The IdentiTi Porous Ti Interbody System is intended for use on patients who have had at least six months of nonoperative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
AMP™ Anti-Migration Plate may be used with IdentiTi LIF interbody spacers to provide integrated fixation. IdentiTi LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. IdentiTi ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.
Type of Use (*Select one or both, as applicable*)
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
#### Device Name
IdentiTi™ NanoTec™ Interbody System
#### K222028
# Indications for Use (Describe)
The IdentiTi Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the IdentiTi NanoTec Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The IdentiTi NanoTec Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
AMP™ Anti-Migration Plate may be used with IdentiTi NanoTec LIF interbody spacers to provide integrated fixation. IdentiTi NanoTec LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
==============================================================================================================================================================================
{4}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
#### Device Name
IdentiTi™ ALIF Standalone Interbody System
K222028
# Indications for Use (Describe)
The IdentiTi ALIF Standalone Interbody System is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.
Additionally, the IdentiTi ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the IdentiTi ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The IdentiTi ALIF Standalone Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.
The IdentiTi ALIF Standalone Interbody System implants of ≤20° are a standalone system. The IdentiTi ALIF Standalone Interbody System implants of >20° must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------|---------------------------------------------|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | |
| | | |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
# 510(k) Number (if known)
#### Device Name
IdentiTiTM NanoTec™ ALIF Standalone Interbody System
#### K222028
## Indications for Use (Describe)
The IdentiTi ALIF Standalone Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.
Additionally, the IdentiTi NanoTec ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the IdentiTi Nanotec ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The IdentiTi NanoTec ALIF Standalone Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.
The IdentiTi NanoTec ALIF Standalone Interbody System implants of ≤20° are a standalone system. The IdentiTi NanoTec ALIF Standalone Interbody System implants of >20° must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{6}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K222028
Device Name Transcend™ PEEK Interbody System
## Indications for Use (Describe)
The Transcend PEEK Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondy lolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the Transcend PEEK Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Transcend PEEK Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
AMP™ Anti-Migration Plate may be used with Transcend LIF interbody spacers to provide integrated fixation. Transcend LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. Transcend ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{7}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
# 510(k) Number (if known) K222028
Device Name Transcend™ NanoTec™ Interbody System
## Indications for Use (Describe)
The Transcend PEEK Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the Transcend NanoTec Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Transcend NanoTec Interbody System is intended for use on patients who have had at least six months of nonoperative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
AMP™ Anti-Migration Plate may be used with Transcend NanoTec LIF interbody spacers to provide integrated fixation. Transcend NanoTec LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| <div><span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "atec". The "a" is green, and the rest of the letters are dark blue. There is a black line underneath the logo.
# 510(k) Summary K222028
| I. | SUBMITTER: | Alphatec Spine, Inc.<br>1950 Camino Vida Roble<br>Carlsbad, CA 92008<br>Phone: (760) 431-9286<br>Fax: (760) 431-0289 |
|----|------------------------|----------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Neha Mohindroo<br>Regulatory Affairs Associate<br>Contact Phone: (760) 356-6596 |
| | Date Summary Prepared: | September 16, 2022 |
#### II. DEVICE
| Name of Device: | IdentiTiTM Porous Ti Interbody System |
|-----------------|-------------------------------------------------------|
| | IdentiTiTM NanoTecTM Interbody System |
| | TranscendTM PEEK Interbody System |
| | TranscendTM NanoTecTM Interbody System |
| | IdentiTiTM ALIF Standalone Interbody System |
| | IdentiTiTM NanoTecTM ALIF Standalone Interbody System |
| Common or Usual Name:<br>Classification Name: | Intervertebral body fusion device<br>Intervertebral fusion device, lumbar<br>Intervertebral fusion device with integrated fixation, lumbar |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Class: | Class II |
| Product Code: | MAX, OVD, PHM |
#### LEGALLY MARKETED PREDICATE DEVICES III.
| 510(k) | Product Code | Trade Name | Manufacturer |
|------------------------------|-----------------------|-------------------------------------------------|----------------|
| Primary Predicate Device | | | |
| K203201 | MAX, PHM, OVD | NuVasive Thoracolumbar<br>Interbody Systems | NuVasive |
| Additional Predicate Devices | | | |
| K220782 | MAX, OVD, PHM,<br>OVE | IdentiTiTM and TranscendTM<br>Interbody Systems | Alphatec Spine |
K222028 510k Summary Page 1 of 7
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the "atec" logo. The "a" is green, and the "tec" is dark blue. There is a horizontal line under the logo.
