XO Score LP Percutaneous Transluminal Angioplasty Scoring Catheter, XO Score LP Catheter
Applicant
Transit Scientific, LLC
Product Code
PNO · Cardiovascular
Decision Date
Oct 18, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XO Score LP Percutaneous Transluminal Angioplasty Scoring Catheter is intended to be used in conjunction with a PTA balloon to facilitate dilation and apposition of the scoring surface to the stenotic material in the iliac, femoral, iliofemoral, popliteal, infra-popliteal, and renal arteries, and for treatment of obstructive lessions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
Device Story
XO Score LP is a low-profile, single-lumen, metal alloy shaft catheter with a distal balloon-expandable scoring element. Used in conjunction with a standard PTA balloon catheter to facilitate lesion dilation and apposition of the scoring surface to stenotic material. Operated by physicians in clinical settings. The device is advanced over a guidewire to the target lesion; the scoring element is expanded via the PTA balloon to score the vessel wall or fistula lesion. This scoring action aims to improve dilation outcomes. The device is provided sterile for single use.
Clinical Evidence
Bench testing only. Performance verification included kink radius, tensile testing, retrieval force, device recovery, burst pressure, torque, cycling, and scoring capability assessment. Biocompatibility verified per ISO 10993-1 for limited exposure (<24hrs) to circulating blood.
Technological Characteristics
Metal alloy shaft, proximal polymer outer jacket, distal balloon-expandable scoring element. 2.2 Fr and 3.8 Fr diameters. 150 cm length. Radiopaque tip with radiopaque filler. Compatible with 0.014" and 0.018" guidewires. Sterilized via ethylene oxide (EO).
Indications for Use
Indicated for patients requiring dilation of stenotic material in iliac, femoral, iliofemoral, popliteal, infra-popliteal, and renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Contraindicated for use in coronary or neuro-vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
XO Score Percutaneous Transluminal Angioplasty Scoring Catheter (K193495)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 18, 2022
Transit Scientific, LLC % Spencer Walker Director of Regulatory Affairs University of Utah 10 North 1900 East, EHSL Rm, 22B Salt Lake City, Utah 84112
Re: K221986
Trade/Device Name: XO Score LP Percutaneous Transluminal Angioplasty Scoring Catheter, XO Score LP Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PNO Dated: September 29, 2022 Received: September 29, 2022
Dear Spencer Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221986
Device Name
XO SCORE LP Percutaneous Transluminal Angioplasty Scoring Catheter, XO Score LP Catheter
#### Indications for Use (Describe)
The XO Score LP Percutaneous Transluminal Angioplasty Scoring Catheter is intended to be used in conjunction with a PTA balloon to facilitate dilation and apposition of the scoring surface to the stenotic material in the iliac, femoral, iliofemoral, popliteal, infra-popliteal, and renal arteries, and for treatment of obstructive lessions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Transit Scientific. The logo is in black and white and features the words "TRANSIT" stacked on top of "SCIENTIFIC". The font is modern and sans-serif, with rounded corners on some of the letters.
# 510(K) SUMMARY
# (21 CFR 807.92)
# GENERAL INFORMATION
| Submitter: | Transit Scientific, LLC |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Spencer Walker, MSc – Director of Regulatory Affairs<br>University of Utah/ Center for Medical Innovation<br>10 North 1900 East, Rm. 22B<br>Salt Lake City, UT 84112<br>(801) 708-2238<br>Spencer.walker@hsc.utah.edu |
| Date Prepared: | October 5, 2022 |
| Trade Name: | XO Score LP Percutaneous Transluminal Angioplasty Scoring<br>Catheter<br>XO Score LP Catheter |
| Classification Name: | Percutaneous Catheter<br>21 CFR §870.1250, Product Code PNO |
| Device Class: | Class II |
| Predicate Device: | 510(k) No.: K193495<br>Model: XO Score Percutaneous Transluminal Angioplasty Scorin<br>Catheter<br>Manufacture: Transit Scientific, LLC<br>Classification: PNO |
| Reference Device: | 510(k) No.: K193420<br>Model: XO Cross Support Catheter<br>Manufacture: Transit Scientific, LLC<br>Classification: DQY |
# Device Description:
The XO Score LP (Low Profile) Percutaneous Transluminal Angioplasty Scoring Catheter is a 2.2 Fr (0.76 mm) and 3.8 Fr (1.27 mm) diameter, single-lumen, metal alloy shaft, a proximal polymer outer jacket and a balloon expandable scoring element at the distal tip. Designed to work with a standard PTA balloon catheter. The XO Score LP catheter essentially adds the lesion scoring / cutting capability to a standard PTA balloon catheter with the scoring element at
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Transit Scientific. The logo is in black and consists of two words, "TRANSIT" on the top line and "SCIENTIFIC" on the bottom line. The letters are all capitalized and have a modern, sans-serif font. The trademark symbol is located to the right of the word "SCIENTIFIC".
