K221965 · Femdx Medsystems, Inc. · MKO · Mar 17, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K221965
Device Name
FemDx FalloView
Applicant
Femdx Medsystems, Inc.
Product Code
MKO · Obstetrics/Gynecology
Decision Date
Mar 17, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Indications for Use
The FemDx FalloView™ is intended for visualization of the cervical canal, uterine cavity, and proximal fallopian tube(s) during diagnostic gynecological procedures.
Device Story
Disposable, hand-held, battery-operated endoscope for diagnostic gynecological visualization. Device comprises 4.7 mm hysteroscopic cannula, 1.2 mm falloposcopic catheter, and integrated visualization system. Catheter contains CMOS camera module and LED illumination; cannula includes saline irrigation port. Operator (trained medical professional) uses device in hospital or clinic to visualize cervical canal, uterine cavity, and proximal fallopian tubes. Real-time images displayed on integrated 3.5-inch LCD handle screen. Device captures screenshots; transmits images via Bluetooth to PC-based 'FalloView™ Image Viewer' software. Enables direct visualization of anatomy to assist clinical assessment of tubal patency or proximal occlusion. Benefits include integrated, portable visualization without external light sources or complex fiber-optic bundles.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included mechanical performance (tensile, torque, bending), optical quality (resolution, field of view, signal-to-noise ratio), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993), and sterilization validation (ISO 11135).
Technological Characteristics
Disposable, battery-operated (9V regulated to 3.3V) endoscope. 4.7 mm cannula, 1.2 mm catheter. CMOS camera module (40,000 pixels, 1.75 um pixel size). LED illumination (378-192 LUX). Integrated 3.5-inch LCD. Connectivity: Bluetooth for image transmission. Sterilization: Ethylene Oxide (EO). Standards: ISO 8600 (endoscopes), ISO 10993 (biocompatibility), IEC 60601-1 (electrical safety), IEC 62471 (photobiological safety).
Indications for Use
Indicated for visualization of the cervical canal, uterine cavity, and proximal fallopian tubes in patients undergoing diagnostic gynecological procedures by trained medical professionals.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
Conceptus Fallopian Tube Catheterization with Microendoscopy (FTCM) System (K962587)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 17, 2023
FemDx Medsystems, Inc. % Sevrina Ciucci Regulatory Consultant Regulatory Consulting Services LLC 2336 Walsh Avenue, Suite A Santa Clara, CA 95051
Re: K221965
Trade/Device Name: FemDx FalloView™ Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: MKO, HIH Dated: February 16, 2023 Received: February 17, 2023
Dear Sevrina Ciucci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221965
Device Name FemDx FalloView™
Indications for Use (Describe)
The FemDx FalloView™ is intended for visualization of the cervical canal, uterine cavity, and proximal fallopian ube(s) during diagnostic gynecological procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
| Date Prepared: | March 15, 2023 |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | FemDx Medsystems, Inc.<br>2336 Walsh Avenue, Suite A<br>Santa Clara, CA 95051 |
| Contact Person: | Xi Francis, CEO<br>Email: Ashlee@femdxmedical.com |
| Trade/Device Name: | FemDx FalloView™ |
| Common Name: | Falloposcope, Hysteroscope |
| Classification Regulation: | 21 CFR §884.1690, Hysteroscope and accessories |
| Product Codes: | MKO, HIH |
| Device Classification: | Class II |
| Classification Panel: | Obstetrics/Gynecology |
| Primary Predicate: | Conceptus Fallopian Tube Catheterization with Microendoscopy (FTCM)<br>System (K962587) |
| Reference Devices: | LiNA Medical ApS LINA Opera Scope (K171113)<br>Cook Incorporated Rösch-Thurmond Fallopian Tube Catheterization Set and<br>Fallopian Tube Catheterization Set (K171604)<br>Aveta System, Aveta Disposable Hysteroscope (Pearl/Opal/Coral), Aveta<br>Disposable Cystoscope (Coral) (K213171) |
#### Device Description
The FemDx FalloView™ is a disposable, hand-held, battery-operated endoscope that contains a 4.7 mm hysteroscopic cannula, a 1.2 mm falloposcopic catheter, and a visualization system. The catheter is integrated inside the cannula center lumen and can be deployed, retracted, or rotated. The cannula includes a lumen and a proximal female luer lock port for saline irrigation. The visualization system includes a camera module positioned at the tip of the catheter, a controller board, an LCD display integrated into the device's handle, and LED/fiber optic illumination. The FemDx FalloView™ contains embedded software designed to display the real time image on the LCD display, control the illumination light level, and capture a screenshot. The captured image can be transmitted to a computer via Bluetooth and then accessed using the FalloView™ Image Viewer, a separate PC based software application.
