Prospective multi-center clinical study (longitudinal and cross-sectional cohorts); Clinical structural disease status (ultrasound, imaging, biopsy)
The study used a longitudinal cohort of patients with differentiated thyroid cancer to establish real-world structural disease prevalence, which was then used to adjust the clinical performance estimates (NPV and PPV) for the device.
Prospective, multi-center study of Elecsys Tg II; Prospective, multi-center, longitudinal and cross-sectional cohorts; Follow-up/Duration: 24 months
Patients with differentiated thyroid cancer (DTC) who have had thyroid surgery; Sample Size: 530 samples (461 longitudinal, 69 cross-sectional); Number of Sites: 9
Not applicable for this study
Sensitivity, Specificity, NPV, PPV relative to structural disease status
Indications for Use
Immunoassay for the in vitro quantitative determination of thyroglobulin in human serum and plasma. Determination of Tg is used as an aid in monitoring for the presence of persistent or recurrent/metastatic disease in patients who have differentiated thyroid cancer (DTC) and have had thyroid surgery (with or without ablative therapy). The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers.
Device Story
Elecsys Tg II is an in vitro diagnostic immunoassay for quantitative thyroglobulin (Tg) measurement in human serum/plasma. Used on cobas e 411 analyzers in clinical laboratories. Principle: two-step, double-antigen sandwich electrochemiluminescence immunoassay (ECLIA). Input: patient serum/plasma sample; biotinylated anti-Tg antibody; ruthenium-labeled anti-Tg antibody; streptavidin-coated microparticles. Process: sandwich complex formation; magnetic capture on electrode; voltage application induces chemiluminescence measured by photomultiplier. Output: Tg concentration (ng/mL) derived from instrument-specific calibration curve. Results aid clinicians in monitoring DTC patients for persistent or recurrent disease by comparing Tg levels against clinical cut-offs (e.g., 0.2 ng/mL). Benefits include standardized monitoring of disease status post-thyroidectomy.
Clinical Evidence
Prospective multi-center study (N=530 samples) evaluated clinical performance against structural disease (SD) status over 24 months. Sensitivity 98.91% (95% CI: 94.10-99.81%), specificity 53.42% (95% CI: 48.74-58.05%), NPV 99.89%, PPV 10.03% at 0.2 ng/mL cut-off. Study included longitudinal and cross-sectional cohorts. Results support use of Tg levels to categorize response to therapy (Excellent, Indeterminate, Biochemical Incomplete).
Technological Characteristics
Sandwich immunoassay using biotinylated monoclonal anti-Tg antibody and ruthenium-labeled monoclonal anti-Tg antibody. Detection via electrochemiluminescence (ECLIA) on cobas e analyzers. Measuring range 0.1–500 ng/mL. Standardized against CRM 457. Reagents include streptavidin-coated microparticles, biotinylated antibody, and ruthenium-labeled antibody. Compatible with serum, Li-Heparin, K2-EDTA, and K3-EDTA plasma. Software-controlled calibration via master curve and 2-point calibration.
Indications for Use
Indicated for monitoring persistent, recurrent, or metastatic disease in patients ≥ 22 years old with differentiated thyroid cancer (DTC) post-thyroid surgery. Contraindicated in patients with medullary or anaplastic thyroid carcinoma, TSH-secreting pituitary adenomas, or presence of thyroglobulin autoantibodies.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
Predicate Devices
Access® Thyroglobulin Reagents on the Access® Immunoassay Systems (k002905)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K221890
B Applicant
Roche Diagnostics
C Proprietary and Established Names
Elecsys Tg II
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MSW | Class II | 21 CFR 866.6010 – Tumor Associated Antigen Immunological Test System | IM - Immunology |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Thyroglobulin
C Type of Test:
Quantitative, Electrochemiluminescence immunoassay (ECLIA)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K221890 - Page 2 of 17
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
Immunoassay for the in vitro quantitative determination of thyroglobulin in human serum and plasma. Determination of Tg is used as an aid in monitoring for the presence of persistent or recurrent/metastatic disease in patients who have differentiated thyroid cancer (DTC) and have had thyroid surgery (with or without ablative therapy).
The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
In Vitro Diagnostic Use Only
The Instruction for Use of the Elecsys Tg II contains the following blackbox statement:
Thyroglobulin (Tg) determinations can be affected by the presence of Tg autoantibodies (anti-Tg) in some patient samples. These autoantibodies may interfere with the assay used to measure Tg, causing false high or false low Tg values.
The measured Tg value of a patient’s sample can also vary depending on the assay used. The laboratory finding must therefore always contain a statement on the Tg assay method used. Tg values determined on patient samples by different assays cannot be directly compared with one another and could be the cause of erroneous medical interpretations. If there is a change in the Tg assay procedure used while patient monitoring, the Tg values obtained upon changing to the new procedure must be confirmed by parallel measurements with both methods.
### D Special Instrument Requirements:
cobas e 411 analyzer (cleared under K062279)
## IV Device/System Characteristics:
### A Device Description:
Materials included in the Elecsys TgII assay
Elecsys Tg II consists of the reagent rack-pack. Each reagent rack-pack is for 100 tests and contains:
- M (6.5 mL): Streptavidin-coated microparticles (transparent cap) 0.72 mg/mL and preservative.
{2}
- R1 (9 mL): Biotinylated monoclonal anti Tg antibody (mouse) (gray cap) 1 mg/L in Bis Tris buffer 50 mmol/L, pH 6.3 and preservative.
