← Product Code [FFX](/productcode/FFX) · K222000

# Transit-Pellets (K222000)

_Medifactia AB · FFX · Jan 18, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222000

## Device Facts

- **Applicant:** Medifactia AB
- **Product Code:** [FFX](/productcode/FFX.md)
- **Decision Date:** Jan 18, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1725
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K222000 · Jan 18, 2023 | Transit-Pellets | Medifactia AB | Post-Market Clinical Follow-Up (PMCF) study; Clinical literature review (18 clinical investigations); User survey | The sponsor utilized a PMCF study, including a literature review of 18 clinical investigations and a user survey, to support the safety and effectiveness of the device in a pediatric population (at least 2 years old). | PMCF; Pediatric; Colonic transit time; Literature review; User survey |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Post-Market Clinical Follow-Up (PMCF) study; Literature review and user survey | Children and young adults (at least 2 years old) with chronic constipation; Sample Size: 1054 children and young adults | Not applicable for this study | Safety and effectiveness of colonic transit time measurement |

## Indications for Use

For evaluation of colonic transit in adult and pediatric patients (at least 2 years old) with chronic constipation and used to aid in differentiating slow and normal transit constipation.

## Device Story

Device consists of radiopaque markers (barium sulfate/silicone rubber) in HPMC capsules; used for colonic transit time evaluation. Patient swallows fixed dose of markers daily for six days; abdominal radiograph taken on day seven. Physician/clinician analyzes radiograph to count retained markers and determine their colonic position; calculates transit time by comparing results to reference values. Used in hospitals/clinics to differentiate slow vs. normal transit constipation. Output assists clinical decision-making for chronic constipation management. Benefits include non-invasive assessment of colonic motility in both adult and pediatric populations.

## Clinical Evidence

Clinical evidence based on literature review of 18 investigations involving 1054 children and young adults using similar radiopaque marker devices. Studies confirmed safety and performance of the marker-based transit time measurement principle. No device-related adverse events reported. User survey confirmed clinical utility and safety in pediatric patients.

## Technological Characteristics

Markers: 22% Barium Sulphate, 78% Elastosil® R 401/60 Silicone rubber. Capsules: HPMC (Hydroxypropylmethylcellulose). Dimensions: 2x4.5mm (ring) and 6x2mm (tube). Radiopacity tested per ASTM F640-12. Biocompatibility per ISO 10993-5, -10, -11. Non-sterile. Shelf life 2.5 years.

## Regulatory Identification

A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.

## Predicate Devices

- Transit-Pellets ([K181760](/device/K181760.md))
- Sitzmarks ([K181750](/device/K181750.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 18, 2023

Medifactia AB % Cherita James Regulatory Consultant M Squared Associates Inc. 127 West 30th Street 9th Floor New York, NY 10001

Re: K222000

Trade/Device Name: Transit-Pellets Regulation Number: 21 CFR 876.1725 Regulation Name: Gastrointestinal motility monitoring system Regulatory Class: Class II Product Code: FFX Dated: December 16, 2022 Received: December 16, 2022

Dear Cherita James:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdri/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

# Sivakami Venkatachalam -S

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K222000

Device Name

Transit-Pellets

Indications for Use (Describe)

For evaluation of colonic transit in adult and pediatic patients (at least 2 years old) with chronic constipation and used to aid in differentiating slow and normal transit constipation.

| Type of Use (Select one or both, as applicable)                                                |                                               |
|------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

In accordance with 21 CFR 807.92, the following information constitutes the Medifactia AB summary for the Transit-Pellets.

#### l. Submission Date

January 11, 2023

#### II. Submitter

Medifactia AB

C/O IOffice Business Center

Kungsgatan 60

Stockholm

Sweden 11122

Contact person: Diana Nyström

Phone number: +46 (0) 31-787 70 77

E-mail: diana.nystrom@medifactia.com

Official Correspondent: Cherita James Regulatory Consultant M Squared Associates, Inc. 127 West 30th St, Floor 9 New York, New York 10001 Ph: 347-954-0624 E-mail: Cjames@Msquaredassociates.com

#### III. Device

Name of device: Transit-Pellets Common or usual name: System, Gastrointestinal Motility Regulatory class: Class II according to 876.1725

{5}------------------------------------------------

### Classification code: FFX

Panel: Gastroenterology/Urology

#### IV. Predicate device

| 510(k)             | Company                | Device          |
|--------------------|------------------------|-----------------|
| Primary: K181760   | Medifactia             | Transit-Pellets |
| Reference: K181750 | Konsyl Pharmaceuticals | Sitzmarks       |

#### V. Device Description

The device is intended to be used for evaluation of colonic transit time in patients with chronic constipation and used to aid in differentiating slow and normal transit constipation in adults and pediatric patients (at least 2 years old). Ten (10) radiopaque markers per day are swallowed for six consecutive days. On day seven an abdominal radiograph is taken. Based on the number of retained markers and the position in colonic transit time is calculated and compared to reference values. Both total transit and segmental transit dysfunction in the colon can be evaluated with the device. By dividing the particle dose on day six by taking five (5) markers in the morning and five (5) markers in the evening, the whole range of transit times (slow, normal, rapid) transit can be measured from the radiograph.

