The Catapult Guide Sheath is indicated to be used for introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.
Device Story
Catapult Guide Sheath is a sterile, single-use, coil-reinforced introducer sheath system; includes dilator with tapered tip, hemostasis valve, and side port with 3-way stopcock. Used in healthcare facilities by clinicians to introduce interventional/diagnostic devices into peripheral/coronary vasculature. Sheath inserted over dilator and guidewire; hemostasis valve minimizes blood leakage. Radiopaque marker aids distal visualization under fluoroscopy. Hydrophilic coating on distal portion reduces friction. Stainless-steel coil reinforcement provides kink resistance in tortuous anatomy. Available in 4F-8F sizes and various lengths/shapes. Device remains in patient for up to 24 hours. Facilitates safe access for diagnostic or interventional procedures.
Clinical Evidence
Bench testing only. Performance verified via mechanical testing (kink resistance, leakage, bond strength, coating integrity), usability study, accelerated aging, and biocompatibility assessment (ISO 10993). No clinical data presented.
Technological Characteristics
Coil-reinforced introducer sheath; materials include Pebax, PTFE liner, stainless steel coil, polycarbonate/silicone hemostasis valve. Hydrophilic coating. Sterilized via ethylene oxide (SAL 10-6). Sizes 4F-8F; lengths 15-90cm. Standards: ISO 10555-1, ISO 11070, ISO 10993, ASTM F1980, ASTM F640.
Indications for Use
Indicated for introduction of interventional and diagnostic devices into peripheral and coronary vasculature in patients requiring percutaneous catheterization.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 29, 2022
Contract Medical International s.r.o. Jan Kubicek Senior Regulatory Specialist Vazni 848 Hradec Kralove. 50011 Czech Republic
Re: K221914
Trade/Device Name: Catapult Guide Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB, DRE Dated: June 27, 2022 Received: June 30, 2022
Dear Jan Kubicek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Finn Donaldson Team Lead DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221914
Device Name Catapult Guide Sheath
Indications for Use (Describe)
The Catapult Guide Sheath is indicated to be used for introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
#### Submitter 1.1
Submitter: Contract Medical GmbH Lauensteiner Strasse 37 01277 Dresden Germany Contact Person: Juliana Vaz Nuernberger, Position: Global Head of Regulatory Affairs +49 351 210 747-19 Phone: Email: juliana.nuernberger@heraeus.com Date Summary Prepared: 28th July 2022
#### 1.2 Device
| Device Trade Name: | Catapult Guide Sheath |
|----------------------|-------------------------------------------------------------------------------|
| Device Common Name: | Introducer Sheath |
| Classification Name: | Introducer Catheter, and<br>Dilator, Vessel, For Percutaneous Catheterization |
| Device Class: | II |
| Product Code: | DYB, and DRE |
#### 1.3 Predicate Device
Predicate device: DuraSheath Introducer Sheath System (K181463)
Reference device: R2P Destination Slender Guiding Sheath (K193125)
#### 1.4 Device Description
The Catapult Guide Sheath consists of a coil reinforced introducer sheath with hemostasis valve and side port, as well as a dilator with a tapered tip and luer lock at the proximal end. The main introducer sheath tubing is connected at the proximal end to a hemostasis valve with side port tubing that is connected to a plastic color coded 3-way stopcock valve. The side port is used for flushing the introducer sheath. The introducer sheath is introduced into the vascular system with the aid of the dilator. The hemostasis valve at the proximal end of the introducer sheath conforms and seals around guide wires and catheters to reduce blood leakage from the introducer sheath. A radiopaque marker helps identify the distal end of the introducer sheath. The introducer sheath has a lubricious hydrophilic coating on the outer surface up to 20cm on the distal portion.