- R2 (9 mL): Monoclonal anti Tg antibodies (mouse) labeled with ruthenium complex (black cap) 3.1 mg/L in Bis Tris buffer 50 mmol/L, pH 6.3 and preservative.
## Materials required (but not supplied)
- Tg II CalSet (4 x 1.0 mL): 2 vials each of Tg at 0.15 ng/mL, and 180 ng/mL
- PreciControl Universal (4 x 3.0 mL): 2 vials each of Tg at 25 ng/mL, and 100 ng/mL
- PreciControl TS (4 x 2.0 mL): 4 vials of 1 ng/mL
- Diluent MultiAssay (2 x 16 mL): sample diluent
- Anti-Tg assay: to verify the presence of antibodies to Tg in patient samples
- Distilled or deionized water
## B Principle of Operation:
The Elecsys Tg II assay is a quantitative, two-step, double antigen sandwich assay. In the first incubation, thyroglobulin (Tg) in the patient sample reacts with a biotinylated monoclonal Tg-specific antibody and monoclonal Tg-specific antibodies labeled with a ruthenium complex to form a sandwich complex. After addition of streptavidin-coated microparticles, the complex bounds to the solid phase via interaction of biotin and streptavidin in second incubation. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell/ProCell M. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier. Results for the Tg concentration of samples are determined via a calibration curve, which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent barcode.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
Access Thyroglobulin Reagents On The Access Immunoassay Systems
### B Predicate 510(k) Number(s):
K002905
### C Comparison with Predicate(s):
K221890 - Page 3 of 17
{3}
K221890 - Page 4 of 17
| Device & Predicate Device(s): | K221890 Candidate Device | K002905 Predicate |
| --- | --- | --- |
| Device Trade Name | Elecsys Tg II | Access Thyroglobulin |
| **General Device Characteristic Similarities** | | |
| Intended Use / Indications For Use | Immunoassay for the in vitro quantitative determination of thyroglobulin in human serum and plasma. Determination of Tg is used as an aid in monitoring for the presence of persistent or recurrent/metastatic disease in patients who have differentiated thyroid cancer (DTC) and have had thyroid surgery (with or without ablative therapy). The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers. | The Access Thyroglobulin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin levels in human serum and plasma, using Access Immunoassay Systems. This device is intended to aid in the monitoring for the presence of local and metastatic thyroid tissue in patients who have had thyroid gland ablation (using thyroid surgery with or without radioactivity) and who lack serum thyroglobulin antibodies. |
| Analyte | Thyroglobulin | Same |
| Test format | Quantitative | Same |
| Traceability / Standardization | Standardized against CRM (Certified Reference Material) 457 of the BCR (Community Bureau of Reference) of the European Union. | Calibrators are traceable to the European Community Bureau of Reference (BCR) CRM 457 thyroglobulin standard |
| Measuring Range | 0.1 ng/mL – 500 ng/mL | Same |
| **General Device Characteristic Differences** | | |
| Sample Type / Matrix | Serum and Plasma (Li-Heparin, K2-EDTA, and K3-EDTA) | Serum and Plasma (heparinized) |
| Calibrator | Tg II CalSet (0.15, and 180 ng/mL) | Access Thyroglobulin Calibrators (0, 1.0, 10, 100, 250, and 500 ng/mL) |
| Calibration Interval | • 28 days when using the same reagent lot
• 7 days (when using the same reagent kit on the analyzer)
• as required: e.g., quality controls outside the defined limits | Calibration is required every 56 days or whenever new lot numbers of reagents are placed into use |
| Controls | • PreciControl Universal
• PreciControl TS
• Other suitable control material | Commercial Control Materials |
| Reagent Stability | • Up to the stated expiration date (unopened at 2–8°C)
• 84 days (12 weeks) after opening, stored at 2–8°C | • Up to stated expiration date (unopened at 2–10°C)
• 28 days (4 weeks) after opening, stored as 2–10°C |
{4}
K221890 - Page 5 of 17
| Detection Limit | LoB: 0.02 ng/mL
LoD: 0.04 ng/mL
LoQ: 0.1 ng/mL | LoB: 0.03 ng/mL
LoD: Not available
LoQ: 0.1 ng/mL |
| --- | --- | --- |
VI Standards/Guidance Documents Referenced:
The following Clinical and Laboratory Standards Institute (CLSI) guidelines were used:
- CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures
- CLSI EP06 2nd Edition, Evaluation of Linearity of Quantitative Measurement Procedures
- CLSI EP07 3rd Edition, Interference Testing in Clinical Chemistry
- CLSI EP09c 3rd Edition, Measurement Procedure Comparison and Bias Estimation Using Patient Samples
- CLSI EP12 3rd Edition, Evaluation of Qualitative, Binary Output Examination Performance
- CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
- CLSI EP37 1st Edition, Supplemental Tables for Interference Testing in Clinical Chemistry
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
The precision of the Elecsys Tg II assay was evaluated according the CLSI EP05-A3 guideline.
Within-Laboratory Precision:
A panel of six native serum samples (Sample #1–6) and one pooled serum sample (Sample #7) and three controls (PreciControl Universal 1 and 2 (PCU1 and PCU2), and PreciControl TS (PC TS)) were tested on one cobas e 411 analyzer using one reagent lot. Each sample was tested with two replicates per run, two runs per day for 21 days, yielding a total of 84 measurements.