| 510(k) | Product Code | Trade Name | Manufacturer |
|---------|-------------------------|-------------------------------------------------------------------|----------------------------------------|
| K183705 | MAX, PHM, OVD, ODP | IdentiTi™ Porous Ti Interbody System | Alphatec Spine |
| K180480 | MAX, PHM | ATEC Universal Spacer System | Alphatec Spine |
| K220782 | MAX, OVD, PHM, OVE | IdentiTi™ ALIF Standalone Interbody System | Alphatec Spine |
| K182746 | MAX, PHM, OVD | ATEC ALIF and LLIF Spacer System | Alphatec Spine |
| K211805 | MAX, ODP, OVD, PHM | IdentiTi and Transcend Interbody Systems | Alphatec Spine |
| K214010 | MAX, OVD, OLO | Anteralign™ Spinal System with Titan nanoLOCK™ Surface Technology | Medtronic |
| K210497 | MAX, ODP, OVD, OVE, MQP | SeaSpine Spacer System NanoMetalene | SeaSpine<br>Orthopedics<br>Corporation |
#### IV. DEVICE DESCRIPTION
The IdentiTi and Transcend Interbody Systems are thoracolumbar intervertebral body fusion systems designed to be inserted through anterior and posterior surgical approaches. The interbody spacers are manufactured from PEEK (polyetheretherketone) Optima LT1 per ASTM F2026, tantalum per ASTM F560, titanium alloy (Ti-6Al-4V ELI), and commercially pure titanium (CPTi Grade 2) per ASTM F67. The interbody spacers are available in the following material options: (1) PEEK (polyetheretherketone) with tantalum and titanium alloy markers, or (2) commercially pure porous titanium (PTi), or (3) a combination of commercially pure porous titanium (CPT) Grade 2) per ASTM F67 and titanium alloy (Ti-6Al-4V ELI) per ASTM F136.
The subject IdentiTi and Transcend Interbody Systems implants consist of various lengths, widths, heights and lordotic options to accommodate individual patient anatomy. To mitigate risk of expulsion, the interbody endplates feature teeth. All interbody spacers feature an internal graft aperture for placement of graft material to promote fusion through the cage. Additionally, the IdentiTi implants are offered with a microstructure due to the layering of material that forms the porous architecture. This porous geometry extends to the superior and inferior surfaces of the device for implant fixation.
The IdentiTi and Transcend NanoTec Interbody Systems implant surfaces have been treated with a 20-40 nanometer thin hydroxyapatite (HA) surface treatment. The surface treatment presents nano-scale topography on the entirety of the implant surface, in addition to macro-/micro-scale topography existing from prior to treatment.
The purpose of this Traditional 510(k) is to receive clearance for expanded indications for the treatment of degenerative spondylolisthesis, multilevel degenerative scoliosis, spinal
> K222028 510k Summary Page 2 of 7
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "atec" in a stylized font. The "a" is green and the "tec" is dark blue. There is a curved line above the "tec" that extends from the top of the "a". There is a trademark symbol to the right of the "c".
stenosis, and sagittal deformity, and use of allogeneic bone consisting of cortical bone, as well as allogeneic bone consisting of demineralized allograft bone with bone marrow aspirate.
#### V. INDICATIONS FOR USE
# IdentiTi Porous Ti Interbody System
The IdentiTi Porous Ti Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the IdentiTi Porous Ti System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The IdentiTi Porous Ti Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
AMP™ Anti-Migration Plate may be used with IdentiTi LIF interbody spacers to provide integrated fixation. IdentiTi LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. IdentiTi ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.
