← Product Code [LZA](/productcode/LZA) · K221887

# Nitrile Examination Glove, Pink Color; Nitrile Examination Glove, Black Color, Nitrile Examination Glove, White Color (K221887)

_Guangdong Kingfa Sci. & Tech.Co., Ltd. · LZA · Sep 24, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K221887

## Device Facts

- **Applicant:** Guangdong Kingfa Sci. & Tech.Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Sep 24, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

## Device Story

Nitrile examination gloves; single-use; non-sterile; powder-free. Available in pink, black, and white colors; sizes ranging from extra-small to extra extra-large. Used by healthcare examiners to provide a barrier against contamination between patient and examiner. Device functions as a physical barrier; material is nitrile rubber. No electronic, software, or algorithmic components. Bench testing confirms compliance with ASTM standards for physical dimensions, tensile strength, elongation, and freedom from holes. Biocompatibility testing confirms non-irritating, non-sensitizing, and non-toxic properties.

## Clinical Evidence

No clinical data. Bench testing only. Testing included physical dimensions (ASTM D6319), tensile strength/elongation (ASTM D412), freedom from holes (ASTM D5151), residual powder (ASTM D6124), and biocompatibility (ISO 10993-10, ISO 10993-11). All results met acceptance criteria.

## Technological Characteristics

Material: Nitrile rubber. Form factor: Disposable, non-sterile, powder-free gloves. Sizes: XS to XXL. Standards: ASTM D6319-19 (specifications), ASTM D3767-03 (dimensions), ASTM D5151-19 (holes), ASTM D6124-06 (powder), ASTM D573-04 (deterioration), ASTM D412-16 (tension). Biocompatibility: ISO 10993-10 (irritation/sensitization), ISO 10993-11 (systemic toxicity). No software or electronic components.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, Pink/Black Color ([K190942](/device/K190942.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

September 24, 2022

Guangdong Kingfa Sci. & Tech. Co., Ltd. Xiaoge Yu Manager No. 28 Delong Ave., Shijiao Town, Qingcheng District Qingyuan, Guangdong 511545 China

Re: K221887

Trade/Device Name: Nitrile Examination Glove, Pink Color, Nitrile Examination Glove, Black Color, Nitrile Examination Glove, White Color Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, Dated: June 9, 2022 Received: June 29, 2022

Dear Xiaoge Yu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Oian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K221887

#### Device Name

Nitrile Examination Glove, Pink Color Nitrile Examination Glove, Black Color Nitrile Examination Glove, White Color

#### Indications for Use (Describe)

The nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K221887-510(k) summary

#### l. Submitter

GUANGDONG KINGFA SCI. & TECH.CO., LTD. No. 28 Delong Avenue, Shijiao Town, Qingcheng District, Qingyuan City, Guangdong Province 511545, China

Contact person: Xiaoge Yu Position: Manager Tel.: +86-13570952157 E-mail: yuxiaoge@kingfa.com.cn

Preparation date: June 09, 2022

## II. Proposed Device

| Device Trade Name  | Nitrile Examination Glove, Pink Color  |
|--------------------|----------------------------------------|
|                    | Nitrile Examination Glove, Black Color |
|                    | Nitrile Examination Glove, White Color |
| Common name:       | Polymer Patient Examination Glove      |
| Regulation Number: | 21 CFR 880.6250                        |
| Regulatory Class:  | Class I                                |
| Product code:      | LZA                                    |
| Review Panel       | General Hospital                       |

#### III. Predicate Devices

| 510(k) Number:  | K190942                                                               |
|-----------------|-----------------------------------------------------------------------|
| Trade name:     | Disposable Powder Free Nitrile Examination Glove, Pink/Black<br>Color |
| Common name:    | Patient Examination Gloves                                            |
| Classification: | Class I                                                               |
| Product Code:   | LZA                                                                   |
| Manufacturer    | Ever Growth (Vietnam) Co. Ltd.                                        |

#### IV. Device description

The propose devices is powder free nitrile patient examination gloves, provided as non-sterile and disposable device. The proposed devices are provided with white, black and pink color. There are six sizes, extra-small, medium, large and extra-large, extra extra-large for optional.