The data were analyzed for repeatability (within-run), between-run, between-day, and within-laboratory precision. The mean (ng/mL), standard deviation (SD) (ng/mL) and percent coefficient of variation (%CV) are summarized in the table below:
{5}
| Sample | N | Mean (ng/mL) | Repeatability (Within Run) | | Between-Run | | Between-Day | | Within-Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 84 | 0.13 | 0.01 | 7.56 | 0.00 | 0.00 | 0.01 | 5.49 | 0.01 | 9.34 |
| 2 | 84 | 0.15 | 0.01 | 5.86 | 0.00 | 0.00 | 0.01 | 6.50 | 0.01 | 8.75 |
| 3 | 84 | 0.23 | 0.01 | 4.00 | 0.00 | 1.81 | 0.01 | 3.59 | 0.01 | 5.67 |
| 4 | 84 | 1.67 | 0.05 | 2.67 | 0.02 | 0.91 | 0.03 | 1.50 | 0.05 | 3.20 |
| 5 | 84 | 38.70 | 0.50 | 1.30 | 0.00 | 0.00 | 0.46 | 1.19 | 0.68 | 1.76 |
| 6 | 84 | 235 | 4.52 | 1.92 | 0.00 | 0.00 | 2.90 | 1.23 | 5.37 | 2.28 |
| 7 | 84 | 459 | 7.76 | 1.69 | 5.00 | 1.09 | 4.39 | 0.96 | 10.20 | 2.23 |
| PCU 1 | 84 | 24.10 | 0.27 | 1.11 | 0.04 | 0.16 | 0.31 | 1.28 | 0.41 | 1.70 |
| PCU 2 | 84 | 84.40 | 0.87 | 1.02 | 0.51 | 0.60 | 1.03 | 1.22 | 1.44 | 1.71 |
| PC TS | 84 | 1.05 | 0.02 | 1.55 | 0.00 | 0.00 | 0.02 | 1.79 | 0.03 | 2.37 |
Lot-to-lot imprecision and reproducibility:
Testing was performed on three cobas e 411 analyzers (at three external sites) with three reagent lots. Two of three lots were distributed to each of the three sites as below:
| Testing Site | Lot 1 | Lot 2 | Lot 3 |
| --- | --- | --- | --- |
| Site 1 | | x | x |
| Site 2 | x | | x |
| Site 3 | x | x | |
Seven serum samples (Sample 1-5: native serum samples, Sample 6-7: pooled native sera) and three controls (PCU1, PCU2, and PC TS) were tested in five replicates per run, one run per day, for five days, yielding $N = 25$ datapoints per sample per lot and $N = 50$ datapoints per sample at each site. The results are summarized in the following table:
| Sample | N* | Mean (ng/mL) | Within-Run | | Between-Day | | Between-Lot | | Between-Site | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 148 | 0.22 | 0.02 | 7.56 | 0.01 | 6.66 | 0.01 | 6.49 | 0.01 | 6.45 | 0.03 | 13.61 |
| 2 | 148 | 0.32 | 0.02 | 6.62 | 0.02 | 7.14 | 0.01 | 3.77 | 0.02 | 6.88 | 0.04 | 12.50 |
| 3 | 146 | 2.00 | 0.08 | 3.79 | 0.09 | 4.52 | 0.01 | 0.62 | 0.07 | 3.34 | 0.14 | 6.81 |
| 4 | 148 | 39.70 | 1.26 | 3.17 | 2.02 | 5.08 | 0.54 | 1.36 | 1.53 | 3.86 | 2.88 | 7.26 |
| 5 | 149 | 261 | 10.97 | 4.20 | 12.10 | 4.64 | 8.02 | 3.07 | 13.43 | 5.15 | 22.62 | 8.67 |
| 6 | 148 | 407 | 13.95 | 3.43 | 17.79 | 4.37 | 10.62 | 2.61 | 20.39 | 5.01 | 32.24 | 7.92 |
| 7 | 149 | 464 | 17.55 | 3.79 | 19.27 | 4.16 | 0.00 | 0.00 | 18.47 | 3.98 | 31.95 | 6.89 |
| PCU 1 | 149 | 24.20 | 1.16 | 4.80 | 1.79 | 7.43 | 0.89 | 3.70 | 1.26 | 5.2 | 2.63 | 10.91 |
| PCU 2 | 148 | 92.99 | 4.31 | 4.63 | 4.62 | 4.97 | 3.32 | 3.57 | 3.10 | 3.34 | 7.78 | 8.36 |
| PC TS | 149 | 1.08 | 0.03 | 3.15 | 0.05 | 4.23 | 0.03 | 3.08 | 0.02 | 1.57 | 0.07 | 6.30 |
* Reason for N being less than 150: Failed quality control, handling errors, or flag from instrument; Outliers were removed for analysis as per CLSI EP05-A3 guideline.