its distal end. It is available in the one catheter length (150 cm), and longitudinal scoring element length (23cm).
| Table 1: XO Score LP Catheter Model Numbers | | |
|---------------------------------------------|-----------|--------------------------------------------------------------------------------------------------------------|
| Product Family | Model No. | Description |
| XO Score LP<br>Catheter | S3.20-150 | XO Score LP Catheter – 2.2 Fr (0.76 mm)<br>and 3.8 Fr (1.27 mm) diameter catheter shaft<br>in 150cm lengths. |
| | S2.15-150 | |
# Indications for Use:
The XO Score LP Percutaneous Transluminal Angioplasty Scoring Catheter is intended to be used in conjunction with a PTA balloon to facilitate dilation and apposition of the scoring surface to the stenotic material in the iliac, femoral, illo-femoral, popliteal, infra-popliteal, and renal arteries, and for treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.
# Comparative Analysis:
Modifications in design, materials, and size range to the previously 510(k) cleared XO Score Catheter (K193495) resulted in new product codes. The following modifications were made to the subject catheter:
- Catheter Shaft Design ●
- Catheter Sizes (Diameter)
- Catheter Effective Length o
- Number of Scoring Struts ●
- Scoring Element Diameter when Expanded
- Scoring Depth
- Balloon Length Compatibility
- Tip Material
- . Accessory Loading Tool
It has been demonstrated that the modified XO Score LP catheter is comparable to the predicate device in fundamental scientific technology, design, materials, principles of operation and functional performance evaluations and is substantial equivalent as summarized in Table 2. Furthermore, the XQ Score LP catheter has been fully assessed within the Transit Scientific Risk Management and Design Controls systems. The differences raise no additional or different questions of safety or effectiveness from that already identified for the predicate device.
{5}------------------------------------------------
| Table 2: Substantial Equivalence Comparison Chart | | | | |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | Predicate - K193495<br>(XO Score Catheter) | Subject Device -<br>(XO Score LP Catheter) | | |
| Ind. for Use | The XO Score Percutaneous Transluminal<br>Angioplasty Scoring Catheter is intended<br>to be used in conjunction with a PTA<br>balloon to facilitate dilation and apposition<br>of the scoring surface to the stenotic<br>material in the iliac, femoral, ilio-femoral,<br>popliteal, infra-popliteal, and renal<br>arteries, and for treatment of obstructive<br>lesions of native or synthetic<br>arteriovenous dialysis fistulae. Not for use<br>in the coronary or neuro-vasculature. | The XO Score LP Percutaneous<br>Transluminal Angioplasty Scoring<br>Catheter is intended to be used in<br>conjunction with a PTA balloon to facilitate<br>dilation and apposition of the scoring<br>surface to the stenotic material in the iliac,<br>femoral, ilio-femoral, popliteal, infra-<br>popliteal, and renal arteries, and for<br>treatment of obstructive lesions of native<br>or synthetic arteriovenous dialysis fistulae.<br>Not for use in the coronary or neuro-<br>vasculature. | | |
| Classification<br>Name | Cardiovascular Device - Percutaneous<br>Catheter<br>21 CFR §870.1250<br>Product Code: PNO<br>Class II | Same | | |
| Single Use | Yes | Same | | |
| Fundamental<br>Scientific<br>Technology | Dilation and treatment of obstructive<br>lesions | Same | | |
| Prescription<br>(Rx Only) | Yes | Same | | |
| Anatomical<br>Access | The iliac, femoral, ilio-femoral, popliteal,<br>infra-popliteal, renal arteries and<br>arteriovenous dialysis fistulae | Same | | |
| Effective Length<br>(cm) | 65 cm, 125 cm | 150 cm | | |
| Shaft Outer<br>Diameter | 6.4Fr (0.085") | D15: 2.2 Fr (0.030")<br>D20: 3.8 Fr (0.050") | | |
| Scoring<br>Length(s) | 5cm, 7cm, 11.5cm, 13.5cm | 23 cm | | |