The device is a convenience package of radiopaque markers (22% Barium Sulphate, 78% Elastosil® R 401/60 Silicone rubber) placed in vegetarian capsules from cellulose (HPMC, Hydroxypropylmethylcellulose), intended for single patient use. The dimension of the markers is 2x4.5mm (ring-formed markers) and 6x2mm (tube-formed markers). The capsules are packed in a blister pack is placed inside a folding box.

### There is no change to the Transit-Pellets previously cleared in K181760 for this expanded patient population.

#### VI. Intended use

For evaluation of colonic transit in adult and pediatric patients (at least 2 years old) with chronic constipation and used to aid in differentiating slow and normal transit constipation.

#### VII. Comparison of the technology characteristics with the predicate devices

{6}------------------------------------------------

|                                        | Subject device                                                                                                                                                                                                                                                                                                                              | Primary Predicate                                                                                                                                                                                                                                                                                                                                    | Reference Predicate                                                                                                                                                                                                                                                                                                                                                                                                                            | SE |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|
|                                        | Medifactia AB<br>Transit-Pellets                                                                                                                                                                                                                                                                                                            | Medifactia AB<br>Transit-Pellets                                                                                                                                                                                                                                                                                                                     | Konsyl<br>Pharmaceuticals<br>Sitzmarks                                                                                                                                                                                                                                                                                                                                                                                                         |    |
| K#                                     | K222000                                                                                                                                                                                                                                                                                                                                     | K181760                                                                                                                                                                                                                                                                                                                                              | K181750                                                                                                                                                                                                                                                                                                                                                                                                                                        |    |
| Indication for<br>use/<br>Intended use | For evaluation of<br>colonic transit in adult<br>and pediatric patients<br>(at least 2 years old)<br>with chronic<br>constipation and used<br>to aid in<br>differentiating slow<br>and normal transit<br>constipation.                                                                                                                      | For evaluation of<br>colonic transit in<br>adult patients with<br>chronic<br>constipation and<br>used to aid in<br>differentiating slow<br>and normal<br>constipation.                                                                                                                                                                               | SITZMARKS capsule<br>is a diagnostic test<br>indicated for aiding<br>in the evaluation of<br>colonic motility in<br>patients with severe<br>constipation, as<br>diagnosed by your<br>healthcare<br>professional, but<br>otherwise negative<br>GI evaluations.<br>SITZMARKS capsule,<br>for use in adult and<br>pediatric patients<br>(at least 2 years<br>old), is to be<br>dispensed only by<br>physicians to<br>patients for oral<br>intake. | X  |
| Application                            | Measuring and using<br>colonic transit time to<br>evaluate patients with<br>chronic (severe)<br>constipation.                                                                                                                                                                                                                               | Measuring and using<br>colonic transit time to<br>evaluate patients<br>with chronic<br>(severe)<br>constipation.                                                                                                                                                                                                                                     | Measuring and<br>using colonic transit<br>time to evaluate<br>patients with<br>chronic (severe)<br>constipation.                                                                                                                                                                                                                                                                                                                               | X  |
| Target<br>population                   | Adult and pediatric<br>patients (at least 2<br>years old) patients<br>with chronic (severe)<br>constipation.                                                                                                                                                                                                                                | Adult patients with<br>chronic (severe)<br>constipation.                                                                                                                                                                                                                                                                                             | Adult and pediatric<br>patients (at least 2<br>years old) patients<br>with chronic<br>(severe)<br>constipation.                                                                                                                                                                                                                                                                                                                                | X  |
|                                        | Subject device                                                                                                                                                                                                                                                                                                                              | Primary Predicate                                                                                                                                                                                                                                                                                                                                    | Reference Predicate                                                                                                                                                                                                                                                                                                                                                                                                                            | SE |
|                                        | Medifactia AB<br>Transit-Pellets                                                                                                                                                                                                                                                                                                            | Medifactia AB<br>Transit-Pellets                                                                                                                                                                                                                                                                                                                     | Konsyl<br>Pharmaceuticals                                                                                                                                                                                                                                                                                                                                                                                                                      |    |
|                                        |                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                      | Sitzmarks                                                                                                                                                                                                                                                                                                                                                                                                                                      |    |
|                                        |                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                |    |
| Anatomical site                        | Gastrointestinal<br>tract/colon                                                                                                                                                                                                                                                                                                             | Gastrointestinal<br>tract/colon                                                                                                                                                                                                                                                                                                                      | Gastrointestinal<br>tract/colon                                                                                                                                                                                                                                                                                                                                                                                                                | X  |
| Where used                             | Hospitals, clinics                                                                                                                                                                                                                                                                                                                          | Hospitals, clinics                                                                                                                                                                                                                                                                                                                                   | Hospitals, clinics                                                                                                                                                                                                                                                                                                                                                                                                                             | X  |
| Design                                 | Ring-formed and<br>tube-formed markers.                                                                                                                                                                                                                                                                                                     | Ring-formed and<br>tube-formed<br>markers.                                                                                                                                                                                                                                                                                                           | Three different<br>shapes of markers:<br>O Rings, Double D<br>and Tri-Chamber.                                                                                                                                                                                                                                                                                                                                                                 | X  |
