A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, poly(vinyl chloride) (PVC) examination glove; beige or clear color. Provides single-use barrier protection for wearer. Used in medical settings by clinicians or patients to prevent cross-contamination. Device functions as a physical barrier; no active components. Performance verified against ASTM D5250-19 standards for tensile strength, elongation, and barrier integrity. Benefits include protection against contamination during medical examinations.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D5250-19, ASTM D3767, ASTM D5151-19, ASTM D412-16, ASTM D573-04, and ASTM D6124-06. Biocompatibility testing performed per ISO 10993-10 (skin irritation/sensitization) and ISO 10993-5 (cytotoxicity). All tests met predetermined acceptance criteria.
Technological Characteristics
Material: Poly(vinyl chloride) (PVC). Non-sterile. Single-use. Complies with ASTM D5250-19. Dimensions: 230 mm min length; palm/finger thickness ≥ 0.08 mm. Tensile strength ≥ 11 MPa; elongation ≥ 300% (before/after aging). Residual powder < 2mg/glove. Biocompatibility: ISO 10993-5 and ISO 10993-10 compliant.
Indications for Use
Indicated for use as a disposable, non-sterile, non-powdered patient examination glove worn on the examiner's hand or finger to prevent contamination between patient and examiner. Suitable for over-the-counter use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 6, 2023
Ever Global (Vietnam) Enterprise Corporation Elizabeth Deng US Agent 5748 Eaglewood Place Rancho Cuamonga, California 91739
Re: K221980
Trade/Device Name: Disposable Powder Free Vinyl Examination Gloves, Beige/Clear Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: January 18, 2023 Received: January 20, 2023
Dear Elizabeth Deng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K221980
Device Name
Disposable Powder Free Vinyl Examination Gloves, Beige/Clear
Indications for Use (Describe)
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'EG' in yellow, with the 'E' being larger than the 'G'. Behind the letters is a blue handprint. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP' is at the bottom.
LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG FAX: (0251)351 4023 024 IGLOVESSUPERIEURVIETNAM
### 510(k) SUMMARY (K221980)
### 0.0 Summary Preparation Date: Jan 31, 2023
### 1.0 Submitter:
| Submitter's name: | Ever Global (Vietnam) Enterprise Corporation |
|-------------------------|--------------------------------------------------------------------------|
| Submitter's address: | Long Thanh Industrial Zone<br>Taman Village Dong Nai Province, VN 810000 |
| Phone number: | 84-61-3514022 |
| Fax number: | 84-61-3514023 |
| Name of contact person: | Jerry Lin |
Elizabeth Deng
909 4659188
5748 Eaglewood Place
Rancho Cucamonga, California Rancho Cucamonga, CA 91739
baxianunited48@yahoo.com
### 2.0 US Agent:
US Representative Name: Company Address:
Telephone Number: Contact Email Address:
### 3.0 Name of the Device
Proprietary/Trade name:
Common Name: Classification Name: Device Classification: Regulation Number: Product Code:
4.0 Predicate device Device Name:
> Company name: 510(K) Number:
Disposable Powder Free Vinyl Examination Gloves, Beige/Clear Vinyl Examination Gloves Non-powdered Patient Examination Glove Class I 21 CFR 880.6250 LYZ
Disposable Powder Free Vinyl Examination Glove, Clear/Yellow Ever Global (Vietnam) Enterprise Corporation K170575
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'E' and 'G' in a stylized, overlapping design in yellow. Behind the letters is a blue hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP' is at the bottom, both in blue.
LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG FAX: (0251)351 4023 02 ACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM
### 510(k) SUMMARY (K221980)
### 5.0 Device Description:
A Disposable Powder Free Vinyl Examination Gloves (with color Beige or Clear) is a patient examination glove made from poly (vinvl chloride), non-sterile (as per 21 CFR 880.6250, Class I). The operation principle of this medical device is to provide single use barrier protection for the wearer and the device meets all the requirement specifications for barrier protection, tensile properties as defined in ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.
### 6.0 Device Indications for use:
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
| Device<br>Characteristic | Predicate Device | Subject Device | Comparison |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Product Name | Disposable Powder Free Vinyl<br>Examination Glove,<br>Clear/Yellow | Disposable Powder Free Vinyl<br>Examination Gloves,<br>Beige/Clear | similar glove |
| 510(K) No. | K170525 | K221980 | n/a |
| Product Owner | Ever Global (Vietnam)<br>Enterprise Corporation | Ever Global (Vietnam)<br>Enterprise Corporation | same |
| Product Code | LYZ | LYZ | same |
| Regulation | 21 CFR 880.6250 | 21 CFR 880.6250 | same |
| Class | I | I | same |
| Indications for<br>Use | A patient examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or finger<br>to prevent contamination<br>between patient and examiner. | A non-powdered patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or finger<br>to prevent contamination<br>between patient and examiner. | same |
| Power free | Yes | Yes | same |
| Size | Small/ Medium/Large/X Large | Small/Medium/Large/X Large | same |
| Over-The-Counter Use | Yes (21 CFR 801 Subpart C) | Yes (21 CFR 801 Subpart C) | same |
| Single Use | YES | YES | same |
| Non-Sterile | YES | YES | same |
### 7.0 Comparison of device technological characteristics:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features a large, stylized "EG" in gold, superimposed on a light blue hand. The words "EVER GLOBAL" are written in blue above the "EG", and "ENTERPRISE CORP" is written in blue below the "EG".