K221890 - Page 6 of 17
{6}
K221890 - Page 7 of 17
## 2. Linearity:
The linearity of the Elecsys Tg II assay was assessed based on CLSI document EP06 2nd Edition. Linearity series were prepared by mixing High and low native serum samples to create 13 levels that span across the analytical measuring interval (AMI). Each level was measured with four replicates using three lots of reagents on the cobas e 411 analyzer. For each level, the mean value of the measured values, predicted value and the deviation from linearity were calculated. Predicted values were calculated using a best fitted straight line by a weighted least squares linear regression analysis. Percent deviations from linearity were calculated as differences between the observed values (mean values) and the predicted values divided by the predicted values. The table below summarizes the linearity data analysis:
| Lot (Sample) | Range (ng/mL) | Slope | Intercept | %Deviation* | Deviation** |
| --- | --- | --- | --- | --- | --- |
| 1 (1) | 0.029 – 568 | 0.926 | -0.009 | -4.4% – 8.0% | -0.0004 – 0.017 |
| 1 (2) | 0.006 – 506 | 0.917 | 0.007 | -9.7% – 9.5% | -0.013 – 0.0002 |
| 2 (1) | 0.022 – 578 | 0.924 | -0.004 | -6.6% – 8.2% | 0.003 – 0.005 |
| 2 (2) | 0.005 – 516 | 0.925 | 0.006 | -7.7% – 8.2% | -0.015 – 0.0001 |
| 3 (1) | 0.035 – 567 | 0.932 | 0.005 | -6.1% – 7.3% | -0.003 – 0.0005 |
| 3 (2) | 0.089 – 521 | 0.936 | -0.015 | -8.8% – 6.8% | 0.002 – 0.026 |
* % Deviation from Linearity for concentrations ≥0.3 ng/mL
** Deviation (ng/mL) from Linearity for concentrations <0.3 ng/mL
The data support the linearity interval from 0.005 to 578 ng/mL with deviations from linearity within ±10% for values ≥ 0.3 ng/mL and ±0.03 ng/mL for values < 0.3 ng/mL. The study results support the linearity of the claimed analytical measuring interval (AMI): 0.1 – 500 ng/mL.
## High Dose Hook Effect:
The high-dose hook effect of the Elecsys Tg II assay was assessed using three samples spiked with purified human Thyroglobulin up to concentrations at 226,021, 229,820, and 244,927 ng/mL. Each sample was diluted with Diluent MultiAssay to produce a set of 11 samples with concentrations spanning between 55.1 and 226,021, 56.1 and 229,820, and 59.8 and 244,927 ng/mL. All samples were tested in singleton using the Elecsys Tg II on one cobas e 411 analyzer. The hook concentration reported corresponds to the highest analyte concentration that generates a signal ≥ 10% above the upper limit of the measuring range. No hook effect was observed for the Elecsys Tg II up to 120,000 ng/mL.
## 3. Analytical Specificity/Interference:
The studies were performed to evaluate potential interference and cross reactivity of the Elecsys Tg II assay following the CLSI EP07 (3rd Edition) and EP37 (1st Edition) guidelines.
## Endogenous Interfering Substances:
Potential endogenous interfering substances were tested for their ability to interfere with the Elecsys Tg II assay using three serum samples with thyroglobulin concentration around 0.2
{7}
ng/mL, 2 ng/mL, and 40 ng/mL. For each interfering substance, test samples were spiked with the test substance and results were compared to control samples spiked with an equal volume of solvent. All samples were measured in five replicates using one Elecsys Tg II assay on one cobas e 411 analyzer. The interference was calculated by comparing the mean measurements of the test and control samples. No-significant interference (≤ ±10% difference of test from control) were observed for the Elecsys Tg II up to the following concentrations for each endogenous substance tested:
| Endogenous Substance | Concentration |
| --- | --- |
| Bilirubin* | 66 mg/dL |
| Hemoglobin | 600 mg/dL |
| Intralipid | 2,000 mg/dL |
| Biotin | 1,200 ng/mL |
| IgG | 2 g/dL |
| Albumin | 7 mg/dL |
*90% unconjugated + 10% conjugated Bilirubin
## Exogenous Interfering Substances:
Potential exogenous interfering substances (drugs) were tested for their ability to interfere with the Elecsys Tg II assay using three serum samples with thyroglobulin concentrations around 0.2 ng/mL, 2 ng/mL, and 50 ng/mL. For each interfering substance, test samples were spiked with the test substance and results were compared to control samples spiked with an equal volume of solvent (blank). All samples were measured in five replicates using one Elecsys Tg II assay on one cobas e 411 analyzer. The interference was calculated by comparing measurements of the test and control samples. No significant interference (≤ ±10% difference of test from control) for the Elecsys Tg II was observed up to the following concentrations of the potential exogenous interfering substances tested:
| Exogenous Substance | Concentration | Exogenous Substance | Concentration |
| --- | --- | --- | --- |
| Acetaminophen | 156 mg/L | Lenvatinib Mesylate | 0.15 mg/L |
| Acetylcysteine | 150 mg/L | Levodopa | 7.5 mg/L |
| Acetylsalicylic Acid | 30 mg/L | Methyldopa +1.5 | 22.5 mg/L |
| Amiodarone | 200 mg/L | Metronidazole | 123 mg/L |
| Ampicillin-Na | 75 mg/L | Octreotide | 0.3 mg/dL |
| Ascorbic acid | 52.5 mg/L | Perchlorate | 2,000 mg/L |
| Cabozantinib-S- | 4.14 mg/L | Phenylbutazone | 107* mg/L |
| Carbimazole | 30 mg/L | Prednisolone | 100 mg/L |
| Cefoxitin | 750 mg/L | Propranolol | 240 mg/L |
| Cyclosporine | 1.8 mg/L | Propylthiouracil | 300 mg/L |
| Doxycycline | 6* mg/L | Rifampicin | 16* mg/L |
| Fluocortolone | 100 mg/L | Theophylline | 60 mg/L |
| Heparin | 3,300 IU/L | Thiamazole | 80 mg/L |
| Hydrocortisone | 200 mg/L | D-T3 | 0.5 mg/L |
| Ibuprofen | 219 mg/L | L-T3 | 0.5 mg/L |
| Iodide | 0.2 mg/L | D-T4 | 5 mg/L |
| Itraconazole | 10* mg/L | L-T4 | 5 mg/L |
*Corresponds to one-time maximum daily dose
K221890 - Page 8 of 17
{8}
K221890 - Page 9 of 17
# Cross-reactivity:
Each potential cross-reacting substance was tested with serum samples at two thyroglobulin concentrations: 5 ng/mL and 50 ng/mL. Test samples were prepared by spiking with the potential cross-reacting substances. Results were compared to control samples which were spiked with an equal volume of solvent (blank) where appropriate. All samples were measured in five replicates using one Elecsys Tg II assay on one cobas e 411 analyzer. The cross-reactivity was calculated by comparing measurements of the test and control samples. No significant cross reactivity (≤ ±10% difference of test from control) for the Elecsys Tg II was observed up to the following concentrations of the potential cross-reacting substances tested:
| Cross-reacting Substance | Concentration |
| --- | --- |
| Thyroxine Binding Globulin (TBG) | 200,000 ng/mL |
| Thyroid Stimulating Hormone (TSH) | 1,000 mIU/mL) |
# Blocking Effect of HAMA/Heterophilic Ab:
The effect of the presence of human anti-mouse antibodies (HAMA) on the Elecsys Tg II assay was assessed on the cobas e 411 analyzer. A serum pool with thyroglobulin concentration of 7 ng/mL was used to spike in HAMA. The spiked sample and the corresponding serum pool without HAMA were tested in four replicates. The recovery of the serum pool containing HAMA compared to the serum pool without HAMA was calculated. No interference was observed up to 805 µg/L of HAMA.