The System consists of the following components:
- A One Introducer Sheath) with hemostasis valve
- A One dilator or
- A Two dilators for 15 cm version
{4}------------------------------------------------
The shaft and hemostasis valve are pre-assembled and packaged together with a dilator. Each shaft and each dilator are placed in a separate protection tube, the protection tubes are fixated onto a packaging pad. Products are sealed inside a medical grade Tyvek pouch on which a product label is placed. Devices are packaged in labeled packers containing five units and one Instruction for Use (IFU) booklet.
The Catapult Guide Sheath is a sterile, disposable device. It is a prescription medical device that is used only in healthcare facilities or hospitals. The device is suitable for placement in patients for up to 24 hours.
Devices are sterilized using ethylene oxide. Maximum levels of remaining tested sterilant residuals meet requirement limits EO < 4mg/24h and ECH < 9mg/24h; while achieving Sterility Assurance Level SAL 10-6.
#### Physical Description 1.5
The Catapult Guide Sheath exists in 4 French, 6 French, 7 French and 8 French sizes, each with effective lengths of 15 cm, 45 cm, 60 cm, or 90 cm.
Selected product versions are manufactured in one or more of the following shapes - Left Internal Mammary Artery (LIMA), Hockey Stick (HS), Multi-Purpose (MP) and Renal Double Curve (RDC). The shape of sheaths is formed by heating of distal part of straight product versions.
| ID<br>(French) | Length | Coating | Straight | Multipurp<br>ose<br>(MP) | Left<br>Internal<br>Mammary<br>Artery<br>(LIMA) | Renal<br>Double<br>Curve<br>(RDC) | Hockey<br>Stick<br>(HS) |
|----------------|--------|---------|----------|--------------------------|-------------------------------------------------|-----------------------------------|-------------------------|
| 4F | 15cm | 5cm | x | | | | |
| 4F | 45cm | 10cm | x | | | | |
| 4F | 60cm | 10cm | x | | | | |
| 4F | 90cm | 20cm | x | | | | |
| 5F | 15cm | 5cm | x | | | | |
| 5F | 45cm | 10cm | x | x | | x | x |
| 5F | 60cm | 10cm | x | | | | |
| 5F | 90cm | 20cm | x | | | | |
| 6F | 15cm | 5cm | x | | | | |
| 6F | 45cm | 10cm | x | x | x | x | x |
| 6F | 60cm | 10cm | x | | | | |
| 6F | 90cm | 20cm | x | x | | x | x |
| 7F | 15cm | 5cm | x | | | | |
| 7F | 45cm | 10cm | x | x | x | x | x |
| 7F | 60cm | 10cm | x | | | | |
#### Device Models 1.6
{5}------------------------------------------------
| 7F | 90cm | 20cm | X | X | | X |
|----|------|------|---|---|--|---|
| 8F | 15cm | 5cm | X | | | |
| 8F | 45cm | 10cm | X | | | |
| 8F | 60cm | 10cm | X | | | |
| 8F | 90cm | 20cm | X | | | |
Table 1 - Catapult Guide Sheath Device Models
#### Indications for Use 1.7
The Catapult Guide Sheath is indicated to be used for introduction of interventional and diagnostic devices into the peripheral (and coronary) vasculature.
#### Device Comparison 1.8
The new versions of the Catapult Guide Sheath are a manually operated, sterile, single patient use sheath system made predominantly of thermoplastic polymers. The sheath is reinforced with a stainless-steel coil in order to provide kink resistance when passed through tortuous paths. Regarding the design, device features, method of sterilization, and mode of operation, the new versions of Catapult Guide Sheath does not differ from the predicate device.
