← Product Code [MHX](/productcode/MHX) · K221904

# EK12 V2 Algorithm (K221904)

_Ge Medical Systems Information Technologies, Inc. · MHX · Dec 20, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221904

## Device Facts

- **Applicant:** Ge Medical Systems Information Technologies, Inc.
- **Product Code:** [MHX](/productcode/MHX.md)
- **Decision Date:** Dec 20, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device, Pediatric

## Indications for Use

EK12 v2 analyzes 10 or more seconds of a previously acquired electrocardiogram (ECG) from physiological ECG signal recording devices for rhythm and measurements. EK12 is used to create reports intended for use by a Qualified Medical Professional, including a trained ECG Technician operating within Independent Diagnostic Testing Facility (IDTF) requirements and performance standards for the review and assessment of an ECG. EK12 is indicated for use on adults and pediatric patients older than 2 years. The device is intended for use in an IDTF or a professional medical facility, such as a hospital, clinic, or physician's office.

## Device Story

Software-only algorithm; analyzes 10+ seconds of previously acquired ECG signals; performs rhythm analysis and measurements; generates reports for qualified medical professionals/technicians. Used in IDTFs, hospitals, clinics, or physician offices. Operates as part of a host system; outputs include text files with pass/fail status, rhythm/measurement evaluations, event logs (arrhythmias/noise), and QRS detection annotations. Does not perform active patient monitoring; no real-time alarms or alerts. Benefits include automated ECG interpretation and standardized reporting to support clinical decision-making.

## Clinical Evidence

No clinical studies were required. Substantial equivalence is supported by non-clinical bench testing, including software verification, performance testing, and compliance with AAMI/ANSI EC57:2012 for cardiac rhythm and ST-segment measurement algorithms.

## Technological Characteristics

Software-only algorithm; utilizes 12SL ECG analysis technology. Inputs: ECG signal data (WFDB format), optional patient age/gender, optional lead names. Outputs: Tab-delimited text files and binary WFDB annotation files. Complies with AAMI/ANSI EC57:2012. Designed for integration into host ECG recording systems.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- Ek12 Algorithm ([K170155](/device/K170155.md))

## Reference Devices

- 12SL ECG Analysis Program ([K141963](/device/K141963.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

December 20, 2022

GE Medical Systems Information Technologies, Inc. Shlomi Deler Regulatory Affairs Director 9900 Innovation Drive Wauwatosa, Wisconsin 53226

Re: K221904

Trade/Device Name: EK12 V2 Algorithm Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) Regulatory Class: Class II Product Code: MHX Dated: November 21, 2022 Received: November 22, 2022

Dear Shlomi Deler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shruti N. Mistry -S

for

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221904

Device Name EK12 V2 Algorithm

#### Indications for Use (Describe)

EK12 V2 analyzes 10 or more seconds of a previously acquired electrocardiogram (ECG) from physiological ECG signal recording devices for rhythm and measurements.

EK12 is used to create reports intended for use by a Qualified Medical Professional, including a trained ECG Technician operating within Independent Diagnosic Testing Facility (IDTF) requirements and performance standards for the review and assessment of an ECG.

EK 12 V2 is indicated for use on adults and pediatric patients older than 2 years. The device is intended for use in an IDTF or a professional medical facility, such as a hospital, clinic, or physician's office.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular frame. The frame has wave-like designs on the top and bottom, giving it a dynamic appearance. The logo is presented in a blue color.

