← Product Code [NUC](/productcode/NUC) · K221908

# JO CUMPLAY Personal Lubricant (K221908)

_CC Wellness, LLC · NUC · Sep 20, 2022 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221908

## Device Facts

- **Applicant:** CC Wellness, LLC
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** Sep 20, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

JO CUMPLAY Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

## Device Story

JO CUMPLAY Personal Lubricant is a non-sterile, water-based lubricant for penile, anal, or vaginal use. It is intended to moisturize and enhance comfort during sexual activity. The device is sold over-the-counter in a 4 fl. oz. polyethylene terephthalate bottle with a polypropylene pump. It is applied topically by the user. The lubricant supplements natural moisture and is compatible with natural rubber latex and polyisoprene condoms, but not polyurethane condoms. It functions as a physical barrier/lubricant to reduce friction. Clinical benefit is derived from increased ease and comfort during intimate activity.

## Clinical Evidence

Bench testing only. Biocompatibility testing conducted per ISO 10993-1 (Cytotoxicity, Sensitization, Vaginal Irritation, Acute Systemic Toxicity). Condom compatibility tested per ASTM D7661-10. Shelf-life validated via real-time testing. Microbial and pathogen testing performed per USP <61> and <62>.

## Technological Characteristics

Water-based lubricant. Ingredients: Water, Propylene Glycol, Dimethicone, Dimethiconol, Lauryl Glucoside, Aloe Barbadensis Leaf Powder, Vitamin E Acetate, Hydroxyethylcellulose, PEG-90M, Hydrogenated Polydecene, Acrylic Copolymer, Sodium Acrylates/Beheneth-25 Metharylate Crosspolymer, Chlorphenesin. Viscosity: 7800-8800 cps; pH: 6.8-8.0; Osmolality: 700-1000 mOsm/kg. Packaging: PET bottle with PP pump. Non-sterile.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Coconut Infused Hybrid Personal Lubricant ([K180712](/device/K180712.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 20, 2022

CC Wellness LLC Marlent Perez Quality/Regulatory Specialist 29000 N. Hancock Pkwy. Valencia, CA 91355

Re: K221908

> Trade/Device Name: JO CUMPLAY Personal Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: June 27, 2022 Received: June 30, 2022

Dear Marlent Perez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Monica D. Garcia Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221908

Device Name JO CUMPLAY Personal Lubricant

#### Indications for Use (Describe)

JO CUMPLAY Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K221908 JO CUMPLAY Personal Lubricant

#### 1. Submitter Information

Applicant: CC Wellness LLC Contact: Marlent Perez Regulatory Affairs Specialist Bruce Albert Chief Scientific Officer Address: 29000 N. Hancock Parkway Valencia, CA 91355 Phone: (661) 481-6390 Phone: (661) 295-1700, ext. 231

## 2. Correspondent Information

Contact: Marlent Perez Address: 29000 N. Hancock Parkway Valencia, CA 91355 Phone: (661) 481-6390 Email: mperez@ccwellness.com

## 3. Date prepared: August 26, 2022

## 4. Subject Device Information

Device Trade Name: JO CUMPLAY Personal Lubricant Common Name: Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Product Code: NUC (lubricant, personal) Device Class: Class II

#### 5. Predicate Device Information

Device Name: Coconut Infused Hybrid Personal Lubricant 510(k) Number: K180712 Manufacturer: United Consortium

The predicate device has not been subject to a design-related recall.

## 6. Device Description

JO CUMPLAY Personal Lubricant is a non-sterile, water-based personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This device is compatible with natural rubber latex and polyisoprene condoms and is not compatible with polyurethane condoms.

Its formulation consists of Water (Aqua), Propylene Glycol, Dimethicone, Dimethiconol, Lauryl Glucoside, Aloe Barbadensis Leaf Powder, Vitamin E Acetate, Hydroxyethylcellulose, PEG-90M, Hydrogenated Polydecene, Acrylic Copolymer, Sodium Acrylates/Beheneth-25 Metharylate Crosspolymer, and Chlorphenesin.