Materials used for manufacture of the new Catapult Guide Sheath are the same or very similar as those contained in the predicate device.
| Characteristics | Subject Device<br>Catapult Guide Sheath | Predicate<br>DuraSheath Introducer Sheath System |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | Catapult Guide Sheath | DuraSheath Introducer Sheath System |
| Indications for use | The Catapult Guide Sheath is indicated to be<br>used for introduction of interventional and<br>diagnostic devices into the peripheral (and<br>coronary) vasculature.<br>* These Indications for Use for the subject<br>device are a subset of those cleared for the<br>Predicate | The DuraSheath Introducer Sheath System<br>is indicated to be used for introduction of<br>interventional and diagnostic devices into<br>the peripheral (and coronary) vasculature.<br>The device is also intended to be used<br>within a pediatric population. |
| Principle of operation | Manual operation where the sheath is<br>inserted and removed over dilator and<br>guidewire | Same |
| French sizes | 4F to 8F | Same |
| Effective lengths | 15, 45, 60 and 90cm | Same |
| Shaft distal shape<br>configuration | Straight and curved | Straight |
| Hemostasis valve | Polycarbonate (Valve body and cap) | Same |
| Characteristics | Subject Device<br>Catapult Guide Sheath | Predicate<br>DuraSheath Introducer Sheath System |
| material | Silicone (Valve disc +lubricant) | Same |
| Hemostasis valve design | Removable hemostasis valve (disk valve)<br>with side port | Same |
| Side port for flushing | Side port tubing with a color coded 3-way<br>stopcock | Same |
| Stopcock material | Polycarbonate (body)<br>Acetal (handle)<br>Polyurethane (extension line) | Same |
| Shaft outer layer (cover)<br>material | Pebax<br>Colorant | Polyurethane (some models)<br>Pebax<br>Colorant |
| Shaft reinforcement<br>material | Stainless steel coil | Same |
| Shaft inner layer (liner)<br>material | PTFE inner layer | Same |
| Shaft proximal connection | Luer connector | Same |
| Shaft / Dilator luer hub | Luer hub (Pebax), colour coded | Same |
| Lubricious coating | Hydrophilic coating | Hydrophobic coating |
| Shaft radiopaque marker | Platinum iridium marker at the distal end | Same |
| Dilator material | Pebax with Barium sulphate (Dilator body)<br>Pebax (Dilator hub) | Same |
| Dilator design | Tapered tip on dilator.<br>Color coded hub.<br>Snap fit between dilator and valve cap.<br>Shape - Straight | Same |
| Guidewire included | No | Same |
| Radiation safety | Fluoroscopy allowed. | Same |
| Sterile package | Pouch | Same |
| Sterilization method /<br>Sterilization Assurance<br>Level (SAL) | Ethylene oxide cycle<br>SAL 10-6 | Same |
| Shelf Life | 3 years | Same |
| Compatibility with other<br>devices | Guidewires up to 0.018 (15cm version only),<br>0,035 for 4F-6F and 0.038 7F-8F inches and<br>instruments up to inner diameter for<br>introducer sheath. | Same |
| French Size/Guidewire<br>compatibility marking | French size of the devices is differentiated<br>by color-coding of sheath and dilator Luer<br>hub. Maximum guide wire size is referenced<br>in labelling or in case of 15cm version<br>additionally on dilator hub. | Same |
{6}------------------------------------------------
Table 2 - Device Comparison
{7}------------------------------------------------
#### Performance Data 1.9
Performance data demonstrate that the new Catapult Guide Sheath is substantially equivalent to the predicate. The following performance data from non-clinical tests support the substantial equivalence determination:
- . Mechanical testing, including tests required under relevant international standards, and usability test performed to verify and validate the design.
- . Biocompatibility Risk Assessment (BRA) and performed biocompatibility device testing to demonstrate biocompatibility.
- . Sterilization information to confirm sterility of the device upon exposure to the selected sterilization EO cycle.
- Accelerated aging testing to confirm product performance at the end of the shelf life.