## 510(k) Summary

|                                              | In accordance with 21 CFR 807.92 the following summary of information is provided:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:                                        | 20-June-2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Submitter:                                   | GE Medical Systems Information Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                              | 9900 Innovation Drive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                              | Wauwatosa, WI 53226                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Primary Contact Person:                      | Shlomi Deler                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                              | Regulatory Affairs Director                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                              | GE Medical Systems Information Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                              | Phone: +(972) 544301056                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Secondary Contact Person:                    | Michael Petrini                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                              | Regulatory Affairs Executive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                              | GE Medical Systems Information Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                              | Phone: +1(414)581-8987                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Trade Name:                           | EK12 V2 Algorithm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common/Usual Name:                           | Arrhythmia Detection Algorithm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification Names:                        | Monitor, Physiological, Patient (With Arrhythmia Detection or<br>Alarms)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code:                                | MHX , 21 CFR 870.1025                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Predicate Device(s):                         | Ek12 Algorithm (K170155)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Reference Device:                            | 12SL ECG Analysis Program (K141963)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Description:                          | EK12 V2 algorithm is a software only product that provides<br>computerized measurements from the ECG parameter data<br>acquired by the host device. EK12 V2 analyzes ECG recordings for<br>rhythm and measurements that is deployed as part of a host<br>system used to generate ECG reports                                                                                                                                                                                                                                                                                                                                                                                               |
| Intended Use:                                | EK12 v2 analyzes 10 or more seconds of a previously acquired<br>electrocardiogram (ECG) from physiological ECG signal recording<br>devices for rhythm and measurements.<br>EK12 is used to create reports intended for use by a Qualified<br>Medical Professional, including a trained ECG Technician operating<br>within Independent Diagnostic Testing Facility (IDTF) requirements<br>and performance standards for the review and assessment of an<br>ECG.<br>EK12 is indicated for use on adults and pediatric patients older than<br>2 years.<br>The device is intended for use in an IDTF or a professional medical<br>facility, such as a hospital, clinic, or physician's office. |
| Technology:                                  | The EK12 V2 Algorithm employs the same fundamental scientific<br>technology as its predicate and the reference devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Determination of Substantial<br>Equivalence: | The EK12 V2 Algorithm is substantially equivalent to the predicate<br>Ek12 Algorithm (K170155) and the reference device 12SL ECG<br>Analysis Program (K141963) as described in the table below:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. There are three water droplet-like shapes evenly spaced around the circle's perimeter, giving the logo a distinctive and recognizable appearance.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and the circle has a white border with a swirling pattern. The logo is simple and recognizable, and it is associated with a well-known company.

