The INTEGRATE™-C Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The INTEGRATE™-C Interbody Fusion is intended for use for anterior cervical interbody fusion in patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The INTEGRATE™-C Interbody Fusion System is intended to be used with supplemental fixation. The INTEGRATE™-C Interbody Fusion System is designed for use with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft and/or demineralized allograft bone with bone marrow aspirate as an adjunct to fusion. Nonoperative treatment prior to treatment with INTEGRATE™-C Interbody Fusion System is six (6) weeks. INTEGRATE™-C Interbody Fusion Systems are to be implemented via an open anterior approach.
Device Story
INTEGRATE™-C Interbody Fusion System is a monolithic, hydroxyapatite-impregnated PEEK interbody fusion implant. Device features porous regions integrated into the body and surface teeth for initial stability; Tantalum markers aid intra- and post-operative imaging. Implanted via open anterior approach by surgeons to treat cervical disc degeneration or instability. Device acts as a structural spacer to facilitate fusion when used with autograft or allogenic bone graft and supplemental fixation. Benefits include spinal stabilization and fusion at contiguous levels from C2-T1.
Clinical Evidence
Bench testing only. Testing included static/dynamic axial compression, compressive shear, and torsion (ASTM F2077); subsidence (ASTM F2267); static tension (ASTM F1147); dynamic shear (ASTM F1160); and static shear (ASTM F1044). A validation cadaver lab was also performed. Results confirm mechanical strength is sufficient for intended use.
Technological Characteristics
Monolithic hydroxyapatite-impregnated PEEK (ASTM F2026 & F1185). Tantalum (ASTM F560) radiopaque markers. Porous regions are integral to the implant body. Features surface teeth for stability. Available in various footprints, lordosis, and heights.
Indications for Use
Indicated for intervertebral body fusion in skeletally mature patients with cervical disc degeneration and/or spinal instability (C2-T1) causing radiculopathy, myelopathy, or pain. Requires 6 weeks of failed nonoperative treatment. Must be used with supplemental fixation and bone graft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
April 28, 2023
HAPPE Spine % Meredith P. Vanderbilt Director of Consulting Empirical Testing Corporation 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K222004
Trade/Device Name: INTEGRATE™-C Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: March 29, 2023 Received: March 29, 2023
Dear Ms. Vanderbilt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Katherine D. Kavlock -S
for
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.
510(k) Number (if known)
K222004
Device Name
INTEGRATE™-C Interbody Fusion System
Indications for Use (Describe)
The INTEGRATE™-C Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The INTEGRATE™-C Interbody Fusion is intended for use for anterior cervical interbody fusion in patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The INTEGRATE™-C Interbody Fusion System is intended to be used with supplemental fixation. The INTEGRATE™-C Interbody Fusion System is designed for use with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft and/or demineralized allograft bone with bone marrow aspirate as an adjunct to fusion. Nonoperative treatment prior to treatment with INTEGRATE™-C Interbody Fusion System is six (6) weeks. INTEGRATE™-C Interbody Fusion Systems are to be implemented via an open anterior approach.
Type of Use (Select one or both, as applicable) 図 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov " An agency may not conduct or sponsor, and a person is not respond to, a collection of information
unless it displays a currently valid OMB number.'
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(K) SUMMARY
| Submitter's Name: | HAPPE Spine, LLC |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address: | 220 Lyon Street, NW, Suite 500<br>Grand Rapids, MI 49503 |
| Submitter's Telephone: | +1-616-294-1026 |
| Contact Person: | Meredith Lee May MS, RAC<br>Empirical Testing Corp.<br>1-719-337-7579<br>MMay@EmpiricalTech.com<br>Image: Empirical Testing Corp. Logo |
| Date Summary was Prepared: | July 6, 2022 |
| Trade or Proprietary Name: | INTEGRATE™-C Interbody Fusion System |
| Common or Usual Name: | Intervertebral Fusion Device with Bone Graft, Cervical |
| Classification: | Class II per 21 CFR §888.3080 |
| Product Code: | ODP |
| Classification Panel: | Spinal Devices (DHT6B) |
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The INTEGRATE™-C Interbody Fusion System implant is made of a single, continuous piece of hydroxyapatite impregnated polyetheretherketone polymer (ASTM F2026 & F1185). The INTEGRATE™-C Interbody Fusion System implant body is monolithic with porous regions derived directly from the implant body, not a sintered or otherwise additive coating, and extended through the device. The device is available in a variety of footprints, lordosis and heights to accommodate variations in the individual pathology and anatomic of the patient. The superior and inferior surfaces of the device contain a pattern of teeth to provide for initial stability. Tantalum (ASTM F560) radiopaque markers are placed in the device to aid in determining the location of the implant intra- and post-operatively.
## INDICATIONS FOR USE
The INTEGRATE™-C Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The INTEGRATE™-C Interbody Fusion is intended for use for anterior cervical interbody fusion in patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The INTEGRATE™-C Interbody Fusion System is intended to be used with supplemental fixation. The INTEGRATE™-C Interbody Fusion System is designed for use with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft and/or demineralized allograft bone with bone marrow aspirate as an adjunct to fusion. Nonoperative treatment prior to treatment with INTEGRATE™-C Interbody Fusion System is six (6) weeks. INTEGRATE™-C Interbody Fusion Systems are to be implemented via an open anterior approach.
HAPPE Spine INTEGRATE™-C Interbody Fusion System
{4}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS
The subject and predicate devices have similar technological characteristics and the differences regarding material composition do not raise any new issues of safety and effectiveness; mechanical and biocompatibility testing have mitigated concerns based on differences in material composition. Specifically, the following characteristics are identical between the subject and predicates:
- Indications for Use ●
- Product Code ●
- Sterility ●
- Sizes ●
- Graft Material
- Device Material ●
#### Predicate Devices
| 510k Number | Trade or Proprietary or Model Name | Manufacturer | Product Code | Predicate Type |
|-------------|------------------------------------|-----------------------------------------|--------------|----------------|
| K173030 | Cohere® | Vertera Spine™ | ODP | Primary |
| K133784 | Vista®-S | Zimmer Trabecular Metal Technology, Inc | ODP | Additional |
| K172484 | A-CIFT™ SoloFuse™ | SpineFrontier (LESpine) | ODP, OVE | Additional |
| K191581 | Endoskeleton® TL | Titan Spine | OVD, MAX | Reference |
## PERFORMANCE DATA
The INTEGRATE™-C Interbody Fusion System has been tested in the following test modes:
- Static and dynamic axial compression per ASTM F2077 ●
- Static and dynamic compressive shear per ASTM F2077
- Static and dynamic torsion per ASTM F2077 ●
- Subsidence per ASTM F2267 ●
- Validation Cadaver Lab ●
- Static Tension Testing per ASTM F1147 ●
- Dynamic Shear Testing per ASTM F1160
- Static Shear Testing per ASTM F1044 ●
The results of this non-clinical testing show that the strength of the INTEGRATE™-C Interbody Fusion System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
{5}------------------------------------------------
# CONCLUSION
The overall technology characteristics and mechanical performance data lead to the conclusion that the INTEGRATE™-C Interbody Fusion System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.