The list of tests performed in support of determination of substantial equivalence is provided in Table 3.
| No. | Verification / Validation<br>Activity | Test Type | Results | Applicable Standard(s)<br>(Recognized Consensus # if<br>applicable] |
|-----|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Visual and Dimensional<br>Evaluation | Visual | Pass | Internal requirement |
| 2 | Sheath/dilator fit test | Mechanical/Visual | Pass | Internal requirement |
| 3 | Sheath pull out test | Mechanical/Visual | Pass | ISO 10555-1:2013 (6-408)<br>IEC 62366-1:2020 (5-129)<br>Internal requirement |
| 4 | Coating integrity | Mechanical/Visual | Pass | Internal requirement<br>Compared to reference device:<br>R2P Destination Slender Guiding<br>Sheath (K193125) |
| 5 | Sheath Liquid Leakage<br>Liquid leakage through<br>hemostasis valve | Mechanical/Visual | Pass | ISO 11070:2014 (N/A)<br>ISO 10555-1:2013 (6-408) |
| 6 | Coating integrity - Particle<br>evaluation test | Mechanical/Visual | Pass | AAMI TIR42 (3-172)<br>ISO 8536-4 (6-447)<br>USP <788> (N/A for medical<br>devices)<br>Compared to reference device:<br>R2P Destination Slender Guiding<br>Sheath (K193125) |
| No. | Verification / Validation<br>Activity | Test Type | Results | Applicable Standard(s)<br>(Recognized Consensus # if<br>applicable] |
| 7 | Sheath kink resistance test | Mechanical | Pass | EN 13868:2002 (N/A)<br>Internal requirement |
| 8 | Sheath force at break test<br>Sheath creep to break test<br>Dilator hub bond strength | Mechanical | Pass | ISO 10555-1:2013 (6-408)<br>ISO 11070:2014 (N/A)<br>Internal requirement |
| 9 | Sheath Stiffness test | Mechanical | Pass | Internal requirement |
| 10 | Usability evaluation<br>Simulated use test | Usability Study | Pass | IEC 62366-1:2020 (5-129)<br>Internal requirement |
| 11 | Accelerated Age Study | Mechanical/ External<br>Laboratory | Pass | ASTM F1980-16 (14-497)<br>Various per performed tests |
| 12 | Biocompatibility testing<br>Cytotoxicity<br>Hemolysis<br>Partial Thromboplastin<br>Time (PTT)<br>Complement activation<br>Biological Risk Assessment | External Laboratory<br>Documented<br>assessment | Pass | ISO 10993-1:2018 (2-258)<br>ISO 10993-4:2017 (2-248)<br>ISO 10993-5:2009 (2-245)<br>ISO 10993-10:2010 (2-174)<br>ISO 10993-11:2017 (2-255)<br>ISO 10993-12:2012 (2-191)<br>ISO 14971:2019 (5-125)<br>FDA Guidance Use of International<br>Standard ISO 10993-1 |
| 13 | Sterilization adoption | Documented<br>assessment /<br>External laboratory<br>testing | Pass | ISO 10993-7:2008 (14-408)<br>ISO 11737-1:2018 (14-514)<br>ISO 11135:2014 (14-529)<br>ISO 10993-1:2018 (2-258)<br>AAMI TIR28:2009 (N/A) |
| 14 | Visibility of the sheath<br>under X-ray | Documented<br>assessment | Leveraged<br>from predicate | ASTM F 640-20 (8-556)<br>ISO 11070:2014 (N/A) |
| 15 | Packaging integrity<br>assessment | Documented<br>assessment | Leveraged<br>from predicate | EN 868-5:2019<br>ASTM F88/F88M-15(14-482)<br>ISO 11607-1 Second edition 2019-<br>02 (14-530) |
{8}------------------------------------------------
Table 3 - Testing overview
{9}------------------------------------------------
## 1.10 Conclusions
Based on the intended use, technological characteristics, and performance data provided, the Catapult Guide Sheath is considered appropriate for the intended use. The technological and material differences between the subject and predicate devices do not raise new questions of safety or effectiveness. The subject Catapult Guide Sheath device is considered to be substantially equivalent to the predicate DuraSheath Introducer Sheath System (K181463).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.