| Dimensions/<br>Capsule size            | 2 x 4.5mm, 6 x 2mm<br>Size -00-                                                                                                                                                                                                                                                                                                             | 2 x 4.5mm, 6 x<br>2mm<br>Size -00-                                                                                                                                                                                                                                                                                                                   | 1mm x 4.5mm<br>Size -00-                                                                                                                                                                                                                                                                                                                                                                                                                       | X  |
| Materials                              | Markers: Barium<br>sulphate, Elastosil® R<br>401/60 Silicone<br>rubber<br>Capsules: HPMC,<br>Hydroxypropyl-<br>methylcellulose                                                                                                                                                                                                              | Markers: Barium<br>sulphate, Elastosil®<br>R 401/60 Silicone<br>rubber<br>Capsules: HPMC,<br>Hydroxypropyl-<br>methylcellulose                                                                                                                                                                                                                       | Radiopaque Marker<br>• Polyvinyl Chloride<br>Resin (PVC)<br>• Bis(2-ethylhexyl)<br>Phthalate (DEHP)<br>• Epoxides Soya Oil<br>• Calcium/Zinc<br>Stabilizers<br>• Phosphate<br>Stabilizer<br>• Fatty Ester &<br>Polyamide<br>Lubricants<br>• Ultramarine<br>Tinting Agents<br>• Barium Sulfate<br>• Dioctyl Phthalate                                                                                                                           | X  |
|                                        | Subject device                                                                                                                                                                                                                                                                                                                              | Primary Predicate                                                                                                                                                                                                                                                                                                                                    | Reference Predicate                                                                                                                                                                                                                                                                                                                                                                                                                            | SE |
|                                        | Medifactia AB<br>Transit-Pellets                                                                                                                                                                                                                                                                                                            | Medifactia AB<br>Transit-Pellets                                                                                                                                                                                                                                                                                                                     | Konsyl<br>Pharmaceuticals<br>Sitzmarks                                                                                                                                                                                                                                                                                                                                                                                                         |    |
|                                        |                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                      | • Mono &<br>Diglycerides<br>Capsules:<br>Hypromellosa                                                                                                                                                                                                                                                                                                                                                                                          |    |
| Principle of<br>operation              | Patient swallows a<br>fixed dose radiopaque<br>marker for a number<br>of days. On day 7 a<br>single abdominal<br>radiograph or<br>fluoroscopy is taken.<br>Based on the number<br>of retained markers<br>on abdominal film and<br>their position in colon<br>a colonic transit time<br>is calculated and<br>compared to<br>reference value. | Patient swallows a<br>fixed dose<br>radiopaque marker<br>for a number of<br>days. On day 7 a<br>single abdominal<br>radiograph or<br>fluoroscopy is<br>taken. Based on the<br>number of retained<br>markers on<br>abdominal film and<br>their position in<br>colon a colonic<br>transit time is<br>calculated and<br>compared to<br>reference value. | Patient swallows a<br>fixed dose<br>radiopaque marker<br>for a number of<br>days. Several<br>abdominal<br>radiographs may be<br>necessary. Based on<br>the number of<br>retained markers on<br>abdominal film(s)<br>and their position in<br>colon a colonic<br>transit time is<br>calculated and<br>compared to<br>reference value.                                                                                                           | x  |
| Performance                            | Colonic Transit Time<br>(CTT/OATT) numerical<br>values reported in<br>days.                                                                                                                                                                                                                                                                 | Colonic Transit<br>Time (CTT/OATT)<br>numerical values<br>reported in days.                                                                                                                                                                                                                                                                          | Colonic Transit Time<br>(CTT/OATT)<br>numerical values<br>reported in hours.                                                                                                                                                                                                                                                                                                                                                                   | x  |
| Biocompatibility                       | Yes, safe for intended<br>use<br>ISO 10993-5, -10, and<br>-11                                                                                                                                                                                                                                                                               | Yes, safe for<br>intended use<br>ISO 10993-5, -10,<br>and -11                                                                                                                                                                                                                                                                                        | No information                                                                                                                                                                                                                                                                                                                                                                                                                                 | x  |
| Packaging                              | Blister and folding box                                                                                                                                                                                                                                                                                                                     | Blister and folding<br>box                                                                                                                                                                                                                                                                                                                           | Folder combo/single<br>pack                                                                                                                                                                                                                                                                                                                                                                                                                    | x  |
| Sterile                                | No                                                                                                                                                                                                                                                                                                                                          | No                                                                                                                                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                                                                                                                                                                                                                                             | x  |
|                                        | Subject device                                                                                                                                                                                                                                                                                                                              | Primary Predicate                                                                                                                                                                                                                                                                                                                                    | Reference Predicate                                                                                                                                                                                                                                                                                                                                                                                                                            | SE |
|                                        | Medifactia AB<br>Transit-Pellets                                                                                                                                                                                                                                                                                                            | Medifactia AB<br>Transit-Pellets                                                                                                                                                                                                                                                                                                                     | Konsyl<br>Pharmaceuticals<br>Sitzmarks                                                                                                                                                                                                                                                                                                                                                                                                         |    |
| Shelf life                             | 2.5 years                                                                                                                                                                                                                                                                                                                                   | 2.5 years                                                                                                                                                                                                                                                                                                                                            | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                        |    |
| Rx Only                                | Yes                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                            | ×  |