| PTA Balloon<br>Length | Compatible with PTA balloon lengths of<br>20-40 mm | Compatible with PTA balloon lengths of<br>20-200 mm | | |
| Scoring<br>Members | D25: 22 Struts,<br>D35: 14 Struts,<br>D50: 14 Struts | D15: 10 Struts<br>D20: 12 Struts | | |
| Expanded<br>Scoring<br>Element<br>Diameter | D25: 4.5 - 8.5 mm<br>D35: 4.7 - 8.7 mm<br>D50: 5 - 9 mm | D15: 1.8 - 4.3 mm<br>D20: 3.4 - 7.4 mm | | |
| Scoring Depth | D25: 0.25 mm scoring depth<br>D35: 0.35 mm scoring depth<br>D50: 0.50 mm scoring depth | D15: 0.15 mm scoring depth<br>D20: 0.20 mm scoring depth | | |
| Integrated<br>Balloon | No | No | | |
| Table 2: Substantial Equivalence Comparison Chart | | | | |
| | Predicate - K193495<br>(XO Score Catheter) | Subject Device -<br>(XO Score LP Catheter) | | |
| Scoring<br>member fixed to<br>Balloon | No | No | | |
| Scoring<br>member Profile<br>(Design) | Rectangular Edges/ Diamond Pattern | Same | | |
| Scoring<br>member size<br>(thickness) | 0.0025" | D15: 0.0025"<br>D20: 0.0030" | | |
| Rated Burst<br>Pressure | Dependent on RBP of Balloon Used | Same | | |
| Visibility | Radiopaque catheter body and tip | Radiopaque catheter body, polymer tip<br>with radiopaque filler | | |
| Guidewire<br>Compatibility | 0.014", 0.018", 0.035" | D15: 0.014"<br>D20: 0.014", 0.018" | | |
| Sheath /<br>Introducer<br>Compatibility | 1 French Size Larger than the PTA<br>labeling | 1 French Size Larger than the PTA<br>labeling | | |
| Sterility | Sterile - EO | Same | | |
| Biocompatibility | ISO 10993 | Same | | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Transit Scientific. The logo is in black and white and features the words "TRANSIT" in large, bold letters on top of the word "SCIENTIFIC" in smaller letters. The logo has a modern and clean design.
# Functional/Safety Testing:
The following testing was conducted to validate and verify that the subject device was substantially equivalent to the predicate devices. All data met pre-determined acceptance criteria.
- Biocompatibility - Biocompatibility of the complete and finished XO Score LP catheter has been verified according to the requirements and testing prescribed in ISO 10993-1 and in accordance with FDA guidance document "Use of International Standard ISO 10993-1" for an external communicating device with limited exposure (<24hrs) to circulating blood.
- Additional material biocompatibility data was leveraged from the K193420 O reference device, to support the modified subject device since the device materials and manufacturing processes are the same.
- Cytotoxicity testing was completed for the Accessory Loading Tool. O
- o Design Verification - Performance bench testing was conducted to ensure that the XO Score LP Catheter met the applicable design and performance requirements throughout its shelf life, verify conformity to applicable standards, and demonstrate substantial equivalence to the predicate system. The following performance testing was performed or fulfilled with the XO Score LP Catheter.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Transit Scientific. The logo is in black and white and features the words "TRANSIT" in large, bold letters on the top line. Below that, the word "SCIENTIFIC" is written in a smaller font.
- Kink Radius O
- O Tensile Testing
- Retrieval Force Testing O
- Device Recovery o
- Accessory PTA Balloon Catheter burst pressure within XO Score LP Catheter O
- Torque O
- XO Score LP Catheter Cycling with Accessory PTA Balloon catheter o
- Scoring Capability Assessment O
- Packaging The proposed changes to the XO Score LP Catheter did not affect the . packaging or its configuration.
- Sterilization The proposed changes to the XO Score LP Catheter did not affect the . sterilization cycle or its parameters.
### Conclusion:
The minor differences between the subject XO Score LP catheter and the predicated device did not raise new questions of safety or effectiveness during the design verification testing. The subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.