{4}------------------------------------------------

JO CUMPLAY Personal Lubricant is sold as an over-the-counter (OTC) product and is provided in a 4 fl. Oz./120 mL polyethylene terephthalate bottle and capped with matte black Polypropylene (PP) pump closure.

Device specifications for the JO CUMPLAY Personal Lubricant are listed in Table 1.

| Property                                              | Specification                                                                                                                                                   |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Appearance                                            | Viscous liquid                                                                                                                                                  |
| Color                                                 | White to off-white                                                                                                                                              |
| Odor                                                  | Odorless                                                                                                                                                        |
| Viscosity (per USP<911>)                              | 7800 - 8800 cps                                                                                                                                                 |
| pH (per USP<791>)                                     | 6.8 – 8.0                                                                                                                                                       |
| Osmolality (per USP<785>)                             | 700 – 1000 mOsm/kg                                                                                                                                              |
| Total Aerobic Microbial Count (TAMC, per<br>USP <61>) | <100 cfu/g                                                                                                                                                      |
| Total Yeast and Mold Count (TYMC, per<br>USP <61>)    | <10 cfu/g                                                                                                                                                       |
| Presence of Pathogens (per USP <62>)                  | Specification                                                                                                                                                   |
| Pseudomonas aeruginosa                                | Absent                                                                                                                                                          |
| Staphylococcus aureus                                 | Absent                                                                                                                                                          |
| Candida albicans                                      | Absent                                                                                                                                                          |
| Escherichia coli                                      | Absent                                                                                                                                                          |
| Salmonella/Shigella                                   | Absent                                                                                                                                                          |
| Antimicrobial Effectiveness Testing (per<br>USP <51>  | Specification                                                                                                                                                   |
| Bacteria                                              | Meets USP <51> criteria for category 2.<br>No less than 2.0 log reduction from initial<br>count at 14 days and no increase from the 14-<br>day count at 28 days |
| Yeast and Molds                                       | No increase from the initial calculated count at<br>14 and 28 days                                                                                              |

## Table 1: Device Specifications for JO CUMPLAY Personal Lubricant

## 7. Indications for Use

JO CUMPLAY Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

## 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below compares the intended use and technological characteristics of the subject and predicate device.