Long Thanh Industrial Zone, Taman Village, Dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNC0.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM
# 510(k) SUMMARY (K221980)
| Device Characteristic | Predicate Device | Subject Device | Comparison |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Dimensions-<br>Length | Complies with ASTM<br>D5250-06<br>230 mm min. | Complies with ASTM<br>D5250-19<br>230 mm min. | same |
| Dimensions<br>-Palm Width | Complies with ASTM<br>D5250-06<br>Small 85 ± 5<br>Medium 95 ± 5<br>Large 105 ±5<br>X large 115 ±5 | Complies with ASTM<br>D5250-19<br>Small 85 ± 5<br>Medium 95 ± 5<br>Large 105 ±5<br>X large 115 ±5 | same |
| Dimensions<br>-Thickness | Complies with ASTM<br>D5250-06<br>Palm - 0.08mm min.<br>Finger - 0.1 mm min. | Complies with ASTM<br>D5250-19<br>Palm - 0.08 mm min.<br>Finger - 0.08 mm min. | same, both<br>are thicker<br>than<br>standard<br>request |
| Physical<br>Properties | Tensile Strength<br>Before Aging: 11 MPa, min.<br>After Aging: 11 MPa, min.<br>Elongation:<br>Before Aging: 300% min.<br>After Aging: 300% min. | Tensile Strength<br>Before Aging: 11 MPa, min.<br>After Aging: 11 MPa, min.<br>Elongation:<br>Before Aging: 300% min.<br>After Aging: 300% min. | same |
| Residual powder | Complies with D5250-06<br>< 2mg per glove | Complies with ASTM<br>D5250-19<br>< 2mg per glove | same |
| Freedom from<br>Holes | In accordance with ASTM<br>D5250-06 (G-1 with AQL 2.5) | In accordance with ASTM<br>D5250 -19 (G-1 with AQL 2.5) | same |
| Biocompatibility | ISO 10993-10<br>Skin Sensitization and Skin<br>Irritation test: Pass<br><br>ISO 10993-5: not applicable | ISO 10993-10<br>Skin Sensitization and Skin<br>Irritation test: Pass<br><br>ISO 10993-5<br>In vitro cytotoxicity test: Pass | similar.<br>Subject<br>device pass<br>cytotoxicity<br>test. |
### 8.0 Assessment of Non-Clinical Performance Data:
The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. Disposable Powder Free Vinyl Examination Gloves, Beige/Clear Color made by Ever Global (Vietnam) Enterprise Corporation met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'E,' 'G,' and 'C' in yellow, superimposed on a blue hand. The words 'EVER GLOBAL' are at the top of the logo, and the words 'ENTERPRISE CORP' are at the bottom.
Long Thanh Industrial Zone, Taman Village, Dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 MAIL: INFO@EGVNC0.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM
# 510(k) SUMMARY (K221980)
| Test | Test Method | Purpose | Acceptance Criteria | Results |
|---------------------------------------------------|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimension | ASTM D3767 | Determine the<br>geometrical dimension<br>of gloves | Length: 230 mm min.<br>Thickness:<br>Palm - 0.08 mm min.<br>Finger - 0.08 mm min.<br>Palm Width:<br>Small 85 ± 5 mm<br>Medium 95± 5 mm<br>Large 105 ± 5 mm<br>X Large 115 ± 5 mm | Pass |
| Freedom from<br>holes (Water leak) | 21 CFR 800.20.<br>&<br>ASTM<br>D5151-19 | Detect the holes on the<br>gloves. | G-I/AQL 2.5 | Pass |
| Tensile strength<br>(Before<br>aging/After aging) | ASTM D412-16<br>& ASTM<br>D573-04 | Evaluate the tensile<br>(tenson) properties of<br>the gloves. In addition,<br>it also determines the<br>influence of elevated<br>temperature on the<br>physical properties of<br>gloves. | Before Aging: 11 MPa,<br>min.<br>After Aging: 11 MPa, min | Pass |
| Elongation<br>(Before<br>aging/After aging) | ASTM D412-16<br>& ASTM<br>D573-04 | temperature on the<br>physical properties of<br>gloves. | Before Aging: 300% min.<br>After Aging: 300% min. | Pass |
| Powder Residual | ASTM<br>D6124-06 | Determine the average<br>powder mass found on<br>the gloves | < 2mg per glove | Pass |
| Biocompatibility-<br>Skin Irritation | ISO<br>10993-10:2010 | determine the potential<br>of glove to promote<br>skin sensitization & | Negative Response | Pass |
| Biocompatibility-<br>Skin Sensitization | ISO<br>10993-10:2010 | irritation reactions after<br>repeated applications | No contact sensitization | Pass |
| Biocompatibility-<br>cytotoxicity | ISO<br>10993-5:2009 | determine the<br>cytotoxicity potential<br>of glove | No in vitro cytotoxicity | Pass |
### 9.0 Assessment of Clinical Performance Data:
Clinical data is not needed for this type of device.
### 10.0 Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the subject device Disposable Powder Free Vinyl Examination Gloves, Beige/Clear is as safe, as effective, and performs as well as or better than the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.