{4}------------------------------------------------

# V. Indication for use

The nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

## VI. Comparison of technological characteristics with the predicate devices

| Item                                     | Proposed device                                                                                                                                                                                                             | Predicate device<br>(K190942)                                                                                                                                                                                                                                | Discussion |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product name                             | Nitrile Examination<br>Glove, Pink Color<br>Nitrile Examination<br>Glove, Black Color<br>Nitrile Examination<br>Glove, White Color                                                                                          | Disposable Powder<br>Free Nitrile<br>Examination Glove,<br>Pink/Black Color                                                                                                                                                                                  | -          |
| Product Code                             | LZA                                                                                                                                                                                                                         | LZA                                                                                                                                                                                                                                                          | Same       |
| Regulation No.                           | 21 CFR 880.6250                                                                                                                                                                                                             | 21 CFR 880.6250                                                                                                                                                                                                                                              | Same       |
| Classification                           | Class I                                                                                                                                                                                                                     | Class I                                                                                                                                                                                                                                                      | Same       |
| Powder free                              | Yes                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                          | Same       |
| Indication for<br>use                    | The nitrile<br>examination glove is<br>intended to be worn<br>on the hands of<br>examiner's to prevent<br>contamination<br>between patient and<br>examiner. This is a<br>single-use,<br>powder-free,<br>non-sterile device. | The Nitrile Powder<br>Free patient<br>examination glove is<br>a non-sterile<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's<br>hands or finger to<br>prevent<br>contamination<br>between patient and<br>examiner. | Same       |
| Main Material                            | Nitrile rubber                                                                                                                                                                                                              | Nitrile rubber                                                                                                                                                                                                                                               | Same       |
| Color                                    | Pink\ Black\ White                                                                                                                                                                                                          | Pink\ Black                                                                                                                                                                                                                                                  | Similar    |
| Size                                     | X-Small, Small,<br>Medium, Large,<br>X-large. XX-large.                                                                                                                                                                     | X-Small, Small,<br>Medium, Large,<br>X-large                                                                                                                                                                                                                 | Similar    |
| Palm width                               | X- Small(70±10mm)                                                                                                                                                                                                           | X- Small(70±10mm)                                                                                                                                                                                                                                            | Similar    |
|                                          | Small (80±10mm)                                                                                                                                                                                                             | Small (80±10mm)                                                                                                                                                                                                                                              |            |
|                                          | Medium (95±10mm)                                                                                                                                                                                                            | Medium (95±10mm)                                                                                                                                                                                                                                             |            |
|                                          | Large (110±10mm)                                                                                                                                                                                                            | Large (110±10mm)                                                                                                                                                                                                                                             |            |
|                                          | X-large (120±10mm)                                                                                                                                                                                                          | X large (120±10mm)                                                                                                                                                                                                                                           |            |
|                                          | XX-large (≥ 120mm)                                                                                                                                                                                                          |                                                                                                                                                                                                                                                              |            |
| Length                                   | XS(220mm min)                                                                                                                                                                                                               | ≥230mm                                                                                                                                                                                                                                                       | Similar    |
|                                          | S (220mm min)                                                                                                                                                                                                               |                                                                                                                                                                                                                                                              |            |
|                                          | M (230mm min)                                                                                                                                                                                                               |                                                                                                                                                                                                                                                              |            |
|                                          | L (230mm min)                                                                                                                                                                                                               |                                                                                                                                                                                                                                                              |            |
|                                          | XL (230mm min)                                                                                                                                                                                                              |                                                                                                                                                                                                                                                              |            |
|                                          | XXL (230mm min)                                                                                                                                                                                                             |                                                                                                                                                                                                                                                              |            |
|                                          | Palm: 0.05mm min                                                                                                                                                                                                            | Palm≥0.05mm                                                                                                                                                                                                                                                  | Similar    |
| Thickness                                | Finger: 0.08mm min                                                                                                                                                                                                          | Finger tip≥0.05mm                                                                                                                                                                                                                                            |            |
| Freedom from                             | Meets requirements                                                                                                                                                                                                          | Meets requirements                                                                                                                                                                                                                                           | Similar    |
| holes                                    | of the ASTM                                                                                                                                                                                                                 | of the ASTM                                                                                                                                                                                                                                                  |            |
|                                          | D6319-19                                                                                                                                                                                                                    | D6319-10                                                                                                                                                                                                                                                     |            |
| Physical                                 | Meets requirements                                                                                                                                                                                                          | Meets requirements                                                                                                                                                                                                                                           | Similar    |
| Properties                               | of the ASTM                                                                                                                                                                                                                 | of the ASTM                                                                                                                                                                                                                                                  |            |
| (before aging)                           | D6319-19                                                                                                                                                                                                                    | D6319-10                                                                                                                                                                                                                                                     |            |
| Physical                                 | Meets requirements                                                                                                                                                                                                          | Meets requirements                                                                                                                                                                                                                                           | Similar    |
| Properties                               | of the ASTM                                                                                                                                                                                                                 | of the ASTM                                                                                                                                                                                                                                                  |            |
| (after aging)                            | D6319-19                                                                                                                                                                                                                    | D6319-10                                                                                                                                                                                                                                                     |            |
| Powder                                   | ≤2.0 mg/gloves                                                                                                                                                                                                              | ≤2.0 mg/gloves                                                                                                                                                                                                                                               | Same       |
| residual                                 |                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                              |            |
| Sterility                                | Non-sterile                                                                                                                                                                                                                 | Non-sterile                                                                                                                                                                                                                                                  | Same       |
| For single use                           | Yes                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                          | Same       |
| Type of use                              | Over the counter use                                                                                                                                                                                                        | Over the counter use                                                                                                                                                                                                                                         | Same       |
| Biocompatibility                         | ISO 10993-10                                                                                                                                                                                                                | ISO 10993-10                                                                                                                                                                                                                                                 | Same       |
|                                          | Under the conditions                                                                                                                                                                                                        | Under the conditions                                                                                                                                                                                                                                         |            |
|                                          | of the study, not an                                                                                                                                                                                                        | of the study, not an                                                                                                                                                                                                                                         |            |
|                                          | irritant and sensitizer                                                                                                                                                                                                     | irritant and sensitizer                                                                                                                                                                                                                                      |            |
|                                          | ISO 10993-11                                                                                                                                                                                                                |                                                                                                                                                                                                                                                              |            |
|                                          | Cytotoxicity is                                                                                                                                                                                                             |                                                                                                                                                                                                                                                              |            |
|                                          | assessed via                                                                                                                                                                                                                |                                                                                                                                                                                                                                                              |            |
|                                          | rationale. Under the                                                                                                                                                                                                        |                                                                                                                                                                                                                                                              |            |
|                                          | condition of acute                                                                                                                                                                                                          |                                                                                                                                                                                                                                                              |            |
|                                          | systemic toxicity test,                                                                                                                                                                                                     |                                                                                                                                                                                                                                                              |            |
|                                          | the test article did not                                                                                                                                                                                                    |                                                                                                                                                                                                                                                              |            |
| show acute systemic<br>toxicity in vivo. |                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                              |            |