{5}------------------------------------------------

## Table 2: Intended Use and Technological Characteristics Comparison of the Subject and Predicate Device

|                                         | JO CUMPLAY Personal<br>Lubricant<br>K221908<br>Subject Device                                                                                                                                                                                                                                                                                                                                                              | Coconut Infused Hybrid<br>Personal Lubricant<br>K180712<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                                             |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                     | JO CUMPLAY Personal Lubricant is a<br>personal lubricant for penile, anal and/or<br>vaginal application, intended to lubricate<br>and moisturize, to enhance the ease and<br>comfort of intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is compatible<br>with natural rubber latex and<br>polyisoprene condoms. This product is<br>not compatible with polyurethane<br>condoms. | Coconut Infused Hybrid Personal<br>Lubricant is a water-based personal<br>lubricant for penile, anal and/or<br>vaginal application, intended to<br>lubricate and moisturize, to enhance<br>the ease and comfort of intimate<br>sexual activity and supplement the<br>body's natural lubrication. This<br>product is compatible with natural<br>rubber latex and polyisoprene<br>condoms. This product is not<br>compatible with polyurethane<br>condoms |
| Base type                               | Water                                                                                                                                                                                                                                                                                                                                                                                                                      | Water                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Primary ingredients                     | Water (Aqua), Propylene Glycol,<br>Dimethicone, Dimethiconol, Lauryl<br>Glucoside, Aloe Barbadensis Leaf<br>Powder, Vitamin E Acetate,<br>Hydroxyethylcellulose, PEG-90M,<br>Hydrogenated Polydecene, Acrylic<br>Copolymer, Sodium<br>Acrylates/Beheneth-25 Metharylate<br>Crosspolymer, Chlorphenesin                                                                                                                     | Water (Aqua), Propylene Glycol,<br>Caprylic/Capric Triglyceride, Cocos<br>Nucifera (Coconut) Oil, Flavor<br>(Aroma), Phenoxyethanol,<br>Polyacrylate 13, Cellulose Gum<br>(sodium carboxymethylcellulose),<br>Raphanus Sativus (Radish) Seed<br>Extract, Polyisobutene, Polysorbate<br>20, PEG-45M                                                                                                                                                      |
| Rx/OTC                                  | OTC                                                                                                                                                                                                                                                                                                                                                                                                                        | OTC                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Sterile                                 | No                                                                                                                                                                                                                                                                                                                                                                                                                         | No                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Appearance                              | Viscous liquid                                                                                                                                                                                                                                                                                                                                                                                                             | Translucent, semi-viscous cream                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Color                                   | White to off-white                                                                                                                                                                                                                                                                                                                                                                                                         | White to off-white                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Odor                                    | Odorless                                                                                                                                                                                                                                                                                                                                                                                                                   | Odorless                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Viscosity                               | 7,800 - 8800 cps                                                                                                                                                                                                                                                                                                                                                                                                           | 20,000 - 31,000 cps                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| pH                                      | 6.8 - 8.0                                                                                                                                                                                                                                                                                                                                                                                                                  | 5.7 - 6.3                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Osmolality                              | 700 - 1000 mOsm/Kg                                                                                                                                                                                                                                                                                                                                                                                                         | 450-900 mOsm/kg                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Total Aerobic Microbial<br>count (TAMC) | <100 cfu/g                                                                                                                                                                                                                                                                                                                                                                                                                 | <10 cfu/g                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Total Yeast and Mold Count<br>(TYMC)    | <10 cfu/g                                                                                                                                                                                                                                                                                                                                                                                                                  | <10 CFU/g                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Absence of Pathogenic<br>Organisms      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Antimicrobial Effectiveness<br>Tested   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Condom Compatibility                    | Compatible with natural rubber latex and polyisoprene condoms                                                                                                                                                                                                                                                                                                                                                              | Compatible with natural rubber latex and polyisoprene condoms                                                                                                                                                                                                                                                                                                                                                                                           |
| Biocompatibility Tested                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Shelf life                              | 2.5 years                                                                                                                                                                                                                                                                                                                                                                                                                  | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{6}------------------------------------------------

The subject and predicate device indications for use and intended use are the subject and predicate device have different technological characteristics, including formulations, shelf-life, and specifications for appearance, TAMC, viscosity, pH, and osmolality. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.

#### 9. Summary of Non-Clinical Performance Testing

#### Biocompatibility

Biocompatibility testing was performed in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process. " The following testing was conducted:

- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010) ●
- Vaginal Irritation (ISO 10993-10:2010) ●
- Acute Systemic Toxicity (ISO 10993-11:2017) ●

The results of testing support the biocompatibility of the device materials.

#### Shelf-Life

The subject device has a shelf-life of 2.5 years. Results from real-time testing demonstrated that the device maintains its specifications (as shown in Table 1) over the duration of its shelf-life.

#### Condom Compatibility

The compatibility of JO CUMPLAY Personal Lubricant with condoms was evaluated in accordance with ASTM D7661-10(R) 2017 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms." The results of this test showed JO CUMPLAY Personal Lubricant to be compatible with natural rubber latex and polyisoprene condoms. Results showed JO CUMPLAY Personal Lubricant not to be compatible with polyurethane condoms.

#### 10. Conclusion

The results of the performance testing described above demonstrate that JO CUMPLAY Personal Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K221908](https://fda-staging.innolitics.com/device/K221908)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).