{4}------------------------------------------------
#### Indications for Use
The FemDx FalloView™ is intended for visualization of the cervical canal, uterine cavity, and proximal fallopian tube(s) during diagnostic gynecological procedures.
#### Substantial Equivalence Discussion
The FemDx FalloView™ is substantially equivalent to Conceptus Fallopian Tube Catheterization with Microendoscopy (FTCM) System (K962587). Table 1 compares the FemDx FalloView™ to the predicate device. The different technological characteristics exist in other previously cleared FDA devices of the same type and do not raise different questions of safety and effectiveness than the predicate device.
| | Predicate Device | Subject Device |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Conceptus FTCM System | FemDx FalloView™ |
| | K962587 | K221965 |
| Device Classification | II | II |
| Classification Name | Falloposcope | Falloposcope, Hysteroscope and Accessories |
| Regulation Number | 884.1690 | 884.1690 |
| Product Code | MKO | MKO, HIH |
| Indications for Use | The Conceptus FTCM System is intended for<br>use in selective catheterization / cannulation<br>of the fallopian tubes in order to evaluate<br>proximal tubal occlusion or patency under<br>direct visualization. | The FemDx FalloView is intended for<br>visualization of the cervical canal, uterine<br>cavity, and proximal fallopian tube(s) during<br>diagnostic gynecological procedures. |
| Components | Soft Seal Cervical Catheter, Soft Torque<br>Uterine Catheter (including Mandrel and<br>Accessory Valve), and/or Articulating Tip<br>Hysteroscope, Variable Softness Catheter,<br>Guidewires (Robust Guidewire, Traveler<br>Guidewire, Supple Guidewire, and Pirouette<br>Guidewire), and a Microendoscope with<br>Eyepiece | •Handle with LCD display unit, battery pull<br>tab, and control button to turn device on and<br>off.<br>•Cannula with stainless -steel hypo-tube that<br>forms the inner lumen of the cannula that<br>houses the catheter. The distal tip<br>incorporates a low intensity LED to illuminate<br>the uterus.<br>•Catheter with a CMOS-chip endoscope that<br>maintains an image of the tissue in contact<br>with its surface. Contains an LED and camera<br>connected to the controller board. |
| Intended User | Trained Medical Professionals | Trained Medical Professionals |
| Site of Use | Hospitals and Physicians Offices | Hospitals and Physicians Offices |
| Anatomical Use | Proximal Fallopian Tube(s) | Cervical Canal, Uterine Cavity, and Proximal<br>Fallopian Tube(s) |
| Cannula Working<br>Length | Unknown | 250 mm |
| Cannula Outer<br>Diameter | Unknown | 4.7 mm |
| | Predicate Device | Subject Device |
| Characteristic | Conceptus FTCM System | FemDx FalloView™ |
| | K962587 | K221965 |
| Cannula Tip Design | Soft Torque Uterine Catheter: curved tip | Rounded polymer tip with angled distal portion (10 mm approx. 15°) |
| Cannula Lumens | Soft Seal Cervical Catheter: dual lumen<br>Soft Torque Uterine Catheter and Variable<br>Softness Catheter: single lumen | Three lumens |
| Inflow / Outflow Design | Inflow: Inflow occurs through a port on the<br>Variable Softness catheter<br>Outflow: No outflow | Inflow: Inflow occurs through a port on the proximal side of the cannula.<br>Outflow: No outflow |
| Inflow Pressure Limit | Unknown, use a syringe, or irrigation tube | 75 mmHg |
| Catheter Working Length | Unknown | 108 mm (extension beyond cannula tip) |
| Catheter Outer Tip Diameter | Variable Softness Catheter: 1.2 to 1.3 mm<br>according to Surrey et.al.1996Error! Bookmark not<br>defined. | 1.2 mm |
| Catheter Tip Design | Unknown | Rounded polymer |
| Imaging Technology | Microendoscope: fiberscope with optical fiber<br>bundle for pixels | CMOS Camera Module |
| LCD Display Size | Not applicable, use eyepiece or external video<br>system for display | 3.5 inches (diagonal) |