| Feature /<br>Function                           | Proposed Device<br>EK12v2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>EK12v1 (K170155)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Change Explanation /<br>Notes |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Intended Use                                    | EK12 analyzes 10 or more seconds of a<br>previously acquired electrocardiogram (ECG)<br>from physiological ECG signal recording<br>devices for rhythm and measurements.<br><br>EK12 is used to create reports intended for<br>use by a Qualified Medical Professional,<br>including a trained ECG Technician operating<br>within Independent Diagnostic Testing Facility<br>(IDTF) requirements and performance<br>standards for the review and assessment of<br>an ECG.<br><br>EK12 is indicated for use on adults and<br>pediatric patients older than 2 years.<br><br>The device is intended for use in an IDTF or a<br>professional medical facility, such as a<br>hospital, clinic, or physician's office. | EK12 analyzes ten or more seconds of a<br>previously acquired electrocardiogram (ECG)<br>from physiological ECG signal recording<br>devices for rhythm and measurements.<br><br>EK12 is used to create reports intended for<br>use by a Qualified Medical Professional,<br>including a trained ECG Technician operating<br>within Independent Diagnostic Testing Facility<br>(IDTF) requirements and performance<br>standards for the review and assessment of<br>an ECG.<br><br>EK12 is indicated for use on adults and<br>pediatric patients older than 2 years.<br><br>The device is intended for use in an IDTF or a<br>professional medical facility, such as a<br>hospital, clinic, or doctor's office. | Identical                     |
| Indications for<br>Use                          | EK12 is software algorithm that analyzes ECGs<br>for rhythm and measurements that is<br>deployed as part of a system used to generate<br>ECG reports / findings.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | EK12 is software algorithm that analyzes ECGs<br>for rhythm and measurements that is<br>deployed as part of a system used to generate<br>ECG reports / findings.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Identical                     |
| Contraindications                               | EK12 cannot be used for active patient<br>monitoring since it can only analyze<br>prerecorded ECG signals that are at least 10<br>seconds long. It does not provide any time-<br>sensitive information, continuous display of<br>information, alarms, or alerts intended to<br>alert a caregiver to take an immediate clinical<br>action.                                                                                                                                                                                                                                                                                                                                                                       | EK12 cannot be used for active patient<br>monitoring since it can only analyze<br>prerecorded ECG signals that are at least 10<br>seconds long. It does not provide any time-<br>sensitive information, continuous display of<br>information, alarms, or alerts intended to<br>alert a caregiver to take an immediate clinical<br>action.                                                                                                                                                                                                                                                                                                                                                                     | Identical                     |
| Patient<br>Population                           | EK12 is indicated for use on adults and<br>pediatric patients older than 2 years.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | EK12 is indicated for use on adults and<br>pediatric patients older than 2 years.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                     |
| Use Environment                                 | The device is intended for use in an IDTF or a<br>professional medical facility, such as a<br>hospital, clinic, or physician's office.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The device is intended for use in an IDTF or a<br>professional medical facility, such as a<br>hospital, clinic, or doctor's office.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                     |
| Result                                          | An output text file containing the pass/fail<br>status (result) of the analysis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | An output text file containing the pass/fail<br>result of the analysis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                     |
| Final report                                    | An output text file containing the final<br>evaluation results for rhythms,<br>measurements, and other information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | An output text file containing the final<br>evaluation results for rhythms,<br>measurements, and other information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                     |
| Event log<br>detected<br>arrhythmias,<br>noise) | A column-oriented tab-delimited output text<br>file annotating the beginning and end of all<br>detected rhythm and noise events.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | A column-oriented tab-delimited output text<br>file annotating the beginning and end of all<br>detected rhythm and noise events.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Identical                     |
| QRS detections                                  | Two output files:<br>1. A column-oriented tab-delimited output<br>text file listing annotations of QRS detection<br>times and beat classification.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Two output files:<br>1. A column-oriented tab-delimited output<br>text file listing annotations of QRS detection<br>times and beat classification.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Identical                     |
| Feature / Function                              | Proposed Device<br>EK12v2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>EK12v1 (K170155)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Change Explanation / Notes    |
|                                                 | 2. A binary file in the WFDB annotation file<br>format containing QRS detection times and<br>beat classification.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 2. A binary file in the WFDB-format<br>annotation file format containing QRS<br>detection times and beat classification.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                               |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circular border. The color of the logo is a light blue.