# 4. Assay Reportable Range:
The analytical measuring interval (AMI) for the Elecsys TgII is 0.1 – 500 ng/mL.
The instrument has a 1:10 auto-dilution feature for the device. The claimed extended measuring range is from 500 ng/mL to 5,000 ng/mL. Samples with concentrations above 5,000 ng/mL are reported as > 5,000 ng/mL.
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
## Traceability:
The Elecsys Tg II was standardized against the Certified Reference Material (CRM) 457 of the Community Bureau of Reference of the European Union.
## Stability:
### Shelf-life stability:
The shelf-life stability of the Elecsys Tg II was conducted. The reagent packs were stored at 2–8°C and the real-time stability was evaluated at baseline and testing timepoints up to 17 months. The data support a shelf-life of the Elecsys Tg II up to 15 months at 2–8°C.
{9}
### Open-pack reagent stability:
Reagent stability at 2–8°C after first opening for the Elecsys Tg II assay was evaluated on one cobas e 411 analyzer. A fresh reagent rack pack was placed on the analyzer and calibrated. Five native single donor serum samples covering the AMI of the assay were tested at baseline and other time points up to 101 days. The data of each sample at each testing time point were compared to the data of initial value tested at baseline. The data support the reagent stability of 84 days at 2–8°C after first opening.
### On-board reagent stability:
On-board reagent stability for the Elecsys Tg II was evaluated on one cobas e 411 analyzer. Five native serum samples covering the AMI of the assay were tested at baseline and other time points up to 44 days. The data of each sample at each testing time point were compared to the data of initial value tested at baseline. The data support the claim of 28 days stability for the Elecsys Tg II reagents stored on-board on the cobas e 411 analyzer.
### Calibration stability:
The Elecsys Tg II was calibrated with a fresh reagent kit on Day 0 using one cobas e 411 analyzer. Five native serum samples covering the AMI of the assay were tested up to 29 days. On each testing day, a new reagent kit of the same lot stored at 2–8°C was used, and recovery of samples was determined using the calibration curve of day 0. The data support the claim of 28 days calibration stability on the cobas e 411.
### On-board calibration stability:
On-board calibration stability for the Elecsys Tg II assay was evaluated on one cobas e 411 immunoassay analyzer. A fresh Elecsys Tg II reagent rack pack was placed on the analyzer. Five serum samples covering the AMI of the assay were tested at baseline and other time points up to 22 days. After specific time point, the same samples were measured using a separate Elecsys Tg II reagent kit kept under on-board condition (20 ± 3°C) without recalibration of the reagent kit. The data support the claim of 7 days for the on-board calibration stability on the cobas e 411.
### Sample stability:
The sample stability of serum, Li-Heparin plasma, K2-EDTA and K3-EDTA plasma was evaluated using samples stored at 2–8°C, 15–25°C, and -20°C (±5°C) (-25 to -15°C) for each sample matrix. Seven samples were prepared covering the AMI of the assay and tested using the Elecsys Tg II on the cobas e 411 analyzer at Day 0 and at various testing timepoints during storage: 2–8°C and 15–25°C up to 15 days, and -20°C (±5°C) up to 41 months. At each testing point, measurements were performed in three replicates and recovery was calculated as percent or absolute deviation compared to the values of Day 0. The data support the stability for all sample types for 14 days at 2–8°C, 14 days at 15–25°C, and 24 months at -20 (± 5)°C.
6. Detection Limit:
The Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) for Elecsys Tg II reagent pack were determined based on the CLSI EP17-A2 guidelines.
K221890 - Page 10 of 17
{10}
LoB:
The LoB was determined by testing five analyte-free serum samples with at least two Elecsys Tg II reagent lots on one cobas e 411 analyzer. Samples were tested in two replicates per run, two runs per day for two days, and one run per day for another two days, for a total of 60 measurements per reagent lot. LoB was defined as the value corresponding to the 95th percentile of the rank position of the distribution of values. The highest LoB value observed across tested reagent lots was 0.017 ng/mL. The claimed LoD for the Elecsys Tg II is 0.02 ng/mL.