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

#### VIII. Performance testing

Transit-Pellets have a size <8mm and are made tube-formed and ring-formed. To test transit characteristics of various types of markers, five types of distinguishable markers in the specific gravity range 1.2-1.6 were examined. The size of the particles was chosen in the range 2-7 mm so that emptying from stomach was likely to occur also with a meal. Transit-Pellets were designed for their size and configurations and are safe and effective for their intended use when compared to the predicate device. Simulated capsule digestion testing in a simulated gastric fluid, dimensional analysis, as well as the clinical literature provided in the predicate 510k submission (K181760), confirms the device performs as intended for the proposed indication for use, i.e., is safe and effective for evaluation of colonic transit in patients with chronic constipation and used to aid in differentiating slow and normal constipation. The Transit-Pellets is substantially equivalent to Sitzmarks regarding material, technological characteristics, and indications for use.

Biocompatibility testing to ISO 10993-5 and -10 of the Transit-Pellets confirm the device is noncytotoxic, non- irritant, and non-sensitizing. Testing to ISO 10993-11 confirmed that the Transit-Pellet extracts did not induce a significantly greater biologic reaction than the control following systemic injection and produced a non-pyrogenic response. Additionally, 14 Day Repeat IV and Intraperitoneal Toxicity in Rats did not demonstrate systemic signs of toxicity over 14 days.

Radiopacity Testing in accordance with ASTM F640-12 confirms visibility of the Transit-Pellets for the duration and environment of their intended use during colonic transit.

Medifactia conducted Bite Force testing with both the Transit-Pellets and the reference predicate device to confirm the safe use of the subject device consumed without a capsule in the proposed pediatric population in the event of accidental biting of the markers. The testing concluded that the Transit-Pellets can withstand similar forces as the reference device, when subjected to simulated bite force testing.

#### IX. Clinical Evidence

A Post-Market Clinical Follow-Up (PMCF) study provided supports the safe and effective use of barium sulphate impregnated particles to measure colonic transit time in pediatric patients at least 2 years old. The literature review carried out during the PMCF study identified 18 clinical investigations performed on similar devices to be of satisfactory quality. In summary, the device performance and safety were successfully measured colonic transit time on 1054 children and young adults. In 12 of the 18 clinical

{10}------------------------------------------------

investigations, the principle of taking markers for six days were applied. No device related adverse event reported in any of the clinical investigations.

Further, a user survey, also part of the PMCF study, confirmed that the Transit-Pellets and the Transit-Pellets principle are used to measure colonic transit time in children and young adults with no adverse events or side-effects identified. The survey also confirmed that some professional users inform patients that they can open the capsules and only swallow the markers. The Transit-Pellets meets of the professional users.

#### X. Conclusion

The subject and primary predicate device are identical. The subject device includes a patient population of both adults and pediatric patients (at least 2 years old) patients with chronic (severe) constipation which is the same as the reference predicate. All devices share the indication for use of measuring and using colonic transit time to evaluate patients with chronic constipation. The original Transit-Pellets submission (K181760) established substantial equivalence to the reference device for use in adults.

The information provided in this 510(k) support that the Transit-Pellets is substantially equivalent in function, composition, and intended use in a pediatric population to the predicate devices. The proposed device raises no new issues of safety and effectiveness.

{11}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K222000](https://fda-staging.innolitics.com/device/K222000)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).