Table 1 Comparison of Nitrile Examination Gloves

{5}------------------------------------------------

{6}------------------------------------------------

#### VII. Non-Clinical Testing

Non clinical tests were conducted in accordance with following standards to verify that the proposed device met all design specifications.

- ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for . Medical Application
- ASTM D3767-03(2020), Practice for rubber-Measurement of Dimensions ●
- ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves ●
- ASTM D6124-06(2017), Standard Test Method for Residual Powder on Medical . Gloves
- ASTM D573-04(2019), Standard Test Method for Rubber—Deterioration in an Air ● Oven
- ASTM D412-16, Standard Test Methods for Vulcanized Rubber and ● Thermoplastic Elastomers—Tension
- . ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- ISO 10993-11:2017, Biological evaluation of medical devices Part11:Tests for . Systemic Toxicity

| Test Method | Purpose                     | Acceptance Criteria                                              | Results                                     |
|-------------|-----------------------------|------------------------------------------------------------------|---------------------------------------------|
| ASTM D6319  | Physical<br>Dimensions Test | Extra-Small:<br>Length: $\geq$ 220mm Width: 70 $\pm$ 10 mm;      | Pass                                        |
|             |                             | Small:<br>Length: $\geq$ 220mm Width: 80 $\pm$ 10mm;             | X-Small:<br>Length: 226 mm<br>Width: 73 mm; |
|             |                             | Medium:<br>Length: $\geq$ 230mm Width: 95 $\pm$ 10mm             | Small:<br>Length: 225 mm<br>Width: 85 mm;   |
|             |                             | Large:<br>Length: $\geq$ 230mm Width: 110 $\pm$ 10mm             | Medium:<br>Length: 235 mm<br>Width: 98 mm;  |
|             |                             | Extra- Large:<br>Length: $\geq$ 230mm Width: 120 $\pm$ 10mm      | Large:<br>Length: 234 mm                    |
|             |                             | Extra- Extra- Large:<br>Length: $\geq$ 230mm Width: $\geq$ 120mm |                                             |

|  |  | Table 2 Summary of Non-Clinical Performance Testing |  |
|--|--|-----------------------------------------------------|--|