LoD:
The LoD was determined by testing five serum samples containing low levels of thyroglobulin with three reagent lots on one cobas e 411 analyzer. Samples were run in two replicates per run, two runs per day for two days, and one run per day for another two days, yielding a total of 60 measurements per lot. The LoD was calculated as a log (total of low analyte samples). The highest LoD value observed across three lots was 0.04 ng/mL. The claimed LoD for Elecsys Tg II test is 0.04 ng/mL.
LoQ:
The LoQ was determined by testing a set of seven low levels of serum samples using three reagent lots on one cobas e 411 analyzer. The samples were analyzed in five replicates on one cobas e 411 analyzer, one run per day over five days, for a total of N=25 determinations per sample per reagent lot. The LoQ was determined to be the lowest analyte concentration that measured with a %CV of within-laboratory precision ≤ 20% and %bias within ±15%. The highest LoQ value observed across lots was 0.057 ng/mL. The claimed LoQ for Elecsys Tg II test is 0.1 ng/mL.
7. Assay Cut-Off:
See Clinical Cut-Off below.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Method comparison study were conducted by testing 141 native serum samples in singleton at one clinical site using one reagent lot of Elecsys Tg II (candidate device) on cobas e 411 analyzer, and one reagent lot of Access Thyroglobulin assay (predicate device) on Beckman Access DxI 800. Among 141 samples, 126 samples had values within the measuring ranges of both candidate and predicate devices and analyzed with Passing-Bablok regression. The results showed a slope of 1.394 with an intercept of -0.102. Such results indicated a systematic difference between the candidate and predicate devices. Therefore, in addition to the method comparison study, the recovery study was conducted as follows:
The recovery study used the master calibrator material traceable to the Certified Reference Material (CRM) for spiking into the Diluent MultiAssay (sample diluent used for the Elecsys Tg II). A dilution series including 17 samples were prepared by diluting the sample with value of 557 ng/mL with Diluent MultiAssay. Samples with spiked values from 1 ng/mL to
K221890 - Page 11 of 17
{11}
557 ng/mL were measured in four replicates and samples with spiked values from 0.0557 ng/mL to 0.195 ng/mL were measured in 10 replicates. All measurements were performed using one Elecsys Tg II reagent lot on one cobas e 411 analyzer. The mean-value of replicates for each sample level was compared to its expected concentration level, and the percent recovery (% recovery) was calculated. The results showed that the % recovery for 17 samples is from 91.7% to 110.5%.
2. Matrix Comparison:
A study was performed to demonstrate that lithium heparin, K2-EDTA, and K3-EDTA plasma matrices yield comparable values as serum with the Elecsys Tg II. The study included 65 matched native single and pooled samples with thyroglobulin concentration covering the measuring range of the assay. Samples were tested in singleton using one reagent lot on cobas e 401. Passing-Bablok regression analyses were performed by comparing the results of samples from different plasma samples (y) to the results of corresponding serum samples (x) for two different ranges of the serum sample concentrations (0.197 to 19.9 ng/mL and 20 to 490 ng/mL). The results are summarized in the tables below:
| Plasma/Serum | N | Range (ng/mL) | Slope (95% CI) | Intercept (95% CI) | R |
| --- | --- | --- | --- | --- | --- |
| Li-heparin plasma vs serum | 49 | 0.197 –19.9 | 0.9752 (0.9553; 0.9940) | -0.0001 (-0.0900; 0.09745 | 0.997 |
| | 16 | 20 – 490 | 1.0133 (0.9962; 1.0452) | -0.9379 (-3.0174; -0.1074) | 1.000 |
| K2-EDTA plasma vs serum | 49 | 0.197 – 19.9 | 0.9904 (0.9774; 1.0059) | 0.0005 (-0.1142; 0.0620) | 0.996 |
| | 16 | 20 – 490 | 1.0070 (0.9546; 1.0196) | -0.5143 (-1.1575; 0.7477) | 1.000 |
| K3-EDTA plasma vs serum | 49 | 0.197 – 19.9 | 0.9535 (0.9350; 0.9731) | 0.0291 (-0.0367; 0.1401) | 0.998 |
| | 16 | 20 – 490 | 0.9919 (0.9219; 1.0043) | -1.0145 (-1.9147; 1.1720) | 0.999 |
| | Li-Heparin vs Serum | K2-EDTA vs Serum | K3-EDTA vs Serum |
| --- | --- | --- | --- |
| Bias at 0.2 ng/mL | -2.5% | -0.7% | 9.9% |
| Bias at 1 ng/mL | -2.5% | -0.9% | -1.7% |
| Bias at 2 ng/mL | -2.5% | -0.9% | -3.2% |
| Bias at 10 ng/mL | -2.5% | -1.0% | -4.4% |
| Bias at 100 ng/mL | 0.4% | 0.2% | -1.8% |
| Bias at 200 ng/mL | 0.9% | 0.4% | -1.3% |
| Bias at 400 ng/mL | 1.1% | 0.6% | -1.1% |
K221890 - Page 12 of 17
{12}
C Clinical Studies:
The effectiveness of the Elecsys Tg II assay as an aid in monitoring of disease status in patients who have differentiated thyroid cancer (DTC) and have had thyroid surgery (with or without ablative therapy) was determined from a prospective, multi-center study. The clinical performance of the Elecsys Tg II at a cut-off of 0.2 ng/mL was evaluated by comparing the test results with the clinical status, the presence of structural disease (SD) over a 24-month period after surgery in subjects with a diagnosis of DTC based on an estimated real-world structural disease prevalence. The clinical study consists of two sets of patients: longitudinal cohort and cross-sectional cohort.