# IdentiTi NanoTec Interbody System
The IdentiTi Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the IdentiTi NanoTec Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The IdentiTi NanoTec Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or
> K222028 510k Summary Page 3 of 7
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for "atec". The "a" is green, and the rest of the letters are dark blue. There is a horizontal line under the logo.
allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
AMP™ Anti-Migration Plate may be used with IdentiTi NanoTec LIF interbody spacers to provide integrated fixation. IdentiTi NanoTec LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation.
# Transcend PEEK Interbody System
The Transcend PEEK Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Additionally, the Transcend PEEK Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Transcend PEEK Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
AMPTM Anti-Migration Plate may be used with Transcend LIF interbody spacers to provide integrated fixation. Transcend LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. Transcend ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.
# Transcend NanoTec Interbody System
The Transcend PEEK Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
> K222028 510k Summary Page 4 of 7
{12}------------------------------------------------
Additionally, the Transcend NanoTec Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Transcend NanoTec Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine.
AMP™ Anti-Migration Plate may be used with Transcend NanoTec LIF interbody spacers to provide integrated fixation. Transcend NanoTec LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation.
# IdentiTi ALIF Standalone Interbody System
The IdentiTi ALIF Standalone Interbody System is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.
Additionally, the IdentiTi ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the IdentiTi ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The IdentiTi ALIF Standalone Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.
The IdentiTi ALIF Standalone Interbody System implants of _20° are a standalone system. The IdentiTi ALIF Standalone Interbody System implants of >20° must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
# IdentiTi NanoTec ALIF Standalone Interbody System
The IdentiTi ALIF Standalone Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as
> K222028 510k Summary Page 5 of 7
{13}------------------------------------------------
back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.
Additionally, the IdentiTi NanoTec ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adiacent levels, the IdentiTi Nanotec ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The IdentiTi NanoTec ALIF Standalone Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.
The IdentiTi NanoTec ALIF Standalone Interbody System implants of ≤20° are a standalone system. The IdentiTi NanoTec ALIF Standalone Interbody System implants of >20° must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
#### VI. TECHNOLOGICAL COMPARISON TO PREDICATES
The technological design features of the subject IdentiTi and Transcend Interbody Systems are substantially equivalent to the primary predicate NuVasive CoRoent Thoracolumbar System (K203201), and additional predicates: IdentiTi™ and Transcend™ Interbody Systems (K220782), IdentiTi Porous Ti Interbody System (K183705), ATEC Universal Spacer System (K180480), ATEC ALIF and LLIF Spacer System (K182746), IdentiTi and Transcend Interbody Systems (K211805), Medtronic Anteralign™ Spinal System with Titan nanoLOCK Surface Technology (K214010), and SeaSpine Spacer System NanoMetalene (K210497).
The technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function, and technology and it was demonstrated that they are substantially equivalent.
#### VII. PERFORMANCE DATA
The following non-clinical testing was performed and included, where appropriate for the design, or referenced in predicate 510(k) submissions to support clearance of IdentiTi™ and Transcend™ Interbody Systems.
- ASTM F2077 static and dynamic axial compression, compression-shear and torsion ●
- ASTM F2267 static subsidence ●
- ASTM Draft F-04.25.02.02 static push-out ●
- . F1714 gravimetric analysis
K222028 510k Summary Page 6 of 7
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for "atec". The "a" is green and the "tec" is dark blue. There is a trademark symbol in the upper right corner of the logo. The logo is simple and modern.
- F1877 particulate analysis ●
- . Screw push-out
Since the technological characteristics of the subject IdentiTi and Transcend systems are substantially equivalent to the predicate systems, no further clinical or non-clinical testing is required to support the expanded indications for use of the subject systems.
# Clinical Information
Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.
# VIII. CONCLUSION
Based upon the information provided in this 510(k) submission, it has been determined that the subject devices are substantially equivalent to legally marketed devices in regard to indications for use, intended use, design, technology, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.