# GE Healthcare 510(k) Premarket Notification Submission

### Substantial Equivalence to the reference device, 12SL ECG Analysis Program (K141963):

| Feature /<br>Function                   | Proposed:<br>EK12v2                                                                                                                                                                                                                                                                                                                                                                                                             | Reference device: 12SL ECG Analysis<br>Program (K141963)                                                                                                                                                                                                                                                                                                                                           | Discussion of Differences                                                                                                                                                                                                                                                                                                                                                                                             |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient age and<br>gender               | Optional inputs. If present, they<br>are contained within the WFDB<br>input record. When present,<br>they are validated and passed<br>directly to the 12SL analysis. If<br>not present, the value for<br>unknown age and/or gender are<br>provided to the 12SL analysis                                                                                                                                                         | Required inputs which are used in the ECG<br>interpretation.<br>Age is provided to the program in years,<br>an encoded value for age less than one<br>year, or an encoded value for age<br>unknown.<br>Gender is provided to the program as<br>one of male, female, or unknown.                                                                                                                    | Substantial equivalent<br>The proposed EK12 V2 adds support for the<br>optional input of patient age and/or gender.                                                                                                                                                                                                                                                                                                   |
| Identification of<br>ECG leads          | Lead names are an optional<br>component of the input WFDB<br>input record and are an optional<br>input of EK12 V2. Any input<br>signals containing a lead name<br>from the set {I, II, V1, V2, V3, V4,<br>V5, V6} will be treated as that<br>lead in the internal 12SL analysis                                                                                                                                                 | Assumes the first 8 channels are ECG<br>leads I, II, V1, V2, V3, V4, V5, V6.<br>Measurements are provided for leads<br>after the first 8 channels but are<br>otherwise not used in any interpretation<br>statements with the exception of lead<br>V4r, which if present, will be examined for<br>right ventricular involvement (ST<br>elevation) if an acute inferior infarct is<br>detected       | Substantial equivalent<br>The proposed EK12 V2 adds support for the<br>optional identification of ECG leads.                                                                                                                                                                                                                                                                                                          |
| 12SL rhythm<br>statement<br>trends      | A column-oriented tab-delimited<br>text file that provides the 12SL<br>interpretation statement codes<br>for rhythm statements from each<br>10-second 12SL analysis at<br>intervals of every 5 seconds. That<br>is, each consecutive analysis has<br>5 seconds of overlapping<br>waveform.<br>The 12SL interpretation<br>statement codes are censored to<br>remove all statement codes not<br>related to rhythm interpretation. | All output is contained in an in-memory<br>data structure; the host product is<br>responsible for creation of the output<br>record and/or report.  There is no concept<br>of "trends" in a single execution of the<br>12SL analysis.<br>The method of creating the 12SL<br>statements are identical to the reference<br>device for the "12SL interpretation<br>statement codes" as described above | Substantial equivalent<br>Interpretation statement codes forming the<br>overall ECG interpretation is identical to the<br>reference 12SL ECG Analysis Program<br>(K141963).<br>The proposed EK12 v2 adds the export of<br>the rhythm statement codes to an output<br>file.                                                                                                                                            |
| Median<br>complex<br>waveform<br>trends | A column-oriented tab-delimited<br>text file that provides the median<br>complexes for each supplied lead<br>from each 10-second 12SL<br>analysis at intervals of every 5<br>seconds. That is, each<br>consecutive analysis has 5<br>seconds of overlapping<br>waveform                                                                                                                                                         | All output is contained in an in-memory<br>data structure; the host product is<br>responsible for creation of the output<br>record and/or report.  There is no concept<br>of "trends" in a single execution of the<br>12SL analysis.<br>The method of creating the 12SL median<br>complexes are identical to the reference<br>device for the "12SL median complexes"<br>as described above         | Substantial equivalent<br>Median complexes were formed as part of<br>the internal 12SL analysis and were the basis<br>for the per-lead measurements provided in<br>the predicate EK12 v1 (K170155). The<br>method for median waveform formation is<br>identical to the reference device 12SL ECG<br>Analysis Program (K141963).<br>The proposed EK12 v2 adds the export of<br>the median complexes to an output file. |

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. There are also several white teardrop shapes surrounding the circle, giving the impression of movement or energy.

| Performance standards:                                 | The EK12 V2 Algorithm complies with the<br>voluntary consensus standard AAMI/ANSI EC57: 2012 –<br>Testing And Reporting Performance Results Of Cardiac<br>Rhythm and St-Segment Measurement Algorithms.                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Determination of<br>Substantial Equivalence:<br>tested | Summary of Non-Clinical Tests<br>The EK12 V2 Algorithm program was designed and<br>for compliance with applicable clauses of the following<br>voluntary standard:<br><br>AAMI/ANSI EC57: 2012 – Testing And Reporting<br>Performance Results Of Cardiac Rhythm and St-Segment<br>Measurement Algorithms. (Cardiovascular)<br><br>The following quality assurance measures were applied to<br>the development of the system:<br>- Risk Analysis<br>- Requirements Reviews<br>- Code Inspections<br>- Software Verification Testing<br>- Performance testing |
| Summary of Clinical Tests:                             | The subject of this premarket submission, EK12 V2<br>Algorithm, did not require clinical studies to support<br>substantial equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Conclusion:                                            | GE Healthcare considers the EK12 V2 Algorithm to be as<br>safe, as effective, and performance is substantially<br>equivalent to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                      |

---

**Source:** [https://fda-staging.innolitics.com/device/K221904](https://fda-staging.innolitics.com/device/K221904)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).