Subject inclusion and exclusion criteria are as follows:
**Inclusion criteria:**
- Subject ≥ 22 years of age at the time of enrollment
- Subject able and willing to provide informed consent
- Histologically confirmed and documented papillary or follicular thyroid carcinoma, including Hurthle and follicular variants of papillary carcinoma
**For Longitudinal Cohort**
- Subject diagnosed with DTC and has undergone total/near- total thyroidectomy within 4 – 12 weeks prior to enrollment
- Subjects who have a total/near total thyroidectomy as a result of a completion thyroidectomy following lobectomy and who meet all other entry requirements may be enrolled once at least 4 weeks have elapsed following the completion surgery
- Subject willing and able to tolerate removal of up to 20 mL of blood per visit
**For Cross-sectional cohort**
- Subject diagnosed with DTC and has undergone total/near total thyroidectomy > 12 weeks prior to enrollment with current evidence of structural disease
- Subject willing and able to tolerate removal of up to 20 mL of blood or in the case of residual samples, approximately 1 mL of serum is available.
**Exclusion criteria:**
- Subjects with medullary or anaplastic histology (including tumors that have any component of poorly differentiated histology if not classified as anaplastic cancer)
- Subjects with TSH-secreting pituitary adenomas
- Presence of Tg antibodies (Elecsys anti-Tg ≥ 22 IU/mL or local anti-Tg test positive)
- Known pregnancy or lactating
- Participation in an investigational medicinal product known to interfere with Tg synthesis or secretion or any other substance known to interfere directly or indirectly with Tg synthesis or secretion within the last 3 months (90 days) before enrollment, if applicable.
From nine different sites in U.S., a total of 772 samples from 219 subjects were collected for longitudinal cohort (Group 1) and a total of 72 samples from 72 subjects were collected from cross-sectional cohort (Group 2). Among them, 242 samples were excluded (due to subject
K221890 - Page 13 of 17
{13}
discontinuation or other reasons) and therefore, a total of 530 samples including 461 samples in Group 1 and 69 samples in Group 2 were used in the study analysis.
For the samples of 530 evaluated in the study, the distributions of demographic and clinical variables are described in the table below:
| N=530 | | | |
| --- | --- | --- | --- |
| Category | Demographic or Clinical Variable | N | % Total |
| Race | White, Caucasian | 430 | 81.1 |
| | African American | 30 | 5.7 |
| | Asian or Asian American | 13 | 2.5 |
| | Native Hawaiian or other Pacific Islander | 4 | 0.8 |
| | Other | 23 | 4.3 |
| | Unknown/ Not reported | 30 | 5.7 |
| Ethnicity | Hispanic or Latino | 58 | 10.9 |
| | Non-Hispanic or Latino | 470 | 88.7 |
| | Unknown | 2 | 0.4 |
| Sex | Female | 363 | 68.5 |
| | Male | 167 | 31.5 |
| Age | Minimum | 23 years | |
| | Median | 53 years | |
| | Maximum | 97 years | |
| RAI*-Treated | Yes | 108 | 20.4 |
| | No | 422 | 79.6 |
| ATA Risk Classification | Low | 237 | 44.7 |
| | Intermediate | 172 | 32.5 |
| | High | 121 | 22.8 |
*RAI: Radioactive iodine
For the Group 1 longitudinal cohort, serum samples were collected from subjects within 4 - 12 weeks following total or near total thyroidectomy but before radioiodine ablation (if planned). Thyroglobulin levels were measured at four additional time points (approximately 6 months, 12 months, 18 months, and 24 months post-surgery/radioiodine ablation) resulting in five planned visits per patient. For the Group 2 cross-sectional cohort, samples from subjects had only one single visit. All samples were tested with the Elecsys Tg II on cobas e 411 analyzer.
For each sample at each visit, structural disease $(\mathrm{SD}+)$ was defined as evidence of disease on ultrasound, cross-sectional or functional imaging, or biopsy proven disease as determined by the investigator.
# Results:
In this study, samples from the cross-sectional cohort (Group 2) were used to increase the number of observations from the patients with structural disease due to the low prevalence of structural disease in the longitudinal cohort. This increased number of observations from patients
K221890 - Page 14 of 17
{14}
with structural disease provides estimations of sensitivity and negative predictive value (NPV) with less uncertainty. Because the longitudinal and the cross-sectional cohorts were combined, the prevalence of structural disease was larger than the real-life prevalence, the estimates of NPV and positive predictive value (PPV) were adjusted using the real-world prevalence of structural disease observed in the longitudinal cohort (4.99% = 23/461).