{7}------------------------------------------------

|                        |                                                                                                    |                                                     |        |                                                                                                | Width: 115 mm;     |
|------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------|--------|------------------------------------------------------------------------------------------------|--------------------|
|                        |                                                                                                    |                                                     |        |                                                                                                | X- Large:          |
|                        |                                                                                                    |                                                     |        |                                                                                                | Length: 235 mm     |
|                        |                                                                                                    |                                                     |        |                                                                                                | Width: 123 mm;     |
|                        |                                                                                                    |                                                     |        |                                                                                                | XX-Large:          |
|                        |                                                                                                    |                                                     |        |                                                                                                | Length: 235 mm     |
|                        |                                                                                                    |                                                     |        |                                                                                                | Width: 124 mm;     |
|                        |                                                                                                    | Thickness (mm):                                     |        |                                                                                                | Pass               |
|                        |                                                                                                    | Finger:≥0.08<br>Palm: ≥0.05                         |        |                                                                                                | X-Small:           |
|                        |                                                                                                    |                                                     |        |                                                                                                | Finger: 0.104 mm   |
|                        |                                                                                                    |                                                     |        |                                                                                                | Palm: 0.059 mm;    |
|                        |                                                                                                    |                                                     |        |                                                                                                | Small:             |
|                        |                                                                                                    |                                                     |        |                                                                                                | Finger: 0.106 mm   |
|                        |                                                                                                    |                                                     |        |                                                                                                | Palm: 0.060 mm;    |
|                        |                                                                                                    |                                                     |        |                                                                                                | Medium:            |
|                        |                                                                                                    |                                                     |        |                                                                                                | Finger: 0.104 mm   |
|                        |                                                                                                    |                                                     |        |                                                                                                | Palm: 0.060 mm;    |
|                        |                                                                                                    |                                                     |        |                                                                                                | Large:             |
|                        |                                                                                                    |                                                     |        |                                                                                                | Finger: 0.105 mm   |
|                        |                                                                                                    |                                                     |        |                                                                                                | Palm: 0.062 mm;    |
|                        |                                                                                                    |                                                     |        |                                                                                                | X- Large:          |
|                        |                                                                                                    |                                                     |        |                                                                                                | Finger: 0.106 mm   |
|                        |                                                                                                    |                                                     |        |                                                                                                | Palm: 0.061mm;     |
|                        |                                                                                                    |                                                     |        |                                                                                                | XX-Large:          |
|                        |                                                                                                    |                                                     |        |                                                                                                | Finger: 0.105 mm   |
|                        |                                                                                                    |                                                     |        |                                                                                                | Palm: 0.058 mm;    |
| Physical<br>properties | Before<br>Aging                                                                                    | Tensile<br>Strength                                 | ≥14MPa |                                                                                                | Pass               |
|                        |                                                                                                    |                                                     |        |                                                                                                | Lot1:17.6MPa, 552% |
|                        |                                                                                                    | Ultimate<br>Elongation                              | ≥500%  |                                                                                                | Lot2:18.5MPa, 573% |
|                        |                                                                                                    |                                                     |        |                                                                                                | Lot3:18.2MPa, 583% |
|                        | After Aging                                                                                        | Tensile<br>Strength                                 | ≥14MPa |                                                                                                | Pass               |
|                        |                                                                                                    |                                                     |        |                                                                                                |                    |
|                        |                                                                                                    | Ultimate<br>Elongation                              | ≥500%  | Lot1:19.8MPa, 521%<br>Lot2:20.3MPa, 518%<br>Lot3:20.1MPa, 526%                                 |                    |
| ASTM D5151             | Freedom from<br>pinholes                                                                           | Meet the requirements of ASTM D5151Test for AQL 2.5 |        | Pass<br>Lot1: ≤ AQL 2.5<br>Lot2: ≤ AQL 2.5<br>Lot3: ≤ AQL 2.5                                  |                    |
| ASTM D6124             | Powder Residue                                                                                     | Meet the requirements of ASTM D6124< 2.0mg          |        | Pass<br>Lot1: 1.2 mg<br>Lot2: 1.5 mg<br>Lot3: 1.2 mg                                           |                    |
| ISO 10993-10           | To determine if the<br>finished device<br>material is an<br>irritant                               | Non-irritating                                      |        | Under the<br>conditions of the<br>study not an<br>irritant/ Pass                               |                    |
| ISO 10993-10           | To determine if the<br>finished device<br>material is a<br>sensitizer                              | Non- sensitizing                                    |        | Under conditions<br>of the study, not a<br>sensitizer. / Pass                                  |                    |
| ISO 10993-11           | To determine if the<br>finished device<br>material extracts<br>pose a systemic<br>toxicity concern | Non-acute systemic toxicity                         |        | Under conditions<br>of the study, did<br>not show acute<br>systemic toxicity<br>in vivo / Pass |                    |

{8}------------------------------------------------

## VIII. Clinical Testing

No clinical study is included in this submission.

#### IX. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K221887](https://fda-staging.innolitics.com/device/K221887)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).