| Image Resolution | Unknown | 40,000 pixels<br>Pixel size 1.75 um |
| Illumination Light Source | External light source.<br>Fiber optics in hysteroscope and<br>microendoscope extend the light energy. | LED at tip of the cannula and internal LED<br>with fiber optics that extends to the tip of the<br>catheter. |
| Light at 20 mm from target. Range high to low level [LUX] | Not applicable | 378 - 192 LUX |
| Image Transmission | Not applicable | Transits images from CMOS camera module<br>to integrated LCD display. |
| Energy Source | Not applicable | Handpiece: 9V battery, regulated to 3.3V<br>internally |
| Sterilization | Unknown | Ethylene Oxide (EO) |
| Image Saving Software | Not applicable | Yes |
#### Table 1: Substantial Equivalence Table
{5}------------------------------------------------
## Non-Clinical Testing
To demonstrate safety and effectiveness of the FemDx FalloView™ and show substantial equivalence to the predicate device, a comprehensive series of non-clinical tests were completed including mechanical dimensional and performance, electrical hardware/software performance, optical performance, and safety testing in accordance with FDA's guidance, "Hysteroscopes and Gynecological Laparoscopes Submission
{6}------------------------------------------------
Guidance for a 510(k)". The FemDx FalloView™ passed all applicable testing in accordance with internal requirements and national and international standards. The results demonstrate that the FemDx FalloView™ satisfies the performance, functional, and safety requirements relative to the product specifications, risk analyses, and Instructions for Use, and does not raise different questions of safety and effectiveness than the predicate device.
The non-clinical tests relied on for a determination of substantial equivalence are listed below.
- . "Power on" testing including battery lifetime.
- . Image quality comparative testing
- Field of view, Direction of view ISO 8600-3 O
- Image color quality о
- Image resolution ISO 8600-5 O
- O Depth of field
- Image signal to noise ratio O
- Dynamic range O
- Image geometric distortion O
- Image intensity uniformity O
- Flow rate and leakage testing ISO 10555-1 ●
- Dimensional testing ●
- Maximum width of insertion portion per ISO 8600-4 O
- Cannula maximum insertion width ISO 8600-4 O
- Catheter shaft maximal outer diameter ISO 8600-4 O
- Mechanical testing for bending (displacement), tensile strength, and torque ●
- Catheter insertion force, deployment force, and retraction force O
- O Cannula and catheter bending force
- Cannula and catheter rotational force O
- Mechanical strength ●
- Catheter sheath tensile strength o
- Catheter stiffness/bending curvature radius O
- Catheter torsional strength o
- Cannula tensile strength O
- Cannula torsional strength O
- Visualization and illumination design validation testing including equipment compatibility.
- Cannula and catheter insertion testing within a uterus and fallopian tube model O
- Illumination testing within a uterus and fallopian tube model O
- LED illumination light level testing
- Thermal safety testing
- . Photobiological safety IEC 62471
- Packaging testing ISO 11607-1
- Simulated shipping conditions ASTM D4169 o
- Climatic conditioning ASTM D4332 O
{7}------------------------------------------------
- Package and seal integrity visual inspection ASTM F1886/F1186M O
- O Bubble leak ASTM F2096
- O Aseptic opening ISO 11607-1
- Seal strength ASTM F88/F88M O
- Seal width ASTM F2203 O
- Biocompatibility ISO 10993-1
- Cytotoxicity ISO 10993-5 O
- o Sensitization ISO 10993-10
- o Irritation ISO 10993-10
- Sterilization ISO 11135 ●
- . EO and ECH residual testing ISO 10993-7
- . Software verification and validation IEC 62304
- . Electrical safety IEC 60601-1
- Electromagnetic compatibility IEC 60601-1-2 (EMC and Immunity)
#### Conclusion
The FemDx FalloView™ is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.