The following tables represent the combined data from two groups and the clinical performance:
| | Disease Status | | Total | |
| --- | --- | --- | --- | --- |
| | | SD+ | | SD- |
| Elecsys Tg II | ≥0.2 ng/mL | 91 | 204 | 295 |
| | <0.2 ng/mL | 1 | 234 | 235 |
| | Total | 92 | 438 | 530 |
| Clinical Performance Measures | Estimate | 95% CI |
| --- | --- | --- |
| Sensitivity | 98.91% (91/92) | (94.10%; 99.81%) |
| Specificity | 53.42% (234/438) | (48.74%; 58.05%) |
| Prevalence | 4.99% (23/461) | (3.35%; 7.37%) |
| NPV | 99.89% | (99.42%; 99.98%) |
| PPV | 10.03% | (9.16%; 11.03%) |
The pre-defined suppressed Tg concentrations according to the ATA Response Classification categories is shown in the table below:
| Excellent Response | Tg < 0.2 ng/mL |
| --- | --- |
| Indeterminate Response | Tg ≥ 0.2 ng/mL AND Tg < 1.0 ng/mL |
| Biochemical Incomplete Response | Tg ≥ 1.0 ng/mL |
Likelihood ratios for the above three different Tg value ranges were calculated for the RAI (radioactive iodine)- treated patients and Not-RAI treated patients separately. Results of the calculations were presented in the tables below:
Combined Cohorts- RAI-Treated
| | Disease Status | | Total | LR* (95%CI) | |
| --- | --- | --- | --- | --- | --- |
| | | SD+ | | | SD- |
| Elecsys Tg II (ng/mL) | Tg < 0.2 | 1 | 27 | 28 | 0.03 (0.00; 0.14) |
| | Tg ≥0.2 and Tg<1.0 | 0 | 11 | 11 | 0.00 (0.00; 0.23) |
| | Tg ≥1.0 | 63 | 6 | 69 | 7.22 (3.67; 15.44) |
| Total | | 64 | 44 | 108 | |
*The confidence intervals for the LR were calculated as confidence intervals for the ratios of two independent binomial proportions by the asymptotic method.
K221890 - Page 15 of 17
{15}
Combined Cohorts- Not RAI-Treated
| | Disease Status | | Total | LR* (95%CI) | |
| --- | --- | --- | --- | --- | --- |
| | | SD+ | | | SD- |
| Elecsys Tg II (ng/mL) | Tg < 0.2 | 0 | 207 | 207 | 0.00 (0.00; 0.23) |
| | Tg ≥0.2 and Tg<1.0 | 0 | 124 | 124 | 0.00 (0.00; 0.39) |
| | Tg ≥1.0 | 28 | 63 | 91 | 6.25 (5.02; 7.88) |
| Total | | 28 | 394 | 422 | |
*The confidence intervals for the LR were calculated as confidence intervals for the ratios of two independent binomial proportions by the asymptotic method.
In 491 samples in the longitudinal study, 49 samples were from RAI-treated patients and 412 samples were from Not-RAI-treated patients. The prevalence of SD+ based on this study cohort was determined as: 10.20% (5/49) for RAI-treated patients and 4.37% (18/412) for Not-RAI-treated patients. The probability of SD+ for three ranges of the Elecsys Tg II described above is presented in the table below for RAI-treated and Not RAI-treated patients separately and these probabilities were adjusted for the SD+ real-world prevalence of 10.20% for RAI-treated and 4.37% for Not RAI-treated patients separately [PV probability of SD+ then PV/(1-PV)=LR*prevalence/(1-prevalence)].
| | Prevalence | Probability of SD+, (95% CI) | | |
| --- | --- | --- | --- | --- |
| | | Tg <0.2 ng/mL | Tg ≥0.2 ng/mL and Tg <1.0 ng/mL | Tg ≥1.0 ng/mL |
| RAI-Treated | 10.20% | 0.29% (0.05%; 1.55%) | 0.00% (0.00%; 2.56%) | 45.07% (29.44%; 63.70%) |
| Not RAI-Treated | 4.37% | 0.00% (0.00%; 1.04%) | 0.00% (0.00%; 1.73%) | 22.22% (18.64%; 26.46%) |
The results from the clinical study demonstrated that at Tg concentrations ≥ 1.0 ng/mL (corresponding to ATA "Biochemical Incomplete Response"), the probability of structural disease (i.e., 45% for RAI-treated patients and 22% for Not RAI-treated patients) was much higher than the probability of structural disease for Tg concentrations < 0.2 ng/mL (corresponding to ATA "Excellent Response").
D Clinical Cut-Off:
The clinical cut-off for the Elecsys Tg II is 0.2 ng/mL. This was set based on 2015 ATA Guidelines which introduced an "Excellent response" with "no clinical, biochemical or structural evidence of disease" and for patients without TSH stimulation is defined by negative imaging and a suppressed Tg < 0.2 ng/mL in the absence of Tg antibodies. Tg ≥ 0.2 AND Tg < 1.0 was used corresponding to the ATA response to therapy classification to determine "Indeterminate Response." Tg > 1.0 ng/mL was used corresponding to the ATA response to therapy classification to determine "Biochemical Incomplete Response."
E Expected Values/Reference Range:
A reference range of the Elecsys Tg II was established by testing serum samples from a total of 463 apparently healthy individuals including 244 males and 219 females. The results are summarized in the following table:
K221890 - Page 16 of 17
{16}
| | Male | Female | Total |
| --- | --- | --- | --- |
| N | 244 | 219 | 463 |
| Age (years) | 22–79 | 22–77 | 22–79 |
| Min – Max (ng/mL) | 2.66–82.5 | 1.65–234 | 1.65–234 |
| Median (ng/mL) | 15.1 | 18.3 | 16.6 |
| 2.5th percentile (ng/mL) | 3.3 | 3.9 | 3.6 |
| 97.5th percentile (ng/mL) | 63.2 | 104 | 77 |
Additionally, the value of the Elecsys Tg II was evaluated in a total of 127 subjects with differentiated thyroid cancer with no evidence of disease for 4 or more years following total/near-total thyroidectomy. The study included 100 female patients aged 26 to 92 years, and 27 male patients aged 34 to 81 years. The results showed that the thyroglobulin concentration ranged 0.1 to 11.6 ng/mL. About 80.3% (102/127) of subjects had thyroglobulin level below 0.1 ng/mL, and 95% of patients had thyroglobulin level ≤ 0.786 ng/mL.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K221890 - Page 17 of 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.